APEX HOY-ADVANCE-ERecalls & reported problems
Joerns Healthcare, LLC · HOYER ADVANCE LIFTER · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00810055880041), received between Mar 5, 2025 and Aug 29, 2025 — classified as 6 injuries.
The most frequently reported specific device problem is “Unintended Movement” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Unintended Movement2
- Appropriate Device Problem Term/Code Not Available1
- Collapse1
- Insufficient Device Problem Information1
- Unintended Collision1
Patient problems
- Fall4
- Abrasion1
- Bruise/Contusion1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryAug 29, 2025Collapse; Collapse
IT WAS REPORTED TO THE MANUFACTURER, BY THE END USER, PER THE END USER, THAT THE AID WAS PUTTING THE PATIENT TO BED, THE SLING WAS ABOUT 4 FT HIGH. THE SID CLOSED LEGS TO UNDER THE BED, THE LEG COLLAPSED, AND THE PATIENT FELL 4 FT BREAKING HER FEMUR. SHE HAD SURGERY AND 3 PINS PLACED. COMPLAINT AND RA#(B)(4) WAS ENTERED INTO OUR SYSTEM TO HAVE THE LIFT RETURNED FOR INVESTIGATION. AS OF THIS WRITING, THE LIFT HAS NOT BEEN RETURNED.
Patient problems: Fall
MAUDE report 22920161 ↗InjuryAug 22, 2025Unintended Movement; Unintended Movement
IT WAS REPORTED TO THE MANUFACTURER, BY THE END USER, PER THE END USER, THAT CUSTOMER CALLED AND SAID PIECE OF LIFT BROKE OFF. MY LIFT IS ONLY 3 MONTHS OLD AND THE MAST BOLT SHEARED CAUSING MY WIFE TO FALL. SHE WEIGHS 170 LBS. WHICH IS WAY BELOW CAPACITY. I HAD TO GET A BOLT FROM HOME DEPOT TO MAKE IT OPERATIONAL BECAUSE I USE IT DAILY. I NEED A REPLACEMENT BECAUSE I AM AFRAID, SHE MAY HAVE ANOTHER FALL. HE SENT IN PICS SHOWING A BENT MAST TIGHTENING KNOB AND SHEERED METAL ON THE MAST OF THIS LIFT UPDATED INFORMATION ON THE INCIDENT PUT IN THE FALL NOTES ABOVE. MEDICAL SUPPLIES NOW HAD ALREADY SENT A REPLACEMENT NOTIFICATION IS NOW COMPLETE. END USER WAS NOT CONSISTENT IN HIS ORIGINAL REPORT FOR THE INCIDENT AS NOTED ABOVE. COMPLAINT # (B)(4) AND RA #(B)(4) WERE ENTERED INTO OUR SYSTEM TO HAVE THE LIFT RETURNED FOR INVESTIGATION.
Patient problems: Fall
MAUDE report 22880126 ↗InjuryJul 28, 2025Unintended Collision; Unintended Collision
IT WAS REPORTED TO THE MANUFACTURER, BY THE END USER, PER THE END USER, THAT REPORTING THE SPREADER BAR TRANSPORT BASKET APPEARS TO BE HANGING ON THE BOLT OR PIN TOO LOOSE, AND ALSO WILL NOT HOLD THE SPREADER BAR WHEN PLACED IN THE BASKET. - ALSO SAYING THE ELASTIC LEG RESTRAINING NEEDS TO BE REPLACED (STATES THE PLASTIC KNOB IS BROKEN) CONTRIBUTING TO AN INJURY. - FYI - FOUND CHG REPL (B)(4) FOR A BATTERY, NO DEFECT UPON RETURN AND INSPECTION - STATES INJURY TO FINGERNAIL ON LEFT MIDDLE FINGER SAW BLOOD UNDER FINGERNAIL (INCIDENT RPT CREATED). - STATES THIS WAS CAUSED FROM THE PLASTIC KNOB ON THE ELASTIC LEG RESTRAINT; BUT THEN SAYS IT WAS CAUSED FROM THE LEG OF THE LIFT HITTING HER FINGER WHEN THE LIFT WAS BEING STORED IN THE FOLDED POSITION. COMPLAINT #(B)(4). WAS ENTERED INTO OUR SYSTEM.
Patient problems: Abrasion
MAUDE report 22627018 ↗InjuryMay 13, 2025Unintended Movement; Unintended Movement
IT WAS REPORTED TO THE MANUFACTURER, BY THE END USER, PER THE END USER, THAT (B)(6) WAS CALLING IN TO TROUBLESHOOT LIFT, REPORTED BATTERY RANDOMLY SHORTS OUT AND WON'T POWER LIFT. WHEN QUESTIONED FURTHER, (B)(6) MENTIONED AN INCIDENT THAT OCCURRED BACK IN "FEBRUARY" WHERE SHE WAS TRYING TO FOLD LIFT OVER FOR STORAGE WHEN THE BLACK ROUND PULL KNOB BROKE IN HER HAND WHILE SHE WAS TRYING TO PULL IT UP AND OVER THE LEGS, HER FINGER GOT BADLY PINCHED UNDER ONE OF THE WHEELS AND THE LIFT FELL OVER DAMAGING ONE OF HER CHAIRS. COMPLAINT #(B)(4) WAS ENTERED INTO OUR SYSTEM.
