Nuvasive Specialized Orthopedics, Inc. · Orthopedic intramedullary limb lengthening rod, antegrade femur (piriformis), 10.7 mm diameter, 305 mm length · Class 2 · Implant

All models of PRECICE Intramedullary Limb Lengthening System (551)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
10
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00812258020279), received between Apr 23, 2021 and Sep 30, 2025 — classified as 1 injury, 9 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 29, 2021Class IIOpen, Classified

    The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.

    Z-0319-2022 · Root cause: Device Design · Recall details · FDA record ↗

  2. Feb 12, 2021Class IIOpen, Classified

    Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.

    Z-1349-2021 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break4
  • Failure to Advance4
  • Mechanical Problem2
  • Material Twisted/Bent1

Patient problems

  • Failure of Implant6
  • No Clinical Signs, Symptoms or Conditions4
  • Pain2

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2025Break; Break

    IMPLANTED PRECISE NAIL BROKEN, REVISION PROCEDURE COMPLETED TO REMOVE AND REPLACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23183495 ↗
  • MalfunctionSep 4, 2025Failure to Advance; Material Twisted/Bent; Failure to Advance; Material Twisted/Bent

    IT WAS REPORTED THAT A REVISION SURGERY TOOK PLACE TO REMOVE AND REPLACE A PRECICE NAIL THAT FAILED TO DISTRACT.

    Patient problems: Failure of Implant

    MAUDE report 22965065 ↗
  • MalfunctionJan 21, 2025Break; Break

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED DUE TO THE PATIENT EXPERIENCING PAIN AND AN X-RAY IMAGE REVEALING A BROKEN NAIL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21200478 ↗
  • MalfunctionJan 23, 2024Failure to Advance; Failure to Advance

    INFORMATION WAS RECEIVED THAT THE NAIL DID NOT WORK UPON IMPLANTATION INSIDE THE PATIENT. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 18566688 ↗
  • MalfunctionOct 12, 2023Failure to Advance; Failure to Advance

    NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 17921711 ↗
  • InjuryAug 29, 2023Failure to Advance; Failure to Advance

    INFORMATION WAS RECEIVED THAT 15 DAYS AFTER THE IMPLANT PROCEDURE, THE SURGEON NOTICED THAT THE NAIL HAD DISTRACTED VERY LITTLE AND DECIDED THAT THE DISTRACTION OF THE NAIL SHOULD BE ACCELERATED; HOWEVER, AFTER FOLLOW UP, IT WAS NOTICED THAT DISTRACTION HAD COMPLETELY STOPPED. IT WAS THEN DECIDED THAT THE PATIENT SHOULD HAVE ANOTHER PROCEDURE WHERE THE OSTEOTOMY LINE WAS CHECKED AND 1CM LENGTHENING WAS ATTEMPTED AND WAS UNSUCCESSFUL. THE NAIL WAS THEN EXPLANTED, LENGTHENING WAS ATTEMPTED IN THE OPERATING ROOM AGAIN AND WAS SUCCESSFUL. THE NAIL WAS WAS THEN IMPLANTED AGAIN INTO THE PATIENT, HOWEVER, THE NAIL WOULD NOT DISTRACT. IT WAS FINALLY DECIDED THAT THEY WOULD IMPLANT A NEW NAIL.

    Patient problems: Failure of Implant

    MAUDE report 17646958 ↗
  • MalfunctionJul 28, 2022Break; Break

    INFORMATION WAS RECEIVED THAT THE PATIENT NOTICED ROTATIONAL SPINNING. THE NAIL HAS BEEN EXPLANTED. THE PATIENT REPORTED EXPERIENCING PAIN.

    Patient problems: Failure of Implant; Pain

    MAUDE report 15125577 ↗
  • MalfunctionJul 28, 2022Break; Break

    INFORMATION WAS RECEIVED THAT THE PATIENT NOTICED ROTATIONAL SPINNING. THE NAIL HAS BEEN EXPLANTED. THE PATIENT REPORTED EXPERIENCING PAIN.

    Patient problems: Failure of Implant; Pain

    MAUDE report 15125499 ↗
  • MalfunctionSep 20, 2021Mechanical Problem; Mechanical Problem

    NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12501931 ↗
  • MalfunctionApr 23, 2021Mechanical Problem; Mechanical Problem

    NO ADDITIONAL EVENT INFORMATION.