Skip to content

ZOOM LATITUDE recallZ-1670-2022 · Boston Scientific Corporation · initiated Jul 7, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is potential for unanticipated software anomalies to occur when using the Model 2868 Application loaded on the 3120 ZOOM LATITUTDE programmer when communicating with certain Boston Scientific transvenous defibrillators.

FDA root cause category: Under Investigation by firm

What the company is doing

Boston Scientific notified their consignees by letter on 07/07/2022. The Important Programmer Information stated the following: "Programmer Software Information July 2022 - Continued use of older Model 2868 software on Model 3120 ZOOM programmers to support transvenous CRT-Ds and ICDs until supply constraints on newer programmers are resolved. Summary- " Global supply chain challenges have constrained availability of the newer programmers. " Until the supply chain constraints are resolved locally, you may continue using older 3120 ZOOM programmers. " The 2868 software on the 3120 ZOOM programmer has been uninstalled or is outdated. As such, users may observe unexpected behaviors when attempting communications with transvenous defibrillators including the potential inability to interrogate or the display of incorrect/missing parameters " A software update has been developed to resolve some of the disruptive behaviors associated with continued use of 2868 software on the 3120 ZOOM programmer. Boston Scientific will release software after local approval. " The risk of insufficient programmer supply outweighs the risks associated with these unexpected behaviors, therefore, Boston Scientific recommends continued use of 3120 ZOOM programmers for transvenous defibrillators until sufficient supply of the newer programmers can be achieved locally." and provided the following recommendations: "Recommendations: Please distribute this letter to all users of the 3120 ZOOM programmer and complete/return the enclosed Customer Acknowledgment Form. Continue using 3120 ZOOM programmers to interrogate Boston Scientific TV defibrillators, as needed, with the following in mind: 1) If unable to interrogate a TV defibrillator with a 3120 ZOOM programmer, use a 3300 LATITUDE programmer instead. If this behavior occurs, STAT PACE, STAT SHOCK, and DIVERT THERAPY programmer operations4 are not available through a 3120 ZOOM programmer, so external pacing/defibrillation or magnet applicat

Recall details

Recall number
Z-1670-2022
Status
Open, Classified
Initiated by the company
Jul 7, 2022
Recalling firm
Boston Scientific Corporation
Quantity
17,172 units in total
Product code
LWS
FDA submissions
P960040
Distribution
Worldwide Distribution

Product description

The PRM is intended to be used as part of the ZOOM LATITUDE Programming System to communicate with Boston Scientific implantable pulse generators. The software in use controls all communication functions for the pulse generator.

Code information (models, lots, serial numbers)

Show the full code information (142k characters)

UPN: 00802526434723, Serial Numbers: 051026, 051075, 051091, 051121, 051146, 051156, 051171, 051182, 051207, 051232, 051236, 051239, 051270, 051320, 051335, 051358, 051367, 051370, 051374, 051413, 051418, 051425, 051432, 051435, 051456, 051459, 051473, 051478, 051486, 051487, 051495, 051501, 051511, 051533, 051535, 051549, 051552, 051558, 051561, 051580, 051584, 051592, 051593, 051612, 051623, 051626, 051632, 051640, 051655, 051696, 051727, 051745, 051763, 051769, 051770, 051784, 051808, 051818, 051846, 051850, 051863, 051906, 051924, 051931, 051989, 051997, 052018, 052022, 052030, 052034, 052088, 052131, 052145, 052202, 052320, 052460, 054970, 054981, 054993, 054996, 055007, 055016, 055020, 055023, 055037, 055041, 055048, 055059, 055065, 055066, 055069, 055070, 055079, 055086, 055089, 055098, 055113, 055117, 055122, 055132, 055142, 055148, 055154, 055157, 055172, 055174, 055191, 055193, 055205, 055211, 055213, 055214, 055219, 055227, 055246, 055248, 055253, 055256, 055270, 055277, 055306, 055330, 055336, 055343, 055344, 055345, 055350, 055360, 055362, 055376, 055386, 055387, 055388, 055416, 055424, 055449, 055454, 055479, 055501, 055507, 055527, 055537, 055541, 055546, 055568, 055570, 055578, 055580, 055586, 055591, 055625, 055641, 055650, 055658, 055662, 055665, 055666, 055675, 055676, 055677, 055689, 055693, 055697, 055703, 055709, 055721, 055726, 055733, 055735, 055743, 055747, 055750, 055755, 055758, 055789, 055791, 055793, 055799, 055800, 055812, 055813, 055818, 055856, 055860, 055861, 055871, 055872, 055873, 055874, 055877, 055878, 055882, 055885, 055915, 055943, 055952, 055959, 055960, 055977, 055978, 055985, 055987, 056010, 056034, 056051, 056055, 056062, 056069, 056075, 056078, 056100, 056118, 056127, 056130, 056143, 056146, 056164, 056180, 056185, 056192, 056213, 056227, 056229, 056248, 056250, 056289, 056295, 056311, 056315, 056318, 056325, 056329, 056330, 056332, 056336, 056342, 056347, 056348, 056366, 056382, 056395, 056405, 056408, 056424, 056439, 056441, 056449, 056451, 056459, 056460, 056461, 056466, 056472, 056492, 056497, 056501, 056506, 056510, 056522, 056553, 056554, 056556, 056570, 056574, 056576, 056578, 056584, 056591, 056609, 056620, 056645, 056652, 056659, 056666, 056671, 056680, 056689, 056691, 056697, 056698, 056699, 056707, 056737, 056743, 056748, 056749, 056752, 056766, 056786, 056787, 056794, 056796, 056805, 056807, 056808, 056813, 056859, 056871, 056877, 056891, 056899, 056904, 056921, 056925, 056927, 056935, 056954, 056958, 056961, 056964, 056965, 056966, 056970, 057002, 057007, 057008, 057020, 057022, 057042, 057047, 057065, 057066, 057078, 057082, 057087, 057101, 057113, 057115, 057124, 057126, 057129, 057131, 057141, 057143, 057147, 057154, 057163, 057172, 057175, 057180, 057186, 057190, 057192, 057210, 057221, 057228, 057237, 057238, 057250, 057290, 057294, 057296, 057301, 057302, 057310, 057330, 057340, 057347, 057361, 057364, 057371, 057374, 057376, 057380, 057384, 057397, 057402, 057404, 057463, 057471, 057478, 057490, 057497, 057501, 057513, 057519, 057520, 057534, 057535, 057548, 057549, 057553, 057558, 057562, 057567, 057572, 057575, 057587, 057590, 057594, 057610, 057619, 057623, 057631, 057635, 057639, 057648, 057655, 057657, 057659, 057662, 057664, 057665, 057667, 057676, 057677, 057687, 057694, 057695, 057703, 057704, 057708, 057711, 057726, 057728, 057729, 057733, 057742, 057743, 057753, 057780, 057790, 057794, 057812, 057814, 057820, 057825, 057829, 057836, 057838, 057843, 057854, 057856, 057864, 057877, 057882, 057887, 057893, 057916, 057925, 057928, 057941, 057944, 057946, 057949, 057953, 057956, 057973, 057979, 057983, 057986, 057996, 057997, 058003, 058014, 058020, 058035, 058037, 058041, 058043, 058047, 058048, 058050, 058053, 058059, 058089, 058103, 058105, 058117, 058123, 058126, 058132, 058133, 058134, 058136, 058145, 058147, 058156, 058163, 058171, 058184, 058192, 058206, 058210, 058212, 058228, 058241, 058248, 058257, 058267, 058268, 058294, 058314, 058319, 058329, 058331, 058341, 058350, 058358, 058368, 058378, 058379, 058381, 058385, 058391, 058393, 058406, 058412, 058425, 058427, 058430, 058431, 058432, 058441, 058455, 058457, 058460, 058463, 058464, 058467, 058472, 058473, 058478, 058484, 058525, 058526, 058527, 058528, 058529, 058539, 058540, 058547, 058554, 058585, 058597, 058601, 058604, 058606, 058615, 058622, 058625, 058626, 058630, 058638, 058639, 058642, 058648, 058655, 058674, 058675, 058680, 058684, 058688, 058694, 058696, 058698, 058700, 058713, 058740, 058743, 058744, 058752, 058753, 058761, 058768, 058776, 058780, 058781, 058783, 058787, 058791, 058793, 058799, 058803, 058807, 058811, 058820, 058822, 058825, 058828, 058848, 058851, 058857, 058889, 058897, 058900, 058910, 058919, 058922, 058929, 058940, 058951, 058954, 058958, 058962, 058966, 058978, 058982, 058985, 058994, 058997, 059002, 059003, 059014, 059019, 059022, 059025, 059054, 059057, 059072, 059082, 059083, 059097, 059106, 059110, 059116, 059123, 059128, 059140, 059142, 059143, 059174, 059176, 059183, 059200, 059209, 059223, 059235, 059236, 059251, 059255, 059256, 059266, 059267, 059276, 059278, 059280, 059282, 059285, 059289, 059295, 059332, 059336, 059339, 059351, 059354, 059355, 059373, 059383, 059388, 059392, 059403, 059408, 059409, 059420, 059421, 059426, 059436, 059439, 059451, 059478, 059482, 059502, 059504, 059508, 059522, 059531, 059538, 059548, 059557, 059560, 059564, 059572, 059591, 059593, 059605, 059612, 059629, 059636, 059643, 059658, 059679, 059681, 059684, 059686, 059689, 059706, 059718, 059757, 059762, 059767, 059771, 059808, 059828, 059831, 059846, 059857, 059880, 059899, 059900, 059902, 059911, 059917, 059918, 059927, 059930, 059942, 059976, 059987, 059988, 059991, 059993, 059995, 060005, 060011, 060029, 060032, 060034, 060037, 060040, 060052, 060053, 060073, 060078, 060080, 060091, 060098, 060099, 060105, 060112, 060124, 060142, 060150, 060163, 060171, 060179, 060180, 060191, 060205, 060209, 060220, 060225, 060227, 060229, 060238, 060241, 060243, 060251, 060256, 060257, 060261, 060264, 060265, 060277, 060280, 060281, 060283, 060290, 060295, 060303, 060310, 060311, 060312, 060321, 060322, 060344, 060357, 060364, 060366, 060367, 060390, 060392, 060402, 060419, 060427, 060428, 060437, 060467, 060468, 060472, 060473, 060478, 060480, 060482, 060497, 060500, 060504, 060509, 060512, 060523, 060524, 060525, 060544, 060545, 060548, 060550, 060557, 060582, 060585, 060589, 060597, 060629, 060633, 060641, 060652, 060653, 060659, 060660, 060675, 060677, 060694, 060702, 060711, 060715, 060718, 060753, 060755, 060759, 060761, 060779, 060781, 060783, 060786, 060802, 060811, 060812, 060818, 060827, 060837, 060843, 060845, 060850, 060852, 060864, 060865, 060876, 060881, 060882, 060891, 060910, 060917, 060918, 060919, 060922, 060923, 060926, 060928, 060940, 060942, 060944, 060956, 060958, 060963, 060964, 061001, 061006, 061012, 061014, 061046, 061050, 061055, 061072, 061813, 061816, 061818, 061831, 061844, 061851, 061891, 061904, 061906, 061945, 061978, 061979, 062021, 062024, 062026, 062030, 062037, 062042, 062047, 062060, 062061, 062069, 062081, 062095, 062103, 062108, 062109, 062114, 062132, 062140, 062144, 062145, 062146, 062159, 062160, 062162, 062174, 062176, 062177, 062183, 062198, 062203, 062204, 062209, 062213, 062214, 062215, 062218, 062229, 062238, 062252, 062262, 062268, 062276, 062281, 062286, 062299, 062304, 062307, 062313, 062352, 062355, 062357, 062358, 062369, 062375, 062377, 062392, 062395, 062401, 062403, 062411, 062417, 062429, 062431, 062459, 062467, 062485, 062490, 062493, 062496, 062508, 062512, 062519, 062520, 062521, 062523, 062529, 062539, 062543, 062554, 062562, 062569, 062570, 062579, 062597, 062599, 062600, 063455, 063456, 063477, 063480, 063490, 063497, 063507, 063508, 063517, 063529, 063531, 063532, 063534, 063542, 063547, 063548, 063560, 063567, 063577, 063578, 063579, 063604, 063605, 063609, 063620, 063622, 063634, 063635, 063641, 063649, 063664, 063672, 063683, 063684, 063689, 063694, 063699, 063707, 063734, 063735, 063740, 063744, 063752, 063753, 063779, 063780, 063781, 063783, 063786, 063788, 063790, 063793, 063805, 063809, 063826, 063832, 063841, 063845, 063846, 063849, 063851, 063862, 063867, 063868, 063869, 063871, 063875, 063888, 063914, 063917, 063931, 063932, 063942, 063963, 063983, 063991, 063992, 064009, 064011, 064019, 064020, 064027, 064036, 064039, 064040, 064042, 064051, 064052, 064053, 064054, 064055, 064056, 064071, 064072, 064079, 064124, 064142, 064144, 064145, 064148, 064152, 064153, 064162, 064166, 064177, 064182, 064190, 064203, 064229, 064234, 064246, 064268, 064280, 064281, 064288, 064301, 064302, 064304, 064309, 064318, 064329, 064340, 064346, 064349, 064354, 064356, 064358, 064368, 064369, 064383, 064389, 064395, 064399, 064401, 064405, 064406, 064411, 064420, 064437, 064443, 064444, 064460, 064472, 064487, 064488, 064507, 064510, 064516, 064526, 064528, 064531, 064533, 064552, 064558, 064562, 064569, 064573, 064574, 064575, 064584, 064591, 064597, 064606, 064616, 064634, 064674, 064676, 064677, 064687, 064710, 064717, 064723, 064730, 064732, 064734, 064735, 064739, 064742, 064743, 064764, 064774, 064781, 064813, 064816, 064817, 064822, 064827, 064842, 064844, 064855, 064860, 064864, 064874, 064879, 064881, 064884, 064890, 064894, 064916, 064919, 064926, 064946, 064948, 064961, 064962, 064971, 064976, 064977, 064993, 064995, 064996, 064997, 065011, 065012, 065013, 065014, 065018, 065035, 065041, 065044, 065045, 065067, 065076, 065081, 065084, 065097, 065115, 065127, 065140, 065141, 065182, 065188, 065190, 065194, 065199, 065200, 065215, 065239, 065241, 065283, 065287, 065291, 065299, 065307, 065310, 065314, 065328, 065333, 065335, 065345, 065349, 065356, 065363, 065364, 065368, 065374, 065390, 065405, 065406, 065418, 065424, 065426, 065439, 065444, 065451, 065453, 065464, 065469, 065474, 065483, 065485, 065490, 065500, 065504, 065508, 065513, 065517, 065520, 065528, 065537, 065539, 065541, 065545, 065558, 065569, 065572, 065585, 065586, 065592, 065602, 065610, 065614, 065624, 065625, 065630, 065636, 065638, 065644, 065645, 065665, 065666, 065672, 065684, 065694, 065698, 065710, 065718, 065725, 065727, 065746, 065751, 065758, 065770, 065772, 065774, 065780, 065794, 065800, 065802, 065803, 065917, 065955, 065957, 065969, 065974, 065975, 065976, 065987, 065991, 066018, 066032, 066034, 066035, 066041, 066043, 066049, 066052, 066053, 066062, 066069, 066081, 066090, 066095, 066097, 066099, 066500, 066504, 066520, 066545, 066565, 066568, 066570, 066579, 066587, 066588, 066591, 066592, 066594, 066612, 066613, 066646, 066660, 066690, 066739, 066754, 066866, 066882, 066884, 066940, 066944, 066946, 066961, 066963, 066973, 066974, 066975, 066976, 066993, 066995, 067004, 067012, 067018, 067020, 067027, 067030, 067031, 067043, 067059, 067364, 067365, 067371, 067379, 067386, 067393, 067401, 067409, 067415, 067418, 067428, 067453, 067456, 067471, 067491, 067497, 067505, 067514, 067534, 067538, 067552, 067554, 067558, 067563, 067565, 067570, 067596, 067601, 067618, 067621, 067624, 067641, 067644, 067647, 067660, 067665, 067673, 067677, 067681, 067691, 067694, 067703, 067713, 067723, 067736, 067738, 067739, 067746, 067747, 067748, 067755, 067808, 067825, 067827, 068198, 068202, 068214, 068220, 068224, 068230, 068234, 068247, 068255, 068269, 068271, 068281, 068297, 068298, 068321, 068323, 068331, 068332, 068334, 068335, 068344, 068349, 068353, 068471, 068475, 068477, 068480, 068492, 068512, 068515, 068516, 068518, 068539, 068560, 068562, 068572, 068576, 068588, 068591, 068592, 068594, 068608, 068612, 068615, 068631, 068632, 068635, 068641, 068644, 068646, 068660, 068667, 068679, 068688, 068708, 068709, 068726, 068730, 068732, 068756, 068758, 068767, 068780, 068806, 068807, 068842, 068846, 068855, 068878, 068885, 068899, 068904, 068921, 068930, 068936, 068946, 068948, 068971, 068978, 068983, 068991, 069000, 069005, 069016, 069021, 069022, 069023, 069025, 069038, 069040, 069046, 069047, 069069, 069080, 069083, 069091, 069093, 069095, 069109, 069125, 069128, 069134, 069135, 069138, 069142, 069143, 069145, 069161, 069166, 069191, 069203, 069215, 069221, 069249, 069270, 069280, 069300, 069314, 069315, 069326, 069352, 069365, 069367, 069370, 069372, 069375, 069376, 069383, 069385, 069392, 069394, 069410, 069414, 069416, 069420, 069426, 069444, 069445, 069458, 069461, 069462, 069477, 069492, 069494, 069497, 069502, 069512, 069513, 069518, 069520, 069532, 069533, 069536, 069537, 069549, 069562, 069579, 069595, 069602, 069604, 069622, 069635, 069637, 069654, 069659, 069663, 069666, 069680, 069740, 069742, 069751, 069753, 069756, 069758, 069763, 069764, 069765, 069772, 069773, 069780, 069785, 069792, 069796, 069810, 069814, 069821, 069829, 069834, 069836, 069838, 069855, 069858, 069866, 069880, 070255, 070297, 070298, 070319, 070346, 070352, 070370, 070374, 070399, 070410, 070415, 070419, 070547, 070551, 070559, 070578, 070583, 070586, 070594, 070596, 070601, 070611, 070621, 070626, 070634, 070650, 070665, 070672, 070695, 070712, 070713, 070737, 070738, 070741, 070742, 070744, 070757, 070760, 070761, 070795, 070805, 070811, 070825, 070830, 070841, 070850, 070859, 070876, 070891, 070919, 070931, 070954, 070961, 071002, 071006, 071038, 071039, 071042, 071047, 071049, 071098, 071105, 071114, 071118, 071127, 071134, 071154, 071164, 071169, 071170, 071171, 071179, 071181, 071191, 071199, 071201, 071203, 071205, 071216, 071225, 071229, 071234, 071236, 071258, 071262, 071269, 071300, 071314, 071488, 071508, 071513, 071526, 071527, 071536, 071538, 071551, 071584, 071589, 071595, 071608, 071614, 071618, 071628, 071639, 071645, 071651, 071668, 071670, 071677, 071689, 071691, 071693, 071695, 071706, 071710, 071721, 071729, 071743, 071761, 071777, 071791, 071794, 071795, 071818, 071825, 071836, 071963, 071967, 071978, 071979, 071990, 071991, 071999, 072017, 072023, 072035, 072057, 072060, 072063, 072068, 072083, 072102, 072397, 072454, 072503; UPN 00802526456961, Serial Numbers: 051625, 051843, 054959, 055034, 055550, 055680, 055681, 055976, 056616, 056639, 056915, 056919, 057173, 057281, 057950, 058471, 058818, 058850, 058990, 059407, 060062, 060458, 060570, 062013, 062405, 062433, 062443, 062549, 063614, 064043, 064139, 064353, 065088, 066639, 066674, 066687, 066815, 066835, 066852, 066869, 066880, 066885, 066898, 066953, 066999, 067022, 067466, 067494, 067615, 067642, 067769, 068196, 068306, 068327, 068479, 068704, 068821, 068830, 068857, 069067, 069085, 069086, 069129, 069216, 069272, 069301, 069332, 069571, 070267, 070282, 070286, 070636, 070658, 070674, 070687, 070692, 070699, 070702, 070703, 070708, 070767, 070786, 070801, 070807, 070821, 070842, 070870, 070877, 070908, 070910, 070918, 070922, 070924, 070944, 070952, 070953, 070959, 070970, 070972, 070976, 070984, 071057, 071062, 071067, 071077, 071092, 071121, 071128, 071133, 071136, 072496, 073187, 073190, 073193, 073210, 073233, 073234, 073248, 073254, 073257, 073258, 073290, 073293, 073304, 073355, 073359, 073374, 073383, 073385, 073387, 073395, 073401, 073422, 073435, 073446, 