Azur Detachable Helical HydroCoilRecalls & reported problems across all models
Microvention, Inc. · 42 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 190
- Deaths reported
- 1
- Injuries reported
- 41
- Recalls
- 0
We found 42 models of Azur Detachable Helical HydroCoil by Microvention, Inc. in the FDA's device registry (GUDID). Together they are named in 190 adverse event reports in the FDA's MAUDE database, received between Feb 16, 2016 and Sep 30, 2026 — classified as 1 death, 41 injuries, 148 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (90 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Azur Detachable Helical HydroCoil problems
Device problems
- Detachment of Device or Device Component90
- Detachment Of Device Component44
- Premature Separation38
- Device-Device Incompatibility20
- Migration10
- Separation Failure9
- Difficult to Remove8
- Component Missing6
Patient problems
- No Clinical Signs, Symptoms or Conditions86
- No Consequences Or Impact To Patient83
- No Known Impact Or Consequence To Patient10
- Aneurysm7
- Chest Tightness/Pressure3
- Vascular System (Circulation), Impaired3
- No Patient Involvement2
- Obstruction/Occlusion1
Azur Detachable Helical HydroCoil models
42 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Azur Detachable Helical HydroCoil 45-480420Device, Vascular, for Promoting Embolization19 reports
- Azur Detachable Helical HydroCoil 45-480410Device, Neurovascular Embolization14 reports
- Azur Detachable Helical HydroCoil 45-452030Device, Vascular, for Promoting Embolization10 reports
- Azur Detachable Helical HydroCoil 45-451230Device, Vascular, for Promoting Embolization9 reports
- Azur Detachable Helical HydroCoil 45-450820Device, Vascular, for Promoting Embolization8 reports
- Azur Detachable Helical HydroCoil 45-480202Device, Neurovascular Embolization8 reports
- Azur Detachable Helical HydroCoil 45-480305Device, Vascular, for Promoting Embolization8 reports
- Azur Detachable Helical HydroCoil 45-480620Device, Vascular, for Promoting Embolization8 reports
- Azur Detachable Helical HydroCoil 45-450620Device, Vascular, for Promoting Embolization7 reports
- Azur Detachable Helical HydroCoil 45-451530Device, Vascular, for Promoting Embolization7 reports
- Azur Detachable Helical HydroCoil 45-480415Device, Vascular, for Promoting Embolization7 reports
- Azur Detachable Helical HydroCoil 45-480405Device, Vascular, for Promoting Embolization5 reports
- Azur Detachable Helical HydroCoil 45-480610Device, Neurovascular Embolization5 reports
- Azur Detachable Helical HydroCoil 45-480815Device, Neurovascular Embolization5 reports
- Azur Detachable Helical HydroCoil 45-482030Device, Vascular, for Promoting Embolization5 reports
- Azur Detachable Helical HydroCoil 45-450815Device, Vascular, for Promoting Embolization4 reports
- Azur Detachable Helical HydroCoil 45-451020Device, Vascular, for Promoting Embolization4 reports
- Azur Detachable Helical HydroCoil 45-451520Device, Vascular, for Promoting Embolization4 reports
- Azur Detachable Helical HydroCoil 45-480615Device, Vascular, for Promoting Embolization4 reports
- Azur Detachable Helical HydroCoil 45-480820Device, Neurovascular Embolization4 reports
- Azur Detachable Helical HydroCoil 45-481015Device, Neurovascular Embolization4 reports
- Azur Detachable Helical HydroCoil 45-481220Device, Vascular, for Promoting Embolization4 reports
- Azur Detachable Helical HydroCoil 45-480204Device, Neurovascular Embolization3 reports
- Azur Detachable Helical HydroCoil 45-480310Device, Vascular, for Promoting Embolization3 reports
- Azur Detachable Helical HydroCoil 45-481020Device, Neurovascular Embolization3 reports
