Skip to content

Azur Detachable Helical HydroCoil 45-481215Recalls & reported problems

Microvention, Inc. · Azur · Class 2 · Implant

All models of Azur Detachable Helical HydroCoil (42)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00810170018022), received between Sep 27, 2018 and Sep 27, 2018 — classified as 2 injuries.

The most frequently reported specific device problem is “Migration” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Migration2

Patient problems

  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 27, 2018Migration; Migration

    IT WAS REPORTED THAT NINE (9) COILS WERE IMPLANTED IN AN ARTERIOVENOUS (AV) FISTULA OFF THE RIGHT ILIAC ARTERY IN (B)(6) 2018. APPROXIMATELY THREE (3) MONTHS LATER IN (B)(6) THE PATIENT HAD SYMPTOMATIC COMPLAINTS IN HER PELVIS AGAIN. ANGIOGRAPHIC EXAMINATION REVEALED THE COIL PACK WAS NO LONGER IN THE AREA OF THE AV FISTULA, AND HAD MIGRATED TO THE PATIENT'S LUNG. ATTEMPTS WERE MADE TO RETRIEVE THE COIL PACK OVER A PERIOD OF 1.5 HOURS; HOWEVER, THEY WERE UNSUCCESSFUL. THE PATIENT HAS HAD NO SYMPTOMS OF COMPLICATIONS IN THE RESPIRATORY SYSTEM. THE PATIENT HAS BEEN REFERRED TO A CT SURGEON. THE PATIENT IS CURRENTLY REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7916755 ↗
  • InjurySep 27, 2018Migration; Migration

    IT WAS REPORTED THAT NINE (9) COILS WERE IMPLANTED IN AN ARTERIOVENOUS (AV) FISTULA OFF THE RIGHT ILIAC ARTERY IN (B)(6) 2018. APPROXIMATELY THREE (3) MONTHS LATER IN (B)(6) THE PATIENT HAD SYMPTOMATIC COMPLAINTS IN HER PELVIS AGAIN. ANGIOGRAPHIC EXAMINATION REVEALED THE COIL PACK WAS NO LONGER IN THE AREA OF THE AV FISTULA, AND HAD MIGRATED TO THE PATIENT'S LUNG. ATTEMPTS WERE MADE TO RETRIEVE THE COIL PACK OVER A PERIOD OF 1.5 HOURS; HOWEVER, THEY WERE UNSUCCESSFUL. THE PATIENT HAS HAD NO SYMPTOMS OF COMPLICATIONS IN THE RESPIRATORY SYSTEM. THE PATIENT HAS BEEN REFERRED TO A CT SURGEON. THE PATIENT IS CURRENTLY REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7916750 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page