Patient problems: Bruise/Contusion
MAUDE report 22015042 ↗InjuryMar 5, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED TO THE MANUFACTURER, BY THE END USER, PER THE END USER, THAT CALL FROM LF REPORTING SHE FELL OUT OF HER LIFT AND NEEDS TO GET A PART. SHE SAYS SHE CALLED PREVIOUSLY BUT WAS TOLD THAT WE DON'T PROVIDE PARTS. THE FALL HAPPENED WHEN THE CAREGIVER WAS PUTTING THE SLING ON, THEN IN THE TRANSFER INTO THE CHAIR AS SHE GOT CLOSER TO HER CLOSET A PIECE OF THE SLING SLIPPED OUT OF THE SPREADER THEN SHE FELL AND HIT HER HEAD ON THE FLOOR. EMS AND FIRE DEPARTMENT WERE CALLED AND THERE WAS A HOSPITAL VISIT BUT THAT WAS A WEEK LATER BECAUSE SHE WAS DIZZY AND THEN HER HEAD STARTED HURTING. THEY DID AN MRI AND ALSO DETERMINED SHE HAD A LUNG INFECTION. COMPLAINT WAS ENTERED INTO OUR SYSTEM. THE TYPE OF SLING USED WAS NOT REPORTED AND END USER DOES NOT HAVE JOERNS SLING ON HER ORIGINAL ORDER.
Patient problems: Fall
MAUDE report 21527207 ↗InjuryMar 5, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED TO THE MANUFACTURER, BY THE END USER, PER THE END USER, THAT PER CAREGIVER SHE GOT MRS UP TO USE THE RESTROOM, AFTERWARDS WHILE LIFTING HER OFF OF TOILET, MRS STARTING FALLING SO CAREGIVER CAUGHT HER AND SET HER DOWN JUIE THEN WHEN TO GET HOYER LIFT TO RAISE HER UP OFF OF THE FLOOR, WAS RAISING HER UP WITH LEGS OF LIFT OPEN WHEN MOVING LIFT, CLOSED LEGS AND HEARD A POP, STARTED LOWERING MRS AS A PRECAUTION WHEN ONE OF THE TIE RODS SNAPPED AND LIFT TILTED OVER. MRS FELL ONLY A SHORT DISTANCE, ONLY INJURY REPORTED AT THIS TIME BEING SOME MILD BRUSING TO HER KNEE. COMPLAINT # (B)(4) AND RA #(B)(6) WERE ENTERED INTO OUR SYSTEM TO HAVE THE LIFT RETURNED. AS OF THIS WRITING, THE LIFT HAS NOT RETURNED.
Patient problems: Fall
MAUDE report 21527206 ↗
FDA clearance and classification
The FDA regulates product code FSA (lift, patient, non-ac-powered) under 21 CFR 880.5510 as a Class I device. 21 CFR 880.5510 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the APEX HOY-ADVANCE-E
Has the APEX HOY-ADVANCE-E been recalled?
No FDA recall record in our data names this exact device identifier (DI 00810055880041), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the APEX HOY-ADVANCE-E?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Mar 5, 2025 and Aug 29, 2025. In those reports, the most common specific device problems are “Unintended Movement” (2), “Collapse” (1) and “Unintended Collision” (1); the most common patient problems are “Fall” (4), “Abrasion” (1) and “Bruise/Contusion” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the APEX HOY-ADVANCE-E?
None of the reports are classified as deaths, 6 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the APEX HOY-ADVANCE-E FDA approved?
Devices of this type (product code FSA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Joerns Healthcare, LLC.
Who makes the APEX HOY-ADVANCE-E, and is it still sold?
The FDA's device database lists Joerns Healthcare, LLC as the company responsible for it (brand APEX, model HOY-ADVANCE-E). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Lift, Patient, Non-Ac-Powered devices
- Likorall™ overhead lift 3123001Liko AB · Lift, patient, non-ac-powered1 recall1 death32 reports
- Likorall™ overhead lift 3121001Liko AB · Lift, patient, non-ac-powered1 recall22 reports
- Likorall™ overhead lift 3124050Liko AB · Lift, patient, non-ac-powered1 recall17 reports
- Viking™ XL mobile lift 2040043Liko AB · Lift, patient, non-ac-powered1 recall14 reports
- Likorall™ overhead lift 3122010Liko AB · Lift, patient, non-ac-powered1 recall13 reports
- Medline MDS400ELBMedline Industries, Inc. · Lift, patient, non-ac-powered59 reports
- Uno™ mobile lift 2010004Liko AB · Lift, patient, non-ac-powered1 recall6 reports
- Likorall™ overhead lift 3122502Liko AB · Lift, patient, non-ac-powered1 recall5 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page