073448, 073453, 073473, 073476, 073481; UPN 00802526467653, Serial Numbers: 051130, 051136, 051571, 051610, 051628, 051639, 051644, 051756, 051772, 051890, 055100, 055118, 055177, 055218, 055244, 055301, 055334, 055370, 055482, 055506, 055511, 055540, 055839, 055922, 055939, 056418, 056478, 056527, 056853, 057025, 057097, 057448, 057674, 057747, 057810, 057842, 057930, 058079, 058096, 058197, 058234, 058336, 058521, 058827, 059036, 059231, 059349, 059536, 059881, 059992, 060135, 060488, 060815, 060888, 060929, 060935, 060981, 061999, 062210, 062296, 062302, 062494, 062495, 062557, 062584, 063766, 063806, 063899, 064226, 064253, 064744, 064761, 064823, 065004, 065023, 065447, 065514, 065596, 065687, 066057, 066068, 066648, 066909, 067049, 067373, 067406, 067462, 067687, 067745, 068266, 068346, 068503, 068521, 068574, 068604, 068610, 068713, 068744, 068757, 068808, 068823, 068976, 068990, 069034, 069182, 069184, 069281, 069284, 069292, 069550, 070340, 070397, 070565, 070606, 070727, 070734, 070792, 070826, 070831, 070868, 070872, 070879, 070916, 070927, 070962, 071150, 071316, 071318, 071454, 071459, 071469, 071470, 071472, 071481, 071591, 071758, 071774, 071798, 071823, 071965, 071985, 071995, 072501, 073191, 073204, 073252, 073302, 073348, 073372, 073421, 073424, 073426, 073449, 073451, 073483, 073484, 073488, 073494, 073495, 073506, 073515, 073523, 073530, 073531, 073542, 073544, 073551, 073562, 073564, 073566, 073570, 073577, 073582, 073596, 073602, 073603, 073614; UPN 00802526468582, Serial Numbers: 051049, 051125, 051160, 051443, 051651, 051761, 052121, 052349, 054975, 055212, 055352, 055512, 055588, 055678, 055724, 055769, 055837, 055984, 056275, 056314, 056388, 056509, 056525, 056563, 056912, 056934, 057014, 057070, 057111, 057187, 057418, 057459, 057491, 057546, 057661, 057774, 057792, 057960, 057987, 058131, 058678, 058788, 058859, 058924, 058941, 058969, 059043, 059138, 059182, 059222, 059239, 059465, 059520, 059843, 059931, 059934, 059965, 060082, 060249, 060470, 060487, 060498, 060936, 061010, 061070, 061969, 061998, 062038, 062233, 062279, 063475, 064344, 064376, 064712, 064780, 065055, 065096, 065154, 065268, 065371, 065452, 065962, 066511, 066830, 066941, 067387, 067477, 067501, 067520, 067583, 067625, 067630, 067697, 067726, 067742, 067753, 067757, 067762, 067766, 067783, 067807, 067809, 068272, 068519, 068525, 068553, 068629, 068737, 068800, 068902, 068957, 068964, 069004, 069058, 069190, 069291, 069295, 069457, 069831, 070380, 070619, 070630, 070763, 070955, 070991, 071045, 071528, 071545, 071575, 071621, 071626, 071684, 071762, 071767, 071775, 071982, 072042, 072069, 072098, 072113, 072114, 072157, 072291, 072330, 072358, 072359, 072365, 072412, 072419, 072428, 072465, 073213, 073282; UPN 00802526473449, Serial Numbers: 051361, 051608, 051749, 055029, 055114, 055194, 055489, 055642, 055847, 055903, 056028, 056067, 056142, 056181, 056241, 056362, 056397, 056789, 056882, 057844, 057926, 058373, 058395, 059020, 059241, 059290, 059301, 059616, 059935, 060198, 061859, 062097, 064167, 064220, 064825, 064843, 064848, 065032, 065107, 065472, 065565, 066638, 066799, 066847, 066943, 066947, 067014, 067451, 067578, 067651, 067828, 068213, 068258, 068303, 068315, 068316, 068484, 068532, 068663, 069053, 069175, 069179, 069181, 069183, 069193, 069317, 069329, 069544, 069665, 069787, 069884, 070305, 070764, 070979, 071180, 071257, 071267, 071296, 071491, 071507, 071519, 071636, 071641, 072033, 073321, 073556; UPN 00802526474859, Serial Numbers: 051572, 051601, 051671, 051844, 051958, 055051, 055076, 055259, 055451, 055536, 055563, 055656, 055682, 055698, 055722, 055883, 055910, 056259, 056734, 056746, 056791, 056929, 056946, 056962, 057314, 057379, 057445, 058088, 058760, 058934, 060021, 060065, 060188, 060353, 060469, 060626, 060630, 062066, 062185, 062248, 062463, 063486, 063573, 064092, 064127, 064231, 064361, 064374, 064463, 064680, 064728, 064857, 065616, 065713, 065766, 066013, 066758, 066936, 067437, 067664, 068580, 068723, 069242, 069258, 069336, 069439, 070249, 070312, 070566, 070627, 070806, 070917, 071095, 071167, 071311, 071313, 071319, 071563, 072455, 073389, 073444, 073572, 073608; UPN 00802526475788, Serial Numbers: 051424, 051509, 051804, 051815, 051928, 052015, 055153, 055159, 055232, 055315, 055401, 055475, 055998, 056016, 056326, 056343, 056361, 056375, 056475, 056488, 056606, 056613, 056854, 056955, 056982, 057034, 057051, 057076, 057178, 057185, 057275, 057324, 057545, 057652, 057772, 057804, 057851, 057865, 057870, 057943, 058142, 058325, 058352, 058523, 058555, 058732, 059015, 059294, 059378, 059387, 059488, 059551, 059595, 059633, 059634, 059641, 059785, 059838, 059884, 059954, 059955, 060035, 060125, 060137, 060248, 060305, 060346, 060373, 060408, 060600, 060695, 060726, 060777, 060862, 061827, 061875, 061910, 061950, 062000, 062219, 062293, 062475, 062480, 062540, 063462, 063629, 063636, 063758, 064278, 064512, 064543, 064561, 064563, 064633, 064903, 064915, 064931, 065132, 065151, 065174, 065202, 065227, 065236, 065378, 065436, 065552, 065740, 066074, 066630, 066634, 066920, 067026, 067050, 067435, 067781, 067820, 068338, 068550, 068671, 068743, 068949, 069006, 069030, 069062, 069176, 069244, 069319, 069547, 069608, 069625, 069643, 069741, 070275, 070367, 070416, 070676, 070808, 070869, 070884, 070892, 070990, 071005, 071276, 071460, 071498, 071534, 071625, 071660, 071701, 071726, 071968, 072000, 072105, 072132, 072134, 073194, 073209, 073312, 073313, 073316, 073397, 073408, 073431, 073454, 073459; UPN 00802526477676, Serial Numbers: 051616, 056600, 057919, 062305, 063553, 064509, 064522, 067524, 067758, 070864; UPN 00802526478239, Serial Numbers: 051214, 051888, 055183, 055291, 055624, 055807, 055892, 055926, 055980, 056201, 056495, 056504, 056626, 056684, 057023, 057127, 057182, 057213, 057354, 057424, 057593, 057715, 057751, 058012, 058207, 058235, 058255, 058277, 058507, 058829, 059050, 059056, 059297, 059298, 059399, 059690, 059856, 059871, 059909, 059996, 060221, 060610, 060772, 060825, 060893, 060976, 061062, 061900, 062166, 062323, 062324, 062451, 063709, 064140, 064165, 064327, 064644, 064751, 064772, 064891, 064900, 064928, 065015, 065185, 065323, 065372, 065650, 066016, 066028, 066061, 066850, 067629, 068232, 068286, 068561, 068571, 068682, 068763, 069348, 069436, 069531, 069593, 069612, 069617, 069621, 069628, 069634, 069638, 069679, 070300, 070797, 071046, 071075, 071086, 071455, 071680, 071802, 072046, 072072, 072131, 072156, 072299, 072390, 072464, 073333; UPN 00802526479663, Serial Numbers: 051024, 051642, 051716, 052009, 055047, 055096, 055138, 055364, 055428, 055606, 055716, 055785, 056012, 056064, 056633, 056719, 056911, 057130, 057489, 057528, 057706, 057762, 057777, 057831, 058064, 058251, 058264, 058306, 058315, 058598, 058759, 059058, 059259, 059382, 059430, 059485, 059517, 059626, 059761, 060060, 060102, 060308, 060318, 060338, 060363, 060369, 060395, 060576, 060622, 062333, 064686, 064787, 065134, 065148, 065235, 065660, 065748, 065977, 065995, 066777, 066958, 067515, 067547, 067568, 067623, 067684, 067718, 068345, 068942, 069008, 069028, 069031, 069298, 069438, 069556, 069647, 069839, 069850, 070784, 070833, 070934, 071266, 071480, 071578, 071617, 072008, 072081, 072094, 072500, 073219, 073318, 073413, 073597; UPN 00802526480638, Serial Numbers: 051101, 051334, 051401, 051461, 051591, 051622, 051723, 051762, 051811, 051822, 055064, 055141, 055144, 055234, 055303, 055339, 055483, 055490, 055518, 055582, 055614, 055628, 055705, 055979, 056042, 056047, 056079, 056285, 056396, 056443, 056750, 056767, 056880, 056932, 056973, 057045, 057086, 057198, 057506, 057554, 057628, 057761, 057830, 057935, 057959, 058153, 058287, 058361, 058384, 058386, 058405, 058543, 058635, 058683, 058686, 059021, 059053, 059152, 059173, 059292, 059328, 059370, 059374, 059499, 059526, 059614, 059733, 059786, 059806, 059821, 059916, 060010, 060070, 060133, 060181, 060192, 060247, 060539, 060552, 060644, 060729, 060822, 060883, 060932, 061043, 061830, 061867, 061955, 062055, 062113, 062188, 062243, 062303, 062412, 062420, 062524, 063603, 063656, 063696, 063729, 063745, 063794, 063821, 063959, 064169, 064222, 064241, 064277, 064297, 064328, 064412, 064525, 064665, 064953, 064964, 064978, 065161, 065167, 065168, 065218, 065281, 065285, 065293, 065317, 065352, 065510, 065518, 065629, 065966, 066101, 066629, 066650, 066858, 066888, 066955, 066972, 066977, 067063, 067425, 067469, 067476, 067525, 067658, 067700, 068295, 068319, 068524, 068587, 068605, 068677, 068844, 068956, 069011, 069148, 069188, 069197, 069271, 069293, 069309, 069313, 069341, 069395, 069467, 069468, 069530, 069598, 069790, 069844, 069853, 069865, 070289, 070303, 070317, 070618, 070704, 070771, 070794, 070974, 071026, 071044, 071123, 071126, 071158, 071194, 071281, 071506, 071632, 071642, 071813, 072050, 072074, 072399, 072457, 072484, 073255, 073272, 073375, 073390, 073436, 073442, 073558, 073565, 074370, 074372, 074380, 074395, 074415, 074467, 074537, 074538, 074548, 074550, 074554, 074559, 074560, 074562, 074565, 074570, 074573, 074583, 074587, 074588, 074590, 074597, 074605, 074606, 074613, 074614, 074615, 074617, 074624, 074628, 074629, 074646, 074650, 074662, 074664, 074668, 074670, 074678, 074697, 074700, 074702, 074730, 074818, 074825, 074839, 074840, 074853, 074866, 074872, 074874, 074877, 074879, 074888, 074896, 074902, 074909, 074911, 074942, 074947, 074959, 074963, 074965, 074968, 074974; UPN 00802526484599, Serial Numbers: 074986, 074988, 074991, 074992, 074995, 075001, 075004, 075006, 075007, 075014, 075016, 075017, 075028, 075031, 075034, 075042, 075045, 075046, 075050, 075057, 075060, 075062, 075064, 075069, 075070, 075084, 075093, 075097, 075102, 075106, 075112, 075113, 075121, 075133, 075141, 075142, 075149, 075152, 075153, 075162, 075165, 075166, 075172, 075177, 075192, 075196, 075199, 075200, 075203, 075204, 075205, 075218, 075221, 075222, 075223, 075227, 075230, 075242, 075245, 075246, 075247, 075252, 075253, 075255, 075256, 075257, 075270, 075272, 075281, 075285, 075288, 075289, 075290, 075292, 075295, 075300, 075310, 075313, 075316, 075320, 075323, 075325, 075327, 075329, 075331, 075333, 075342, 075345, 075348, 075352, 075357, 075361, 075370, 075371, 075388, 075390, 075391, 075396, 075408, 075411, 075414, 075415, 075416, 075419, 075425, 075429, 075432, 075445, 075457, 075458, 075462, 075469, 075471, 075477, 075478, 075479, 075482, 075483, 075514, 075516, 075522, 075534, 075548, 075554, 075559, 075567, 075574, 075581, 075584, 075596, 075601, 075609, 075613, 075617, 075626, 075627, 075633, 075637, 075661, 075671, 075680, 075688, 075689, 075693, 075696, 075702, 075710, 075717, 075719, 075720, 075723, 075737, 075751, 075763, 075768, 075778, 075781, 075785, 075801, 075806, 075812, 075846, 075858, 075990, 075993, 076010, 076021, 076060, 076061, 076100; UPN 00802526486906, Serial Numbers: 051057, 051105, 051247, 051394, 051441, 051603, 051872, 052185, 054955, 054976, 054989, 055014, 055195, 055222, 055290, 055294, 055314, 055322, 055340, 055379, 055410, 055418, 055508, 055604, 055612, 055688, 055740, 055803, 056001, 056043, 056097, 056189, 056203, 056226, 056265, 056429, 056505, 056517, 056564, 056587, 056637, 056714, 056735, 056803, 056881, 056992, 057003, 057015, 057028, 057083, 057125, 057171, 057244, 057387, 057423, 057788, 057846, 057880, 057899, 057901, 057971, 058066, 058165, 058219, 058253, 058260, 058261, 058288, 058374, 058456, 058656, 058832, 058895, 058911, 058914, 058972, 058977, 058980, 059093, 059168, 059281, 059296, 059304, 059372, 059450, 059561, 059615, 059625, 059652, 059666, 059707, 059772, 059790, 059859, 059869, 059889, 059972, 060044, 060055, 060131, 060182, 060334, 060383, 060459, 060489, 060532, 060534, 060590, 060591, 060602, 060667, 060691, 060721, 060787, 060835, 060840, 060859, 060937, 061862, 061987, 062014, 062083, 062104, 062139, 062171, 062197, 062211, 062415, 062532, 063483, 063496, 063523, 063527, 063608, 063742, 063775, 063778, 063829, 063853, 063886, 063919, 064012, 064058, 064086, 064164, 064178, 064179, 064228, 064427, 064445, 064523, 064621, 064641, 064691, 064792, 064835, 064877, 064975, 065091, 065304, 065384, 065421, 065431, 065476, 065618, 065671, 065678, 065680, 065726, 065938, 066003, 066734, 066800, 066831, 067024, 067039, 067388, 067412, 067426, 067444, 067486, 067513, 067528, 067636, 068195, 068304, 068370, 068627, 068733, 068895, 068898, 069012, 069020, 069032, 069071, 069147, 069156, 069255, 069264, 069388, 069506, 069539, 069581, 069657, 069788, 069819, 070330, 070391, 070401, 070549, 070562, 070591, 070681, 070827, 070878, 070980, 071008, 071043, 071058, 071087, 071263, 071305, 071447, 071501, 071543, 071553, 071585, 071611, 071734, 072012, 072374, 073200, 073221, 073232, 073278, 073288, 073339, 073469, 073471, 074414, 074546, 074584, 074673, 074709, 074842, 074907, 074924, 074928, 074937, 074940, 074973, 075015, 075073, 075077, 075091, 075135, 075168, 075174, 075233, 075294, 075315, 075317, 075344, 075395, 075401, 075424, 075433, 075436, 075446, 075454, 075461, 075484, 075501, 075504, 075523, 075625, 075715, 075724, 075730, 075734, 075755, 075767, 075770, 075793, 075817, 075825, 075836, 075844, 075847, 075850, 075852, 075854, 075857, 075864, 075865, 075869, 075873, 075885, 075892, 075895, 075900, 075905, 075911, 075912, 075916, 075923, 075989, 075994, 076007, 076008, 076014, 076022, 076023, 076025, 076030, 076035, 076039, 076049, 076059, 076068, 076070, 076078, 076088, 076091, 076101, 076103, 076370, 076408, 076416, 076450, 076452, 076456, 076462, 076464, 076471, 076481, 076483, 076484, 076488, 076496, 076497, 076501, 076502, 076503, 076511, 076517, 076520, 076523, 076533, 076541, 076545, 076555, 076560, 076566, 076569, 076576, 076581, 076597, 076602, 076615, 076629, 076630, 076651, 076654, 076659, 076660, 076664, 076678, 076682, 076689, 076695, 076696, 076697, 076708, 076709, 076714, 076718, 076722, 076727, 076729, 076737, 076738, 076741, 076745, 076752, 076759, 076760, 076761, 076775, 076777, 076780, 076782, 076790, 076794, 076797, 076801, 076806, 076810, 076813, 076814, 076817, 076833, 076834, 076837, 076840, 076843, 076846, 076847, 076850, 076851, 076852, 076856, 076858, 076875, 076878, 076880, 076882, 076883, 076885, 076897, 076899, 076900, 076916, 076922, 076935, 076936, 076976, 076989, 076991, 077004, 077006, 077009, 077014, 077017, 077021, 077026, 077028, 077029, 077030, 077031, 077032, 077035, 077038, 077045, 077048, 077049, 077053, 077054, 077062, 077087, 077090, 077098, 077100, 077117, 077119, 077132, 077133, 077144, 077146, 077149, 077162, 077172, 077181, 077200, 077243; UPN 00802526494406, Serial Numbers: 051557, 051568, 051681, 051797, 055561, 055599, 055787, 056039, 056057, 056368, 056515, 056701, 056724, 056835, 056883, 057054, 057169, 057217, 057227, 057286, 057300, 057525, 057537, 057660, 057738, 057757, 058191, 058245, 058400, 058402, 058504, 058545, 058575, 058770, 058873, 059153, 059330, 059360, 059389, 059509, 059834, 059861, 059975, 060145, 060276, 060389, 060425, 060481, 060526, 060634, 060684, 060754, 060871, 060904, 061804, 061850, 061928, 061963, 061974, 062237, 062257, 062308, 062418, 062565, 063484, 063644, 063697, 063760, 063854, 063946, 063957, 064098, 064364, 064390, 064417, 064452, 064515, 064580, 064681, 064840, 065089, 065149, 065312, 065375, 065913, 066620, 066694, 066792, 066844, 067003, 067414, 067496, 067544, 067649, 067657, 067689, 067816, 068542, 068647, 068653, 068690, 068734, 068786, 068820, 068860, 068866, 068915, 068919, 068927, 069099, 069151, 069170, 069198, 069320, 069480, 069519, 069557, 069752, 069806, 070332, 070622, 070648, 070652, 070746, 070765, 070802, 070816, 070982, 071069, 071109, 071140, 071147, 071223, 071273, 071494, 071627, 071656, 072026, 072066, 072164, 072349, 072385, 073261, 073358, 073396, 073498, 073598, 074539, 074578, 074586, 074594, 074627, 074669, 074875, 074930, 074955, 074989, 075056, 075240, 075305, 075332, 075407, 075447, 075485, 075512, 075518, 075532, 075638, 075704, 075764, 075782, 075868, 076410, 076472, 076549, 076605, 077007, 077180, 077236; UPN 00802526497605, Serial Numbers: 051100, 051139, 051579, 051726, 051742, 051776, 051902, 051957, 052037, 052439, 055102, 055201, 055221, 055435, 055491, 055528, 055549, 055584, 055767, 055801, 055830, 055884, 055921, 055933, 056000, 056066, 056129, 056166, 056202, 056305, 056323, 056491, 056682, 056849, 056874, 057013, 057030, 057148, 057193, 057262, 057383, 057516, 057586, 057746, 057766, 057806, 057965, 057992, 058326, 058410, 058444, 058458, 058522, 058721, 058971, 059044, 059197, 059254, 059279, 059308, 059319, 059393, 059414, 059442, 059539, 059549, 059635, 059682, 059868, 059878, 059948, 059977, 060009, 060051, 060103, 060217, 060375, 060396, 060561, 060587, 060669, 060728, 060884, 061015, 061036, 061956, 061983, 061996, 062034, 062052, 062064, 062085, 062098, 062247, 062297, 062337, 062408, 063618, 063655, 063704, 063754, 063773, 063776, 063927, 063971, 063990, 064047, 064193, 064205, 064357, 064393, 064451, 064473, 064527, 064625, 064740, 065034, 065040, 065054, 065101, 065131, 065145, 065222, 065271, 065415, 065417, 065661, 065721, 065959, 066044, 066577, 066673, 066840, 066854, 066915, 066950, 067370, 067599, 067686, 067803, 068285, 068584, 068595, 068597, 068609, 068642, 068655, 068815, 068952, 069055, 069113, 069123, 069167, 069254, 069450, 069459, 069517, 069671, 069779, 069784, 069883, 070278, 070371, 070815, 070832, 070862, 070942, 071187, 071272, 071307, 071453, 071579, 071606, 071634, 071700, 071977, 071983, 072006, 072010, 072141, 072373, 072421, 073283, 073334, 073344, 073391, 074545, 074847, 074860, 074890, 074957, 074975, 075041, 075154, 075217, 075244, 075335, 075351, 075423, 075444, 075460, 075475, 075529, 075541, 075580, 075789, 075878, 075883, 075884, 075898, 075906, 075907, 075908, 075909, 075917, 075920, 075987, 075988, 075991, 076005, 076016, 076018, 076034, 076037, 076041, 076042, 076047, 076066, 076097, 076098, 076561, 076586, 076665, 076667, 076703, 076854, 076879, 077042, 077167; UPN 00802526516009, Serial Numbers: 051035, 051098, 051264, 051438, 051780, 051789, 051798, 051848, 