NUVASIVE PRECICE Intramedullary Limb Lengthening (IMLL) and… recallZ-1349-2021 · Nuvasive Specialized Orthopedics Inc · initiated Feb 12, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Nuvasive Specialized Orthopedics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.
FDA root cause category: Device Design
What the company is doing
On 02/11/2021, the firm started emailing an "URGENT FIELD SAFETY NOTICE" Notification to customers informing them that due to not having the full complement of biological assessments, the devices are not indicated for use in individuals under the age of 18 years old. Customers are informed to: -An Recalling Firm representative will be contacting your office or you to help with any questions or concerns. -Review, complete, sign and return the attached Consignee Confirmation Form accompanying this notification in accordance with the directions on the form. -If a patient has been previously implanted with a listed device and was under the age of 18, consultation may be warranted, at the discretion of the provider. -If a patient has been previously implanted with a listed device and is pregnant, becomes pregnant, or intends to become pregnant, consultation may be warranted, at the discretion of the provider. -Forward this notice to anyone in your facility that needs to be informed. -Direct any additional manufacturer inquiries to [email protected] -Report to NSO any adverse effect or product complaints related to the use of these devices to [email protected], whether or not those adverse effects are related to this FSN. As a reminder, the following guidelines according to the Instructions for Use, including, but not limited to: -The IFU should be consulted on an ongoing basis before and throughout patient treatment with Precice products. -These Precice family of devices are contraindicated in patients in which the Precice devices would cross joint spaces or open epiphyseal growth plates. -These Precice family of devices are contraindicated in patients unwilling or incapable of following postoperative care instructions. -Precice System devices should be removed after an implantation period of no more than one year. -Once the physician determines that the Precice/Short/Freedom/Unyte System has achieved its intended use and is no longer requir
Recall details
- Recall number
- Z-1349-2021
- Status
- Open, Classified
- Initiated by the company
- Feb 12, 2021
- Posted by the FDA
- Apr 5, 2021
- Recalling firm
- Nuvasive Specialized Orthopedics Inc
- Quantity
- 10,657 units
- Product code
- HSB
- FDA submissions
- K151131, K101997
- Distribution
- Worldwide distribution, including U.S. Nationwide.
Product description
NUVASIVE PRECICE Intramedullary Limb Lengthening (IMLL) and (Short) System End Cap for PRECICE Intramedullary Limb Lengthening System Implant; Packaged End Cap 8.5/10.7mm, +0mm; Packaged End Cap 8.5/10.7mm, +5mm; Packaged End Cap 8.5/10.7mm, +10mm; Packaged End Cap 8.5/10.7mm, +15mm; Packaged End Cap 8.5/10.7mm, +20mm; Packaged End Cap 12.5mm, +0mm; Packaged End Cap 12.5mm, +5mm; Packaged End Cap 12.5mm, +10mm; Packaged End Cap 12.5mm, +15mm; Packaged End Cap 12.5mm, +20mm; 3.5 mm Locking Screw, 20 mm Length; 3.5 mm Locking Screw 25 mm Length; 3.5 mm Locking Screw 30 mm Length; 3.5 mm Locking Screw 35 mm Length; 3.5 mm Locking Screw 40 mm Length; 3.5 mm Locking Screw 45 mm Length; 3.5 mm Locking Screw 50 mm Length; 3.5 mm Locking Screw 55 mm Length; 3.5 mm Locking Screw 60 mm Length; Locking Screw, 3.5x65mm; Locking Screw, 3.5x70mm; Locking Screw, 3.5x75mm; 4.0 mm Locking Screw 20 mm Length; 4.0 mm Locking Screw 25 mm Length; 4.0 mm Locking Screw 30 mm Length; 4.0 mm Locking Screw 35 mm Length; 4.0 mm Locking Screw 40 mm Length; 4.0 mm Locking Screw 45 mm Length; 4.0 mm Locking Screw 50 mm Length; 4.0 mm Locking Screw 55 mm Length; 4.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 20 mm Length; 5.0 mm Locking Screw 25 mm Length; 5.0 mm Locking Screw 30 mm Length; 5.0 mm Locking Screw 35 mm Length; 5.0 mm Locking Screw 40 mm Length; 5.0 mm Locking Screw 45 mm Length; 5.0 mm Locking Screw 50 mm Length; 5.0 mm Locking Screw 55 mm Length; 5.