ZOOM® LATITUDE® 3120 (UDI 00802526575303)Recalls & reported problems
Boston Scientific Corporation · programmer/recorder/monitor · Class 3
All models of ZOOM LATITUDE (9)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 4
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00802526575303), received between May 6, 2019 and Jun 3, 2022 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Defective Device” (3 reports).
4 FDA recall records name this device's identifier, 4 still open.
8 other recalls cover devices cleared under the same FDA submission (P910077); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Nov 30, 2023Class IIOpen, Classified
The software application on the Programmer/Recorder/Monitor (PRM) for Implantable Pulse Generators will display certain diagnostic dates incorrectly, with a year in the 1990s.
Z-0673-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jul 7, 2022Class IIOpen, Classified
There is potential for unanticipated software anomalies to occur when using the Model 2868 Application loaded on the 3120 ZOOM LATITUTDE programmer when communicating with certain Boston Scientific transvenous defibrillators.
Z-1670-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jul 7, 2022Class IIOpen, Classified
There is potential for unanticipated software anomalies to occur when using the Model 2868 Application loaded on the 3120 ZOOM LATITUTDE programmer when communicating with certain Boston Scientific transvenous defibrillators.
Z-1671-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jul 7, 2022Class IIOpen, Classified
There is potential for unanticipated software anomalies to occur when using the Model 2868 Application loaded on the 3120 ZOOM LATITUTDE programmer when communicating with certain Boston Scientific transvenous defibrillators.
Z-1669-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Defective Device3
- Material Integrity Problem1
Patient problems
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
- Syncope/Fainting1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJun 3, 2022Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT WHILE PERFORMING A THREHSOLD TEST THE SCREEN OF THIS PROGRAMMER FROZE PREVENTING THE HALTING OF THIS TEST. THE PATIENT WAS DEPENDENT AND DUE TO LOSS OF CAPTURE (LOC), THE PATIENT LOST CONSCIOUSNESS. THE PROGRAMMER WAS REMOVED FROM SERVICE. TECHNICAL SERVICES (TS) EDUCATED THE CALLER THAT HAD THE PROGRAMMER WAND BEEN REMOVED FROM THE PATIENT THIS TOO WOULD HAVE ENDED THE THRESHOLD TEST.
Patient problems: Syncope/Fainting
MAUDE report 14595160 ↗MalfunctionFeb 4, 2020Defective Device; Defective Device
IT WAS REPORTED THAT THIS PROGRAMMER HEATED UP AND MELTED. THE PROGRAMMER WAS RETURNED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 9667947 ↗MalfunctionJan 31, 2020Defective Device; Defective Device
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PROGRAMMER HAD A POPPING SOUND WHEN IT TURNED ON AND IT HAD SMOKE. THIS PROGRAMMER WILL BE RETURNED FOR ANALYSIS. NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9653794 ↗MalfunctionMay 6, 2019Defective Device; Defective Device
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PROGRAMMER STARTED SMOKING AND SMELLED BURNT UPON INTERROGATING A PATIENT. THE FIELD REPRESENTATIVE WILL RETURN THE PROGRAMMER BACK. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8581825 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P910077. One submission often covers many models, so check the product and lot details in each record.
- Z-1216-2023 · Feb 2, 2023 · Open, ClassifiedLATITUDE NXT Remote Patient Management System, LATITUDE NXT System Server Software Model 6460.
- Z-3260-2018 · Sep 10, 2018 · TerminatedModel 3300 LATITUDE(TM) Programming System Product Usage - The Boston Scientific LATITUDE Programming System is intended for use in hospital and clinical environments to communic
- Z-1445-2012 · Feb 24, 2012 · TerminatedBoston Scientific LATITUDE Patient Management System Communicator, Model 6476, Distributed by Boston Scientific, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, Made in Malaysi
- Z-1096-2012 · Dec 19, 2011 · TerminatedLATITUDE Patient management System, Model 6488, software version 7.2. It is indicated for use to remotely communicate with a compatible Boston Scientific pulse generator and transf
- Z-0333-2012 · Oct 12, 2011 · TerminatedLATITUDE Patient Management System, Models 6441*, 6442*, 6465*, 6488 (USA). There are two software applications: the programmer software and the LATITUDE remote monitoring system
- Z-0037-2012 · Aug 12, 2011 · TerminatedLATITUDE Paceart Integration Software, version 1.00 Model 6472. LATITUDE Paceart Integration is software and therefore does not have any serial numbers. However, the software ver
- Z-1631-2010 · Sep 10, 2009 · TerminatedBoston Scientific, LATITUDE Communicator, Distributed by Boston Scientific, 4100 Hamline Avenue North, St. Paul, MN 55112. Models 6476 (USA) and 6468 (OUS)
- Z-0002-2007 · May 19, 2006 · TerminatedGuidant Zoom Latitude Programming System, Model 3120, Programmer/Recorder/Monitor (PRM). A portable cardiac rhythm management system designed to be used with Guidant implantable p
Other Implantable cardioverter defibrillator (non-CRT) devices
- EMBLEM™ MRI S-ICD A219Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)2 recalls36 deaths15,404 reports
- EMBLEM™ S-ICD A209Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)1 recall7 deaths6,788 reports
- EMBLEM™ S-ICD 3501Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)7 deaths4,936 reports
- Cobalt™ XT DR MRI SureScan™ DDPA2D4Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)3 recalls30 deaths1,603 reports
- ENDOTAK RELIANCE® G 0185Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,632 reports
- VIGILANT™ EL ICD DR D233Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,563 reports
- Evera MRI™ XT DR SureScan® DDMB1D4Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)1 recall35 deaths1,426 reports
- Durata™ 7122St. Jude Medical, Inc. · Implantable cardioverter defibrillator (non-CRT)27 deaths1,377 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page