051904, 051992, 055045, 055251, 055368, 055393, 055517, 055520, 055596, 055717, 055762, 055916, 055981, 056060, 056200, 056376, 056458, 056480, 056657, 056823, 056840, 056843, 056986, 057016, 057021, 057043, 057134, 057199, 057320, 057335, 057352, 057707, 057808, 057918, 057966, 058040, 058217, 058279, 058320, 058321, 058491, 058627, 058765, 058823, 058901, 058915, 059104, 059136, 059149, 059312, 059424, 059570, 059584, 059590, 059674, 059704, 059727, 059742, 059770, 059842, 059851, 059875, 059905, 059914, 059962, 060000, 060046, 060161, 060275, 060350, 060503, 060520, 060606, 060706, 060736, 060867, 060870, 060962, 061052, 061941, 061982, 062076, 062116, 062193, 062241, 062289, 062306, 062353, 062373, 062541, 062566, 062580, 063458, 063600, 063934, 063989, 063997, 064000, 064048, 064212, 064243, 064682, 064839, 064865, 064924, 065026, 065113, 065229, 065316, 065556, 065724, 066027, 066046, 066060, 066574, 066656, 066738, 066855, 067369, 067375, 067458, 067549, 067594, 067608, 067612, 067685, 067727, 068326, 068536, 068817, 068886, 068941, 069037, 069078, 069090, 069398, 069404, 069633, 070276, 070284, 070304, 070628, 070686, 070777, 070812, 071020, 071189, 071239, 071479, 071620, 071657, 071688, 071819, 071994, 072426, 073244, 073400, 073465, 073482, 073485, 075010, 075053, 075212, 075311, 075364, 075600, 075643, 075731, 075792, 075833, 076076, 076434, 076535, 076791, 077001, 077151; UPN 00802526523212, Serial Numbers: 051148, 051430, 051643, 055056, 055121, 055135, 055167, 055171, 055458, 055504, 055965, 056044, 056217, 056571, 056810, 056855, 056937, 056987, 056999, 057036, 057052, 057257, 057305, 057367, 057470, 057612, 057626, 058222, 058293, 058318, 058322, 058600, 058723, 058930, 058964, 058976, 059139, 059175, 059219, 059398, 059541, 059874, 059920, 059953, 059979, 060083, 060090, 060129, 060580, 060782, 060785, 060799, 060879, 060943, 061882, 062039, 062434, 062462, 063556, 063565, 063736, 063817, 063981, 064010, 064360, 064554, 064590, 064789, 064902, 065049, 065065, 065071, 065108, 065225, 065377, 065423, 065434, 065465, 065554, 065609, 065657, 065702, 065709, 065979, 066004, 066598, 066906, 066911, 066926, 066937, 067045, 067678, 067699, 067750, 068575, 068603, 068711, 068724, 068795, 068896, 068996, 069121, 069146, 069154, 069359, 069423, 070378, 070656, 070669, 071212, 071468, 071727, 071969, 072351, 073188, 073201, 073202, 073343, 073376, 073493, 073501, 074651, 074966, 075134, 075266, 075439, 075703, 075784, 075849, 076020, 076485, 076528, 076787, 076835, 077071, 079041, 079155, 079173, 079178, 079191; UPN 00802526535406, Serial Numbers: 051067, 051144, 051337, 051402, 051500, 051689, 051710, 051719, 051929, 052126, 055026, 055053, 055058, 055085, 055108, 055136, 055165, 055203, 055377, 055399, 055457, 055600, 055734, 055948, 055989, 055995, 056235, 056255, 056333, 056335, 056420, 056496, 056531, 056532, 056685, 056723, 056865, 056884, 056885, 057080, 057084, 057151, 057259, 057270, 057304, 057312, 057316, 057441, 057450, 057624, 057672, 057678, 057702, 058046, 058052, 058104, 058109, 058144, 058236, 058273, 058434, 058485, 058599, 058646, 058661, 058722, 058738, 058742, 058778, 058779, 058824, 058887, 058961, 058975, 058988, 059029, 059037, 059102, 059240, 059242, 059247, 059411, 059427, 059434, 059587, 059589, 059710, 059713, 059714, 059740, 059769, 059825, 059890, 060117, 060298, 060301, 060490, 060499, 060611, 060613, 060685, 060707, 060738, 060790, 060909, 060931, 061045, 061825, 061929, 061968, 062082, 062136, 062147, 062272, 062278, 062294, 062356, 062394, 062446, 062469, 062486, 063469, 063473, 063487, 063491, 063570, 063585, 063653, 063747, 063857, 063880, 064024, 064111, 064155, 064227, 064233, 064284, 064307, 064416, 064442, 064482, 064601, 064708, 064984, 064988, 065142, 065163, 065264, 065269, 065404, 065450, 065491, 065529, 065536, 065992, 066010, 066038, 066056, 066531, 066590, 066697, 066712, 066913, 067044, 067470, 067518, 067548, 067616, 067717, 067744, 067764, 067804, 068206, 068211, 068221, 068291, 068307, 068501, 068702, 068797, 068833, 068849, 068851, 068867, 068870, 068874, 068887, 068897, 068968, 069017, 069107, 069253, 069288, 069306, 069405, 069483, 069540, 069867, 070266, 070283, 070334, 070341, 070360, 070375, 070390, 070548, 070576, 070664, 070730, 070739, 070799, 070836, 070880, 071014, 071139, 071153, 071287, 071544, 071577, 071580, 071690, 071828, 071992, 071997, 072003, 072018, 072329, 072384, 072388, 072481, 073222, 073230, 073242, 073363, 073393, 073409, 073455, 073537, 073586, 073600, 073611, 074637, 074794, 074949, 075061, 075080, 075108, 075248, 075291, 075363, 075381, 075402, 075499, 075598, 075629, 075713, 075726, 075797, 075851, 075875, 075899, 075919, 076057, 076062, 076090, 076099, 076552, 076670, 076672, 076675, 076711, 076728, 076736, 076828, 076893, 077018, 077055, 077140, 077240; UPN 00802526554209, Serial Numbers: 051073, 051241, 051490, 051536, 051604, 051614, 051633, 051816, 051840, 051905, 051935, 051951, 055225, 055245, 055308, 055391, 055533, 055587, 055691, 055736, 055751, 055821, 055844, 055849, 055854, 055870, 055936, 056106, 056113, 056120, 056152, 056174, 056371, 056435, 056476, 056523, 056540, 056615, 056640, 056642, 056730, 056852, 056875, 056914, 056922, 057116, 057211, 057315, 057348, 057408, 057436, 057539, 057588, 057602, 057651, 057684, 057723, 057801, 057869, 057897, 057923, 057939, 058002, 058252, 058299, 058337, 058477, 058502, 058518, 058587, 058611, 058645, 058695, 058841, 058894, 058906, 058927, 058928, 058998, 059023, 059055, 059100, 059262, 059277, 059466, 059487, 059630, 059678, 059784, 059810, 059873, 059888, 060002, 060004, 060067, 060172, 060320, 060328, 060368, 060382, 060384, 060387, 060447, 060477, 060601, 060624, 060768, 061004, 061008, 061809, 061920, 061973, 062123, 062165, 062195, 062261, 062321, 062332, 062382, 062419, 062422, 062481, 062537, 062573, 063454, 063482, 063505, 063513, 063599, 063606, 063611, 063613, 063637, 063685, 063718, 063785, 063859, 063904, 063908, 063976, 063987, 064003, 064049, 064063, 064078, 064121, 064126, 064143, 064245, 064249, 064384, 064436, 064481, 064485, 064491, 064628, 064651, 064775, 064980, 064991, 065126, 065179, 065252, 065254, 065259, 065286, 065292, 065295, 065555, 065679, 065744, 065760, 065942, 066011, 066017, 066031, 066059, 066086, 066643, 066791, 066925, 067431, 067523, 067634, 067675, 067702, 067741, 068205, 068263, 068276, 068292, 068293, 068486, 068488, 068599, 068606, 068705, 068772, 068889, 068901, 069063, 069251, 069308, 069442, 069466, 069559, 069615, 069801, 070258, 070417, 070641, 070644, 070725, 070773, 070858, 070873, 070903, 070947, 070966, 071011, 071116, 071165, 071172, 071177, 071185, 071213, 071243, 071290, 071520, 071566, 071803, 071998, 072078, 072087, 072101, 072461, 073246, 073263, 073281, 073287, 073291, 073299, 073347, 073350, 073368, 073370, 073384, 073386, 073402, 073522, 073560, 073590, 073593, 073609, 074696, 074817, 074841, 074889, 074983, 074990, 075103, 075105, 075186, 075261, 075263, 075358, 075383, 075441, 075452, 075508, 075533, 075605, 075800, 075838, 075921, 076001, 076013, 076052, 076087, 076455, 077015, 077034, 077130, 079758, 079764, 079767, 079768, 079769, 079778, 079791, 079794, 079796, 079797, 079798, 079807, 079812, 079823, 079829, 079836, 079838, 079844, 079866, 079888, 079916, 079920, 079933, 079940, 079947, 079953, 079970, 079995, 080000, 080002, 080007, 080011, 080013, 080015, 080021, 080023, 080026, 080032, 080041, 080045, 080047, 080063, 080071, 080080, 080082, 080085, 080091, 080094, 080108, 080131, 080139, 080141; UPN 00802526562600, Serial Numbers: 051041, 051145, 051442, 051570, 051736, 052000, 055054, 055347, 055351, 055566, 055603, 055737, 055834, 055880, 055974, 055991, 056003, 056056, 056084, 056282, 056310, 056355, 056416, 056469, 056539, 056543, 056544, 056547, 056566, 056820, 056821, 056918, 057106, 057375, 057427, 057434, 057607, 057613, 057796, 057951, 058387, 058433, 058448, 058497, 058711, 058735, 058949, 059079, 059085, 059134, 059275, 059315, 059497, 059550, 059617, 059697, 059739, 059908, 060076, 060167, 060269, 060462, 060558, 060683, 060885, 061003, 061842, 061856, 062141, 062170, 062189, 062275, 062322, 062330, 062389, 062441, 063720, 063804, 063938, 064005, 064017, 064097, 064150, 064160, 064256, 064438, 064565, 064703, 064795, 064863, 065183, 065319, 065370, 065413, 065505, 065603, 065681, 065741, 065763, 065765, 065905, 065945, 065971, 066664, 066741, 066783, 066872, 066900, 067008, 067021, 067378, 067639, 068252, 068278, 068348, 068537, 068651, 068674, 068784, 068818, 068879, 068888, 068986, 069112, 069120, 069163, 069346, 069469, 069487, 069776, 069777, 069823, 070581, 070671, 070783, 070788, 070853, 070888, 070948, 071088, 071143, 071160, 071237, 071288, 071312, 071473, 071664, 071831, 071972, 072356, 072376, 073231, 073250, 073308, 073378, 073463, 074580, 074626, 074954, 075297, 075410, 075527, 075744, 075827, 076048, 076054, 076102, 076445, 076461, 076536, 076593, 076877, 076925, 077103, 077111, 077182, 077217, 079811, 079937, 080301, 080302, 080314, 080319, 080322, 080325, 080327, 080331, 080335, 080336, 080339, 080340, 080343, 080349, 080357, 080358, 080360, 080364, 080366, 080372, 080377, 080381, 080387, 080391, 080397, 080418, 080422, 080423, 080425, 080429, 080439, 080442, 080443, 080445, 080447, 080455, 080456, 080457, 080461, 080462, 080476, 080478, 080486, 080489, 080494, 080495, 080502, 080503, 080504, 080505, 080507, 080508, 080512, 080515, 080519, 080523, 080526, 080527, 080529, 080530, 080531, 080533, 080535, 080536, 080538, 080539; UPN 00802526564703, Serial Numbers: 051018, 051050, 051104, 051503, 055258, 055420, 055619, 056162, 056271, 056561, 056608, 056732, 056777, 057067, 057071, 057276, 057453, 057821, 057892, 057902, 057924, 057961, 058081, 058415, 059264, 059732, 059800, 060148, 060451, 060609, 060824, 060920, 060968, 060970, 061021, 061026, 061060, 061846, 061897, 062240, 063660, 063761, 063814, 063956, 063964, 063978, 064029, 064290, 064415, 064448, 064461, 064632, 064870, 065571, 065664, 065689, 065939, 065943, 066036, 066582, 066894, 067403, 067410, 067457, 068311, 068322, 068834, 068955, 069027, 069150, 069228, 069340, 069378, 069443, 069478, 069479, 069526, 069578, 069600, 069818, 070252, 070343, 070683, 070736, 070785, 070967, 071040, 071056, 071071, 071252, 071265, 071592, 071594, 071776, 071806, 072037, 072038, 072151, 072362, 072416, 072505, 073298, 073474, 074366, 074958, 075354, 075448, 075557, 075645, 079806, 079808, 080067, 080310, 080446, 080475, 080479, 080482, 080492, 080493, 080498, 080500, 080521, 080524, 080525, 080543, 080549, 080558, 080559, 080565, 080567, 080569, 080576, 080580, 080581, 080585, 080586, 080589, 080595, 080598, 080602, 080603, 080605, 080607, 080612, 080613, 080616, 080619, 080624, 080633, 080634, 080641, 080648, 080700, 080715; UPN 00802526575303, Serial Numbers: 051359, 052422, 055819, 056879, 058470, 058658, 059394, 060006, 060206, 061949, 062223, 062236, 062287, 063659, 064141, 064998, 065401, 065676, 066601, 067669, 068209, 068290, 068774, 068871, 069075, 069105, 069297, 069661, 070989, 073273, 074746; UPN 00802526579707, Serial Numbers: 051262, 051305, 051421, 051439, 051709, 052177, 054988, 055025, 055030, 055380, 055445, 055544, 055552, 055894, 056107, 056231, 056296, 056299, 056306, 056331, 056363, 056489, 056502, 056592, 056662, 056677, 056727, 056866, 056870, 056897, 056963, 057011, 057018, 057235, 057264, 057328, 057355, 057360, 057475, 057560, 057564, 057663, 057782, 057828, 058019, 058077, 058301, 058305, 058372, 058553, 058593, 058699, 058737, 058800, 058821, 058834, 058903, 059048, 059063, 059114, 059181, 059248, 059261, 059406, 059455, 059489, 059552, 059574, 059687, 059702, 059776, 059822, 059946, 060043, 060089, 060138, 060176, 060228, 060335, 060355, 060616, 060686, 060697, 060872, 060894, 061011, 061057, 061835, 062092, 062258, 062454, 062567, 062603, 063520, 063564, 063725, 063820, 064188, 064192, 064204, 064317, 064394, 064423, 064457, 064504, 064521, 064556, 064706, 064770, 064812, 064912, 064913, 064936, 064939, 064954, 065119, 065177, 065230, 065294, 065302, 065327, 065339, 065342, 065382, 065458, 065484, 065530, 065538, 065605, 065654, 065662, 065754, 065912, 065937, 065972, 066079, 066517, 066610, 066621, 066641, 066672, 066684, 066769, 066904, 067484, 067495, 067572, 067576, 067645, 067690, 067714, 067817, 068373, 068465, 068555, 068640, 068659, 068696, 068707, 068735, 068882, 068928, 068935, 068993, 069076, 069082, 069122, 069246, 069430, 069489, 069565, 069601, 069616, 069769, 069781, 069816, 069830, 069841, 069859, 070344, 070363, 070561, 070590, 070642, 070680, 070701, 070717, 070809, 070866, 070935, 070937, 071017, 071032, 071052, 071073, 071104, 071120, 071253, 071286, 071309, 071524, 071540, 071658, 071659, 071673, 071728, 071733, 071787, 072029, 072054, 072073, 072085, 072091, 072130, 072154, 072437, 072483, 073207, 073352, 073595, 074556, 074690, 074699, 074716, 074749, 074846, 074977, 074984, 075025, 075032, 075036, 075039, 075146, 075147, 075187, 075188, 075225, 075284, 075303, 075355, 075406, 075417, 075474, 075506, 075511, 075546, 075547, 075558, 075560, 075562, 075588, 075634, 075646, 075669, 075706, 075728, 075743, 075759, 075786, 075819, 075826, 075862, 075879, 075889, 075894, 075914, 076546, 076563, 076601, 076657, 076694, 076748, 076857, 079177, 079832, 079841, 079862, 079923, 080311, 080375, 080469, 080485, 080556, 080572, 080593; UPN 00802526582004, Serial Numbers: 051020, 051039, 051054, 051084, 051087, 051099, 051123, 051162, 051185, 051187, 051237, 051283, 051286, 051406, 051434, 051437, 051467, 051475, 051485, 051508, 051527, 051537, 051538, 051540, 051550, 051567, 051573, 051582, 051602, 051649, 051693, 051701, 051750, 051778, 051783, 051809, 051810, 051813, 051829, 051830, 051868, 051870, 051879, 051892, 051937, 051938, 051943, 051955, 051987, 052014, 052045, 052057, 052150, 052162, 052362, 052458, 054953, 054980, 054987, 054998, 055019, 055038, 055040, 055044, 055063, 055072, 055073, 055091, 055107, 055131, 055134, 055146, 055155, 055162, 055181, 055190, 055199, 055224, 055241, 055266, 055273, 055289, 055292, 055304, 055337, 055342, 055400, 055404, 055406, 055414, 055423, 055455, 055471, 055499, 055521, 055526, 055545, 055559, 055562, 055579, 055627, 055638, 055669, 055699, 055746, 055784, 055794, 055797, 055816, 055827, 055859, 055902, 055908, 055931, 055945, 055961, 055963, 056006, 056008, 056011, 056013, 056014, 056037, 056102, 056128, 056133, 056137, 056163, 056171, 056176, 056197, 056220, 056222, 056236, 056243, 056254, 056260, 056268, 056273, 056301, 056304, 056324, 056345, 056364, 056379, 056380, 056400, 056410, 056447, 056455, 056467, 056470, 056546, 056551, 056555, 056559, 056569, 056588, 056599, 056612, 056636, 056653, 056687, 056715, 056747, 056757, 056816, 056832, 056834, 056839, 056860, 056862, 056933, 056938, 056941, 056942, 056950, 056952, 056979, 056998, 057046, 057068, 057077, 057093, 057109, 057128, 057202, 057247, 057256, 057283, 057306, 057336, 057339, 057343, 057363, 057393, 057407, 057411, 057416, 057432, 057461, 057592, 057595, 057622, 057632, 057647, 057669, 057690, 057701, 057713, 057721, 057735, 057744, 057745, 057763, 057764, 057775, 057826, 057848, 057850, 057855, 057860, 057872, 057886, 057920, 057929, 057934, 057955, 058005, 058007, 058008, 058009, 058015, 058026, 058030, 058042, 058063, 058072, 058082, 058154, 058161, 058166, 058172, 058187, 058188, 058193, 058196, 058198, 058204, 058238, 058249, 058265, 058276, 058291, 058303, 058308, 058311, 058347, 058363, 058376, 058380, 058407, 058416, 058423, 058429, 058459, 058475, 058479, 058481, 058505, 058510, 058516, 058530, 058550, 058565, 058567, 058609, 058629, 058637, 058654, 058667, 058670, 058672, 058689, 058692, 058728, 058747, 058772, 058830, 058845, 058858, 058876, 058879, 058885, 058890, 058916, 058926, 058957, 058989, 059059, 059060, 059066, 059077, 059081, 059084, 059117, 059124, 059125, 059131, 059147, 059165, 059189, 059191, 059196, 059202, 059232, 059253, 059287, 059300, 059306, 059311, 059325, 059341, 059342, 059350, 059361, 059380, 059386, 059395, 059410, 059459, 059493, 059505, 059532, 059542, 059553, 059562, 059563, 059576, 059581, 059585, 059597, 059618, 059637, 059640, 059647, 059671, 059685, 059725, 059728, 059738, 059745, 059746, 059755, 059758, 059798, 059813, 059816, 059818, 059823, 059839, 059848, 059858, 059866, 059870, 059882, 059903, 059904, 059921, 059922, 059926, 059940, 059960, 059968, 059986, 059998, 059999, 060015, 060042, 060095, 060104, 060108, 060126, 060155, 060164, 060173, 060189, 060200, 060232, 060235, 060274, 060302, 060306, 060315, 060316, 060345, 060349, 060359, 060388, 060401, 060414, 060510, 060522, 060528, 060531, 060571, 060575, 060594, 060608, 060662, 060678, 060698, 060747, 060769, 060796, 060798, 060803, 060804, 060810, 060813, 060823, 060826, 060831, 060833, 060838, 060854, 060875, 060877, 060946, 060947, 060984, 060987, 061019, 061037, 061044, 061051, 061812, 061837, 061840, 061841, 061857, 061863, 061865, 061883, 061886, 061889, 061916, 061917, 061930, 061938, 061940, 061946, 061952, 061962, 061970, 061990, 061991, 062009, 062016, 062019, 062054, 062062, 062080, 062091, 062110, 062112, 062118, 062134, 062154, 062169, 062222, 062228, 062251, 062282, 062309, 062327, 062349, 062365, 062368, 062371, 062383, 062386, 062398, 062402, 062409, 062421, 062423, 062460, 062477, 062482, 062487, 062498, 062502, 062505, 062515, 062526, 062534, 062536, 062538, 062544, 062545, 062546, 062574, 062578, 062593, 063459, 063460, 063488, 063492, 063494, 063503, 063504, 063510, 063511, 063518, 063626, 063646, 063654, 063674, 063675, 063677, 063692, 063693, 063726, 063731, 063757, 063771, 063792, 063799, 063813, 063833, 063850, 063856, 063872, 063887, 063897, 063912, 063925, 063941, 063960, 063969, 063993, 064001, 064013, 064025, 064035, 064066, 064101, 064115, 064117, 064131, 064133, 064159, 064163, 064183, 064217, 064223, 064239, 064242, 064282, 064314, 064319, 064372, 064377, 064433, 064453, 064464, 064475, 064478, 064502, 064535, 064538, 064539, 064572, 064594, 064599, 064610, 064611, 064613, 064619, 064631, 064649, 064671, 064673, 064685, 064695, 064696, 