Skip to content

TriLeap Lower Extremity Anatomic Plating System 2.5 mm Ti Variable Angle Screw Length 34 mm, MagentaRecalls & reported problems

Crossroads Extremity Systems · 2.5 mm Ti Variable Angle Screw Length 34 mm, Magenta · Class 2 · Implant

All models of TriLeap Lower Extremity Anatomic Plating System (57)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00810112593570), received between Oct 23, 2024 and Oct 23, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 23, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED ON (B)(6) 2024, THE SURGEON WAS USING THE TRILEAP CUBOID LOCKING PLATE FOR A COMMINUTED CUBOID FRACTURE. HE DETERMINED LOCKING SCREWS WOULD BE THE ONLY SCREWS NECESSARY TO USE AFTER REDUCING THE FRACTURE. 6 OUT OF THE 7 LOCKING SCREWS DID NOT EFFECTIVELY LOCK INTO THE PLATE. SCREWS WERE LEFT IN. PATIENT WILL BE NON-WEIGHT BEARING FOR A WHILE DUE TO ADDITIONAL POLY TRAUMA INJURIES. THERE WAS NO SURGERY DELAYED DUE TO THE REPORTED EVENT.

    Patient problems: Insufficient Information

    MAUDE report 20514578 ↗