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 65 mm Length; 5.0 mm Locking Screw 70 mm Length; 5.0 mm Locking Screw 75 mm Length; 4.0 mm Locking Screw 20 mm Length; 4.0 mm Locking Screw 25 mm Length; 4.0 mm Locking Screw 30 mm Length; 4.0 mm Locking Screw 35 mm Length; 4.0 mm Locking Screw 40 mm Length; 4.0 mm Locking Screw 45 mm Length; 4.0 mm Locking Screw 50 mm Length; 4.0 mm Locking Screw 55 mm Length; 4.0 mm Locking Screw 60 mm Length; 4.0 mm X 65 mm Locking Screw, Peg; 4.0 mm X 70 mm Locking Screw, Peg; 4.0 mm X 75 mm Locking Screw, Peg; 4.0 mm X 80 mm Locking Screw, Peg; 4.0 mm X 85 mm Locking Screw, Peg; 4.0 mm X 90 mm Locking Screw, Peg; 4.0 mm X 95 mm Locking Screw, Peg; 4.0 mm X 100 mm Locking Screw, Peg; 5.0 mm Locking Screw 20 mm Length; 5.0 mm Locking Screw 25 mm Length; 5.0 mm Locking Screw 30 mm Length; 5.0 mm Locking Screw 35 mm Length; 5.0 mm Locking Screw 40 mm Length; 5.0 mm Locking Screw 45 mm Length; 5.0 mm Locking Screw 50 mm Length; 5.0 mm Locking Screw 55 mm Length; 5.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 65 mm Length; 5.0 mm Locking Screw 70 mm Length; 5.0 mm Locking Screw 75 mm Length; 5.0 mm X 80 mm Locking Screw, Peg; 5.0 mm X 85 mm Locking Screw, Peg; 5.0 mm X 90 mm Locking Screw, Peg; 5.0 mm X 95 mm Locking Screw, Peg; 5.0 mm X 100 mm Locking Screw, Peg; Orthopedic intramedullary limb lengthening rod, retrograde femur 10 degree, 10.7 mm diameter, 170 mm length; Orthopedic intramedullary limb lengthening rod, antegrade tibia, 10.7 mm diameter, 160 mm length; Orthopedic intramedullary limb lengthening rod, antegrade femur trochanteric, 10.7 mm diameter, 170 mm length; Orthopedic intramedullary limb lengthening rod, universal femur, 10.7 mm diameter, 170 mm length; Orthopedic intramedullary limb lengthening rod, retrograde femur, 10 degree, 10.7 mm diameter, 215 mm length; Orthopedic intramedullary limb lengthening rod, retrograde femur, 10 degree, 10.7 mm diameter, 230 mm length; Orthopedic intramedullary limb lengthening rod, antegrade femur (piriformis), 10.7 mm diameter, 215 mm length; Orthopedic intramedullary limb lengthening rod, antegrade femur (piriformis), 10.7 mm diameter, 230 mm length; Orthopedic intramedullary limb lengthening rod, antegrade tibia, 10.7 mm diameter, 195 mm length; Orthopedic intramedullary limb lengthening rod, antegrade tibia, 10.7 mm diameter, 215 mm length; Orthopedic intramedullary limb lengthening
Code information (models, lots, serial numbers)
Show the full code information (20k characters)
All Lots manufactured between April 2015 to February 2021. PRECICE Intramedullary Limb Lengthening (IMLL) Product Codes: End caps: CPA1-000; CPA2-000; CPA2-005; CPA2-010; CPA2-015; CPA2-020; CPA3-000; CPA3-005; CPA3-010; CPA3-015; CPA3-020; Locking screws: LSB3-020; LSB3-025; LSB3-030; LSB3-035; LSB3-040; LSB3-045; LSB3-050; LSB3-055; LSB3-060; LSB3-065; LSB3-070; LSB3-075; LSB4-020; LSB4-025; LSB4-030; LSB4-035; LSB4-040; LSB4-045; LSB4-050; LSB4-055; LSB4-060; LSB5-020; LSB5-025; LSB5-030; LSB5-035; LSB5-040; LSB5-045; LSB5-050; LSB5-055; LSB5-060; LSB5-065; LSB5-070; LSB5-075; LSC4-020; LSC4-025; LSC4-030; LSC4-035; LSC4-040; LSC4-045; LSC4-050; LSC4-055; LSC4-060; LSC4-065; LSC4-070; LSC4-075; LSC4-080; LSC4-085; LSC4-090; LSC4-095; LSC4-100; LSC5-020; LSC5-025; LSC5-030; LSC5-035; LSC5-040; LSC5-045; LSC5-050; LSC5-055; LSC5-060; LSC5-065; LSC5-070; LSC5-075; LSC5-080; LSC5-085; LSC5-090; LSC5-095; LSC5-100; Limb lengthening intramedullary rods 10.7 mm diameter: P10.7-30H170; P10.7-30J160; P10.7-30K170; P10.7-30U170; P10.7-50A215; P10.7-50A230; P10.7-50B215; P10.7-50B230; P10.7-50C195; P10.7-50C215; P10.7-50C230; P10.7-50D215; P10.7-50D230; P10.7-50E215; P10.7-50E230; P10.7-50H190; P10.7-50J180; P10.7-50K190; P10.7-50M175; P10.7-50N160; P10.7-50P175; P10.7-50Q155; P10.7-50U190; P10.7-50V215; P10.7-50V230; P10.7-50X215; P10.7-50X230; P10.7-80A245; P10.7-80A275; P10.7-80A305; P10.7-80A335; P10.7-80A365; P10.7-80B245; P10.7-80B275; P10.7-80B305; P10.7-80B335; P10.7-80B365; P10.7-80C245; P10.7-80C275; P10.7-80C305; P10.7-80C335; P10.7-80C365; P10.7-80D245; P10.7-80D275; P10.7-80D305; P10.7-80D335; P10.7-80D365; P10.7-80E245; P10.7-80E275; P10.7-80E305; P10.7-80E335; P10.7-80E365; P10.7-80V245; P10.7-80V275; P10.7-80V305; P10.7-80V335; P10.7-80V365; P10.7-80X245; P10.7-80X275; P10.7-80X305; P10.7-80X335; P10.7-80X365; Limb