Skip to content

Valkyrie Thoracic Fixation System 01-020 (UDI 00810071390531)Recalls & reported problems

J.M. Longyear Manufacturing, LLC · X-Plate · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00810071390531), received between Aug 8, 2023 and Oct 3, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Failure of Implant1
  • Wound Dehiscence1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 3, 2023Fracture; Fracture

    IT WAS REPORTED THAT A PATIENT UNDERWENT A REVISION PROCEDURE DUE TO EVIDENCE OF STERNAL INSTABILITY FOLLOWING A SEVERE COUGHING FIT. IT WAS OBSERVED THAT STERNAL WIRES HAD PULLED THROUGH THE BONE AND AN X-PLATE HAD FRACTURED AT A SINGLE LOCATION. IT WAS NOTED THAT THE SCREWS WERE STILL SECURELY LOCKED INTO THE PLATE AS WELL AS FULLY INTACT TO BONE SUBSTRATE. THE DEVICE WAS EXPLANTED, RETAINED BY THE HOSPITAL, AND SENT BACK TO THE MANUFACTURER FOR FURTHER INVESTIGATION. A NEW DEVICE WAS IMPLANTED TO RE-CLOSE THE CHEST-WALL AND THERE WAS NO FURTHER HARM TO THE PATIENT.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions; Wound Dehiscence

    MAUDE report 17863579 ↗
  • MalfunctionAug 8, 2023Fracture; Fracture

    IT WAS REPORTED THAT AN X-PLATE FRACTURED AROUND A SINGLE SCREW HOLE DURING SCREW INSERTION. IT WAS NOTED THAT THE SCREWS WERE INSERTED MULTIPLE TIMES, THE CADDY GUIDE WAS NOT USED, AND SCREWS WERE INSERTED FREE-HANDED INTO THE CONSTRUCT. THE FRACTURED PLATE WAS REMOVED, AND ANOTHER PLATE WAS IMPLANTED. THERE WAS NO HARM TO THE PATIENT NOR WAS THERE A SIGNIFICANT DELAY TO THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17480350 ↗