Skip to content

N/A ASY-00133Recalls & reported problems

Integrity Implants Inc. · Repeater Bone Funnel · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00810004721890), received between Jan 10, 2025 and Jan 10, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Use of Device Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem1

Patient problems

  • Numbness1
  • Paresis1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 10, 2025Use of Device Problem; Use of Device Problem

    ON (B)(6), ACCELUS WAS MADE AWARE OF AN EVENT WHEN DR. (B)(6) INFORMED US THAT THEY WERE NEEDING TO PERFORM SURGICAL INTERVENTION FROM A CASE ON (B)(6)2024. THE PATIENT HAD A TLIF USING FLAREHAWK7. THE SURGEON SAID THEY WERE NEEDING TO REVISE THE CASE DUE TO L5 PARESIS. HE SAID THAT BONE CHIPS WERE BLOCKING THE FORAMEN. HE REPORTED HAVING TROUBLE WITH THE BONE FUNNEL BECAUSE THE BONE GRAFT HAD BIG CHUNKS SO THEY GOT STUCK IN THE FH7 BONE FUNNEL. THUS, THE COMPLAINT IS ALLEGING AN ISSUE WITH GRAFT DELIVERY THROUGH THE TAMP AND FUNNEL AND THERE WAS NO ALLEGED ISSUE WITH THE IMPLANTS FROM THE CASE. ON (B)(6), ADDITIONAL INFORMATION WAS PROVIDED AFTER SPEAKING WITH DR. (B)(6), HE REPORTED THAT THE PATIENT'S REVISION WENT FINE AND THE PATIENT WAS DOING WELL. NO ADDITIONAL INFORMATION WAS ABLE TO BE PROVIDED FOR THE DEVICES INVOLVED. AT THE REVISION, THEY TOOK ALL THE BONE GRAFT THAT WAS MISPLACED. PATIENT HAD MUCH BETTER CONTROL YESTERDAY BUT STILL WITH NUMBNESS IN A TOLERABLE RADICULAR PAIN (IN REGRESSION).

    Patient problems: Numbness; Paresis

    MAUDE report 21123962 ↗