lengthening intramedullary rods 12.5 mm diameter: P12.5-30H170; P12.5-30J160; P12.5-30K170; P12.5-30U170; P12.5-50A215; P12.5-50A230; P12.5-50B215; P12.5-50B230; P12.5-50C195; P12.5-50C215; P12.5-50C230; P12.5-50D215; P12.5-50D230; P12.5-50E215; P12.5-50E230; P12.5-50H190; P12.5-50J180; P12.5-50K190; P12.5-50U190; P12.5-50V215; P12.5-50V230; P12.5-50X215; P12.5-50X230; P12.5-80A245; P12.5-80A275; P12.5-80A305; P12.5-80A335; P12.5-80A365; P12.5-80B245; P12.5-80B275; P12.5-80B305; P12.5-80B335; P12.5-80B365; P12.5-80C245; P12.5-80C275; P12.5-80C305; P12.5-80C335; P12.5-80C365; P12.5-80D245; P12.5-80D275; P12.5-80D305; P12.5-80D335; P12.5-80D365; P12.5-80E245; P12.5-80E275; P12.5-80E305; P12.5-80E335; P12.5-80E365; P12.5-80V245; P12.5-80V275; P12.5-80V305; P12.5-80V335; P12.5-80V365; P12.5-80X245; P12.5-80X275; P12.5-80X305; P12.5-80X335; P12.5-80X365; Limb lengthening intramedullary rods 8.5 mm diameter: P8.5-30H170; P8.5-30J160; P8.5-30K170; P8.5-30U170; P8.5-50A215; P8.5-50A230; P8.5-50B215; P8.5-50B230; P8.5-50C195; P8.5-50C215; P8.5-50C230; P8.5-50D215; P8.5-50D230; P8.5-50E215; P8.5-50E230; P8.5-50H190; P8.5-50J180; P8.5-50K190; P8.5-50M165; P8.5-50N150 P8.5-50P165; P8.5-50Q155; P8.5-50U190; P8.5-50V215; P8.5-50V230; P8.5-50X215; P8.5-50X230; P8.5-80A245; P8.5-80A275; P8.5-80A305; P8.5-80A335; P8.5-80A365; P8.5-80B245; P8.5-80B275; P8.5-80B305; P8.5-80B335; P8.5-80B365; P8.5-80C245; P8.5-80C275; P8.5-80C305; P8.5-80C335; P8.5-80C365; P8.5-80D245; P8.5-80D275; P8.5-80D305; P8.5-80D335; P8.5-80D365; P8.5-80E245; P8.5-80E275; P8.5-80E305; P8.5-80E335; P8.5-80E365; P8.5-80V245; P8.5-80V275; P8.5-80V305; P8.5-80V335; P8.5-80V365; P8.5-80X245; P8.5-80X275; P8.5-80X305; P8.5-80X335; P8.5-80X365; Limb lengthening intramedullary rods 10.5 mm diameter: PR10.5-50A215; PR10.5-50A230; PR10.5-50B215; PR10.5-50B230; PR10.5-50C195; PR10.5-50C215; PR10.5-50C230; PR10.5-50D215; PR10.5-50D230; PR10.5-50E215; PR10.5-50E230; PR10.5-80A245; PR10.5-80A275; PR10.5-80A305; PR10.5-80A335; PR10.5-80A365; PR10.5-80B245; PR10.5-80B275; PR10.5-80B305; PR10.5-80B335; PR10.5-80B365; PR10.5-80C245; PR10.5-80C275; PR10.5-80C305; PR10.5-80C335; PR10.5-80C365; PR10.5-80D245; PR10.5-80D275; PR10.5-80D305; PR10.5-80D335; PR10.5-80D365; PR10.5-80E245; PR10.5-80E275; PR10.5-80E305; PR10.5-80E335; PR10.5-80E365; Limb lengthening intramedullary rods 11.5 mm diameter: PR11.5-50A215; PR11.5-50A230; PR11.5-50B215; PR11.5-50B230; PR11.5-50C195; PR11.5-50C215; PR11.5-50C230; PR11.5-50D215; PR11.5-50D230; PR11.5-50E215; PR11.5-50E230; PR11.5-80A245; PR11.5-80A275; PR11.5-80A305; PR11.5-80A335; PR11.5-80A365; PR11.5-80B245; PR11.5-80B275; PR11.5-80B305; PR11.5-80B335; PR11.5-80B365; PR11.5-80C245; PR11.5-80C275; PR11.5-80C305; PR11.5-80C335; PR11.5-80C365; PR11.5-80D245; PR11.5-80D275; PR11.5-80D305; PR11.5-80D335; PR11.5-80D365; PR11.5-80E245; PR11.5-80E275; PR11.5-80E305; PR11.5-80E335; PR11.5-80E365; Limb lengthening intramedullary rods 12.5 mm diameter: PR12.5-50A215; PR12.5-50A230; PR12.5-50B215; PR12.5-50B230; PR12.5-50C195; PR12.5-50C215; PR12.5-50C230; PR12.5-50D215; PR12.5-50D230; PR12.5-50E215; PR12.5-50E230; PR12.5-80A245; PR12.5-80A275; PR12.5-80A305; PR12.5-80A335; PR12.5-80A365; PR12.5-80B245; PR12.5-80B275; PR12.5-80B305; PR12.5-80B335; PR12.5-80B365; PR12.5-80C245; PR12.5-80C275; PR12.5-80C305; PR12.5-80C335; PR12.5-80C365; PR12.5-80D245; PR12.5-80D275; PR12.5-80D305; PR12.5-80D335; PR12.5-80D365; PR12.5-80E245; PR12.5-80E275; PR12.5-80E305; PR12.5-80E335; PR12.5-80E365; Limb lengthening intramedullary rods 9.0 mm diameter: