TriLeap Lower Extremity Anatomic Plating System Small Ratchet Handle w/ Quick CouplingRecalls & reported problems
Crossroads Extremity Systems · Small Ratchet Handle w/ Quick Coupling · Class 1
All models of TriLeap Lower Extremity Anatomic Plating System (57)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00810112592726), received between Nov 12, 2024 and Oct 16, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Device-Device Incompatibility2
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 16, 2025Break; Material Deformation; Break; Material Deformation
IT WAS REPORTED ON AN UNKNOWN DATE, THAT STERILE PROCESSING DEPARTMENTS STATES THE DEVICES WERE WARPED OR BROKEN. NO PATIENT INVOLVEMENT OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23314138 ↗MalfunctionAug 8, 2025Break; Break
IT WAS REPORTED IN AN UNKNOWN DATE, THE INSTRUMENT BROKE DURING SPD. THE EVENT OCCURRED DURING STERILE PROCESSING, FIELD INSPECTION AND DURING CALIBRATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22755096 ↗MalfunctionJan 8, 2025Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING AN ORIF OF A FEMUR, A DHS 6H WAS USED AND THE PLATE DID NOT SEAT ON THE T HANDLE DHS INSERTER. UNSURE WHAT WAS WRONG WITH PLATE. TSP WAS USED AND THE SM HEXAGONAL PELVIC SCRDRIVER/LRG HNDL WAS USED TO BEND THE PLATE AND THE SHAFT ENDED UP BENDING AS WELL. TWO OF THE DEVICES WERE BENT. THE CABLES WERE USED AND THE CABLE LOCK WOULD NOT LOCK. ALSO, THE RATCHETING DRIVER HAD A 3.5 HEX STUCK AND UNABLE TO TAKE OUT. THERE WAS NO PATIENT CONSEQUENCE. IT WAS FURTHER REPORTED THAT THE COUPLING MECHANISM USED WERE THE JEWELERS CAP HANDLE WITH QUICK COUPLING AND THE SMALL RATCHET HANDLE WITH QUICK COUPLING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21101986 ↗MalfunctionNov 12, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING AN ORIF OF A FEMUR, A DHS 6H WAS USED AND THE PLATE DID NOT SEAT ON THE T HANDLE DHS INSERTER. UNSURE WHAT WAS WRONG WITH PLATE. TSP WAS USED AND THE SD314.001 WAS USED TO BEND THE PLATE AND THE SHAFT ENDED UP BENDING AS WELL. TWO OF THE DEVICES WERE BENT. THE CABLES WERE USED AND THE CABLE LOCK WOULD NOT LOCK. ALSO, THE RATCHETING DRIVER HAD A 3.5 HEX STUCK AND UNABLE TO TAKE OUT. THERE WAS NO PATIENT STATUS, OUTCOME, OR CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20668973 ↗
Other Orthopedic Manual Surgical Instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page