IntelliBlate Ximitry Probe Assembly, 13G x 15cm IB-XPA-1315Recalls & reported problems

Varian Medical Systems, Inc. · IntelliBlate Ximitry probe assembly includes cassette with saline tubing, laser alignment disc, and microwave ablation probe 13G/15cm Probe QTY 1 · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00810563022292), received between Nov 19, 2025 and Mar 17, 2026 — classified as 3 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Appropriate Device Problem Term/Code Not Available1
  • Insufficient Device Problem Information1
  • Use of Device Problem1

Patient problems

  • Air Embolism1
  • Hematoma1
  • Liver Damage/Dysfunction1
  • Pneumothorax1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 17, 2026Appropriate Device Problem Term/Code Not Available

    SEE COMPLETED MW-2026-00009, MFR REPORT NUMBER 3008262715-2026-00006.

    Patient problems: Hematoma; Liver Damage/Dysfunction

    MAUDE report 24621355 ↗
  • InjuryFeb 3, 2026Use of Device Problem; Insufficient Device Problem Information

    DURING A MICROWAVE ABLATION PROCEDURE USING AN INTELLIBLATE IB-XPA-1315 PROBE, THE SYSTEM TRIGGERED A COOLING SYSTEM ALARM APPROXIMATELY ONE MINUTE AFTER ABLATION BEGAN AT 60 W. UPON RESPONDING TO THE ALARM, THE PHYSICIAN PERFORMED IMAGING AND IDENTIFIED A BOLUS OF AIR WITHIN THE LIVER AND IN THE RIGHT VENTRICLE OF THE HEART. THE PHYSICIAN IMMEDIATELY STOPPED THE SYSTEM AND REMOVED THE PROBE, AT WHICH POINT IT WAS OBSERVED THAT THE DISTAL TIP OF THE PROBE HAD DETACHED AND REMAINED INSIDE THE PATIENT. SUBSEQUENT IMAGING CONFIRMED THAT THE DETACHED TIP WAS RETAINED WITHIN THE LIVER AND APPEARED BROKEN JUST AFTER THE FEEDING POINT, EXHIBITING A SLIGHT BURNT ASPECT. BECAUSE THE COOLING PUMP CONTINUED TO OPERATE FOLLOWING THE TIP DETACHMENT, AIR ENTERED THROUGH THE BROKEN SECTION OF THE PROBE AND MIGRATED INTO THE LIVER VASCULATURE, ULTIMATELY REACHING THE RIGHT VENTRICLE. THE PATIENT RECEIVED ANTIBIOTIC TREATMENT AND WAS MONITORED IN THE ICU FOR TWO DAYS WITHOUT FURTHER INCIDENT. THE DETACHED TIP REMAINS IN THE PATIENT. NO LONG-TERM HEALTH CONSEQUENCES ARE ANTICIPATED. VARIAN HAS REQUESTED ADDITIONAL INFORMATION; THE DAMAGED PROBE WILL BE RETURNED FOR ANALYSIS. THE INVESTIGATION IS ONG

    Patient problems: Air Embolism

    MAUDE report 24250684 ↗
  • InjuryNov 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DR. WAS USING SIEMENS MY NEEDLE LASER TO PLAN AND PLACE THE [MICROWAVE] NEEDLES. AFTER PLACEMENT, THE SCAN SHOW THAT THE PLACEMENT WAS NOT WHERE THE PHYSICIAN HAD INTENDED AND HAD CAUSED A PARTIAL PNEUMOTHORAX OF THE RIGHT LUNG. ISSUES WAS NOT WITH [MICROWAVE] SYSTEM. DR. PLACED A 10F DRAIN AND CASE PROCEEDED WITHOUT FURTHER INCIDENT.

    Patient problems: Pneumothorax

    MAUDE report 23597670 ↗

Other SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES devices

All SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES devices →