PR9.0-50A215; PR9.0-50A230; PR9.0-50B215; PR9.0-50B230; PR9.0-50C195; PR9.0-50C215; PR9.0-50C230; PR9.0-50D215; PR9.0-50D230; PR9.0-50E215; PR9.0-50E230; PR9.0-80A245; PR9.0-80A275; PR9.0-80A305; PR9.0-80A335; PR9.0-80A365; PR9.0-80B245; PR9.0-80B275; PR9.0-80B305; PR9.0-80B335; PR9.0-80B365; PR9.0-80C245; PR9.0-80C275; PR9.0-80C305; PR9.0-80C335; PR9.0-80C365; PR9.0-80D245; PR9.0-80D275; PR9.0-80D305; PR9.0-80D335; PR9.0-80D365; PR9.0-80E245; PR9.0-80E275; PR9.0-80E305; PR9.0-80E335; PR9.0-80E365; Limb lengthening intramedullary rods 9.5 mm diameter: PR9.5-50A215; PR9.5-50A230; PR9.5-50B215; PR9.5-50B230; PR9.5-50C195; PR9.5-50C215; PR9.5-50C230; PR9.5-50D215; PR9.5-50D230; PR9.5-50E215; PR9.5-50E230; PR9.5-80A245; PR9.5-80A275; PR9.5-80A305; PR9.5-80A335; PR9.5-80A365; PR9.5-80B245; PR9.5-80B275; PR9.5-80B305; PR9.5-80B335; PR9.5-80B365; PR9.5-80C245; PR9.5-80C275; PR9.5-80C305; PR9.5-80C335; PR9.5-80C365; PR9.5-80D245; PR9.5-80D275; PR9.5-80D305; PR9.5-80D335; PR9.5-80D365; PR9.5-80E245; PR9.5-80E275; PR9.5-80E305; PR9.5-80E335; PR9.5-80E365; Trauma intramedullary rods 10.7 mm diameter: T10.7-50A205; T10.7-50A225; T10.7-50A240; T10.7-50B205; T10.7-50B225; T10.7-50B240; T10.7-50C205; T10.7-50C225; T10.7-50C240; T10.7-50D205; T10.7-50D225; T10.7-50D240; T10.7-50E205; T10.7-50E225; T10.7-50E240; T10.7-80A255; T10.7-80A285; T10.7-80A315; T10.7-80A345; T10.7-80A375; T10.7-80B255; T10.7-80B285; T10.7-80B315; T10.7-80B345; T10.7-80B375; T10.7-80C255; T10.7-80C285; T10.7-80C315; T10.7-80C345; T10.7-80C375; T10.7-80D255; T10.7-80D285; T10.7-80D315; T10.7-80D345; T10.7-80D375; T10.7-80E255; T10.7-80E285; T10.7-80E315; T10.7-80E345; T10.7-80E375; Trauma intramedullary rods 12.5 mm diameter: T12.5-50A205; T12.5-50A225; T12.5-50A240; T12.5-50B205; T12.5-50B225; T12.5-50B240; T12.5-50C205; T12.5-50C225; T12.5-50C240; T12.5-50D205; T12.5-50D225; T12.5-50D240; T12.5-50E205; T12.5-50E225; T12.5-50E240; T12.5-80A255; T12.5-80A285; T12.5-80A315; T12.5-80A345; T12.5-80A375; T12.5-80B255; T12.5-80B285; T12.5-80B315; T12.5-80B345; T12.5-80B375; T12.5-80C255; T12.5-80C285; T12.5-80C315; T12.5-80C345; T12.5-80C375; T12.5-80D255; T12.5-80D285; T12.5-80D315; T12.5-80D345; T12.5-80D375; T12.5-80E255; T12.5-80E285; T12.5-80E315; T12.5-80E345; T12.5-80E375; Trauma intramedullary rods 8.5 mm diameter: T8.5-50A205; T8.5-50A225; T8.5-50A240; T8.5-50B205; T8.5-50B225; T8.5-50B240; T8.5-50C205; T8.5-50C225; T8.5-50C240; T8.5-50D205; T8.5-50D225; T8.5-50D240; T8.5-50E205; T8.5-50E225; T8.5-50E240; T8.5-80A255; T8.5-80A285; T8.5-80A315; T8.5-80A345; T8.5-80A375; T8.5-80B255; T8.5-80B285; T8.5-80B315; T8.5-80B345; T8.5-80B375; T8.5-80C255; T8.5-80C285; T8.5-80C315; T8.5-80C345; T8.5-80C375; T8.5-80D255; T8.5-80D285; T8.5-80D315; T8.5-80D345; T8.5-80D375; T8.5-80E255; T8.5-80E285; T8.5-80E315; T8.5-80E345; T8.5-80E375; Locking screws: TSA4.5-020; TSA4.5-025; TSA4.5-030; TSA4.5-035; TSA4.5-040; TSA4.5-045; TSA4.5-050; TSA4.5-055; TSA4.5-060; TSA4.5-065; TSA4.5-070; TSA4.5-075; TSA4-020; TSA4-022.5; TSA4-025; TSA4-027.5; TSA4-030; TSA4-032.5; TSA4-035; TSA4-037.5; TSA4-040; TSA4-042.5; TSA4-045; TSA4-047.5; TSA4-050; TSA4-055; TSA4-060; TSA4-065; TSA4-070; TSA4-075; TSA4-080; TSA4-085; TSA4-090; TSA4-095; TSA4-100; TSA4-100; TSA5-020; TSA5-020; TSA5-022.5; TSA5-022.5; TSA5-025; TSA5-025; TSA5-027.5; TSA5-027.5; TSA5-030; TSA5-030; TSA5-032.5; TSA5-032.5; TSA5-035; TSA5-035; TSA5-037.5; TSA5-037.5; TSA5-040; TSA5-040; TSA5-042.5; TSA5-042.5; TSA5-045; TSA5-045; TSA5-047.5; TSA5-047.5; TSA5-050; TSA5-050; TSA5-052.5; TSA5-055; TSA5-055; TSA5-057.5; TSA5-060; TSA5-060; TSA5-062.5; TSA5-065; TSA5-065; TSA5-067.5; TSA5-070; TSA5-070; TSA5-072.5; TSA5-075; TSA5-075; TSA5-080; TSA5-080; TSA5-085; TSA5-085; TSA5-090; TSA5-090; TSA5-095; TSA5-095; TSA5-100; TSA5-100; TSA5-105; TSA5-105; TSA5-110; TSA5-110; GTIN Numbers: 00812258020002; 00887517001139; 00887517001146; 00887517001153; 00887517001160; 00887517001177; 00887517001184; 00887517001191; 00887517001207; 00887517001214; 00887517001221; 00812258021337; 00812258021344; 00812258021351; 00812258021368; 00812258021375; 00812258021382; 00812258021399; 00812258021405; 00812258021412; 00887517936233; 00887517936240; 