IntelliBlate Ximitry Probe Assembly, 13G x 15cm IB-XPA-1315Recalls & reported problems
Varian Medical Systems, Inc. · IntelliBlate Ximitry probe assembly includes cassette with saline tubing, laser alignment disc, and microwave ablation probe 13G/15cm Probe QTY 1 · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00810563022292), received between Nov 19, 2025 and Mar 17, 2026 — classified as 3 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Appropriate Device Problem Term/Code Not Available1
- Insufficient Device Problem Information1
- Use of Device Problem1
Patient problems
- Air Embolism1
- Hematoma1
- Liver Damage/Dysfunction1
- Pneumothorax1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryMar 17, 2026Appropriate Device Problem Term/Code Not Available
SEE COMPLETED MW-2026-00009, MFR REPORT NUMBER 3008262715-2026-00006.
Patient problems: Hematoma; Liver Damage/Dysfunction
MAUDE report 24621355 ↗InjuryFeb 3, 2026Use of Device Problem; Insufficient Device Problem Information
DURING A MICROWAVE ABLATION PROCEDURE USING AN INTELLIBLATE IB-XPA-1315 PROBE, THE SYSTEM TRIGGERED A COOLING SYSTEM ALARM APPROXIMATELY ONE MINUTE AFTER ABLATION BEGAN AT 60 W. UPON RESPONDING TO THE ALARM, THE PHYSICIAN PERFORMED IMAGING AND IDENTIFIED A BOLUS OF AIR WITHIN THE LIVER AND IN THE RIGHT VENTRICLE OF THE HEART. THE PHYSICIAN IMMEDIATELY STOPPED THE SYSTEM AND REMOVED THE PROBE, AT WHICH POINT IT WAS OBSERVED THAT THE DISTAL TIP OF THE PROBE HAD DETACHED AND REMAINED INSIDE THE PATIENT. SUBSEQUENT IMAGING CONFIRMED THAT THE DETACHED TIP WAS RETAINED WITHIN THE LIVER AND APPEARED BROKEN JUST AFTER THE FEEDING POINT, EXHIBITING A SLIGHT BURNT ASPECT. BECAUSE THE COOLING PUMP CONTINUED TO OPERATE FOLLOWING THE TIP DETACHMENT, AIR ENTERED THROUGH THE BROKEN SECTION OF THE PROBE AND MIGRATED INTO THE LIVER VASCULATURE, ULTIMATELY REACHING THE RIGHT VENTRICLE. THE PATIENT RECEIVED ANTIBIOTIC TREATMENT AND WAS MONITORED IN THE ICU FOR TWO DAYS WITHOUT FURTHER INCIDENT. THE DETACHED TIP REMAINS IN THE PATIENT. NO LONG-TERM HEALTH CONSEQUENCES ARE ANTICIPATED. VARIAN HAS REQUESTED ADDITIONAL INFORMATION; THE DAMAGED PROBE WILL BE RETURNED FOR ANALYSIS. THE INVESTIGATION IS ONG
Patient problems: Air Embolism
MAUDE report 24250684 ↗InjuryNov 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DR. WAS USING SIEMENS MY NEEDLE LASER TO PLAN AND PLACE THE [MICROWAVE] NEEDLES. AFTER PLACEMENT, THE SCAN SHOW THAT THE PLACEMENT WAS NOT WHERE THE PHYSICIAN HAD INTENDED AND HAD CAUSED A PARTIAL PNEUMOTHORAX OF THE RIGHT LUNG. ISSUES WAS NOT WITH [MICROWAVE] SYSTEM. DR. PLACED A 10F DRAIN AND CASE PROCEEDED WITHOUT FURTHER INCIDENT.
Patient problems: Pneumothorax
MAUDE report 23597670 ↗
Other SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES devices
- Emprint CA20L2Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES2 recalls1 death43 reports
- Emprint CA15L2Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES2 recalls1 death41 reports
- Emprint CA30L2Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES2 recalls2 reports
- Emprint CA15L1Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES1 recall49 reports
- Solero 700106001-USAngiodynamics, Inc. · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES92 reports
- NEUWAVE PR15XTNeuwave Medical, Inc. · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES90 reports
- Emprint CA20L1Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES1 recall18 reports
- Emprint CA30L1Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES1 recall1 death9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page