064697, 064756, 064759, 064828, 064852, 064925, 064937, 064941, 064987, 064989, 065001, 065007, 065009, 065020, 065021, 065024, 065036, 065043, 065074, 065079, 065092, 065105, 065117, 065121, 065129, 065138, 065139, 065155, 065158, 065212, 065213, 065228, 065233, 065234, 065242, 065255, 065274, 065297, 065311, 065329, 065361, 065366, 065445, 065448, 065456, 065457, 065466, 065477, 065526, 065527, 065542, 065564, 065575, 065577, 065581, 065604, 065619, 065626, 065628, 065635, 065647, 065652, 065658, 065677, 065686, 065701, 065703, 065707, 065739, 065773, 065783, 065909, 065914, 065983, 065989, 065996, 066001, 066022, 066025, 066040, 066064, 066070, 066071, 066072, 066102, 066569, 066606, 066615, 066637, 066645, 066655, 066751, 066767, 066821, 066832, 066841, 066862, 066864, 066874, 066903, 066948, 066951, 066971, 066988, 067009, 067016, 067041, 067392, 067394, 067411, 067419, 067433, 067472, 067481, 067500, 067502, 067517, 067553, 067560, 067581, 067619, 067635, 067671, 067672, 067701, 067706, 067710, 067715, 067721, 067731, 067794, 068191, 068212, 068239, 068245, 068308, 068309, 068312, 068314, 068317, 068325, 068337, 068347, 068485, 068502, 068508, 068544, 068545, 068568, 068578, 068579, 068581, 068582, 068611, 068617, 068625, 068630, 068684, 068691, 068700, 068706, 068754, 068778, 068793, 068798, 068802, 068810, 068862, 068917, 068937, 068940, 068960, 068977, 068984, 068988, 068995, 069029, 069066, 069097, 069106, 069111, 069119, 069171, 069201, 069222, 069237, 069241, 069267, 069302, 069307, 069327, 069347, 069356, 069390, 069397, 069418, 069421, 069424, 069490, 069493, 069495, 069496, 069516, 069534, 069548, 069580, 069587, 069611, 069632, 069645, 069646, 069667, 069669, 069670, 069673, 069677, 069746, 069750, 069760, 069797, 069837, 069849, 069854, 069870, 070257, 070293, 070309, 070326, 070338, 070342, 070345, 070350, 070351, 070357, 070388, 070394, 070395, 070420, 070582, 070589, 070605, 070615, 070617, 070632, 070637, 070666, 070693, 070707, 070740, 070745, 070747, 070753, 070758, 070762, 070766, 070834, 070839, 070844, 070852, 070874, 070881, 070886, 070896, 070897, 070928, 070939, 070951, 070960, 070964, 070971, 070985, 070987, 070993, 071000, 071013, 071021, 071025, 071051, 071054, 071066, 071072, 071082, 071084, 071090, 071112, 071117, 071138, 071145, 071162, 071174, 071217, 071221, 071247, 071256, 071270, 071285, 071298, 071299, 071310, 071458, 071493, 071500, 071503, 071505, 071510, 071516, 071523, 071547, 071552, 071570, 071574, 071593, 071596, 071597, 071603, 071604, 071612, 071615, 071637, 071638, 071652, 071681, 071698, 071707, 071708, 071711, 071718, 071739, 071769, 071779, 071780, 071804, 071829, 071971, 071986, 071987, 071988, 072001, 072007, 072011, 072025, 072044, 072055, 072067, 072089, 072126, 072135, 072142, 072167, 072174, 072295, 072322, 072337, 072383, 072423, 072443, 072467, 072472, 073189, 073199, 073203, 073212, 073236, 073245, 073247, 073269, 073277, 073294, 073297, 073300, 073310, 073317, 073329, 073336, 073345, 073392, 073412, 073420, 073425, 073433, 073460, 073468, 073497, 073508, 073510, 073524, 073550, 073559, 073567, 073575, 073592, 073612, 073613, 074258, 074323, 074351, 074362, 074369, 074551, 074553, 074585, 074611, 074622, 074693, 074704, 074753, 074769, 074812, 074852, 074855, 074887, 074892, 074898, 074945, 074972, 074993, 074999, 075018, 075020, 075023, 075054, 075058, 075059, 075082, 075087, 075089, 075107, 075116, 075120, 075123, 075139, 075150, 075170, 075178, 075195, 075197, 075202, 075207, 075208, 075213, 075258, 075259, 075265, 075275, 075278, 075280, 075282, 075287, 075301, 075321, 075328, 075337, 075339, 075368, 075375, 075378, 075384, 075389, 075397, 075404, 075421, 075427, 075434, 075450, 075468, 075473, 075500, 075502, 075510, 075521, 075524, 075525, 075543, 075549, 075556, 075566, 075583, 075587, 075592, 075616, 075619, 075664, 075665, 075672, 075673, 075675, 075682, 075687, 075700, 075708, 075712, 075714, 075765, 075776, 075798, 075802, 075803, 075809, 075840, 075856, 075877, 075901, 075902, 075915, 076000, 076015, 076038, 076040, 076074, 076104, 076105, 076378, 076438, 076454, 076500, 076513, 076532, 076558, 076578, 076633, 076663, 076683, 076698, 076699, 076713, 076743, 076750, 076768, 076778, 076789, 076812, 076845, 076860, 076861, 076943, 077010, 077011, 077022, 077027, 077082, 077088, 077089, 077096, 077114, 077116, 077118, 077131, 077137, 077143, 077155, 077161, 077166, 079031, 079168, 079185, 079786, 079788, 079813, 079816, 079817, 079830, 079909, 079919, 079964, 079983, 080005, 080008, 080020, 080049, 080081, 080089, 080100, 080135, 080308, 080346, 080368, 080433, 080448, 080459, 080464, 080470, 080481, 080488, 080501, 080516, 080520, 080563, 080566, 080574, 080588, 080615, 080631, 080645, 080651, 080667, 080687, 080697, 080698, 080702; UPN 00802526602900, Serial Numbers: 056713; UPN 00802526604584, Serial Numbers: 051055, 051525, 051566, 051597, 051669, 051921, 051979, 051982, 052333, 055031, 055095, 055369, 055690, 055777, 056116, 056357, 056399, 056421, 056433, 056474, 056579, 056648, 056706, 056811, 056858, 056947, 057104, 057155, 057280, 057444, 057499, 057797, 057976, 058108, 058159, 058181, 058375, 058487, 058612, 058640, 058677, 059001, 059092, 059161, 059169, 059198, 059218, 059257, 059365, 059467, 059469, 059657, 059699, 059827, 059847, 059925, 059951, 060201, 060272, 060410, 060421, 060486, 060535, 060547, 060621, 060640, 060650, 060829, 060995, 061038, 061864, 061877, 062028, 062045, 062094, 062150, 062163, 062393, 062410, 063509, 063559, 063652, 063702, 063777, 063784, 063858, 063865, 063939, 064123, 064385, 064392, 064434, 064497, 064737, 064758, 064972, 065153, 065321, 065381, 065414, 065425, 065519, 065544, 065642, 065722, 065788, 066956, 066968, 067060, 067543, 067590, 067604, 068242, 068250, 068749, 068920, 068980, 068994, 069041, 069045, 069205, 069211, 069213, 069229, 069407, 069525, 069863, 070313, 070608, 070616, 070678, 070682, 070720, 071113, 071176, 071208, 071499, 071533, 071822, 072106, 072408, 073224, 073394, 073518, 073534, 073535, 074277, 074843, 074946, 075122, 075143, 075193, 075387, 075493, 075495, 075540, 075589, 075611, 076011, 076012, 076585, 076839, 076999, 077102, 077124, 079127; UPN 00802526606366, Serial Numbers: 051015, 051086, 051103, 051124, 051161, 051304, 051332, 051341, 051349, 051450, 051451, 051524, 051574, 051586, 051596, 051684, 051690, 051734, 051740, 051767, 051807, 051852, 051882, 051922, 052233, 054982, 054983, 055004, 055010, 055017, 055024, 055088, 055110, 055287, 055288, 055310, 055346, 055353, 055398, 055437, 055466, 055513, 055529, 055660, 055712, 055728, 055759, 055775, 055815, 055826, 055876, 055958, 056009, 056053, 056076, 056086, 056139, 056160, 056170, 056173, 056182, 056209, 056212, 056256, 056258, 056283, 056309, 056321, 056487, 056507, 056528, 056550, 056580, 056679, 056686, 056705, 056728, 056731, 056783, 056828, 056856, 056863, 056930, 056959, 057001, 057026, 057044, 057072, 057123, 057133, 057136, 057191, 057321, 057337, 057344, 057386, 057389, 057395, 057429, 057457, 057479, 057502, 057543, 057565, 057591, 057668, 057670, 057693, 057734, 057739, 057784, 057878, 057909, 057910, 057915, 057917, 058016, 058031, 058033, 058038, 058049, 058073, 058080, 058099, 058135, 058152, 058266, 058297, 058340, 058343, 058367, 058388, 058389, 058418, 058511, 058572, 058589, 058595, 058701, 058757, 058758, 058763, 058805, 058852, 058854, 058860, 058874, 058892, 058938, 058970, 059009, 059033, 059047, 059132, 059163, 059187, 059234, 059245, 059303, 059353, 059364, 059377, 059390, 059391, 059397, 059461, 059462, 059473, 059479, 059483, 059498, 059543, 059556, 059582, 059639, 059712, 059730, 059731, 059763, 059789, 059791, 059793, 059804, 059849, 059885, 059886, 059893, 059898, 059959, 059966, 059971, 060033, 060088, 060106, 060127, 060139, 060143, 060158, 060242, 060259, 060291, 060313, 060325, 060331, 060379, 060449, 060501, 060515, 060530, 060543, 060574, 060631, 060642, 060682, 060687, 060704, 060709, 060832, 060887, 060939, 060941, 060974, 061016, 061064, 061629, 061807, 061819, 061896, 061931, 061943, 061948, 061993, 062084, 062127, 062128, 062187, 062234, 062244, 062265, 062319, 062406, 062437, 062466, 062497, 062516, 062530, 062564, 063522, 063590, 063594, 063657, 063658, 063739, 063797, 063838, 063861, 063870, 063873, 063910, 063923, 063929, 063955, 064045, 064046, 064069, 064100, 064181, 064199, 064283, 064305, 064316, 064321, 064331, 064339, 064343, 064345, 064397, 064414, 064428, 064450, 064456, 064511, 064545, 064567, 064660, 064679, 064716, 064733, 064738, 064783, 064784, 064845, 064868, 064895, 064910, 064938, 065056, 065164, 065165, 065198, 065201, 065330, 065373, 065379, 065391, 065411, 065480, 065502, 065522, 065580, 065668, 065693, 065737, 065781, 065929, 065951, 066045, 066055, 066082, 066083, 066596, 066623, 066654, 066749, 066875, 066889, 066952, 067017, 067372, 067461, 067475, 067571, 067600, 067661, 067711, 067733, 067760, 067800, 067830, 067831, 068246, 068248, 068274, 068352, 068363, 068368, 068517, 068520, 068600, 068622, 068669, 068676, 068752, 068779, 068783, 068785, 068789, 068819, 068854, 069050, 069074, 069110, 069117, 069210, 069269, 069455, 069569, 069583, 069591, 069606, 069675, 069828, 069835, 069864, 070268, 070281, 070288, 070328, 070358, 070421, 070553, 070557, 070613, 070623, 070639, 070709, 070718, 070719, 070732, 070755, 070769, 070780, 070791, 070793, 070857, 070861, 070863, 070865, 070963, 070998, 071028, 071053, 071059, 071078, 071089, 071102, 071156, 071190, 071231, 071248, 071291, 071457, 071517, 071549, 071561, 071586, 071605, 071622, 071631, 071633, 071649, 071713, 071747, 071793, 071821, 071826, 071837, 072005, 072049, 072065, 072070, 072137, 072138, 072305, 072392, 072498, 073182, 073183, 073251, 073275, 073284, 073369, 073398, 073428, 073440, 073479, 073499, 073528, 073561, 073568, 073579, 073606, 074564, 074609, 074648, 074657, 074819, 074873, 074918, 074921, 074967, 074976, 074998, 075096, 075171, 075224, 075347, 075431, 075437, 075453, 075572, 075612, 075653, 075654, 075667, 075705, 075716, 075756, 075760, 075761, 075780, 075790, 075821, 075835, 075842, 075843, 075861, 075886, 075887, 076036, 076543, 076580, 076661, 076690, 076719, 076981, 077016, 079834, 079942, 080022, 080140, 080304, 080450, 080491, 080532, 080537, 080551, 080555, 080570, 080664, 080672, 080701; UPN 00802526606403, Serial Numbers: 056709; UPN 00802526611322, Serial Numbers: 051354, 051363, 051981, 052036, 055052, 055103, 055158, 055200, 055402, 055519, 055551, 055605, 055710, 055749, 056032, 056287, 056341, 056499, 056641, 056742, 057089, 057161, 057176, 057370, 057603, 057615, 057765, 057942, 058006, 058365, 058792, 058891, 058999, 059088, 059484, 059535, 059659, 059956, 059990, 060068, 060136, 060342, 060599, 060716, 060897, 060927, 060954, 060993, 061821, 062036, 062044, 062192, 062316, 062442, 062518, 062522, 063701, 063973, 063982, 064106, 064112, 064447, 064804, 064851, 064983, 065085, 065109, 065191, 065251, 065258, 065403, 065493, 065515, 065524, 065531, 066087, 066644, 066887, 066954, 066983, 067421, 067535, 067582, 067610, 067709, 067729, 067798, 068546, 068804, 068850, 068947, 068981, 069162, 069200, 069504, 069553, 069585, 069618, 069826, 070414, 070587, 070673, 070899, 071640, 071654, 072475, 073241, 074577, 075115, 075579, 076063, 076067, 076474, 076557, 077121, 080052, 080054, 080088, 080329, 080583, 080695. UPN: 00802526473449, Serial Numbers: 066847; UPN 00802526435430, Serial Numbers: 051122, 051197; UPN 00802526437458, Serial Numbers: 051088, 051138, 051250, 051267, 051338, 051365, 051369, 051398, 051930, 051948, 051976, 052089, 052098, 052138, 052205, 052217, 052220, 052247, 052249, 052289, 052331, 052499, 052513, 052557, 052570, 052617, 052624, 052641, 052663, 052674, 052676, 052708, 052712, 052716, 052721, 052722, 052734, 052757, 052760, 052769, 052802, 052804, 052813, 052818, 052881, 052889, 052891, 052894, 052932, 052937, 052944, 052948, 052950, 052958, 052959, 052960, 052965, 052980, 052999, 053010, 053013, 053048, 053133, 053134, 053140, 053147, 053200, 053206, 053271, 053294, 053299, 053318, 053396, 053401, 053431, 053433, 053484, 053538, 053600, 053609, 053614, 053615, 053617, 053632, 053637, 053641, 053643, 053644, 053645, 053657, 053659, 053662, 053663, 053665, 053676, 053677, 053678, 053680, 053688, 053694, 053701, 053706, 053709, 053722, 053724, 053727, 053729, 053730, 053747, 053767, 053769, 053772, 053773, 053778, 053781, 053790, 053798, 053800, 053802, 053812, 053814, 053816, 053823, 053829, 053830, 053835, 053837, 053857, 053877, 053878, 053879, 053892, 053895, 053897, 053906, 053909, 053921, 053922, 053926, 053951, 053958, 053974, 053977, 053980, 053992, 053995, 054002, 054015, 054016, 054019, 054034, 054059, 054064, 054082, 054088, 054095, 054097, 054106, 054115, 054132, 054147, 054162, 054173, 054191, 054196, 054197, 054202, 054218, 054219, 054227, 054230, 054233, 054234, 054242, 054247, 054249, 054267, 054276, 054280, 054281, 054284, 054290, 054298, 054303, 054305, 054308, 054311, 054318, 054326, 054350, 054352, 054358, 054359, 054363, 054364, 054366, 054371, 054372, 054374, 054381, 054385, 054386, 054388, 054389, 054391, 054397, 054398, 054399, 054401, 054402, 054404, 054408, 054409, 054415, 054418, 054422, 054425, 054426, 054430, 054431, 054437, 054442, 054444, 054453, 054455, 054457, 054459, 054480, 054484, 054485, 054489, 054492, 054498, 054505, 054506, 054513, 054515, 054516, 054521, 054537, 054540, 054569, 054570, 054573, 054575, 054582, 054590, 054592, 054595, 054596, 054602, 054608, 054623, 054631, 054641, 054645, 054648, 054653, 054656, 054665, 054680, 054685, 054704, 054708, 054712, 054717, 054718, 054725, 054730, 054731, 054732, 054735, 054741, 054742, 054750, 054756, 054763, 054772, 054776, 054779, 054783, 054784, 054786, 054788, 054791, 054814, 054815, 054817, 054827, 054828, 054845, 054861, 054864, 054869, 054875, 054878, 054879, 054885, 054896, 054909, 054912, 054922, 054926, 054940, 054943, 054944, 054946; UPN 00802526441974, Serial Numbers: 051912, 051926, 051963, 051965, 051994, 052011, 052012, 052031, 052033, 052035, 052047, 052050, 052052, 052102, 052110, 052113, 052116, 052124, 052132, 052135, 052139, 052152, 052155, 052159, 052168, 052169, 052180, 052181, 052183, 052189, 052196, 052197, 052200, 052204, 052206, 052210, 052216, 052222, 052223, 052227, 052228, 052232, 052235, 052243, 052246, 052251, 052264, 052265, 052271, 052272, 052279, 052284, 052286, 052288, 052291, 052293, 052300, 052303, 052310, 052311, 052315, 052319, 052341, 052347, 052351, 052352, 052354, 052368, 052369, 052370, 052371, 052372, 052374, 052381, 052383, 052387, 052388, 052389, 052400, 052402, 052404, 052407, 052410, 052421, 052428, 052434, 052440, 052464, 052468, 052471, 052478, 052481, 052485, 052493, 052495, 052497, 052500, 052512, 052514, 052521, 052522, 052524, 052532, 052533, 052534, 052543, 052544, 052547, 052550, 052554, 052558, 052563, 052567, 052568, 052573, 052577, 052578, 052586, 052588, 052592, 052595, 052596, 052602, 052608, 052610, 052615, 052618, 052619, 052629, 052633, 052643, 052647, 052650, 052652, 052657, 052662, 052664, 052666, 052669, 052672, 052681, 052685, 052691, 052695, 052696, 052703, 052713, 052714, 052719, 052725, 052727, 052743, 052744, 052748, 052749, 052750, 052761, 052767, 052768, 052775, 052783, 052786, 052788, 052789, 052800, 052803, 052821, 052824, 052829, 052832, 052841, 052860, 052861, 052862, 052863, 052867, 052870, 052871, 052873, 052892, 052901, 052905, 052907, 052908, 052911, 052917, 052919, 052940, 052947, 052952, 052972, 052974, 052978, 052987, 052992, 053001, 053016, 053024, 053031, 053041, 053044, 053047, 053055, 053065, 053066, 053067, 053069, 053070, 053073, 053079, 053081, 053096, 053104, 053105, 053106, 053111, 053116, 053120, 053122, 053123, 053128, 053129, 053132, 053135, 053143, 053145, 053146, 053160, 053163, 053168, 053171, 053174, 053176, 053179, 053191, 053197, 053199, 053207, 053213, 053222, 053223, 053224, 053232, 053233, 053237, 053238, 053247, 053248, 053252, 053253, 053255, 053261, 053266, 053278, 053280, 053284, 053291, 053304, 053305, 053308, 053314, 053315, 053319, 053323, 053347, 053354, 053365, 053367, 053368, 053370, 053373, 053377, 053378, 053380, 053388, 053391, 053392, 053404, 053414, 053419, 053429, 053430, 053438, 053439, 053440, 053441, 053456, 053457, 053459, 053461, 053468, 053474, 053486, 053494, 053497, 053503, 053507, 053509, 053511, 053513, 053514, 053519, 053527, 053529, 053533, 053537, 053543, 053549, 053550, 053553, 053557, 053565, 053580, 053584, 053591, 053601, 053603, 053625, 053647, 053684, 053726, 053751, 053758, 053761, 053768, 053779, 053784, 053838, 053845, 053849, 053863, 053866, 053867, 053869, 053908, 053913, 053918, 053923, 053927, 053930, 053955, 053960, 053972, 053994, 054006, 054007, 054017, 054023, 054026, 054030, 054037, 054041, 054047, 054051, 054061, 054070, 054085, 054096, 054107, 054109, 054111, 054125, 054130, 054137, 054150, 054152, 054154, 054156, 054171, 054176, 054183, 054188, 054211, 054235, 054237, 054260, 054261, 054273, 054275, 054291, 054321, 054328, 054357, 054367, 054368, 054387, 054392, 054412, 054434, 054461, 054468, 054472, 054493, 054495, 054502, 054550, 054574, 054599, 054600, 054618, 054636, 054669, 054689, 054744, 054748, 054749, 054751, 054753, 054755, 054768, 054771, 054780, 054792, 054804, 054811, 054820, 054826, 054832, 054844, 054846, 054857, 054862, 054863, 054871, 054882, 054911, 054913, 054935, 054938, 054948, 054951, 061077, 061078, 061079, 061092, 061096, 061113, 061118, 061120, 061122, 061125, 061129, 061130, 061131, 061132, 061133, 061143, 061148, 061152, 061153, 061157, 061158, 061159, 061164, 061170, 061175, 061188, 061193, 061196, 061198, 061202, 061203, 061205, 061206, 061213, 061215, 061216, 061217, 061226, 061233, 061234, 061238, 061242, 061246, 061247, 061253, 061256, 061265, 061272, 061278, 061282, 061283, 061284, 061286, 061288, 061293, 061294, 061295, 061304, 061306, 061310, 061317, 061324, 061325, 061326, 061329, 061337, 061343, 061349, 061350, 061351, 061354, 061356, 061358, 061366, 061372, 061378, 061387, 061395, 061396, 061401, 061404, 061405, 061407, 061414, 061421, 