00887517936257; 00856719002763; 00856719002770; 00856719002787; 00856719002794; 00856719002800; 00856719002817; 00856719002824; 00856719002831; 00856719002848; 00856719002886; 00856719002893; 00856719002909; 00856719002916; 00856719002923; 00856719002930; 00856719002947; 00856719002954; 00856719002961; 00856719002978; 00856719002985; 00856719002992; 00812258021429; 00812258021436; 00812258021443; 00812258021450; 00812258021467; 00812258021474; 00812258021481; 00812258021498; 00812258021504; 00887517000958; 00887517000965; 00887517000972; 00887517000989; 00887517000996; 00887517001009; 00887517001016; 00887517001023; 00812258021511; 00812258021528; 00812258021535; 00812258021542; 00812258021559; 00812258021566; 00812258021573; 00812258021580; 00812258021597; 00812258021603; 00812258021610; 00812258021627; 00887517001030; 00887517001047; 00887517001054; 00887517001061; 00887517001078; 00812258026073; 00812258026134; 00812258026196; 00812258026257; 00812258020057; 00812258020064; 00812258020231; 00812258020248; 00812258020422; 00812258020439; 00812258020446; 00812258020620; 00812258020637; 00812258020804; 00812258020811; 00812258026080; 00812258026141; 00812258026202; 00812258028244; 00812258028268; 00812258028282; 00812258028220; 00812258026264; 00887517934819; 00887517934826; 00887517934833; 00887517934840; 00812258020071; 00812258020088; 00812258020095; 00812258020101; 00812258020118; 00812258020255; 00812258020262; 00812258020279; 00812258020286; 00812258020293; 00812258020453; 00812258020460; 00812258020477; 00812258020484; 00812258020491; 00812258020644; 00812258020651; 00812258020668; 00812258020675; 00812258020682; 00812258020828; 00812258020835; 00812258020842; 00812258020859; 00812258020866; 00887517934857; 00887517934864; 00887517934871; 00887517934888; 00887517934895; 00887517935502; 00887517935519; 00887517935526; 00887517935533; 00887517935540; 00812258026097; 00812258026158; 00812258026219; 00812258026271; 00812258020125; 00812258020132; 00812258020309; 00812258020316; 00812258020507; 00812258020514; 00812258020521; 00812258020699; 00812258020705; 00812258020873; 00812258020880; 00812258026103; 00812258026165; 00812258026226; 00812258026288; 00887517935557; 00887517935564; 00887517935571; 00887517935588; 00812258020149; 00812258020156; 00812258020163; 00812258020170; 00812258020187; 00812258020323; 00812258020330; 00812258020347; 00812258020354; 00812258020361; 00812258020538; 00812258020545; 00812258020552; 00812258020569; 00812258020576; 00812258020712; 00812258020729; 00812258020736; 00812258020743; 00812258020750; 00812258020897; 00812258020903; 00812258020910; 00812258020927; 00812258020934; 00887517935595; 00887517935601; 00887517935618; 00887517935625; 00887517935632; 00887517935649; 00887517935656; 00887517935663; 00887517935670; 00887517935687; 00812258026059; 00812258026110; 00812258026172; 00812258026233; 00812258020019; 00812258020026; 00812258020194; 00812258020200; 00812258020378; 00812258020385; 00812258020392; 00812258020583; 00812258020590; 00812258020767; 00812258020774; 00812258026066; 00812258026127; 00812258026189; 00812258028237; 00812258028251; 00812258028275; 00812258028213; 00812258026240; 00887517935694; 00887517935700; 00887517935717; 00887517935724; 00812258020033; 00812258020040; 00812258022358; 00812258022365; 00812258022372; 00812258020217; 00812258020224; 00812258022389; 00812258022396; 00812258022402; 00812258020408; 00812258020415; 00812258022419; 00812258022426; 00812258022433; 00812258020606; 00812258020613; 00812258022440; 00812258022457; 00812258022464; 00812258020781; 00812258020798; 00812258022471; 00812258022488; 00812258022495; 00887517935731; 00887517935748; 00887517935755; 00887517935762; 00887517935779; 00887517935786; 00887517935793; 00887517935809; 00887517935816; 00887517935823; 00812258025359; 00812258025366; 00812258024253; 00812258024260; 00812258024956; 00812258024963; 00812258024970; 00812258024604; 00812258024611; 00812258025700; 00812258025717; 00812258025373; 00812258025380; 00812258025397; 00812258025403; 00812258025410; 00812258024277; 