Azur Detachable Helical HydroCoil 45-480410Recalls & reported problems
Microvention, Inc. · Azur · Class 2 · Implant
All models of Azur Detachable Helical HydroCoil (42)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00810170017902), received between Sep 27, 2016 and Jul 15, 2026 — classified as 2 injuries, 12 malfunctions.
The most frequently reported specific device problem is “Detachment Of Device Component” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment Of Device Component6
- Premature Separation5
- Device-Device Incompatibility4
- Component Missing2
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Consequences Or Impact To Patient5
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 15, 2026Premature Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25830032 ↗InjuryAug 22, 2024Premature Separation; Premature Separation
IT WAS REPORTED THAT DURING A HEPATIC ARTERY EMBOLIZATION PROCEDURE, THE EMBOLIZATION COIL PREMATURELY DETACHED WITH HALF OF THE COIL OUTSIDE THE CATHETER. THE PHYSICIAN ATTEMPTED TO PUSH THE COIL IN INTENDED SITE WITH USE OF GUIDEWIRE AND COIL PUSHER WITH NO SUCCESS. THE COIL WAS RETRIEVED BY PUSHING THE 2.5FR MICROCATHETER OVER ABOUT 80% AND THEN PULLED OUT THROUGH THE MOTHER CATHETER WITH ASPIRATION (WITH EVERYTHING CROSSED). THERE WAS NO HARM OR INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20027189 ↗MalfunctionMar 28, 2024Premature Separation; Premature Separation
IT WAS REPORTED THAT THE COIL WAS RELEASED SUDDENLY INSIDE THE PATIENT WITHOUT ACTIVATING THE HANDLE BEFORE ARRIVING AT THE EXPECTED LOCATION. THE COIL WAS REMOVED AND REPLACED WITH ANOTHER COIL. THERE WAS NO HARM TO THE PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19001112 ↗MalfunctionDec 19, 2023Component Missing; Component Missing
IT WAS REPORTED, UPON OPENING THE COIL DURING THE CASE, THERE WAS NOT A COIL ON THE PUSHER WIRE. THERE WAS NO PATIENT INTERACTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18358477 ↗MalfunctionDec 19, 2023Component Missing; Component Missing
IT WAS REPORTED, UPON OPENING THE COIL DURING THE CASE, THERE WAS NOT A COIL ON THE PUSHER WIRE. THERE WAS NO PATIENT INTERACTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18358476 ↗MalfunctionSep 5, 2023Premature Separation; Premature Separation
IT WAS REPORTED THAT AN EMBOLIZATION COIL IMPLANT DETACHED UNEXPECTEDLY WITH THE LUMEN OF A MICROCATHETER. THE COIL WAS NOT USED. THERE WAS NO PATIENT INJURY OR INTERVENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17686539 ↗MalfunctionJun 24, 2021Premature Separation; Premature Separation
IT WAS REPORTED THAT DURING AN EMBOLIZATION TREATMENT, DURING MANIPULATION OF THE EMBOLIZATION COIL, IT DETACHED WITHIN THE CATHETER. THE COIL COULD NOT BE ADVANCE FROM THE CATHETER. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO HARM OR INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12062089 ↗MalfunctionFeb 25, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING TREATMENT, THE EMBOLIZATION COIL IMPLANT DETACHED. THE IMPLANT WAS PUSHED INTO POSITION. THERE WAS NO REPORTED PATIENT INJURY OR INTERVENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11382723 ↗InjuryMar 2, 2018Detachment Of Device Component; Device-Device Incompatibility; Detachment Of Device Component; Device-Device Incompatibility
IT WAS REPORTED THAT AFTER ATTEMPTING TO POSITION AN EMBOLIZATION COIL IN THE ARTERIOVENOUS MALFORMATION, THE PHYSICIAN PASSED A WIRE DOWN THE LEFT ANTERIOR DESCENDING ARTERY. RESISTANCE WAS ENCOUNTERED WHEN ATTEMPTING TO ADVANCE THE COIL IN THE MICROCATHETER. THE PUSHER BROKE 1CM PROXIMAL TO THE COIL ATTACHMENT POINT. THE DETACHED COIL WAS SUCCESSFULLY RETRIEVED WITH A SNARE DEVICE. THERE WAS NO REPORTED PATIENT INJURY. THE PATIENT WAS REPORTED TO BE STABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7311572 ↗MalfunctionJan 5, 2018Detachment Of Device Component; Device-Device Incompatibility; Detachment Of Device Component; Device-Device Incompatibility
IT WAS REPORTED THAT RESISTANCE WAS ENCOUNTERED DURING ADVANCEMENT OF THE COIL. DURING THE REMOVAL ATTEMPT, THE COIL DETACHED IN THE MICROCATHETER. BOTH SEGMENTS OF THE COIL WERE REMOVED TOGETHER WITH THE MICROCATHETER FROM THE PATIENT. THERE WAS NO REPORTED INTERVENTION OR PATIENT INJURY. THE PATIENT WAS REPORTED TO BE STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7170288 ↗MalfunctionSep 28, 2017Detachment Of Device Component; Detachment Of Device Component
IT WAS REPORTED THAT DURING AN EMBOLIZATION PROCEDURE, THE COIL UNEXPECTEDLY DETACHED IN THE MICROCATHETER. BOTH SEGMENTS OF THE COIL WERE REMOVED FROM THE PATIENT. THERE WAS NO REPORTED INTERVENTION OR PATIENT INJURY. THE PATIENT'S STATUS IS REPORTED TO BE FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6903098 ↗MalfunctionMay 16, 2017Detachment Of Device Component; Device-Device Incompatibility; Detachment Of Device Component; Device-Device Incompatibility
IT WAS REPORTED THAT AFTER PLACEMENT OF AN EMBOLIZATION COIL IN AN ANEURYSM LOCATED AT THE LOWER POLAR SPLENIC BRANCH, RESISTANCE AND FRICTION WERE ENCOUNTERED WITHIN THE MICROCATHETER. DURING AN ATTEMPT TO REPOSITION, THE COIL UNEXPECTEDLY DETACHED. THERE WAS NO REPORTED INTERVENTION. THE COIL REMAINS IMPLANTED IN THE ANEURYSM. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6568832 ↗MalfunctionMar 1, 2017Detachment Of Device Component; Device-Device Incompatibility; Detachment Of Device Component; Device-Device Incompatibility
IT WAS REPORTED THAT COIL EMBOLIZATION TREATMENT WAS ATTEMPTED ON A SPLENIC ARTERY ANEURYSM. DURING ADVANCEMENT OF THE COIL IN THE CATHETER, RESISTANCE WAS ENCOUNTERED AND THE COIL UNEXPECTEDLY DETACHED IN THE CATHETER BEFORE THE COIL HAD EXITED INTO THE PATIENT. THE ENTIRE COIL WAS REMOVED TOGETHER WITH THE CATHETER WITHOUT INCIDENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6370959 ↗MalfunctionSep 27, 2016Detachment Of Device Component; Detachment Of Device Component
IT WAS REPORTED THAT DURING AN ATTEMPT TO PLACE A COIL IN THE GASTRODUODENAL ARTERY, THE COIL DISENGAGED AND GOT STUCK IN THE MICROCATHETER. THE COIL AND THE MICROCATHETER WERE REMOVED TOGETHER, AND ANOTHER MICROCATHETER AND COILS WERE USED FOR THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5982023 ↗
Other Device, Neurovascular Embolization devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page