Skip to content

PRECICE® Intramedullary Limb Lengthening System P12.5-80B305Recalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · Orthopedic intramedullary limb lengthening rod, antegrade femur (piriformis), 12.5 mm diameter, 305 mm length · Class 2 · Implant

All models of PRECICE Intramedullary Limb Lengthening System (551)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00812258020347), received between Mar 8, 2022 and Apr 23, 2025 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 29, 2021Class IIOpen, Classified

    The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.

    Z-0319-2022 · Root cause: Device Design · Recall details · FDA record ↗

  2. Feb 12, 2021Class IIOpen, Classified

    Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.

    Z-1349-2021 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break6
  • Failure to Advance3
  • Fracture2
  • Detachment of Device or Device Component1
  • Mechanical Problem1
  • Unintended System Motion1

Patient problems

  • Failure of Implant7
  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 23, 2025Break; Fracture; Break; Fracture

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE OCCURRED DUE TO THE NAIL BEING DAMAGED SURROUNDING THE MALE-FEMALE JUNCTION SITE. THE REVISION PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21894440 ↗
  • MalfunctionApr 23, 2025Break; Fracture; Break; Fracture

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE OCCURRED DUE TO THE NAIL BEING DAMAGED SURROUNDING THE MALE-FEMALE JUNCTION SITE. THE REVISION PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21894324 ↗
  • MalfunctionAug 24, 2023Failure to Advance; Failure to Advance

    NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 17612905 ↗
  • MalfunctionJul 19, 2023Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    INFORMATION WAS RECEIVED THAT THE NAIL BROKE AFTER ONE MONTH. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 17352924 ↗
  • MalfunctionJun 2, 2023Break; Unintended System Motion; Break; Unintended System Motion

    INFORMATION WAS RECEIVED THAT A FEW MONTHS AGO SOMETHING SNAPPED IN THE PATIENT'S LEG WHILE STRETCHING AND THE PHYSICIAN SUSPECTS A CROWN BREAKAGE. IT WAS ADDITIONALLY REPORTED THAT THE NAIL HAS RECOILED AND CAUSED A ROTATIONAL DEFORMITY. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 17052303 ↗
  • MalfunctionJan 18, 2023Break; Failure to Advance; Break; Failure to Advance

    IT WAS REPORTED THAT THE NAIL DID NOT EXTEND AND WAS REPLACED WITH A NEW ONE.

    Patient problems: Failure of Implant

    MAUDE report 16196538 ↗
  • MalfunctionJan 18, 2023Break; Failure to Advance; Break; Failure to Advance

    NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 16196533 ↗
  • MalfunctionMar 8, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE FEMALE PART OF THE MALE/FEMALE JUNCTION BROKE ON THE NAIL. NO PATIENT ADVERSE EVENT WAS REPORTED. THE PATIENT RECEIVED A NEW NAIL WITH NO COMPLICATIONS AND IS CONTINUING HIS LENGTHENING SESSIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13694771 ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →

Sources and updates