Elekta Medical Linear Accelerator 1552093Recalls & reported problems

Elekta Solutions AB · Versa HD™ · Class 2

All models of Elekta Medical Linear Accelerator (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
5

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05060191071574), received between Jan 17, 2023 and Aug 26, 2026 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Circuit Failure” (1 report).

5 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 14, 2025Class IIOpen, Classified

    Due to a manufacturing issue and electrical grounding of systems.

    Z-1164-2026 · Root cause: Process control · Recall details · FDA record ↗

  2. Oct 25, 2023Class IIOpen, Classified

    Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.

    Z-0445-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. May 5, 2023Class IIOpen, Classified

    A component within a limited number of Elekta Linacs has been realized to be non-compliant to IEC 60601

    Z-1665-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. Sep 28, 2021Class IITerminated

    There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.

    Z-0115-2022 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  5. Sep 17, 2021Class IITerminated

    There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.

    Z-0116-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Circuit Failure1
  • Computer Software Problem1
  • Detachment of Device or Device Component1
  • Gas/Air Leak1
  • Imprecision1
  • Melted1
  • Positioning Problem1
  • Radiation Output Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information3
  • Unintended Radiation Exposure1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • OtherAug 26, 2026Positioning Problem

    The customer asked which tolerances are applied for isocentric couch rotation, Gantry rotation and Collimator rotation during Beam interruption.

    Patient problems: Insufficient Information

    MAUDE report 50450719 ↗
  • MalfunctionMay 7, 2025Gas/Air Leak; Gas/Air Leak

    THE CUSTOMER REPORTED A SF6 LEAKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21974631 ↗
  • InjuryMar 18, 2025Circuit Failure; Circuit Failure

    THE CUSTOMER REPORTED THAT WHEN DELIVERING STEP AND SHOOT IMRT BEAMS ON THIS MACHINE, THE BEAM STARTED PRIOR TO THE MLC LEAVES BEING IN THEIR FINAL POSITION.

    Patient problems: Unintended Radiation Exposure

    MAUDE report 21632534 ↗
  • MalfunctionMar 6, 2025Imprecision; Imprecision

    THE CUSTOMER REPORTED THAT THE DOSE DISPLAYED INCORRECTLY DURING THE CALIBRATION PROCESS.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21534420 ↗
  • MalfunctionMay 17, 2023Computer Software Problem; Detachment of Device or Device Component; Computer Software Problem; Detachment of Device or Device Component

    THE CUSTOMER REPORTED A RATTLING SOUND WHEN THE GANTRY ROTATES.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16944784 ↗
  • MalfunctionJan 17, 2023Melted; Radiation Output Failure

    ADULT MALE NEEDING CHEMORADIATION FOR GLIOBLASTOMA; OUTPATIENT RADIATION ONCOLOGY; UNABLE TO RADIATE- VERSA TREATMENT MACHINE WENT DOWN. A WIRE MELTED INSIDE THE MACHINE, UNABLE TO TREAT PATIENTS. PATIENT WAS TAKEN FROM TREATMENT AREA AND ASKED TO WAIT IN WAITING AREA. VENDOR SERVICE WAS NOTIFIED ALONG WITH PHYSICIST. DETERMINED PART/REPAIR WAS NEEDED. PATIENTS WERE CANCELLED FOR REMAINDER OF DAY. ELEKTA WAS CALLED, FSE NOTIFIED, FSE IS ON HIS WAY FROM (B)(4). PART IS ORDERED. MANUFACTURER RESPONSE FOR ACCELERATOR, LINEAR, MEDICAL, ELEKTA MEDICAL LINEAR ACCELERATOR (PER SITE REPORTER). UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16183471 ↗