- Azur Detachable Helical HydroCoil 45-450415Device, Vascular, for Promoting Embolization2 reports
- Azur Detachable Helical HydroCoil 45-450610Device, Vascular, for Promoting Embolization1 death2 reports
- Azur Detachable Helical HydroCoil 45-451015Device, Vascular, for Promoting Embolization2 reports
- Azur Detachable Helical HydroCoil 45-451220Device, Vascular, for Promoting Embolization2 reports
- Azur Detachable Helical HydroCoil 45-452020Device, Vascular, for Promoting Embolization2 reports
- Azur Detachable Helical HydroCoil 45-481215Device, Vascular, for Promoting Embolization2 reports
- Azur Detachable Helical HydroCoil 45-481520DEVICE, NEUROVASCULAR EMBOLIZATION2 reports
- Azur Detachable Helical HydroCoil 45-481530Device, Vascular, for Promoting Embolization2 reports
- Azur Detachable Helical HydroCoil 45-480510DEVICE, NEUROVASCULAR EMBOLIZATION2 reports
- Azur Detachable Helical HydroCoil 45-480515Device, Neurovascular Embolization2 reports
- Azur Detachable Helical HydroCoil 45-480520Device, Neurovascular Embolization2 reports
- Azur Detachable Helical HydroCoil 45-450410Device, Vascular, for Promoting Embolization1 report
- Azur Detachable Helical HydroCoil 45-451215Device, Vascular, for Promoting Embolization1 report
- Azur Detachable Helical HydroCoil 45-480302Device, Neurovascular Embolization1 report
- Azur Detachable Helical HydroCoil 45-480810Device, Neurovascular Embolization1 report
- Azur Detachable Helical HydroCoil 45-481010Device, Neurovascular Embolization1 report
- Azur Detachable Helical HydroCoil 45-481515Device, Neurovascular Embolization1 report
Recent Azur Detachable Helical HydroCoil adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Premature Separation
As reported, the patient had an abdominal aortic pseudoaneurysm (diameter: 50 mm). During the procedure, a 5F angiographic catheter was used to deliver the first AZUR 20-30 coil. Significant resistance was encountered during advancement, resulting in separation of the delivery wire from the coil. The angiographic catheter was subsequently withdrawn. A new angiographic catheter was then used, and two AZUR 20-30 coils were successfully delivered without issue. A covered stent was subsequently deployed, and the procedure was completed successfully. The patient returned safely to the ward, and no harm or adverse clinical impact occurred. Patient condition, stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50941629 ↗MalfunctionJul 15, 2026Premature Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25830032 ↗MalfunctionSep 16, 2025Premature Separation; Premature Separation
IT WAS REPORTED THAT THE EMBOLIZATION COIL IMPLANT WAS USED TO TREAT A PATIENT WITH AN UNRUPTURED ABDOMINAL INTERNAL ILIAC ARTERY ANEURYSM. THE COIL WAS DEPLOYED AS THIRD COIL VIA AN ANGIOGRAPHIC CATHETER WITHOUT THE USE OF A MICROCATHETER. DURING PLACEMENT OF THE THIRD COIL, SIGNIFICANT RESISTANCE WAS ENCOUNTERED DURING ADVANCEMENT. DESPITE REPOSITIONING THE CATHETER, FURTHER ADVANCEMENT REMAINED IMPOSSIBLE. WHILE ATTEMPTING TO WITHDRAW, THE COIL PREMATURELY DETACHMENT IN ANGIOGRAPHIC CATHETER. THE PROCEDURE WAS DISCONTINUED DUE TO THE INABILITY TO COMPLETE EMBOLIZATION. PRIOR TO THE DETACHMENT OF THIS COIL, TWO COILS OF THE SAME MODEL HAD ALREADY BEEN DEPLOYED FOR EMBOLIZATION. ANGIOGRAPHY CONFIRMED THAT THE EMBOLIZATION EFFECT HAD BEEN ACHIEVED AND NO ADDITIONAL COILS WERE IMPLANTED THEREAFTER. THE VASCULAR ANATOMY MAY HAVE CONTRIBUTED TO THE DETACHMENT OF THE COIL. THE PATIENT WAS IN STABLE CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23065119 ↗InjuryAug 15, 2025Premature Separation; Premature Separation