061424, 061426, 061436, 061439, 061440, 061442, 061443, 061445, 061446, 061449, 061455, 061457, 061461, 061463, 061464, 061466, 061468, 061469, 061472, 061474, 061478, 061480, 061488, 061493, 061496, 061498, 061500, 061501, 061505, 061506, 061508, 061513, 061519, 061520, 061522, 061531, 061533, 061544, 061551, 061564, 061565, 061566, 061570, 061574, 061576, 061579, 061582, 061590, 061594, 061597, 061598, 061600, 061607, 061615, 061619, 061620, 061622, 061623, 061624, 061628, 061632, 061633, 061634, 061635, 061637, 061638, 061643, 061644, 061651, 061654, 061655, 061664, 061672, 061675, 061679, 061683, 061687, 061692, 061694, 061697, 061698, 061700, 061701, 061707, 061712, 061723, 061728, 061731, 061736, 061743, 061746, 061747, 061748, 061755, 061762, 061763, 061766, 061767, 061776, 061777, 061779, 061787, 061791, 061795, 061800, 062626, 062627, 062638, 062649, 062659, 062670, 062706, 062719, 062725, 062732, 062742; UPN 00802526442827, Serial Numbers: 061204, 061477, 061567, 061616; UPN 00802526444807, Serial Numbers: 062624, 062651, 062667, 062679, 062692, 062870, 062872, 062879, 062909, 062910, 062960, 063027, 063028, 063088, 063437, 063445, 063450, 065817, 065827, 065828, 065837, 065853, 065892, 065901, 066112, 066118, 066123, 066215, 066226, 066266, 066281, 066287, 066292, 066299, 066301, 066311, 066339, 066343, 066373, 066388, 066393, 066763, 066781, 066782, 066786, 067067, 067171, 067894, 067898, 067946, 067947, 068061, 068075, 068089, 068394, 068396, 068425, 068434, 068437, 068449, 068458, 068463, 071331, 071386, 072943, 073041; UPN 00802526444814, Serial Numbers: 051959, 052267, 052273, 052305, 052470, 052594, 052670, 052770, 052833, 052904, 052910, 052918, 052934, 053008, 053234, 053362, 053387, 053400, 053476, 053698, 053770, 053777, 053872, 054005, 054020, 054068, 054075, 054094, 054140, 054239, 054265, 054304, 054334, 054463, 054496, 054657, 054920, 058272, 060432, 060559, 061260, 061285, 061292, 061332, 061418, 061419, 061545, 061589, 061665, 062175, 062625, 062628, 062630, 062633, 062657, 062658, 062674, 062676, 062678, 062685, 062689, 062702, 062704, 062708, 062716, 062718, 062721, 062727, 062739, 062747, 062750, 062751, 062755, 062761, 062763, 062764, 062766, 062769, 062774, 062775, 062777, 062780, 062781, 062784, 062785, 062787, 062789, 062797, 062798, 062799, 062801, 062805, 062807, 062810, 062817, 062821, 062828, 062832, 062839, 062847, 062849, 062850, 062851, 062854, 062860, 062862, 062869, 062871, 062874, 062880, 062882, 062892, 062894, 062904, 062906, 062919, 062921, 062923, 062926, 062930, 062934, 062940, 062955, 062959, 062961, 062968, 062974, 062980, 062984, 062985, 062988, 062993, 062997, 062998, 063003, 063007, 063016, 063017, 063018, 063019, 063029, 063031, 063043, 063046, 063052, 063056, 063058, 063059, 063061, 063062, 063066, 063072, 063074, 063077, 063081, 063082, 063085, 063086, 063087, 063096, 063102, 063103, 063109, 063113, 063115, 063116, 063117, 063121, 063129, 063130, 063136, 063137, 063138, 063140, 063142, 063146, 063154, 063155, 063156, 063163, 063166, 063168, 063170, 063173, 063174, 063176, 063177, 063181, 063182, 063187, 063190, 063192, 063195, 063197, 063198, 063199, 063203, 063204, 063206, 063211, 063214, 063217, 063221, 063222, 063228, 063236, 063242, 063243, 063245, 063247, 063251, 063252, 063255, 063260, 063264, 063266, 063269, 063270, 063272, 063281, 063283, 063286, 063287, 063288, 063294, 063299, 063304, 063305, 063307, 063308, 063315, 063317, 063318, 063324, 063334, 063341, 063342, 063344, 063345, 063348, 063349, 063352, 063354, 063358, 063360, 063367, 063368, 063370, 063376, 063388, 063389, 063391, 063393, 063395, 063396, 063399, 063407, 063409, 063410, 063412, 063413, 063415, 063417, 063418, 063419, 063421, 063429, 063432, 063434, 063453, 064830, 065723, 065808, 065816, 065832, 065838, 065841, 065857, 065872, 065893, 065894, 065897, 065965, 066006, 066021, 066108, 066116, 066129, 066130, 066135, 066138, 066140, 066144, 066156, 066157, 066159, 066160, 066161, 066166, 066175, 066177, 066178, 066179, 066181, 066185, 066187, 066189, 066190, 066191, 066193, 066196, 066197, 066199, 066200, 066201, 066203, 066206, 066207, 066210, 066214, 066217, 066219, 066223, 066234, 066237, 066238, 066240, 066249, 066252, 066257, 066260, 066261, 066265, 066267, 066270, 066276, 066282, 066286, 066297, 066305, 066338, 066342, 066349, 066350, 066351, 066352, 066354, 066362, 066363, 066365, 066366, 066368, 066375, 066385, 066390, 066394, 066396, 066399, 066403, 066405, 066413, 066416, 066421, 066423, 066426, 066430, 066436, 066444, 066451, 066455, 066457, 066460, 066461, 066464, 066467, 066471, 066474, 066475, 066477, 066483, 066485, 066487, 066491, 066493, 066494, 066496, 066498, 066502, 066506, 066507, 066509, 066510, 066516, 066518, 066522, 066525, 066529, 066530, 066536, 066538, 066540, 066542, 066543, 066546, 066547, 066550, 066551, 066553, 066554, 066557, 066558, 066559, 066561, 066662, 066679, 066695, 066710, 066713, 066716, 066730, 066733, 066735, 066737, 066750, 066756, 066775, 066796, 066801, 066809, 067064, 067070, 067078, 067079, 067084, 067086, 067090, 067094, 067098, 067099, 067100, 067102, 067107, 067108, 067110, 067112, 067114, 067127, 067128, 067137, 067139, 067142, 067145, 067151, 067155, 067158, 067164, 067170, 067173, 067174, 067175, 067182, 067185, 067187, 067188, 067189, 067191, 067192, 067195, 067196, 067198, 067199, 067204, 067205, 067212, 067215, 067217, 067225, 067229, 067231, 067240, 067242, 067243, 067250, 067259, 067261, 067263, 067266, 067273, 067274, 067276, 067279, 067288, 067289, 067293, 067295, 067296, 067300, 067310, 067311, 067314, 067316, 067318, 067324, 067325, 067326, 067328, 067335, 067338, 067339, 067341, 067342, 067344, 067345, 067351, 067352, 067353, 067359, 067360, 067843, 067853, 067860, 067861, 067862, 067863, 067868, 067870, 067874, 067876, 067878, 067879, 067881, 067886, 067895, 067896, 067900, 067902, 067904, 067907, 067908, 067910, 067914, 067918, 067919, 067921, 067924, 067925, 067926, 067931, 067937, 067938, 067942, 067943, 067945, 067953, 067954, 067957, 067958, 067959, 067960, 067961, 067962, 067966, 067968, 067973, 067974, 067978, 067981, 067984, 067985, 067995, 068002, 068003, 068005, 068006, 068008, 068009, 068017, 068019, 068025, 068028, 068030, 068032, 068034, 068042, 068045, 068050, 068052, 068053, 068065, 068068, 068073, 068083, 068095, 068098, 068099, 069685, 069688, 069689, 069691, 069692, 069695, 069697, 069701, 069704, 069708, 069710, 069711, 069713, 069714, 069715, 069717, 069720, 069721, 069722, 069724, 069734, 070425, 070428, 070429, 070430, 070432, 070436, 070444, 070446, 070448, 070449, 070458, 070459, 070463, 070472, 070473, 070475, 070481, 070483, 070486, 070487, 070489, 070493, 070494, 070495, 070498, 070502, 070506, 070507, 070510, 070511, 070512, 070514, 070516, 070517, 070518, 070524, 070526, 070528, 070529, 070530, 070531, 070535, 070538, 070540, 070544, 070546, 071324, 071326, 071328, 071329, 071332, 071337, 071344, 071346, 071347, 071348, 071350, 071353, 071355, 071356, 071358, 071360, 071363, 071366, 071371, 071372, 071375, 071376, 071384, 071387, 071394, 071397, 071399, 071401, 071403, 071404, 071407, 071408, 071412, 071416, 071420, 071421, 071422, 071423, 071424, 071425, 071427, 071431, 071432, 071433, 071437, 071443, 071839, 071843, 071844, 071846, 071847, 071858, 071885, 071886, 071887, 071888, 071890, 071891, 071895, 071896, 071899, 071900, 071901, 071904, 071907, 071908, 071915, 071917, 071918, 071921, 071923, 071926, 071934, 071938, 071943, 071946, 071953, 072181, 072187, 072193, 072198, 072202, 072209, 072222, 072227, 072231, 072239, 072242, 072246, 072248, 072249, 072252, 072256, 072269, 072273, 072279, 072280, 072285, 072519, 072522, 072524, 072525, 072532, 072533, 072540, 072550, 072559, 072566, 072570, 072573, 072584, 072587, 072588, 072595, 072608, 072609, 072627, 072631, 072636, 072645, 072650, 072653, 072660, 072674, 072676, 072690, 072701, 072757, 072759, 072793, 072806, 072834, 072837, 072842, 072855, 072866, 072904, 073011, 073067, 073085, 073152; UPN 00802526445262, Serial Numbers: 066539; UPN 00802526454301, Serial Numbers: 053708, 054036, 054086, 068057, 072859, 072872, 072877; UPN 00802526454950, Serial Numbers: 052914, 054302, 057108, 057398, 059170, 061735, 065875, 066312, 066425, 067140, 067167, 067264, 068408, 068417, 071868, 071947, 071955, 071958, 072236, 072827, 073124; UPN 00802526467288, Serial Numbers: 051203, 051261, 051280, 052638, 052847, 053006, 053058, 053203, 053531, 053606, 053744, 056105, 058216, 061186, 061416, 062710, 065850, 066198, 066703, 066715, 066717, 066726, 066736, 066753, 066757, 066770, 066797, 066813, 066816, 066818, 066820, 066823, 066825, 072220, 072245, 072521, 072535, 072538, 072551, 072576, 072580, 072586, 072602, 072613, 072615, 072620, 072622, 072625, 072632, 072637, 072651, 072655, 072665, 072667, 072668, 072669, 072672, 072684, 072688, 072691, 072693, 072695, 072699, 072702, 072709, 072712, 072714, 072730, 072731, 072738, 072742, 072744, 072754, 072758, 072769, 072770, 072775, 072782, 072792, 072794, 072796, 072800, 072803, 072804, 072807, 072810, 072815, 072817, 072820, 072821, 072830, 072838, 072858, 072867, 072871, 072873, 072879, 072893, 072900, 072912, 072915, 072916, 072917, 072919, 072923, 072928, 072932, 072934, 072938, 072939, 072940, 072948, 072960, 072962, 072964, 072973, 072974, 072977, 072980, 072992, 072995, 073004, 073014, 073016, 073017, 073019, 073020, 073021, 073025, 073029, 073033, 073040, 073046, 073051, 073056, 073066, 073068, 073069, 073070, 073073, 073075, 073076, 073079, 073082, 073090, 073094, 073095, 073098, 073106, 073113, 073115, 073117, 073122, 073125, 073130, 073134, 073138, 073143, 073145, 073154, 073157, 073161, 073165, 073167, 073173, 073175, 073176, 073177, 073178; UPN 00802526467295, Serial Numbers: 053290, 053884, 054214, 054240, 072968, 072971, 072976, 073055; UPN 00802526467301, Serial Numbers: 072990, 072997; UPN 00802526469343, Serial Numbers: 051652, 051675, 052175, 053322, 055011, 055795, 055975, 056252, 056538, 056623, 057209, 057282, 057511, 057598, 060673, 060800, 062035, 062200, 063812, 064258, 064834, 066020, 067036, 067062, 069403, 070484, 070988, 071004, 071278, 071542, 071824, 072020, 072048, 072528, 072539, 073305; UPN 00802526469350, Serial Numbers: 054320, 060742, 067282; UPN 00802526469374, Serial Numbers: 057896, 059210, 060636, 069766; UPN 00802526474170, Serial Numbers: 072863, 073044; UPN 00802526475450, Serial Numbers: 051523, 056095, 057216, 057482, 057727, 059423, 059736, 060212, 062260, 064630, 066969, 067705, 069843, 069856, 071015, 072736, 072869; UPN 00802526475467, Serial Numbers: 051127, 051481, 051530, 051609, 053775, 053888, 055425, 055723, 056089, 056279, 056956, 056974, 057181, 057773, 058229, 058795, 062511, 067054, 068528, 069847, 070890; UPN 00802526475474, Serial Numbers: 072225; UPN 00802526475757, Serial Numbers: 051583, 055182, 055231, 055390, 055396, 055403, 055602, 055778, 056281, 056457, 056960, 057617, 057969, 058062, 058078, 058194, 058736, 058797, 058967, 059309, 059454, 059611, 059850, 060030, 060215, 060429, 060834, 062230, 062473, 062488, 062556, 063220, 063239, 063642, 064272, 064587, 064707, 065072, 065178, 065332, 065525, 065648, 065960, 066803, 066863, 066890, 067011, 068755, 069377, 070262, 070269, 070302, 071539, 072606, 073206, 073214, 073472, 073519; UPN 00802526475764, Serial Numbers: 057100, 071166; UPN 00802526475771, Serial Numbers: 056900, 062948, 065907, 066432, 066731, 069296, 072447; UPN 00802526477669, Serial Numbers: 051070, 051235, 051917, 052107, 052298, 052426, 052846, 053090, 054113, 054390, 054960, 055363, 055382, 055725, 055850, 055946, 056038, 057099, 057309, 057717, 057752, 058018, 058991, 059052, 059693, 060013, 060431, 060713, 060724, 063219, 065060, 066026, 066667, 066789, 066811, 066828, 066845, 067023, 067536, 067577, 067603, 067631, 067725, 067782, 068324, 068355, 068494, 069153, 069206, 069226, 069260, 069324, 069400, 069564, 070392, 071477, 071502, 072294, 072366, 072378; UPN 00802526477676, Serial Numbers: 055649, 055718, 055760, 055805, 056045, 057268, 058538, 060001, 069283, 070969, 071085, 071686; UPN 00802526477683, Serial Numbers: 062815, 062945, 066230; UPN 00802526478246, Serial Numbers: 051116, 052060, 052163, 052815, 053023, 053269, 053463, 053668, 054701, 054899, 055683, 056678, 057132, 057466, 058246, 059577, 061095, 061115, 061150, 061517, 061640, 062759, 062877, 062946, 066768, 066829, 066899, 067916, 067939, 068049, 068567, 069555, 069623, 071411, 071898, 072616, 072671, 072808, 072844, 072895; UPN 00802526478253, Serial Numbers: 055651, 056406; UPN 00802526479670, Serial Numbers: 051159, 051210, 051240, 051447, 051528, 051751, 051821, 051954, 051985, 052103, 052141, 052165, 052187, 052209, 052219, 052236, 052245, 052301, 052308, 052367, 052394, 052403, 052406, 052462, 052463, 052525, 052538, 052555, 052648, 052667, 052668, 052694, 052707, 052747, 052751, 052772, 052801, 052806, 052893, 052903, 052933, 052976, 052977, 053003, 053007, 053025, 053045, 053139, 053173, 053183, 053185, 053192, 053230, 053241, 053249, 053267, 053303, 053313, 053317, 053335, 053338, 053375, 053379, 053415, 053451, 053455, 053465, 053483, 053555, 053594, 053595, 053597, 053602, 053611, 053613, 053631, 053691, 053717, 053754, 053806, 053810, 053822, 053852, 053876, 053925, 053949, 054018, 054024, 054049, 054067, 054083, 054178, 054446, 054503, 054583, 054610, 054640, 054652, 054687, 054723, 054746, 054764, 054775, 054805, 054838, 054867, 054894, 054936, 055169, 055539, 056063, 057158, 057485, 057585, 057802, 057977, 058051, 058237, 058256, 058621, 058861, 059078, 059431, 059506, 059512, 059721, 059780, 059894, 059978, 060166, 060339, 060474, 060649, 060666, 060836, 061061, 061093, 061094, 061103, 061119, 061121, 061147, 061166, 061227, 061248, 061268, 061269, 061315, 061345, 061348, 061389, 061420, 061422, 061434, 061444, 061458, 061481, 061518, 061532, 061540, 061550, 061577, 061605, 061636, 061641, 061646, 061656, 061668, 061685, 061693, 061703, 061709, 061714, 061730, 061758, 061797, 061927, 061965, 062027, 062291, 062461, 062617, 062643, 062672, 062677, 062686, 062687, 062697, 062714, 062729, 062733, 062736, 062770, 062791, 062800, 062841, 062846, 062878, 062907, 062931, 062943, 062965, 062977, 062987, 063050, 063063, 063123, 063128, 063131, 063133, 063134, 063149, 063202, 063205, 063274, 063330, 063346, 063355, 063378, 063384, 064135, 064244, 064363, 064603, 065249, 065437, 065567, 065633, 065704, 065812, 065859, 065910, 065981, 066009, 066105, 066106, 066107, 066141, 066150, 066209, 066247, 066406, 066428, 066439, 066479, 066484, 066495, 066532, 066552, 066666, 066682, 066728, 066761, 066839, 067010, 067061, 067121, 067147, 067203, 067224, 067232, 067251, 067315, 067337, 067346, 067423, 067810, 067864, 067866, 067875, 067912, 067913, 067930, 067936, 067965, 067982, 067986, 067998, 068022, 068033, 068038, 068064, 068072, 068076, 068079, 068200, 068443, 068593, 068816, 069220, 069417, 069471, 069609, 069707, 069712, 069727, 070402, 070408, 070442, 070466, 070503, 070542, 070545, 070958, 071277, 071381, 071405, 071413, 071415, 071419, 071435, 072116, 072201, 072247, 072259, 072281, 072325, 072379, 072410, 072488, 072514, 072518, 072520, 072526, 072541, 072553, 072554, 072565, 072629, 072633, 072689, 072707, 072723, 072741, 072760, 072776, 072779, 072833, 072931, 072947, 072957, 072986, 073000, 073061, 073071, 073081, 073096, 073099, 073126, 073132, 073141, 073153, 073155, 073512, 073628, 073887, 073891, 073896, 073904, 073906, 073915, 073919, 073941, 073954, 073971, 073976, 073978, 073981, 073984, 074069, 074075, 074098, 074129, 074165, 074191, 074205, 074211, 074220, 074234, 074236, 074237, 074238, 074248, 074250, 074251, 074254, 074255, 074263, 074266, 074267, 074269, 074272, 074274, 074275, 074276, 074279, 074282, 074283, 074290, 074297, 074300, 074304, 074306, 074317, 074318, 074325, 074327, 074331, 074334, 074340, 074341, 074342, 074345, 074352, 074354, 074358, 074363, 074365, 074367, 074368, 074371, 074383, 074388, 074392, 074393, 074397, 074401, 074406, 074408, 074416, 074425, 074426, 074433, 074434, 074448, 074450, 074462, 074463, 074464, 074465, 074468, 074469, 074470, 074471, 074475, 074487, 074493, 074494, 074497, 074501, 074502, 074503, 074510, 074516, 074519, 074520, 074521, 074522, 074525, 074526, 074528, 074529, 074541, 074552, 074568, 074598, 074602, 074603, 074610, 074630, 074652, 074654, 074659, 074663, 074687, 074714, 074715, 074723, 074724, 074729, 074741, 074743, 074750, 074752, 074754, 074756, 074761, 074762, 074765, 074781, 074784, 074803, 074809, 074814, 074822, 074823; UPN 00802526479687, Serial Numbers: 051452, 051555, 051674, 053181, 057263, 057936, 058650, 058709, 059193, 064873, 065046, 066400, 067707, 068918, 070247, 072678, 072949, 072954, 073128, 073617, 073621, 073622, 073623, 073631, 073635, 073653, 073874, 073895, 073920, 073929, 073942, 073986, 073990, 074005, 074007, 074015, 074156, 074202, 074218, 074240, 074310, 074315; UPN 00802526479694, Serial Numbers: 067101; UPN 00802526480652, Serial Numbers: 060270, 060647, 062661, 062665, 063314, 066290, 066309, 066346, 067605, 067941, 068405, 068416, 068431, 071863, 071884, 072870, 072942, 074680, 074815, 074900; UPN 00802526484605, Serial Numbers: 051980, 052519, 052530, 053942, 054395, 055348, 061082, 066255, 067177, 072814, 072913, 075925, 075931, 075932, 075938, 075941, 075943, 075946, 075947, 075951, 075952, 075956, 075957, 075962, 075963, 075970, 075971, 075972, 075979, 075981, 076110, 076111, 076113, 076117, 076119, 076124, 076126, 076128, 076133, 076134, 076135, 076136, 076137, 076143, 076147, 076149, 076155, 076161, 076162, 076163, 076165, 076173, 076178, 076184, 076185, 076187, 076191, 076192, 076198, 076199, 076201, 076203, 076208, 076210, 076212, 076218, 076220, 076221, 076228, 076233, 076234, 076237, 076239, 076242, 076243, 076248, 076254, 076258, 076259, 076260, 076261, 076264, 076265, 076267, 076279, 076281, 076283, 076286, 076292, 076297, 076298, 076301, 076302, 076308, 076317, 076323, 076324, 076325, 076331, 076334, 076338, 076343, 076345; UPN 00802526484803, Serial Numbers: 061719, 067254, 071956; UPN 00802526486081, Serial