Skip to content

iFuse TORQ Implant System 11555TRecalls & reported problems

Si-Bone Inc. · 11.5 mm x 55 mm iFuse-TORQ · Class 2 · Implant

All models of iFuse TORQ Implant System (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00810055520312), received between Oct 21, 2021 and Oct 21, 2021 — classified as 1 injury.

The most frequently reported specific device problem is “Malposition of Device” (1 report).

1 other recall cover devices cleared under the same FDA submission (K203247); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Malposition of Device1

Patient problems

  • Undesired Nerve Stimulation1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 21, 2021Malposition of Device; Malposition of Device

    THE PATIENT HAD RIGHT SIDE SI JOINT ARTHRODESIS IN (B)(6) 2021 WHERE THREE IMPLANTS WERE INSTALLED. THE PATIENT COMPLAINED OF PAIN IMMEDIATELY AFTER THE INITIAL SURGERY. THE SURGEON DETERMINED THAT THE SUPERIOR POSITIONED IMPLANT HAD BREACHED THE NEUROFORAMEN CAUSING THE PAIN. IN (B)(6) 2021, THE SURGEON PERFORMED A REVISION PROCEDURE WHERE THE SUPERIOR POSITIONED IMPLANT WAS REMOVED. NO OTHER PREEXISTING IMPLANTS WERE ADJUSTED OR REMOVED. THE STATUS OF THE PATIENT FOLLOWING THE REVISION PROCEDURE IS NOT KNOWN.

    Patient problems: Undesired Nerve Stimulation

    MAUDE report 12674689 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K203247. One submission often covers many models, so check the product and lot details in each record.

  • Z-0146-2027 · Aug 21, 2026 · Open, Classified
    iFuse TORQ Implant System Model/Catalog Number: 10055T(10x55mm), 10060T(10x60mm), 10065T(10x65mm), 10080T(10x55mm), 11580T(11.5x80mm)

Other Sacroiliac Joint Fixation devices

All Sacroiliac Joint Fixation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page