00812258024284; 00812258024291; 00812258024307; 00812258024314; 00812258024987; 00812258024994; 00812258025007; 00812258025014; 00812258025021; 00812258024628; 00812258024635; 00812258024642; 00812258024659; 00812258024666; 00812258025724; 00812258025731; 00812258025748; 00812258025755; 00812258025762; 00812258025427; 00812258025434; 00812258024321; 00812258024338; 00812258025038; 00812258025045; 00812258025052; 00812258024673; 00812258024680; 00812258025779; 00812258025786; 00812258025441; 00812258025458; 00812258025465; 00812258025472; 00812258025489; 00812258024345; 00812258024352; 00812258024369; 00812258024376; 00812258024383; 00812258025069; 00812258025076; 00812258025083; 00812258025090; 00812258025106; 00812258024697; 00812258024703; 00812258024710; 00812258024727; 00812258024734; 00812258025793; 00812258025809; 00812258025816; 00812258025823; 00812258025830; 00812258025632; 00812258025649; 00812258024390; 00812258024406; 00812258025113; 00812258025120; 00812258025137; 00812258024741; 00812258024758; 00812258025984; 00812258025991; 00812258025656; 00812258025663; 00812258025670; 00812258025687; 00812258025694; 00812258024413; 00812258024420; 00812258024437; 00812258024444; 00812258024451; 00812258025144; 00812258025151; 00812258025168; 00812258025175; 00812258025182; 00812258024765; 00812258024772; 00812258024789; 00812258024796; 00812258024802; 00812258026004; 00812258026011; 00812258026028; 00812258026035; 00812258026042; 00812258025496; 00812258025502; 00812258024468; 00812258024475; 00812258025199; 00812258025205; 00812258025212; 00812258024819; 00812258024826; 00812258025847; 00812258025854; 00812258025519; 00812258025526; 00812258025533; 00812258025540; 00812258025557; 00812258024482; 00812258024499; 00812258024505; 00812258024512; 00812258024529; 00812258025229; 00812258025236; 00812258025243; 00812258025250; 00812258025267; 00812258024833; 00812258024840; 00812258024857; 00812258024864; 00812258024871; 00812258025861; 00812258025878; 00812258025885; 00812258025892; 00812258025908; 00812258025564; 00812258025571; 00812258024536; 00812258024543; 00812258025274; 00812258025281; 00812258025298; 00812258024888; 00812258024895; 00812258025915; 00812258025922; 00812258025588; 00812258025595; 00812258025601; 00812258025618; 00812258025625; 00812258024550; 00812258024567; 00812258024574; 00812258024581; 00812258024598; 00812258025304; 00812258025311; 00812258025328; 00812258025335; 00812258025342; 00812258024901; 00812258024918; 00812258024925; 00812258024932; 00812258024949; 00812258025939; 00812258025946; 00812258025953; 00812258025960; 00812258025977; 00812258023836; 00812258023843; 00812258023850; 00812258023119; 00812258023126; 00812258023133; 00812258023591; 00812258023607; 00812258023614; 00812258023355; 00812258023362; 00812258023379; 00812258024079; 00812258024086; 00812258024093; 00812258023867; 00812258023874; 00812258023881; 00812258023898; 00812258023904; 00812258023140; 00812258023157; 00812258023164; 00812258023171; 00812258023188; 00812258023621; 00812258023638; 00812258023645; 00812258023652; 00812258023669; 00812258023386; 00812258023393; 00812258023409; 00812258023416; 00812258023423; 00812258024109; 00812258024116; 00812258024123; 00812258024130; 00812258024147; 00812258023911; 00812258023928; 00812258023935; 00812258023195; 00812258023201; 00812258023218; 00812258023676; 00812258023683; 00812258023690; 00812258023430; 00812258023447; 00812258023454; 00812258024154; 00812258024161; 00812258024178; 00812258023942; 00812258023959; 00812258023966; 00812258023973; 00812258023980; 00812258023225; 00812258023232; 00812258023249; 00812258023256; 00812258023263; 00812258023706; 00812258023713; 00812258023720; 00812258023737; 00812258023744; 00812258023461; 00812258023478; 00812258023485; 00812258023492; 00812258023508; 00812258024185; 00812258024192; 00812258024208; 00812258024215; 00812258024222; 00812258023751; 00812258023768; 00812258023775; 00812258023034; 00812258023041; 00812258023058; 00812258023515; 00812258023522; 00812258023539; 00812258023270; 00812258023287; 