Numbers: 067269, 072153, 075011, 075021, 075127, 075175; UPN 00802526486913, Serial Numbers: 051137, 051245, 052054, 052257, 052283, 052304, 052359, 052363, 052399, 052424, 052537, 052574, 052718, 052820, 052823, 052831, 052925, 052938, 052962, 052994, 053082, 053097, 053144, 053215, 053239, 053352, 053372, 053446, 053464, 053505, 053559, 053618, 053713, 053789, 053801, 053883, 053896, 053916, 053963, 053989, 053991, 054009, 054033, 054065, 054087, 054244, 054272, 054331, 054382, 054400, 054469, 054581, 054876, 054883, 054907, 055738, 056205, 056427, 056916, 057120, 057224, 059540, 059865, 060418, 061100, 061135, 061200, 061221, 061225, 061275, 061297, 061298, 061417, 061456, 061460, 061560, 061563, 061572, 061613, 061618, 061715, 061757, 062793, 062794, 062804, 062816, 062830, 062837, 062899, 062938, 063023, 063048, 063064, 063080, 063090, 063125, 063178, 063223, 063233, 063301, 063321, 063326, 063335, 063336, 063416, 063420, 066119, 066133, 066155, 066167, 066243, 066254, 066408, 066415, 066478, 066497, 066501, 066519, 066526, 066555, 066701, 066881, 067106, 067126, 067160, 067227, 067840, 068021, 068046, 068081, 068339, 068658, 068859, 069560, 069576, 069694, 069725, 070431, 070435, 070451, 070491, 070541, 070543, 071377, 071385, 071396, 071864, 071872, 071931, 071939, 072064, 072185, 072263, 072470, 072506, 072511, 072569, 072611, 072617, 072662, 072809, 072835, 072891, 072899, 073100, 073120, 074055, 074244, 074394, 074407, 074428, 074474, 074495, 074523, 074633, 075954, 075974, 076127, 076280, 076336, 076350, 076352, 076354, 076355, 076356, 076365, 076368, 076394, 076426, 076427, 076428, 076433, 076435, 076448, 076521, 076556, 076666, 076726, 076765, 076770, 076793, 076799, 076800, 076802, 076804, 076807, 076815, 076816, 076821, 076822, 076823, 076825, 076853, 076865, 076876, 076886, 076887, 076888, 076890, 076902, 076904, 076908, 076914, 076915, 076919, 076921, 076923, 076927, 076929, 076938, 076940, 076941, 076947, 076953, 076958, 076959, 076969, 076972, 076973, 076982, 076988, 076990, 076997, 077005, 077019, 077020, 077023, 077024, 077025, 077037, 077050, 077076, 077078, 077081, 077091, 077092, 077099, 077105, 077126, 077129, 077136, 077141, 077159, 077169, 077173, 077176, 077177, 077184, 077185, 077187, 077188, 077195, 077196, 077197, 077198, 077212, 077215, 077216, 077219, 077576, 077583, 077590, 077601, 077602, 077632, 077650, 077708, 077739, 077744, 077748, 077749, 077752, 077753, 077759; UPN 00802526486920, Serial Numbers: 052390, 076447, 076507, 076516, 076525, 076574, 076622, 076623, 076631, 076642, 076676, 076701; UPN 00802526486937, Serial Numbers: 062908, 063320, 066443, 066691, 068407, 072195, 072936, 076069, 076739, 076751, 077003; UPN 00802526486944, Serial Numbers: 053545, 056672, 066696, 068082, 074881, 076640, 077070, 077093; UPN 00802526496608, Serial Numbers: 051974, 052041, 052262, 052269, 052366, 052446, 052591, 052697, 052745, 052869, 052902, 053050, 053062, 053229, 053403, 053577, 054004, 054081, 054148, 054427, 054443, 054544, 054752, 055124, 059709, 061172, 061183, 061277, 061303, 061383, 061448, 061479, 061490, 061546, 061548, 061642, 061658, 062813, 062836, 062853, 062962, 062967, 062982, 063022, 063091, 063094, 063258, 063292, 064607, 065576, 065868, 065902, 066302, 066535, 066794, 066848, 067143, 067197, 067334, 067901, 068040, 068086, 069738, 070474, 070537, 072178, 072391, 072773, 073026, 073169, 073870, 074180, 074511, 076219, 076328, 076975, 077592, 077599, 077600, 077615, 077617, 077627, 077661, 077662, 077664, 077686, 077690, 077709, 077750; UPN 00802526496615, Serial Numbers: 052302, 052401, 052560, 052620, 052810, 052998, 053072, 053924, 053948, 053967, 053987, 054245, 054277, 054829, 054850, 054949, 056448, 058344, 059250, 060536, 060874, 061144, 061151, 061475, 061514, 061680, 062728, 063238, 063306, 063333, 063401, 063406, 063438, 065806, 066333, 066389, 066441, 067235, 067909, 068362, 069538, 069706, 070447, 070455, 070508, 071414, 072228, 072283, 072545, 072546, 072574, 072661, 072765, 072898, 072982, 073022, 074223, 074374, 074384, 076337, 076340, 076911, 076995, 077666, 077870, 077873, 077875, 077880, 077885, 077886, 077895, 077911, 077912, 077914, 077921, 077924, 077926, 077934, 077935, 077938, 077939, 077942, 077943, 077950, 077952, 077956, 077963, 077966, 077968, 077994; UPN 00802526496622, Serial Numbers: 077562, 077593, 077594, 077821; UPN 00802526496639, Serial Numbers: 055500, 059781, 059799, 060898, 063791, 072910, 073399, 077280, 077287, 077309, 077312, 077314, 077315, 077318, 077876, 077878, 077915, 077925, 077929, 077930, 077932, 078214, 078218, 078250, 078303; UPN 00802526497629, Serial Numbers: 068426, 072257; UPN 00802526497643, Serial Numbers: 053574, 060989, 062886, 065851, 065883, 066310, 071336, 071848, 078227, 078241, 078262, 078271, 078276, 078296, 078298, 078327, 078349, 078351; UPN 00802526506703, Serial Numbers: 051013, 051302, 051562, 052392, 052405, 052486, 052542, 052699, 052762, 052865, 052877, 052898, 052909, 052924, 053034, 053053, 053137, 053187, 053188, 053259, 053343, 053366, 053385, 053522, 053697, 053827, 053880, 054021, 054167, 054168, 054212, 054294, 054338, 054344, 054362, 054686, 054928, 054941, 055175, 057095, 057469, 060663, 061214, 061258, 061276, 061361, 061382, 061432, 061462, 061585, 061647, 061661, 061717, 061744, 061752, 061792, 061977, 062660, 062863, 062966, 062983, 062986, 062994, 063108, 063147, 063208, 063371, 063404, 063440, 065863, 066113, 066262, 066308, 066316, 066856, 066914, 067200, 067268, 067321, 067889, 067911, 067915, 067920, 067980, 068023, 068092, 069577, 071922, 072188, 072268, 072818, 072944, 072951, 072963, 073043, 073050, 073123, 074284, 076175, 076197, 076229, 076313, 076319, 076369, 076818, 076881, 076956, 076967, 076986, 077108, 077249, 077258, 077270, 077295, 077299, 077300, 077301, 077302, 077324, 077443, 077461, 077503, 077527, 077536, 077547, 077559, 077584, 077596, 077606, 077614, 077618, 077622, 077625, 077631, 077642, 077669, 077677, 077680, 077699, 077783, 077784, 077841, 077853, 077854, 077865, 077869, 077883, 077894, 077897, 077904, 077907, 077910, 077920, 077927, 077928, 077937, 077960, 077973, 077976, 077985, 077986, 077993, 077998, 077999, 078000, 078002, 078012, 078015, 078021, 078026, 078027, 078030, 078034, 078035, 078036, 078037, 078040, 078043, 078044, 078046, 078051, 078060, 078063, 078065, 078067, 078068, 078069, 078072, 078073, 078079, 078082, 078104, 078106, 078110, 078113, 078116, 078129, 078135, 078136, 078151, 078157, 078158, 078161, 078166, 078177; UPN 00802526515903, Serial Numbers: 051132, 051542, 051862, 052518, 052609, 052679, 052878, 053178, 053240, 053242, 053316, 053350, 053448, 053524, 053569, 053582, 053803, 053854, 053998, 054177, 054216, 054288, 054354, 054551, 054591, 054766, 054890, 056087, 056147, 061257, 061281, 061534, 061549, 061649, 061737, 062514, 062613, 062806, 062887, 062954, 063114, 063257, 063263, 064050, 066111, 066183, 066216, 066473, 066676, 067073, 067129, 067233, 067292, 067309, 067836, 067869, 067890, 067948, 068029, 068055, 069261, 069573, 071246, 071877, 072058, 072216, 072589, 072789, 073009, 073048, 074271, 074376, 076223, 076238, 076306, 076966, 077013, 077741, 077909, 078070, 078092, 078234, 078283, 078285, 078307, 078308, 078325, 078335, 078341, 078346, 078360, 078366, 078374, 078379, 078385, 078387, 078424, 078426, 078428, 078431, 078432, 078433, 078435, 078441, 078449, 078458, 078459, 078463, 078466, 078468, 078478, 078483, 078494, 078497, 078499, 078500, 078502, 078504, 078507, 078509, 078510, 078514, 078516, 078523, 078525, 078527, 078532, 078588, 078607, 078622, 078625, 078627, 078629, 078631, 078633, 078639, 078640, 078646, 078647, 078649, 078650, 078653, 078654, 078655, 078659, 078661, 078664, 078667, 078670, 078671, 078672, 078674, 078677, 078679, 078681, 078686, 078687, 078689, 078691, 078696, 078697, 078698, 078700, 078701, 078703, 078706, 078712, 078714, 078732, 078733, 078756, 078760, 078763, 078769, 078778; UPN 00802526515910, Serial Numbers: 078423, 078438, 078485, 078489, 078570, 078577, 078595, 078598, 078600; UPN 00802526515927, Serial Numbers: 052983, 066459, 066740, 072314, 073010, 073180, 078442, 078456, 078699, 078723, 078735, 078745, 078748, 078750; UPN 00802526517112, Serial Numbers: 051853, 051860, 052510, 052590, 052660, 052763, 052852, 052929, 052939, 053118, 053251, 053309, 053349, 053359, 053363, 053402, 053551, 053638, 053661, 053740, 053742, 053832, 053946, 054165, 054379, 054490, 054552, 054587, 054638, 054859, 056664, 057243, 058413, 060069, 060801, 061091, 061106, 061178, 061305, 061309, 061319, 061328, 061331, 061339, 061369, 061413, 061433, 062576, 062645, 062724, 062771, 063256, 063322, 063365, 063383, 063386, 065818, 066148, 066280, 067080, 067088, 067136, 067152, 067220, 067221, 067287, 067327, 068448, 069380, 069382, 070920, 071426, 071874, 072675, 072937, 073119, 073140, 073889, 073953, 074421, 074482, 074774, 076132, 076142, 076151, 076171, 076193, 076226, 076252, 076266, 076273, 076431, 076920, 077616, 077695, 077982, 078154, 078417, 078476, 078506, 078543, 078544, 078549, 078551, 078558, 078568, 078572, 078574, 078576, 078580, 078582, 078584, 078585, 078587, 078589, 078605, 078606, 078608, 078783, 078813, 078817, 078822, 078823, 078824, 078832, 078837, 078838, 078857, 078859, 078888, 078895, 078908, 078954, 078956, 078959, 078980, 078982, 078984, 078986, 078989, 078993, 079017, 079021, 079038, 079047, 079063, 079066, 079079, 079083, 079086, 079090, 079091; UPN 00802526517136, Serial Numbers: 078652, 078796, 078806, 078856, 078880, 078892, 078893, 078896, 078903, 078912, 078937, 078945, 078950, 078952, 078966, 078968, 078975, 078978, 078985; UPN 00802526517174, Serial Numbers: 052306, 062861, 068377, 068409, 074861, 078324, 078642, 079194, 079201, 079204; UPN 00802526526602, Serial Numbers: 051152, 051969, 052106, 052263, 052270, 052379, 052382, 052503, 052549, 052553, 052785, 052857, 052973, 053059, 053155, 053328, 053442, 053526, 053739, 053792, 053900, 054060, 054142, 054166, 054300, 054478, 054585, 054639, 054843, 054919, 056208, 057273, 057392, 057809, 060268, 060314, 061087, 061123, 061127, 061165, 061190, 061261, 061262, 061273, 061385, 061541, 061670, 061690, 061695, 061796, 061799, 061880, 062620, 062754, 062973, 063049, 063089, 063153, 063180, 063265, 063447, 063878, 064652, 065216, 065855, 065896, 066114, 066142, 066220, 066472, 067074, 067075, 067119, 067156, 067193, 067202, 067248, 067312, 067871, 067873, 067891, 067988, 068509, 069214, 069287, 069702, 069731, 069736, 070975, 071327, 071364, 071855, 071882, 072119, 072593, 072604, 072663, 072706, 072902, 072920, 072984, 073103, 073104, 073110, 074337, 074360, 074533, 074634, 075948, 075977, 075982, 076109, 076122, 076129, 076131, 076140, 076180, 076194, 076204, 076214, 076217, 076222, 076299, 076339, 076901, 076905, 076924, 076930, 076955, 076977, 077069, 077203, 077326, 077439, 077451, 077506, 077529, 077588, 077603, 077645, 077738, 077857, 077887, 077916, 077955, 077977, 078031, 078050, 078052, 078141, 078313, 078373, 078436, 078464, 078492, 078666, 078720, 078737, 078761, 078827, 078905, 078910, 078914, 078917, 078922, 078930, 078934, 078949, 078961, 078974, 078995, 078996, 078998, 079002, 079003, 079005, 079006, 079007, 079008, 079009, 079011, 079012, 079016, 079019, 079022, 079024, 079025, 079028, 079029, 079032, 079033, 079039, 079040, 079043, 079049, 079053, 079054, 079055, 079057, 079058, 079064, 079067, 079068, 079070, 079071, 079074, 079075, 079076, 079078, 079080, 079084, 079085, 079087, 079089, 079092, 079093, 079094, 079095, 079096, 079098, 079100, 079101, 079103, 079105, 079106, 079107, 079109, 079110, 079111, 079112, 079113, 079114, 079115, 079116, 079117, 079119, 079120, 079121, 079123, 079125, 079126, 079129, 079132, 079136, 079140, 079144, 079145, 079148, 079150, 079156, 079158, 079159, 079160, 079161, 079170, 079175, 079189; UPN 00802526526800, Serial Numbers: 053756, 064154; UPN 00802526526817, Serial Numbers: 062935, 063196, 065880, 067971, 071841, 071842, 071854, 074665, 078302; UPN 00802526527609, Serial Numbers: 051092, 051284, 051347, 051714, 051782, 051857, 051859, 051999, 052079, 052109, 052112, 052182, 052211, 052231, 052253, 052429, 052443, 052511, 052583, 052611, 052627, 052654, 052715, 052731, 052759, 052781, 052808, 052849, 052883, 052884, 052956, 052957, 052996, 053017, 053039, 053075, 053221, 053331, 053407, 053420, 053422, 053458, 053490, 053504, 053516, 053518, 053520, 053540, 053630, 053646, 053702, 053733, 053734, 053735, 053743, 053748, 053752, 053797, 053815, 053855, 053870, 053937, 053973, 053997, 054008, 054010, 054014, 054025, 054027, 054069, 054175, 054192, 054199, 054210, 054217, 054229, 054263, 054299, 054341, 054405, 054407, 054447, 054586, 054589, 054696, 054709, 054728, 054734, 054745, 054773, 054809, 054853, 054881, 055763, 055843, 058186, 058582, 058720, 058731, 059268, 059701, 060278, 061081, 061105, 061161, 061169, 061201, 061336, 061352, 061370, 061399, 061465, 061483, 061504, 061523, 061527, 061539, 061617, 061667, 061688, 061689, 061706, 061720, 061749, 061793, 062605, 062711, 062768, 062795, 062812, 062933, 062950, 062958, 062963, 063011, 063065, 063167, 063191, 063210, 063249, 063277, 063278, 063327, 063357, 063385, 063414, 063424, 063863, 064156, 064821, 065646, 065807, 066124, 066136, 066164, 066171, 066173, 066242, 066258, 066279, 066326, 066429, 066488, 066548, 066624, 067071, 067123, 067135, 067148, 067222, 067245, 067307, 067362, 067591, 067666, 067845, 067883, 067885, 067999, 068026, 068027, 068043, 069698, 069709, 069726, 070294, 070456, 070457, 070515, 071340, 071361, 071362, 071388, 071395, 071398, 071406, 071418, 071869, 071889, 071932, 072513, 072531, 072560, 072596, 072619, 072634, 072729, 072787, 072824, 072862, 072875, 072883, 072927, 072930, 072946, 072950, 072959, 072975, 073038, 073060, 073074, 073164, 073172, 073529, 074154, 074201, 074209, 074260, 074373, 074507, 074647, 076144, 076216, 076255, 076314, 076376, 076820, 076891, 076910, 076913, 076937, 076949, 076957, 076971, 076992, 077112, 077186, 077227, 077563, 077652, 077667, 077696, 077764, 077881, 077900, 077901, 077931, 077965, 078047, 078340, 078495, 078641, 078662, 078665, 078774, 078926, 078931, 078979, 078990, 079034, 079048, 079134, 079141, 079192, 079195, 079197, 079198, 079199, 079202, 079212, 079213, 079214, 079218, 079219, 079220, 079221, 079222, 079225, 079227, 079228, 079231, 079232, 079240, 079242, 079243, 079244, 079245, 079246, 079250, 079251, 079252, 079256, 079259, 079260, 079262, 079263, 079265, 079266, 079267, 079269, 079274, 079275, 079278, 079281, 079282, 079286, 079288, 079289, 079290, 079291, 079297, 079302, 079307, 079308, 079315, 079317, 079320, 079321, 079323, 079326, 079328, 079332, 079334, 079337, 079341, 079345, 079349, 079354, 079358, 079364, 079368, 079370, 079372, 079373, 079379, 079380, 079382, 079385, 079386, 079390, 079392, 079393, 079396, 079399, 079400, 079402, 079405, 079408, 079415, 079417, 079420, 079424, 079425, 079427, 079429, 079432, 079433, 079437, 079441, 079444, 079446, 079451, 079452, 079454, 079456, 079466, 079467, 079471, 079474, 079475, 079476, 079477, 079480, 079487, 079492, 079493, 079496, 079502, 079512, 079513, 079514, 079516, 079517, 079518, 079519, 079522, 079523, 079530, 079531, 079535, 079543, 079544, 079550, 079551, 079552, 079556, 079560, 079561, 079563, 079564, 079566, 079568, 079569, 079570, 079571, 079574, 079576, 079577, 079578, 079579, 079580, 079582, 079586, 079589, 079591, 079599, 079600, 079602, 079604, 079614, 079641, 079642, 079643, 079648, 079654, 079655, 079656, 079664, 079669, 079673, 079675, 079677, 079678, 079692, 079695, 079716, 079719, 079721, 079722, 079729, 079733, 079741, 079743, 079759, 079774, 079885; UPN 00802526527616, Serial Numbers: 052140, 053201, 079247, 079338, 079351, 079360, 079384, 079472; UPN 00802526527623, Serial Numbers: 053125, 059480, 062668, 063010, 063045, 065815, 065826, 065876, 066304, 066331, 068062, 071902, 078675, 078762; UPN 00802526532832, Serial Numbers: 059150, 060795, 063339, 065318, 067184, 071871, 072215, 078505; UPN 00802526554810, Serial Numbers: 077261; UPN 00802526557002, Serial Numbers: 059135, 064896, 065824, 079993, 079994, 079998, 080034, 080059, 080115, 080118; UPN 00802526557019, Serial Numbers: 061581, 063038, 063669, 068454, 074910, 078329, 078752, 078858, 080001, 080046, 080055, 080056, 080058, 080084, 080095, 080097, 080112, 080124, 080134, 080136; UPN 00802526561900, Serial Numbers: 051996, 052337, 052455, 052603, 052683, 052795, 052895, 053037, 053124, 053310, 053333, 053469, 053568, 053590, 053633, 053682, 053755, 053941, 053965, 054160, 054189, 054228, 054451, 054477, 054572, 054611, 054720, 054747, 054759, 054833, 054854, 056486, 056985, 058292, 059269, 060016, 061145, 061181, 061307, 061334, 061470, 062773, 062833, 062875, 063008, 063040, 063075, 063099, 063127, 063271, 065831, 066121, 066180, 066361, 066528, 066779, 067089, 067257, 067280, 070423, 070439, 070452, 070501, 071930, 072507, 072692, 072846, 073349, 074423, 076120, 076141, 076819, 076961, 077174, 077199, 077242, 078042, 078164, 078477, 079015, 079208, 079248, 080142, 080144, 080146, 080148, 080149, 080152, 080153, 080156, 080157, 080160, 080162, 080163, 080164, 080167, 080170, 080172, 080174, 080177, 080178, 080179, 080184, 080189, 080191, 080192, 080193, 080194, 080198, 080200, 080202, 080206, 080207, 080209, 080211, 080212, 080216, 080218, 080222, 080224, 080233, 080235, 080236, 080237, 080238, 080239, 080240, 080247, 080248, 080250, 080253, 080257, 080258, 080259, 080262, 080264, 080266, 080267, 080269, 080278, 080279, 080281, 080285, 080287, 080290, 080291; UPN 00802526564802, Serial Numbers: 051107, 051224, 051504, 051939, 051991, 052046, 052056, 052062, 052085, 052137, 052153, 052268, 052290, 052317, 052340, 052373, 052380, 052425, 052517, 052653, 052665, 052778, 052779, 052855, 052875, 052882, 052945, 052946, 052979, 052985, 053033, 053063, 053068, 053117, 053126, 053150, 053154, 053177, 053205, 053246, 053329, 053355, 053364, 053398, 053406, 053413, 053487, 053492, 053493, 053500, 053541, 053542, 053552, 053620, 053653, 053704, 053707, 053711, 053853, 053864, 053871, 