PLEASE SEE SECTION H11 FOR CORRECTION TO INCLUDE SERIOUS INJURY, AS THE COIL DETACHED WITHOUT USING THE DETACHMENT CONTROLLER BEFORE REACHING THE TARGET SITE AND THEREFORE, EMBOLIZATION OF AN ARTERY OTHER THAN THE ONE INITIALLY INTENDED OCCURRED AS THE COIL WAS LEFT IN PLACE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22814999 ↗MalfunctionAug 6, 2025Premature Separation; Premature Separation
IT WAS REPORTED THAT DURING THE ATTEMPTED DEPLOYMENT OF THE COIL THE COIL DETACHED PREMATURELY WHILE BEING RETRACTED BUT REMAINED INSIDE THE CATHETER. NO HARM OCCURRED TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22729105 ↗MalfunctionJul 29, 2025Premature Separation; Premature Separation
IT WAS REPORTED THAT A COMPETITOR BASE CATHETER, THEN A 2.4 MICROCATHETER CATHETER WAS USED IN THE CELIAC THEN GDA. PUSHED COIL THROUGH CATHETER, AFTER PLACING 2 COMPETITOR COILS. THE COIL OF CONCERN GOT HUNG UP IN THE CATHETER AND PREMATURELY DETACHED. WAS ABLE TO REVERSE SUCTION COIL PARTWAY INTO THE MICROCATHETER THEN PULL THE MICROCATHETER INTO THE BASE CATHETER AND REVERSE SUCTION AND PULL THE BASE CATHETER OUT WITH THE COIL. NO HARM TO PATIENT. ENDED UP USING GELFOAM EMBOLIC AGENT TO COMPLETE THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22657102 ↗InjuryJun 10, 2025Entrapment of Device; Difficult to Remove; Unraveled Material; Difficult or Delayed Separation; Entrapment of Device; Difficult to Remove; Unraveled Material; Difficult or Delayed Separation
IT WAS REPORTED THAT DURING A RENAL EMBOLIZATION PROCEDURE, THE COIL UNSUCCESSFULLY DEPLOYED AFTER MULTIPLE (NORMAL) CYCLES WITH CONTROLLER. PHYSICIAN TRIED TO MANUALLY DEPLOY VIA PUSHER WIRE ROTATION. COIL DEPLOYED BUT SUPPORT MICROCATHETER WAS A DISTANCE AWAY FROM THE END OF COIL. UPON DEPLOYMENT RECOIL OF THE ADHESIVE STRIP (CONDUIT FOR THE THERMO / ELECTRIC DISCHARGE THAT IS AFFIXED ON THE WIRE) WRAPPED THE COIL. ADHESIVE STRIP FINALLY DISENGAGED WITH A SMALL TAG END LEFT THAT WAS STILL IN THE TARGET ARTERY. THE PUSHER WIRE WAS DETACHED BUT ROTATION AND WAS REMOVED W/O INCIDENT. NO INTERVENTION WAS USED TO REMOVE THE COIL. THERE WAS NO HARM TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22181540 ↗MalfunctionMar 28, 2025Premature Separation; Premature Separation
AS REPORTED IN AN INTERVENTIONAL RADIOLOGY, AN AZUR COIL WAS USED DURING THE PROCEDURE. THE COIL DETACHED SPONTANEOUSLY FROM ITS DELIVERY CATHETER INTO THE MICROCATHETER, WITHOUT MANIPULATION ERROR. THE DEVICE WAS REMOVED, AND A SECOND COIL WAS USED. THERE WERE NO CLINICAL CONSEQUENCES FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21724168 ↗MalfunctionMar 11, 2025Premature Separation; Premature Separation
ADDITIONAL PATIENT DEMOGRAPHIC INFORMATION RECEIVED IN SECTIONS: A2, A3A, A4. SEE H11.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21579639 ↗MalfunctionFeb 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT AN IN INTERVENTIONAL RADIOLOGY, THE DEVICE'S RELEASE WIRE BROKE DURING USE, PREVENTING USE OF THE EQUIPMENT WITH THE RELEASE HANDLE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO CLINICAL CONSEQUENCES FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21360458 ↗MalfunctionNov 14, 2024Premature Separation; Premature Separation
IT WAS REPORTED THAT AN EMBOLIZATION COIL IMPLANT DETACHED UNEXPECTEDLY. THE IMPLANT WAS REMOVED AND THE CASE WAS COMPLETED WITH ANOTHER IMPLANT COIL. THERE WAS NO REPORTED INTERVENTION OR PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20692156 ↗InjuryAug 22, 2024Premature Separation; Premature Separation
