Skip to content

Radixact™ Treatment Delivery System 1060000Recalls & reported problems

Accuray Incorporated · Radixact Treatment Delivery System · Class 2

All models of Radixact Treatment Delivery System (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00811376030009), received between Jun 7, 2022 and Aug 15, 2025 — classified as 4 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Computer Software Problem” (4 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 3, 2023Class IICompleted

    It is possible the upper roller bracket in the Radixact System gantry enclosure may not provide enough clearance for service personnel to use a jib crane for moving components while servicing the system, resulting in possible injury to the service personnel.

    Z-2489-2023 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Computer Software Problem4
  • Unintended System Motion4
  • Use of Device Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Circuit Failure1
  • Fracture1
  • Use of Incorrect Control/Treatment Settings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Insufficient Information3
  • Bone Fracture(s)1
  • Limb Fracture1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 15, 2025Fracture; Fracture

    BEAM STOP CRADLE HAS VISIBLE CRACK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22819803 ↗
  • InjuryMar 25, 2024Use of Incorrect Control/Treatment Settings; Use of Incorrect Control/Treatment Settings

    THERE WAS AN ALLEGATION THAT A PATIENT'S SCAN POSITION WAS DIFFERENT FROM THEIR TREATMENT POSITION.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18975247 ↗
  • InjuryFeb 9, 2024Use of Device Problem; Use of Device Problem

    THE PATIENT WAS INADVERTENTLY EXPOSED TO THE TREATMENT FRACTION FOR 5.2 SECONDS WHEN IMAGING WAS INTENDED.

    Patient problems: Insufficient Information

    MAUDE report 18677907 ↗
  • InjuryDec 15, 2023Use of Device Problem; Use of Device Problem

    THE PATIENT WAS INADVERTENTLY EXPOSED TO THE TREATMENT FRACTION FOR 5.2 SECONDS WHEN IMAGING WAS INTENDED.

    Patient problems: Insufficient Information

    MAUDE report 18336039 ↗
  • MalfunctionJun 29, 2023Computer Software Problem; Unintended System Motion; Computer Software Problem; Unintended System Motion

    THE TREATMENT DELIVERY CONSOLE (TDC) FROZE AT THE END OF KVCT AND THE GANTRY PANELS SHOWED "RADIATING".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17226678 ↗
  • MalfunctionJun 8, 2023Circuit Failure; Circuit Failure

    THE STATUS CONSOLE IS NOT STOPPING THE RADIATION FOLLOWING THE RX 3.0.2.1 UPDATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17086990 ↗
  • MalfunctionJun 6, 2023Computer Software Problem; Unintended System Motion; Computer Software Problem; Unintended System Motion

    THE TDC (TREATMENT DELIVERY CONSOLE) FROZE AT THE END OF KVCT AND THE GANTRY PANELS SHOWED "RADIATING".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17074756 ↗
  • MalfunctionMay 30, 2023Computer Software Problem; Unintended System Motion; Computer Software Problem; Unintended System Motion

    CUSTOMER ALLEGED THAT THE COUCH PROCEEDED WITH 99% DISPLAY ON TDC (TREATMENT DELIVERY CONSOLE) DURING KVCT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17021099 ↗
  • MalfunctionApr 18, 2023Computer Software Problem; Unintended System Motion; Computer Software Problem; Unintended System Motion

    GANTRY PANELS SHOW "RADIATING" WITH PATIENT IN THE GANTRY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16765518 ↗
  • InjuryJun 7, 2022Adverse Event Without Identified Device or Use Problem

    A FIELD SERVICE ENGINEER IN JAPAN SUFFERED A BROKEN FINGER WHEN USING THE GANTRY-MOUNTED CRANE/HOIST. THE FIELD SERVICE ENGINEER TRIED TO LIFT THE JIB CRANE AND HIS HAND GOT CAUGHT BETWEEN THE GANTRY AND THE JIB, CAUSING THE INJURY.

    Patient problems: Bone Fracture(s); Limb Fracture

    MAUDE report 14626006 ↗

Other Accelerator, Linear, Medical devices

All Accelerator, Linear, Medical devices →

Sources and updates