Skip to content

Elekta Medical Linear Accelerator 1552092Recalls & reported problems

Elekta Solutions AB · Elekta Infinity™ · Class 2

All models of Elekta Medical Linear Accelerator (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
5

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

5 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Dec 14, 2025Class IIOpen, Classified

    Due to a manufacturing issue and electrical grounding of systems.

    Z-1164-2026 · Root cause: Process control · Recall details · FDA record ↗

  2. Oct 25, 2023Class IIOpen, Classified

    Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.

    Z-0446-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. May 5, 2023Class IIOpen, Classified

    A component within a limited number of Elekta Linacs has been realized to be non-compliant to IEC 60601

    Z-1665-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. Sep 28, 2021Class IITerminated

    There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.

    Z-0115-2022 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  5. Sep 17, 2021Class IITerminated

    There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.

    Z-0116-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Other Accelerator, linear, medical devices

All Accelerator, linear, medical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page