PRECICE® Intramedullary Limb Lengthening System P8.5-50D215Recalls & reported problems
Nuvasive Specialized Orthopedics, Inc. · Orthopedic intramedullary limb lengthening rod, antegrade femur (Trochanteric), 8.5 mm diameter, 215 mm length · Class 2 · Implant
All models of PRECICE Intramedullary Limb Lengthening System (551)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00812258020583), received between Apr 15, 2020 and Nov 10, 2023 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 29, 2021Class IIOpen, Classified
The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.
Z-0319-2022 · Root cause: Device Design · Recall details · FDA record ↗
- Feb 12, 2021Class IIOpen, Classified
Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.
Z-1349-2021 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Break3
- Detachment of Device or Device Component1
- Mechanical Problem1
- Migration1
Patient problems
- Failure of Implant5
- No Clinical Signs, Symptoms or Conditions1
- Pain1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 10, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
INFORMATION WAS RECEIVED THAT DURING EXPLANTATION, THE NAIL SEPARATED AND THE DISTAL PORTION WAS INITIALLY LEFT IN THE PATIENT'S BONE AS THE BONE CORTEX APPEARED TO HAVE CHANGED, WHICH RESULTED IN A REDUCED INTRAMEDULLARY CANAL SIZE. THE DISTAL PORTION WAS SUBSEQUENTLY RETRIEVED ANTEGRADE; HOWEVER, IT TOOK APPROXIMATELY 45 MINUTES TO RETRIEVE THE DISTAL PORTION. THERE WAS NO REPORT OF PATIENT INJURY OR FURTHER PROCEDURE IMPACT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 18112539 ↗MalfunctionApr 24, 2023Migration; Migration
INFORMATION WAS RECEIVED THAT THERE WAS DISLOCATION OF THE NAIL. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 16800977 ↗MalfunctionJan 31, 2023Break; Break
INFORMATION WAS RECEIVED THAT DURING A PLANNED REMOVAL PROCEDURE THE NAIL WAS FOUND TO BE CRACKED IN THE DISTAL ASPECT OF THE OUTER HOUSING. THE CRACK ON THE NAIL WAS NOTED TO EXTEND PROXIMAL ANOTHER 2-4CM. THE PATIENT REPORTED NO FALLS, IMPACTS, PAIN, OR OTHER NOTABLE EVENTS THAT WOULD CONTRIBUTE TO THE BREAKAGE. A PIECE OF THE HOUSING WAS MISSING FROM THE NAIL AND WAS LEFT INSIDE THE PATIENT'S MEDULLARY CANAL. THE SURGEON ATTEMPTED TO RETRIEVE THE REMAINING METAL PIECE IN THE MEDULLARY CANAL, BUT DECIDED TO PROCEED WITH INSERTING THE NEW TRAUMA NAIL WITHOUT SUCCESSFULLY REMOVING THE BROKEN PIECE.
Patient problems: Failure of Implant
MAUDE report 16273137 ↗MalfunctionJul 29, 2022Break; Break
INFORMATION WAS RECEIVED THAT THE NAIL LENGTHENED ABOUT 3.5CM BEFORE IT WAS FOUND TO BE BENT. ALTHOUGH THE NAIL WAS FOUND TO BE BENT, THE PATIENT COMPLETED EXTENSION. HOWEVER, THE PATIENT IS NOW EXPERIENCING PAIN AS THE NAIL IS BROKEN AFTER THE PATIENT JUMPED INTO WATER. THE NAIL WAS EXPLANTED.
Patient problems: Failure of Implant; Pain
MAUDE report 15129605 ↗MalfunctionApr 18, 2022Mechanical Problem; Mechanical Problem
INFORMATION WAS RECEIVED THAT THE PATIENT PREMATURELY CONSOLIDATED AND ANOTHER OSTEOTOMY WAS PERFORMED. THE NAIL WAS REPORTED TO HAVE FAILED TO DISTRACT INTRA-OPERATIVELY. NO PATIENT ADVERSE EVENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14137645 ↗InjuryApr 15, 2020Break; Break
INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2020. AS PER THE REPORTER, THE NAIL WAS BROKEN. REPORTEDLY, THE PATIENT WAS FULL WEIGHT BEARING BEYOND INSTRUCTIONS FOR USE.
Patient problems: Failure of Implant
MAUDE report 9965487 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page