PRECICE® Intramedullary Limb Lengthening System P12.5-80B305Recalls & reported problems
Nuvasive Specialized Orthopedics, Inc. · Orthopedic intramedullary limb lengthening rod, antegrade femur (piriformis), 12.5 mm diameter, 305 mm length · Class 2 · Implant
All models of PRECICE Intramedullary Limb Lengthening System (551)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00812258020347), received between Mar 8, 2022 and Apr 23, 2025 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 29, 2021Class IIOpen, Classified
The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.
Z-0319-2022 · Root cause: Device Design · Recall details · FDA record ↗
- Feb 12, 2021Class IIOpen, Classified
Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.
Z-1349-2021 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Break6
- Failure to Advance3
- Fracture2
- Detachment of Device or Device Component1
- Mechanical Problem1
- Unintended System Motion1
Patient problems
- Failure of Implant7
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2025Break; Fracture; Break; Fracture
INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE OCCURRED DUE TO THE NAIL BEING DAMAGED SURROUNDING THE MALE-FEMALE JUNCTION SITE. THE REVISION PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions
MAUDE report 21894440 ↗MalfunctionApr 23, 2025Break; Fracture; Break; Fracture
INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE OCCURRED DUE TO THE NAIL BEING DAMAGED SURROUNDING THE MALE-FEMALE JUNCTION SITE. THE REVISION PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions
MAUDE report 21894324 ↗MalfunctionAug 24, 2023Failure to Advance; Failure to Advance
NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 17612905 ↗MalfunctionJul 19, 2023Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
INFORMATION WAS RECEIVED THAT THE NAIL BROKE AFTER ONE MONTH. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 17352924 ↗MalfunctionJun 2, 2023Break; Unintended System Motion; Break; Unintended System Motion
INFORMATION WAS RECEIVED THAT A FEW MONTHS AGO SOMETHING SNAPPED IN THE PATIENT'S LEG WHILE STRETCHING AND THE PHYSICIAN SUSPECTS A CROWN BREAKAGE. IT WAS ADDITIONALLY REPORTED THAT THE NAIL HAS RECOILED AND CAUSED A ROTATIONAL DEFORMITY. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 17052303 ↗MalfunctionJan 18, 2023Break; Failure to Advance; Break; Failure to Advance
IT WAS REPORTED THAT THE NAIL DID NOT EXTEND AND WAS REPLACED WITH A NEW ONE.
Patient problems: Failure of Implant
MAUDE report 16196538 ↗MalfunctionJan 18, 2023Break; Failure to Advance; Break; Failure to Advance
MalfunctionMar 8, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE FEMALE PART OF THE MALE/FEMALE JUNCTION BROKE ON THE NAIL. NO PATIENT ADVERSE EVENT WAS REPORTED. THE PATIENT RECEIVED A NEW NAIL WITH NO COMPLICATIONS AND IS CONTINUING HIS LENGTHENING SESSIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13694771 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page