053915, 053920, 053953, 053968, 053985, 054045, 054103, 054187, 054190, 054195, 054316, 054319, 054336, 054370, 054383, 054413, 054423, 054532, 054534, 054542, 054547, 054558, 054593, 054650, 054677, 054760, 054765, 054794, 054798, 054806, 054837, 054906, 054925, 054942, 055415, 057157, 057446, 058175, 059159, 060048, 060638, 060906, 061111, 061139, 061141, 061189, 061209, 061270, 061291, 061312, 061341, 061353, 061355, 061375, 061459, 061526, 061528, 061557, 061571, 061595, 061702, 061708, 061724, 061759, 062119, 062290, 062644, 062648, 062664, 062703, 062723, 062746, 062818, 062840, 062852, 062978, 063015, 063053, 063148, 063200, 063209, 063328, 063351, 063400, 064999, 065512, 065846, 066152, 066248, 066384, 066492, 066563, 066772, 066804, 067138, 067201, 067253, 067320, 067882, 067951, 067983, 068007, 068037, 068067, 068090, 069639, 069730, 070488, 070490, 071379, 071417, 071446, 071716, 072217, 072258, 072282, 072287, 072568, 072679, 072681, 072697, 072733, 072777, 072826, 072884, 072908, 073024, 073057, 073059, 073170, 073264, 074329, 074796, 075935, 075983, 076121, 076182, 076215, 076224, 076315, 076700, 076872, 076948, 076951, 077171, 077191, 077211, 077304, 077575, 077682, 077941, 077947, 077992, 078388, 078418, 078429, 078496, 078519, 078690, 078871, 079026, 079027, 079146, 079183, 079234, 079299, 079333, 079340, 079404, 079528, 079558, 079584, 079590, 079609, 079632, 079647, 079686, 079699, 079724, 079728, 079734, 079747, 079752, 079754, 079763, 079775, 079803, 079848, 079870, 079874, 080147, 080168, 080173, 080181, 080185, 080186, 080187, 080195, 080196, 080203, 080204, 080205, 080208, 080210, 080214, 080220, 080227, 080231, 080245, 080246, 080261, 080265, 080277, 080284, 080288, 080300, 080307, 080315, 080318, 080320, 080323, 080332, 080344, 080348, 080351, 080353, 080355, 080359, 080361, 080363, 080370, 080374, 080376, 080380, 080407, 080411, 080415, 080426, 080431, 080432, 080434, 080438, 080618, 080620, 080621, 080623, 080625, 080626, 080627, 080628, 080629, 080630, 080635, 080638, 080639, 080644, 080647, 080653, 080655, 080657, 080661, 080663, 080665, 080671, 080675, 080676, 080677, 080678, 080684, 080686, 080689, 080690, 080691, 080716, 080721, 080735, 080736, 080741, 080744, 080753, 080754, 080758, 080759, 080763, 080764, 080773, 080778, 080781, 080782, 080784, 080785, 080787, 080789, 080795, 080799, 080802, 080803, 080807, 080814; UPN 00802526564819, Serial Numbers: 052942, 053208, 064458, 069465, 076373, 078427, 078556, 078569, 078632, 078645, 078812, 078897, 078946, 080367, 080378, 080421, 080435, 080436; UPN 00802526565304, Serial Numbers: 053353, 059516, 062868, 062897, 065813, 065864, 066169, 066244, 066306, 066718, 068415, 068418, 068450, 068462, 070445, 072935, 078282, 078315, 078843, 079042, 080009, 080060, 080092; UPN 00802526565427, Serial Numbers: 056386, 059274, 063126, 063224, 063313, 065990, 066271, 066289, 066327, 066335, 066437, 067183, 067246, 068390, 068436, 072614, 072640, 073049, 075075, 078486, 078740, 080116; UPN 00802526574801, Serial Numbers: 051420, 051885, 052068, 052221, 052309, 052339, 052356, 052445, 052447, 052461, 052508, 052575, 052598, 052826, 052836, 052868, 052874, 052887, 052888, 052988, 053153, 053165, 053182, 053210, 053216, 053256, 053405, 053449, 053460, 053558, 053567, 053593, 053692, 053721, 053723, 053759, 053859, 053917, 053957, 053959, 053964, 053966, 053976, 054022, 054071, 054078, 054121, 054224, 054255, 054287, 054432, 054450, 054456, 054488, 054510, 054563, 054672, 054754, 054785, 054884, 054923, 054931, 054937, 055522, 056545, 056984, 057845, 058093, 058616, 060674, 060756, 061099, 061180, 061187, 061267, 061384, 061430, 061431, 061515, 061704, 061738, 061741, 061753, 061769, 061814, 062201, 062636, 062673, 062696, 062743, 062802, 062826, 062922, 062944, 062956, 062970, 063026, 063032, 063132, 063169, 063284, 063296, 063402, 063441, 064608, 065835, 065941, 066073, 066115, 066186, 066272, 066450, 066513, 066760, 066787, 066814, 067066, 067115, 067141, 067223, 067238, 067244, 067851, 067899, 067922, 068010, 068066, 068491, 069189, 069343, 069684, 069693, 069699, 069700, 069716, 070438, 070453, 070465, 070468, 070497, 070983, 071149, 071365, 071866, 072267, 072309, 072601, 072612, 072618, 072635, 072644, 072694, 072696, 072717, 072737, 072819, 072848, 072911, 072945, 073006, 073015, 073030, 073039, 073054, 073080, 073084, 073086, 073150, 073601, 074203, 074333, 074336, 074349, 074386, 074461, 074675, 076189, 076288, 076309, 076422, 077290, 077586, 077633, 077638, 077663, 077665, 077891, 077917, 078008, 078121, 078613, 078657, 078742, 078894, 078971, 079000, 079020, 079088, 079133, 079137, 079277, 079301, 079303, 079325, 079336, 079412, 079422, 079435, 079447, 079494, 079508, 079526, 079529, 079630, 079659, 079667, 079772, 080145, 080155, 080234, 080249, 080252, 080272, 080282, 080297, 080347, 080350, 080369, 080393, 080743, 080745, 080746, 080747, 080748, 080750, 080755, 080762, 080765, 080766, 080768, 080769, 080770, 080771, 080774, 080777, 080779, 080780, 080790, 080792, 080793, 080794, 080797, 080800, 080801, 080810, 080811, 080812, 080813, 080817, 080818, 080822, 080823, 080824, 080830, 080835; UPN 00802526574818, Serial Numbers: 074138, 076475, 076588, 078207; UPN 00802526579905, Serial Numbers: 051381, 051903, 052170, 052212, 052299, 052314, 052342, 052384, 052396, 052416, 052477, 052494, 052515, 052531, 052541, 052565, 052796, 052828, 052935, 052975, 053020, 053038, 053060, 053119, 053151, 053152, 053167, 053190, 053382, 053395, 053416, 053417, 053425, 053434, 053475, 053485, 053495, 053566, 053596, 053598, 053664, 053683, 053750, 053753, 053771, 053799, 053813, 053861, 053862, 053873, 053899, 053934, 053971, 053979, 053984, 053996, 054013, 054031, 054043, 054153, 054201, 054205, 054339, 054355, 054380, 054394, 054527, 054606, 054612, 054733, 054831, 054924, 056413, 056799, 056926, 057081, 057318, 058486, 059607, 059829, 060039, 060916, 060930, 061098, 061104, 061177, 061191, 061266, 061313, 061321, 061340, 061429, 061511, 061671, 061786, 061798, 062071, 062675, 062737, 062757, 062827, 062848, 062865, 062893, 063067, 063106, 063172, 063309, 063332, 063343, 063362, 063947, 065104, 065277, 065839, 065873, 066145, 066153, 066334, 066433, 066562, 066658, 066714, 066727, 067278, 067285, 067319, 067336, 067357, 067620, 067888, 067932, 067977, 068014, 068805, 069626, 070482, 071345, 071351, 071359, 071378, 071444, 071550, 071927, 071945, 072253, 072516, 072544, 072591, 072592, 072598, 072607, 072610, 072698, 072750, 072822, 072880, 072890, 072894, 072994, 073032, 073107, 073174, 073503, 074257, 076153, 076177, 076293, 076296, 076724, 076979, 077209, 077288, 077761, 077771, 077948, 077996, 078049, 078284, 078350, 078488, 078663, 078739, 078775, 078820, 078877, 078948, 079001, 079018, 079062, 079138, 079147, 079210, 079230, 079255, 079314, 079357, 079369, 079389, 079403, 079464, 079469, 079478, 080150, 080223, 080229, 080256, 080270, 080636, 080681, 080752, 080757, 080804, 080815, 080836, 080837, 080839, 080840, 080844, 080845, 080846, 080847, 080855, 080856, 080857, 080858, 080859, 080860, 080861, 080862, 080864, 080865, 080866, 080867, 080870, 080871, 080872, 080875, 080877, 080878, 080879, 080880, 080881, 080884, 080886, 080887, 080888, 080894, 080897, 080899, 080900, 080902, 080903, 080904, 080905, 080907, 080910, 080911, 080912, 080913, 080916, 080918, 080922, 080923, 080929, 080935, 080939, 080940, 080942, 080943, 080945, 080946, 080948, 080954, 080959, 080960, 080963, 080968, 080970, 080971, 080973, 080976, 080977, 080978, 080979, 080981, 080982, 080986, 080989, 080991, 080992, 080993, 080994, 080995, 080996, 080999, 081000, 081001, 081002, 081004, 081005, 081006, 081007, 081008, 081010, 081011, 081012, 081014, 081016, 081017, 081018, 081020, 081022, 081023, 081025, 081026, 081028, 081029, 081031, 081033, 081034, 081035, 081037, 081038, 081041, 081042, 081044, 081046, 081047, 081048, 081049, 081050, 081051, 081053, 081054, 081055, 081056, 081057, 081062, 081064, 081066, 081067, 081068, 081069, 081073, 081074, 081078, 081079, 081081, 081082, 081083, 081084, 081085, 081086, 081089, 081091, 081092, 081094, 081095, 081097, 081099, 081100, 081101, 081102, 081104, 081105, 081107, 081108, 081110, 081111, 081112, 081113, 081114, 081115, 081117, 081118, 081125, 081129, 081130, 081131, 081132, 081135, 081146, 081150, 081200, 081214, 081227, 081231, 081238, 081244, 081254, 081255, 081256, 081429, 081430, 081443; UPN 00802526579929, Serial Numbers: 060169, 061524, 062604, 062680, 062760, 062969, 063068, 063811, 065829, 065849, 065888, 065903, 066110, 066222, 068411, 068424, 068440, 071892, 072952, 074838, 074858, 074969, 074970, 074982, 078306, 080889, 080987, 080988, 080998, 081032, 081123, 081133, 081140, 081151; UPN 00802526581021, Serial Numbers: 051653, 053793, 078593, 080863, 080908, 081177, 081181, 081192, 081201, 081204, 081841; UPN 00802526582011, Serial Numbers: 051078, 051085, 051109, 051154, 051190, 051209, 051285, 051291, 051312, 051318, 051321, 051336, 051342, 051372, 051378, 051385, 051399, 051426, 051444, 051488, 051526, 051559, 051581, 051606, 051615, 051641, 051656, 051661, 051678, 051728, 051748, 051831, 051875, 051894, 051914, 051915, 051923, 051947, 051975, 051977, 051998, 052006, 052020, 052027, 052064, 052066, 052091, 052111, 052143, 052161, 052191, 052199, 052213, 052215, 052224, 052229, 052238, 052240, 052242, 052254, 052256, 052266, 052274, 052297, 052316, 052321, 052326, 052327, 052329, 052335, 052345, 052358, 052361, 052364, 052377, 052386, 052393, 052408, 052415, 052419, 052423, 052433, 052435, 052436, 052438, 052444, 052448, 052451, 052452, 052454, 052456, 052502, 052504, 052506, 052507, 052523, 052527, 052556, 052561, 052562, 052571, 052572, 052580, 052589, 052593, 052605, 052628, 052630, 052634, 052645, 052655, 052661, 052675, 052678, 052698, 052732, 052739, 052741, 052754, 052755, 052776, 052797, 052798, 052807, 052811, 052814, 052822, 052825, 052835, 052837, 052845, 052858, 052864, 052876, 052885, 052890, 052900, 052912, 052913, 052921, 052922, 052926, 052961, 052963, 052969, 052982, 052984, 052986, 053012, 053014, 053018, 053019, 053021, 053022, 053027, 053028, 053036, 053042, 053043, 053051, 053080, 053083, 053084, 053088, 053092, 053094, 053095, 053102, 053103, 053107, 053112, 053114, 053131, 053175, 053184, 053186, 053202, 053211, 053214, 053220, 053226, 053231, 053244, 053250, 053260, 053263, 053268, 053270, 053275, 053276, 053306, 053307, 053311, 053312, 053320, 053326, 053332, 053339, 053346, 053356, 053357, 053360, 053384, 053386, 053390, 053399, 053418, 053421, 053427, 053462, 053470, 053472, 053481, 053502, 053515, 053517, 053544, 053547, 053548, 053560, 053563, 053575, 053608, 053616, 053634, 053655, 053689, 053693, 053695, 053705, 053714, 053720, 053738, 053741, 053749, 053763, 053764, 053782, 053791, 053794, 053796, 053804, 053809, 053831, 053834, 053836, 053839, 053843, 053851, 053868, 053890, 053898, 053919, 053928, 053929, 053931, 053933, 053935, 053938, 053939, 053944, 053947, 053961, 053982, 053986, 053993, 054011, 054028, 054029, 054048, 054050, 054053, 054055, 054079, 054104, 054105, 054110, 054112, 054116, 054138, 054184, 054198, 054209, 054226, 054231, 054232, 054238, 054268, 054271, 054296, 054297, 054309, 054310, 054315, 054324, 054333, 054335, 054351, 054353, 054373, 054428, 054441, 054445, 054452, 054467, 054473, 054476, 054479, 054486, 054487, 054514, 054523, 054539, 054543, 054545, 054557, 054562, 054594, 054601, 054607, 054614, 054619, 054624, 054625, 054629, 054632, 054635, 054658, 054664, 054671, 054675, 054679, 054688, 054694, 054699, 054736, 054774, 054777, 054778, 054781, 054782, 054789, 054790, 054793, 054795, 054800, 054803, 054808, 054813, 054818, 054822, 054835, 054840, 054848, 054849, 054851, 054852, 054856, 054860, 054866, 054870, 054873, 054877, 054886, 054903, 054908, 054910, 054918, 054939, 054945, 054947, 054977, 054994, 055060, 055077, 055126, 055216, 055264, 055272, 055349, 055434, 055436, 055467, 055548, 055646, 055706, 055741, 055768, 055911, 055914, 056015, 056031, 056048, 056136, 056161, 056175, 056188, 056210, 056251, 056276, 056286, 056327, 056337, 056372, 056393, 056440, 056453, 056454, 056493, 056498, 056530, 056557, 056628, 056651, 056658, 056804, 056817, 056850, 056864, 056876, 056893, 056943, 057041, 057103, 057226, 057236, 057385, 057462, 057709, 057858, 057898, 057933, 057978, 057994, 058011, 058056, 058139, 058143, 058202, 058242, 058254, 058262, 058281, 058286, 058327, 058332, 058348, 058354, 058383, 058409, 058447, 058519, 058548, 058569, 058574, 058592, 058664, 058706, 058764, 058769, 058782, 058802, 058817, 058902, 058917, 059008, 059028, 059040, 059042, 059051, 059129, 059201, 059213, 059230, 059307, 059333, 059358, 059385, 059422, 059433, 059477, 059491, 059495, 059500, 059578, 059583, 059592, 059683, 059796, 059797, 059807, 059853, 059872, 059891, 059896, 059952, 060020, 060047, 060049, 060195, 060213, 060255, 060284, 060287, 060324, 060351, 060352, 060372, 060398, 060399, 060461, 060505, 060514, 060555, 060612, 060661, 060746, 060775, 060792, 060899, 060945, 060973, 061002, 061059, 061065, 061083, 061085, 061110, 061112, 061114, 061116, 061117, 061124, 061134, 061146, 061149, 061155, 061156, 061163, 061173, 061184, 061199, 061222, 061223, 061237, 061245, 061254, 061279, 061289, 061290, 061296, 061316, 061320, 061323, 061330, 061333, 061335, 061344, 061357, 061359, 061363, 061367, 061368, 061371, 061374, 061381, 061392, 061402, 061412, 061437, 061450, 061451, 061467, 061476, 061485, 061486, 061494, 061499, 061509, 061510, 061529, 061530, 061535, 061542, 061543, 061553, 061556, 061561, 061573, 061575, 061578, 061580, 061591, 061601, 061602, 061603, 061608, 061611, 061621, 061630, 061657, 061662, 061666, 061673, 061684, 061686, 061711, 061713, 061716, 061742, 061745, 061750, 061751, 061760, 061764, 061771, 061772, 061775, 061778, 061780, 061782, 061783, 061803, 061858, 061860, 061885, 061936, 061967, 062003, 062033, 062058, 062149, 062231, 062283, 062361, 062366, 062432, 062444, 062478, 062525, 062550, 062582, 062631, 062639, 062647, 062663, 062682, 062691, 062694, 062707, 062709, 062722, 062726, 062730, 062735, 062744, 062745, 062756, 062762, 062782, 062783, 062788, 062796, 062811, 062822, 062857, 062859, 062884, 062890, 062898, 062925, 062929, 062941, 062947, 063002, 063005, 063030, 063037, 063039, 063041, 063051, 063054, 063070, 063076, 063079, 063084, 063097, 063101, 063105, 063110, 063112, 063118, 063120, 063139, 063145, 063158, 063159, 063165, 063175, 063179, 063186, 063194, 063230, 063246, 063261, 063267, 063268, 063289, 063300, 063319, 063331, 063369, 063375, 063381, 063398, 063408, 063423, 063427, 063428, 063436, 063439, 063443, 063515, 063526, 063617, 063633, 063662, 063719, 063769, 063828, 063940, 064038, 064044, 064068, 064096, 064118, 064130, 064157, 064185, 064198, 064230, 064252, 064306, 064400, 064404, 064449, 064503, 064549, 064560, 064609, 064614, 064655, 064672, 064689, 064713, 064736, 064788, 064796, 064805, 064831, 064849, 064897, 064901, 064986, 065237, 065313, 065324, 065338, 065357, 065359, 065388, 065468, 065540, 065690, 065711, 065731, 065820, 065821, 065830, 065836, 065874, 065881, 065982, 066096, 066117, 066132, 066137, 066146, 066147, 066154, 066172, 066195, 066211, 066224, 066232, 066246, 066250, 066329, 066330, 066356, 066358, 066360, 066370, 066398, 066409, 066417, 066420, 066442, 066458, 066463, 066470, 066476, 066480, 066489, 066490, 066514, 066521, 066534, 066537, 066541, 066544, 066611, 066729, 066766, 066836, 067051, 067068, 067069, 067072, 067076, 067081, 067082, 067092, 067093, 067095, 067104, 067109, 067120, 067134, 067146, 067153, 067159, 067169, 067194, 067206, 067208, 067214, 067236, 067247, 067260, 067262, 067267, 067270, 067290, 067297, 067299, 067304, 067313, 067332, 067347, 067348, 067349, 067350, 067355, 067617, 067676, 067732, 067734, 067774, 067778, 067815, 067859, 067872, 067877, 067887, 067893, 067897, 067905, 067928, 067934, 067963, 067972, 067976, 067987, 067990, 068012, 068015, 068020, 068035, 068048, 068063, 068070, 068071, 068074, 068091, 068093, 068094, 068277, 068342, 068441, 068514, 068540, 068656, 068664, 068689, 068692, 068826, 068861, 068913, 069224, 069498, 069552, 069686, 069687, 069719, 069723, 069871, 069878, 070251, 070426, 070434, 070437, 070440, 070443, 070450, 070460, 070464, 070467, 070471, 070479, 070500, 070505, 070527, 070534, 070539, 070598, 070689, 070743, 070813, 070828, 071333, 071341, 071342, 071357, 071383, 071390, 071392, 071393, 071410, 071429, 071430, 071436, 071440, 071445, 071600, 071840, 071852, 071859, 071873, 071878, 071894, 071909, 071912, 071919, 071925, 071940, 071941, 071949, 072115, 072190, 072218, 072224, 072243, 072275, 072288, 072355, 072509, 072512, 072517, 072536, 072537, 072572, 072577, 072578, 072581, 072594, 072600, 072605, 072621, 072623, 072628, 072630, 072638, 072639, 072641, 072642, 072647, 072670, 072700, 072708, 072715, 072734, 072743, 072746, 072763, 072767, 072797, 072852, 072853, 072856, 072892, 072897, 072907, 072956, 072961, 072969, 072988, 072989, 072998, 073002, 073008, 073013, 073018, 073042, 073045, 073062, 073092, 073121, 073181, 073319, 073338, 073452, 073690, 073867, 073905, 073912, 073992, 074058, 074119, 074179, 074259, 074281, 074330, 074335, 074338, 074344, 074355, 074357, 074382, 074387, 074398, 074400, 074409, 074410, 074419, 074432, 074440, 074455, 074480, 074485, 074490, 074499, 074514, 074524, 074607, 074676, 074744, 074802, 074806, 074807, 075928, 075933, 075953, 075980, 076112, 076114, 076116, 076118, 076146, 076152, 076156, 076158, 076159, 076164, 076167, 076168, 076213, 076225, 076227, 076241, 076268, 076270, 076277, 076294, 076346, 076348, 076372, 076412, 076506, 076785, 076862, 076952, 076962, 076968, 077059, 077086, 077094, 077101, 077110, 077204, 077208, 077241, 077281, 077298, 077571, 077574, 077587, 077595, 077619, 077721, 077723, 077727, 077766, 077770, 077775, 077787, 077828, 077845, 077848, 077872, 077874, 077884, 077888, 077892, 077896, 077906, 077923, 077933, 077944, 077958, 077979, 078013, 078061, 078066, 078075, 078077, 078078, 078085, 078102, 078128, 078133, 078148, 078160, 078334, 078339, 