Valkyrie Thoracic Fixation System 01-020 (UDI 00810071390012)Recalls & reported problems
J.M. Longyear Manufacturing, LLC · X-Plate · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00810071390012), received between Mar 3, 2023 and Mar 3, 2023 — classified as 1 injury.
The most frequently reported specific device problem is “Fracture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fracture1
Patient problems
- Failure of Implant1
- Wound Dehiscence1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMar 3, 2023Fracture; Fracture
IT WAS REPORTED BY A DISTRIBUTOR THAT A VALKYRIE RE OPERATION PROCEDURE OCCURRED IN A PATIENT. THE PATIENT WAS SEEN 11 DAYS POST OPERATIVELY AFTER EXPERIENCING A FALL AND COUGHING FIT. A RE OPERATION WAS INITIATED ON THE PATIENT AND NOTED STERNAL DEHISCENCE WITH THE OBSERVATION OF THE STERNAL CONSTRUCT FAILURE (PULL THROUGH OF WIRES AND FRACTURE OF PLATES). AS A FOLLOW UP, INTERVIEW WITH (B)(6), PA (HCP PRESENT IN THE ORIGINAL AND REVISION CASE) WAS CONDUCTED ON (B)(6) 2023. IT WAS NOTED THAT THE PATIENT'S BONE QUALITY WAS EXTREMELY POOR. THE X-PLATES AND THE V-PLATE FRACTURED AT THE MID-LINE, SCREWS REMAINED ATTACHED TO THE PLATES AND THE BONE. BLUNT FORCE TRAUMA MAY HAVE CONTRIBUTED TO THE EVENT. DURING THE INITIAL CLOSURE OF THE STERNUM, IT WAS NOTED THAT THE STERNUM BONE QUALITY WAS VERY POOR. PATIENT RECEIVED AN ADDITIONAL PROCEDURE TO REMOVE THE DEVICE, A WOUND VAC. AND MUSCLE FLAP WAS PERFORMED.
Patient problems: Failure of Implant; Wound Dehiscence
MAUDE report 16483153 ↗
Other Plate, Fixation, Bone devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page