00812258023294; 00812258023997; 00812258024000; 00812258024017; 00812258023782; 00812258023799; 00812258023805; 00812258023812; 00812258023829; 00812258023065; 00812258023072; 00812258023089; 00812258023096; 00812258023102; 00812258023546; 00812258023553; 00812258023560; 00812258023577; 00812258023584; 00812258023300; 00812258023317; 00812258023324; 00812258023331; 00812258023348; 00812258024024; 00812258024031; 00812258024048; 00812258024055; 00812258024062; 00887517936011; 00887517936028; 00887517936035; 00887517936042; 00887517936059; 00887517936066; 00887517936073; 00887517936080; 00887517936097; 00887517936202; 00887517936219; 00887517936226; 00887517000033; 00887517000040; 00887517000057; 00887517000064; 00887517000071; 00887517000088; 00887517000095; 00887517000101; 00887517000118; 00887517000125; 00887517000132; 00887517000149; 00887517000156; 00887517000163; 00887517000170; 00887517000187; 00887517000194; 00887517000200; 00887517000217; 00887517000224; 00887517000231; 00887517000248; 00887517000255; 00887517000255; 00812258027728; 00887517000262; 00812258027735; 00887517000279; 00812258027742; 00887517000286; 00812258027759; 00887517000293; 00812258027766; 00887517000309; 00812258027773; 00887517000316; 00812258027780; 00887517000323; 00812258027797; 00887517000330; 00812258027803; 00887517000347; 00812258027810; 00887517000354; 00812258027827; 00887517000361; 00812258027834; 00887517000378; 00812258027841; 00887517000385; 00812258027858; 00812258027865; 00887517000392; 00812258027872; 00812258027889; 00887517000408; 00812258027896; 00812258027902; 00887517000415; 00812258027919; 00812258027926; 00887517000422; 00812258027933; 00812258027940; 00887517000439; 00812258027957; 00887517000446; 00812258027964; 00887517000453; 00812258027971; 00887517000460; 00812258027988; 00887517000477; 00812258027995; 00887517000484; 00812258028008; 00812258028008; 00812258028015; 00812258028015;
Devices named in this recall
625 devices in the FDA device registry whose identifier appears in the recall's code information; the 100 most reported are shown.
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-020Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-025Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-030Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-035Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-055Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-060Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-065Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-070Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-020Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-025Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-030Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-035Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-055Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-060Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-065Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-070Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-080Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-085Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-090Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-095Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-100Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Trauma Nail System T10.7-80A375Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System TSA5-065Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System TSA5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System CPA2-005Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System