IT WAS REPORTED THAT DURING A HEPATIC ARTERY EMBOLIZATION PROCEDURE, THE EMBOLIZATION COIL PREMATURELY DETACHED WITH HALF OF THE COIL OUTSIDE THE CATHETER. THE PHYSICIAN ATTEMPTED TO PUSH THE COIL IN INTENDED SITE WITH USE OF GUIDEWIRE AND COIL PUSHER WITH NO SUCCESS. THE COIL WAS RETRIEVED BY PUSHING THE 2.5FR MICROCATHETER OVER ABOUT 80% AND THEN PULLED OUT THROUGH THE MOTHER CATHETER WITH ASPIRATION (WITH EVERYTHING CROSSED). THERE WAS NO HARM OR INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20027189 ↗InjuryMay 8, 2024Inadequacy of Device Shape and/or Size; Detachment of Device or Device Component; Migration; Inadequacy of Device Shape and/or Size; Detachment of Device or Device Component; Migration
IT WAS REPORTED THAT DURING TREATMENT OF A GASTRODUODENAL ARTERY (GDA) BLEED, THE THIRD EMBOLIZATION COIL IMPLANT TO BE IMPLANTED WAS NOT TAKING SHAPE IMMEDIATELY AND SO THE MICROCATHETER WAS USED TO COMPACT IT. WHEN ADVANCEMENT OF THE PUSHER WAS RESUMED, IT WAS NOTICED THAT THE IMPLANT HAD DETACHED UNEXPECTEDLY. THE CATHETER WAS FLUSHED WITH SALINE, WHICH MANUALLY PUSHED THE COIL OUT OF IT BUT ALSO CAUSED THE CATHETER TO LOSE POSITION AND BACK UP INTO THE HEPATIC ARTERY. THE IMPLANT COIL WAS THEN REPORTED TO HANGING INTO THE HEPATIC ARTERY, AT WHICH POINT THE BACK HALF OF THE COIL MIGRATED UP AND INTO THE LEFT HEPATIC ARTERY, WHERE IT STARTED TO TAKE ITS HELICAL SHAPE. IT WAS REPORTED TO HAVE A DUMBBELL SHAPE. ONE END OF THE IMPLANT WAS IN THE GDA, THEN IT STRAIGHTENED OUT AND UP INTO THE LEFT HEPATIC ARTERY WHERE THE OTHER END COILED AGAIN. THE GDA WAS OCCLUDED AND ACHIEVED THE RESULT DESIRED BY THE PHYSICIAN REGARDING THE EXISTING GDA BLEED. IT WAS REPORTED THAT, UPON FOLLOW UP, THE PATIENT HAD BEEN CONTINUING TO BLEED AND EXHIBITED SIGNS OF LIVER SHOCK. THE PATIENT LATER PASSED AWAY. IT WAS REPORTED THAT THE PHYSICIAN STATED THAT THE IMPLANT COIL DID NOT CONTRIBUTE TO THE PATIE
Patient problems: Obstruction/Occlusion
MAUDE report 19271597 ↗MalfunctionApr 22, 2024Premature Separation; Premature Separation
IT WAS REPORTED THAT THE PHYSICIAN WAS PERFORMING A VENOUS COLLATERAL EMBOLIZATION. AFTER PLACING MULTIPLE COILS, THIS HYDRO COIL WAS SELECTED, WHICH WAS PREPPED AS NORMAL AND WAS INTRODUCED INTO THE MICROCATHETER WITHOUT INCIDENT. AS THE PHYSICIAN WAS DEPLOYING THE COIL, IT GOT STUCK ABOUT 1/3 OF THE WAY OUT OF THE MICROCATHETER. THE PHYSICIAN TRIED TO RETRACT AND THE COIL PREMATURELY DEPLOYED INSIDE THE CATHETER. PHYSICIAN WAITED ABOUT 5-8 MINUTES FOR THE HYDRO TO EXPAND AND TOOK COIL AND MICROCATHETER OUT TOGETHER. THE PHYSICIAN WAS ABLE TO REINTRODUCE THE CATHETER AND FINISHED CASE WITHOUT INCIDENT. THE EMBOLIZATION PROCEDURE WAS SUCCESSFUL WITHOUT REPORT OF HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19159192 ↗MalfunctionApr 1, 2024Packaging Problem; Packaging Problem
AS REPORTED: THE REPRESENTATIVE PULLED A HYDROCOIL OUT OF THE BOX PRIOR TO A CASE. THE HYDROCOIL PACKAGE HAD A RED REJECTED STICKER ON THE BACK OF THE COIL PLASTIC PACKING. THERE WAS NO OTHER STICKER OR LABEL NOTED ON THE BOX OR EXTERIOR PACKAGING OF THE COIL. NO OTHER PACKAGES WERE FOUND WITH A SIMILAR STICKER. THE STERILE PLASTIC PACKING WAS INTACT. THE COIL WAS NEVER PRESENTED TO A CUSTOMER OR USED WITH A PATIENT, NO PATIENT INVOLVEMENT. ALTHOUGH REQUESTED, NO OTHER INFORMATION WAS MADE AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19015160 ↗
Other Microvention, Inc. products
- Azur CX Detachable22 models · 509 reports · 1 recall
- Stent63 models · 237 reports · 2 recalls
- V-Trak75 models · 147 reports
- WEB14 models · 134 reports
- AZUR14 models · 104 reports
- HydroSoft 3D26 models · 101 reports
- MicroPlex Helical Soft9 models · 8 reports · 1 recall
- ERIC Retrieval Device3 models · 5 reports · 1 recall
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page