078368, 078386, 078389, 078450, 078474, 078521, 078548, 078567, 078615, 078626, 078634, 078635, 078636, 078637, 078651, 078688, 078692, 078711, 078715, 078746, 078765, 078770, 078825, 078834, 078852, 078876, 078924, 078927, 078977, 078999, 079030, 079037, 079045, 079056, 079061, 079065, 079072, 079077, 079081, 079099, 079128, 079142, 079217, 079224, 079241, 079254, 079272, 079292, 079304, 079306, 079312, 079313, 079316, 079318, 079322, 079327, 079330, 079339, 079343, 079346, 079347, 079353, 079363, 079387, 079391, 079395, 079407, 079419, 079426, 079431, 079436, 079453, 079468, 079481, 079483, 079484, 079489, 079555, 079562, 079565, 079572, 079573, 079575, 079581, 079594, 079636, 079640, 079683, 079689, 079700, 079705, 079708, 079713, 079714, 079720, 079770, 079782, 079800, 079855, 079883, 079895, 079908, 079922, 079930, 080159, 080176, 080180, 080190, 080215, 080228, 080230, 080241, 080242, 080251, 080260, 080263, 080273, 080275, 080283, 080286, 080289, 080298, 080309, 080317, 080328, 080354, 080400, 080401, 080428, 080637, 080642, 080680, 080696, 080738, 080749, 080761, 080767, 080788, 080791, 080796, 080798, 080805, 080808, 080816, 080819, 080826, 080827, 080829, 080831, 080833, 080834, 080841, 080843, 080848, 080849, 080852, 080869, 080874, 080893, 080896, 080898, 080901, 080906, 080917, 080919, 080921, 080924, 080925, 080926, 080928, 080930, 080932, 080933, 080938, 080941, 080944, 080947, 080949, 080953, 080956, 080962, 080966, 080967, 080975, 080980, 080985, 081013, 081024, 081030, 081040, 081052, 081076, 081087, 081098, 081103, 081109, 081124, 081241, 081327, 081333, 081379, 081402, 081408, 081409, 081424, 081444, 081465, 081512, 081527, 081528, 081534, 081535, 081544, 081554, 081558, 081560, 081563, 081564, 081565, 081567, 081572, 081578, 081580, 081583, 081589, 081593, 081594, 081601; UPN 00802526582028, Serial Numbers: 051664, 051956, 052324, 052526, 052692, 053156, 053189, 053635, 054117, 054135, 054914, 055297, 056123, 056643, 056667, 056717, 058496, 059708, 060092, 060120, 060202, 060376, 060391, 060452, 060592, 060627, 060655, 060780, 062046, 062311, 062577, 063803, 064219, 064617, 065936, 066618, 066680, 066861, 066916, 066965, 067688, 068590, 068728, 068922, 069115, 070336, 070901, 071624, 072649, 072929, 073619, 073638, 073935, 073940, 073944, 074025, 074029, 074114, 074246, 075124, 075936, 075976, 076358, 076392, 076396, 076398, 076424, 076437, 076515, 076658, 077254, 077655, 077675, 077754, 077810, 077843, 078088, 078091, 078096, 078124, 078156, 078171, 078173, 078202, 078215, 078239, 078259, 078261, 078878, 079196, 079235, 079249, 079253, 079271, 079283, 079355, 079375, 079459, 079593, 079665, 079696, 079711, 079762, 079977, 079988, 080326, 080362, 080952, 081195, 081220, 081416, 081433, 081439, 081699, 081706, 081835, 081999; UPN 00802526582301, Serial Numbers: 059192, 062900, 063006, 063359, 063558, 063663, 064134, 066268, 066274, 067091, 067122, 067252, 067302, 068414, 069434, 071905, 072207, 072624, 075657, 076987, 078501, 078508, 078511, 078513, 078755, 080029, 080050, 080125, 080130; UPN 00802526583803, Serial Numbers: 051730, 052280, 053089, 053488, 054546, 055012, 055164, 055465, 055493, 055532, 055534, 055554, 056354, 056373, 056596, 056842, 056995, 057031, 057142, 057430, 057731, 058183, 060219, 060279, 060400, 061126, 061495, 061872, 061903, 062051, 062184, 062256, 062749, 062867, 062916, 062927, 062975, 063184, 063864, 063900, 064076, 065834, 065847, 065852, 065860, 065866, 065867, 065871, 066158, 066168, 066241, 066256, 066295, 066320, 066323, 066336, 066702, 066773, 067839, 067917, 068382, 068392, 068393, 068413, 068430, 068442, 068445, 068460, 070572, 071903, 072646, 073581, 074449, 074857, 074956, 077150, 078299, 078300, 078348, 078358, 078531, 078850, 078855, 079207, 079718, 080093, 080099, 080104, 080132, 080895, 081096, 081365, 081367, 081375, 081384, 081400, 081410, 081413, 081545, 081557, 081586, 081590, 081603, 081643, 081652, 081660, 081665; UPN 00802526587405, Serial Numbers: 057576, 058943, 061255, 062844, 064468, 066278, 066315, 066745, 067176, 067857, 068374, 071893, 071897, 072179, 072199, 072278, 072495, 072771, 075094, 075101, 078757, 078759, 078780, 079943, 080121; UPN 00802526602917, Serial Numbers: 051128, 051149, 051292, 051635, 051713, 051801, 051911, 051925, 051946, 051950, 052001, 052003, 052032, 052061, 052069, 052071, 052092, 052174, 052218, 052226, 052313, 052420, 052432, 052437, 052457, 052559, 052564, 052569, 052631, 052651, 052687, 052704, 052723, 052730, 052737, 052764, 052765, 052780, 052930, 052989, 053005, 053100, 053254, 053287, 053302, 053344, 053371, 053389, 053426, 053473, 053479, 053607, 053670, 053732, 053737, 053746, 053780, 053785, 053817, 053881, 053885, 053907, 053912, 053932, 054003, 054012, 054073, 054080, 054089, 054145, 054246, 054250, 054278, 054285, 054314, 054360, 054499, 054508, 054509, 054526, 054548, 054556, 054604, 054613, 054617, 054634, 054668, 054707, 054797, 054915, 054933, 055378, 055585, 056101, 056233, 056270, 058777, 059177, 060670, 060861, 061084, 061167, 061182, 061208, 061218, 061219, 061249, 061252, 061322, 061346, 061377, 061380, 061397, 061408, 061447, 061516, 061538, 061586, 061599, 061610, 061648, 061659, 061663, 061710, 061722, 061725, 061761, 061788, 061790, 061884, 062619, 062656, 062662, 062712, 062715, 062740, 062741, 062748, 062843, 062858, 062866, 062896, 062953, 063020, 063111, 063213, 063275, 063279, 063356, 063392, 063425, 064200, 064207, 064323, 064646, 064982, 065175, 065797, 065854, 065900, 066122, 066182, 066212, 066218, 066419, 066446, 066448, 066891, 067116, 067125, 067132, 067163, 067165, 067181, 067284, 067854, 067944, 067994, 068013, 068039, 068527, 069705, 069735, 070323, 071354, 071374, 071911, 072200, 072284, 072352, 072555, 072658, 072666, 072751, 072774, 072780, 072790, 072921, 073012, 073158, 073274, 074206, 074273, 074326, 074356, 074417, 074492, 074686, 074764, 074787, 074793, 075929, 076160, 076172, 076176, 076196, 076838, 076848, 076889, 076917, 076933, 077109, 077218, 077325, 077328, 077634, 077692, 077964, 077967, 078011, 078076, 078566, 078705, 078718, 078793, 078862, 078981, 078997, 079118, 079151, 079310, 079331, 079366, 079553, 079559, 079588, 079679, 079887, 080165, 080213, 080221, 080226, 080254, 080312, 080419, 080876, 081015, 081121, 081340, 081350, 081351, 081354, 081385, 081388, 081390, 081392, 081415, 081417, 081418, 081426, 081427, 081428, 081451, 081452, 081454, 081458, 081467, 081477, 081480, 081510, 081668, 081669, 081670, 081672, 081676, 081681, 081690, 081692, 081693; UPN 00802526602924, Serial Numbers: 063021, 066259, 066372, 067130, 068383, 068435, 072196, 073156, 073417, 076758, 078512; UPN 00802526602931, Serial Numbers: 056230, 060860, 061631, 062131, 062224, 063092, 063373, 063374, 063787, 063816, 064585, 065814, 065819, 065879, 065891, 065895, 066347, 066391, 068389, 068433, 070424, 070499, 071428, 071883, 072996, 074916, 077077, 077154, 078815, 079203; UPN 00802526605406, Serial Numbers: 051787, 052134, 052171, 052207, 052338, 052581, 052819, 052906, 052955, 053015, 053052, 053236, 053262, 053397, 053521, 053586, 053626, 053679, 053687, 053774, 053811, 054118, 054163, 054248, 054560, 054580, 054660, 054726, 054952, 061347, 061592, 062701, 063171, 063444, 065822, 067180, 067186, 067209, 067841, 067950, 067996, 068016, 068097, 070476, 071325, 071389, 072203, 072542, 072590, 072652, 074350, 075959, 076211, 076926, 077175, 079352, 080832, 080920, 080927, 080950, 081827, 081831, 081852, 081854, 081856, 081866, 081867, 081869, 081871, 081872, 081879, 081881, 081883, 081885, 081886, 081889, 081890, 081892, 081894, 081895, 081899, 081904, 081908, 081913, 081919, 081920, 081924; UPN 00802526605437, Serial Numbers: 067488, 067609, 070279, 072747, 073879, 073965, 073967, 074024, 074194, 076384, 076444, 077294, 077672, 077742, 078019, 078084, 078196, 078279, 079165, 079270, 079284, 079605, 079610, 079712, 081186; UPN 00802526605468, Serial Numbers: 062063, 063014, 063044, 063889, 068438, 069231, 076564; UPN 00802526606304, Serial Numbers: 051343, 051694, 051967, 054579, 054986, 054997, 055082, 057551, 058017, 058468, 059490, 059660, 059820, 060144, 061034, 061154, 061625, 061976, 062126, 062608, 062666, 062688, 062885, 062912, 062920, 062939, 063035, 063995, 064041, 065620, 065810, 065825, 065833, 065840, 065843, 065856, 065861, 065862, 065885, 065886, 065890, 066162, 066170, 066235, 066291, 066294, 066296, 066324, 066348, 066387, 066392, 066445, 068386, 068404, 068406, 068412, 068421, 071881, 072499, 072922, 072966, 072979, 074361, 074782, 074783, 074828, 074850, 074926, 074961, 075264, 075586, 077107, 078217, 079157, 079205, 079342, 080006, 080066, 080117, 080957, 081019, 081144, 081182, 081595; UPN 00802526606311, Serial Numbers: 051685, 052059, 053340, 053757, 053847, 054100, 054257, 054259, 054406, 054507, 055558, 056080, 056154, 057088, 057225, 058021, 059203, 059756, 060289, 060411, 060450, 060664, 064748, 065017, 065696, 066959, 068318, 071764, 072360, 072745, 073630, 073645, 073649, 073997, 074017, 074043, 074052, 074219, 074225, 076362, 076388, 076405, 076425, 076480, 076610, 077273, 077313, 077605, 077636, 077643, 077735, 078056, 078105, 078119, 078174, 078189, 078191, 078206, 078251, 078260, 078401, 078540, 078846, 078887, 078915, 078962, 079261, 079376, 079547, 079606, 079631, 079674, 079693, 079731, 080882, 080969, 081139, 081145, 081170, 081194, 081210, 081225; UPN 00802526606342, Serial Numbers: 051157, 051166, 051234, 051259, 051298, 051317, 051373, 051380, 051405, 051589, 051600, 051920, 051927, 051932, 051936, 051986, 052002, 052007, 052029, 052040, 052055, 052090, 052129, 052133, 052148, 052188, 052190, 052192, 052194, 052208, 052255, 052258, 052276, 052278, 052281, 052296, 052307, 052378, 052411, 052450, 052480, 052491, 052529, 052536, 052540, 052545, 052546, 052582, 052584, 052587, 052613, 052614, 052626, 052636, 052640, 052682, 052689, 052702, 052705, 052711, 052735, 052740, 052753, 052766, 052771, 052773, 052784, 052809, 052812, 052838, 052839, 052850, 052872, 052879, 052897, 052916, 052931, 052968, 052993, 052997, 053004, 053026, 053029, 053030, 053054, 053056, 053057, 053061, 053071, 053101, 053115, 053148, 053164, 053172, 053204, 053217, 053225, 053235, 053264, 053273, 053286, 053289, 053293, 053298, 053376, 053412, 053447, 053499, 053506, 053530, 053534, 053561, 053564, 053579, 053605, 053640, 053658, 053850, 053889, 053901, 053903, 053936, 053950, 053952, 053975, 054032, 054038, 054039, 054066, 054136, 054139, 054164, 054206, 054241, 054243, 054252, 054289, 054313, 054327, 054330, 054348, 054349, 054376, 054419, 054420, 054462, 054517, 054530, 054533, 054535, 054541, 054565, 054605, 054620, 054626, 054630, 054644, 054654, 054727, 054737, 054812, 054839, 054842, 054874, 054880, 054889, 054892, 054893, 054900, 055008, 055018, 055081, 055257, 055281, 055438, 055713, 055753, 055774, 055881, 055966, 056134, 056346, 056438, 056500, 056521, 056536, 056607, 056708, 057159, 057183, 057223, 057503, 057512, 057771, 057849, 058013, 058071, 058214, 058269, 058307, 058421, 058749, 058838, 058899, 058907, 059476, 059604, 059619, 059749, 059819, 060847, 061066, 061075, 061080, 061088, 061137, 061140, 061162, 061168, 061228, 061230, 061235, 061259, 061280, 061364, 061373, 061390, 061425, 061454, 061471, 061473, 061497, 061502, 061507, 061521, 061525, 061547, 061555, 061559, 061583, 061584, 061639, 061653, 061674, 061691, 061699, 061705, 061774, 061784, 061789, 061801, 061806, 061893, 061898, 061934, 062173, 062212, 062226, 062390, 062414, 062681, 062683, 062695, 062717, 062765, 062767, 062792, 062831, 062889, 062905, 062942, 062957, 062964, 062971, 062981, 062989, 063024, 063034, 063135, 063143, 063144, 063150, 063183, 063227, 063254, 063262, 063297, 063311, 063323, 063329, 063353, 063363, 063394, 063433, 063451, 063452, 063586, 063682, 064172, 064559, 064694, 064797, 064892, 065204, 065712, 065858, 065870, 066120, 066149, 066208, 066231, 066236, 066344, 066377, 066380, 066418, 066434, 066447, 066468, 066578, 066683, 066689, 066707, 066755, 066808, 066989, 066991, 067033, 067065, 067085, 067097, 067103, 067117, 067133, 067178, 067226, 067230, 067256, 067258, 067265, 067291, 067333, 067354, 067356, 067834, 067838, 067858, 067867, 067903, 067906, 067923, 067952, 067967, 067991, 067993, 068000, 068011, 068056, 068058, 068203, 068215, 068366, 068500, 068925, 068926, 069172, 069174, 069524, 069737, 069820, 070274, 070400, 070407, 070454, 070509, 070521, 070525, 070684, 070883, 071010, 071349, 071370, 071400, 071402, 071439, 071475, 071709, 071845, 071860, 071862, 071867, 071870, 071906, 071928, 071937, 071950, 071959, 072009, 072184, 072204, 072210, 072221, 072241, 072260, 072264, 072274, 072477, 072543, 072549, 072567, 072583, 072705, 072716, 072726, 072735, 072766, 072783, 072786, 072791, 072811, 072829, 072847, 072860, 072885, 072906, 072941, 072993, 073089, 073102, 073105, 073114, 073131, 073447, 073458, 073491, 073509, 073521, 073894, 073925, 073961, 073964, 074235, 074287, 074298, 074309, 074399, 074405, 074411, 074418, 074420, 074466, 074517, 074518, 074527, 074579, 074683, 075318, 075509, 075677, 075924, 075949, 075975, 076115, 076130, 076154, 076179, 076181, 076190, 076195, 076206, 076231, 076244, 076284, 076305, 076311, 076327, 076395, 076430, 076710, 076746, 076811, 076863, 076867, 076909, 076928, 076932, 076954, 076970, 077044, 077079, 077147, 077165, 077194, 077275, 077297, 077310, 077498, 077541, 077691, 077762, 077798, 077823, 077855, 077893, 077908, 077945, 077954, 077957, 077959, 077962, 078010, 078054, 078336, 078356, 078370, 078472, 078479, 078539, 078604, 078617, 078656, 078676, 078708, 078709, 078719, 078743, 078898, 078933, 078938, 079036, 079050, 079082, 079131, 079153, 079193, 079268, 079294, 079319, 079442, 079455, 079511, 079666, 079676, 079959, 079996, 080154, 080169, 080183, 080217, 080232, 080243, 080338, 080395, 080685, 080756, 080776, 080783, 080806, 080820, 080955, 080965, 080974, 080984, 081003, 081065, 081122, 081268, 081381, 081459, 081463, 081464, 081466, 081471, 081475, 081478, 081486, 081868, 081888, 081902; UPN 00802526606359, Serial Numbers: 053983, 055249, 055905, 056775, 059171, 062778, 063468, 064725, 065920, 066404, 066849, 067416, 067844, 067855, 067969, 068085, 068087, 068380, 068381, 068395, 068397, 068401, 068419, 068452, 068459, 068596, 069817, 071962, 072183, 072197, 072233, 072251, 072265, 072478, 075496, 075655, 076073, 076749, 076983, 078555, 078619, 078728, 080048, 080074, 080083; UPN 00802526611308, Serial Numbers: 051294, 051306, 052049, 052230, 052234, 052237, 052259, 052292, 052332, 052355, 052365, 052385, 052391, 052417, 052418, 052427, 052449, 052492, 052505, 052539, 052579, 052585, 052623, 052642, 052658, 052729, 052758, 052774, 052844, 052943, 052954, 052971, 053064, 053086, 053087, 053099, 053170, 053228, 053437, 053443, 053444, 053452, 053508, 053573, 053585, 053642, 053649, 053660, 053715, 053819, 053893, 053905, 054072, 054120, 054127, 054131, 054170, 054179, 054204, 054279, 054306, 054325, 054417, 054424, 054518, 054597, 054647, 054681, 054693, 054758, 054761, 054799, 054891, 054895, 054916, 054927, 055338, 055831, 056061, 056085, 057569, 058280, 058981, 059547, 059608, 059638, 060162, 060767, 061086, 061174, 061179, 061243, 061301, 061308, 061338, 061403, 061453, 061492, 061554, 061558, 061587, 061606, 061614, 061652, 061721, 061794, 061954, 062468, 062720, 062779, 062819, 062835, 062873, 062903, 063055, 063071, 063073, 063122, 063151, 063157, 063193, 063240, 063250, 063291, 063310, 063337, 063338, 063449, 063576, 064670, 065848, 066008, 066104, 066143, 066225, 066227, 066273, 066456, 066512, 066533, 066659, 066807, 067087, 067207, 067249, 067317, 067358, 067801, 067997, 068004, 068077, 068909, 068954, 069207, 069683, 069696, 070433, 070462, 070519, 070533, 070710, 071409, 071531, 071876, 071879, 072177, 072189, 072191, 072192, 072286, 072548, 072582, 072683, 072722, 072778, 072832, 072881, 072933, 072972, 073142, 073147, 073987, 074169, 074375, 074381, 074385, 074431, 074437, 074484, 074557, 074604, 075961, 075967, 076145, 076148, 076150, 076166, 076169, 076188, 076200, 076246, 076256, 076272, 076380, 076415, 076827, 076946, 076963, 076994, 077084, 077193, 077284, 077568, 077577, 077623, 077640, 077670, 077795, 077879, 077989, 077997, 078007, 078125, 078163, 078337, 078484, 078545, 078553, 078638, 078673, 078767, 078819, 079073, 079398, 079440, 079557, 079585, 080175, 080199, 080274, 080292, 080640, 080646, 080683, 080688, 080725, 080726, 080838, 080934, 080972, 081058, 081061; UPN 00802526612411, Serial Numbers: 051470, 051773, 051800, 051972, 054729, 054802, 055217, 055305, 056187, 056831, 057645, 061881, 062553, 063379, 065805, 065842, 066184, 066204, 066213, 066239, 066283, 066293, 066332, 066438, 066609, 068047, 068391, 068444, 071849, 074486, 074851, 074913, 075699, 076918, 078359, 080101, 081382; UPN 00802526612428, Serial Numbers: 052928, 061848, 064904, 070936, 073943, 077317, 077877, 079710, 081160, 081169, 081345, 081348, 081349, 082018; UPN 00802526612442, Serial Numbers: 051855, 053074, 057442, 059775, 064186, 065590, 066221, 066288, 066371, 066379, 066386, 066402, 066454, 067301, 067323, 067343, 067846, 068378, 068385, 068428, 069799, 071913, 071948, 072213, 072219, 072235, 072276, 072277, 072685, 072903, 075607, 075668, 075686, 075995, 076638, 077051, 078487, 078563, 078658, 078721, 078747, 078768, 080040, 080138; UPN 00802526615603, Serial Numbers: 053453, 079460; UPN 00802526615627, Serial Numbers: 051995, 056292, 056436, 060983, 061596, 062814, 062913, 062918, 063095, 063316, 065865, 066163, 066263, 066300, 066322, 066325, 066700, 068398, 068447, 068457, 068461, 072186, 074863, 077213, 078412, 081088, 081134, 081394; UPN 00802526615634, Serial Numbers: 053149, 053699, 054282, 054301, 062325, 069803, 071935, 072255, 072343, 072889, 073641, 075683, 076526, 078117, 078222, 078240, 079206, 079279, 080077, 080406, 081163.

Devices named in this recall

9 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

45 devices cleared under P960040 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 90524).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page