CPA1-000Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSB3-020Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSB3-025Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSB3-030Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSB3-035Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSB3-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSB3-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSB3-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSB3-055Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSB3-060Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-020Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-025Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-030Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-035Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-055Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-060Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T8.5-50B205Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T8.5-50B225Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T8.5-50B240Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T8.5-80B255Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T8.5-80B285Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T8.5-80B315Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T8.5-80B345Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T8.5-80B375Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T10.7-50B205Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T10.7-50B225Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T10.7-50B240Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T10.7-80B255Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T10.7-80B285Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T10.7-80B315Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T10.7-80B345Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T10.7-80B375Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T12.5-50B205Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T12.5-50B225Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T12.5-50B240Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T12.5-80B255Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T12.5-80B285Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T12.5-80B315Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T12.5-80B345Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T12.5-80B375Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T8.5-50D205Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T8.5-50D225Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T8.5-50D240Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T8.5-80D255Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T8.5-80D285Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T8.5-80D315Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T8.5-80D345Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T8.5-80D375Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T10.7-50D205Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T10.7-50D225Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T10.7-50D240Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T10.7-80D255Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T10.7-80D285Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T10.7-80D315Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T10.7-80D345Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T10.7-80D375Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T12.5-50D205Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T12.5-50D225Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T12.5-50D240Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T12.5-80D255Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T12.5-80D285Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T12.5-80D315Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T12.5-80D345Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Trauma Nail System T12.5-80D375Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 87353).
- Z-1350-2021 · Open, ClassifiedNUVASIVE PRECICE UNYTE System (Humerus and Tibia): PRECICE Trauma Nail System/PRECICE Unyte-Humerus Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 165 mm lengt…
- Z-1351-2021 · Open, ClassifiedNUVASIVE FREEDOM: End Cap for PRECICE Intramedullary Limb Lengthening System Implant; Packaged End Cap 8.5/10.7mm, +0mm; Packaged End Cap 8.5/10.7mm, +5mm; Packaged End Cap 8…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page