PRECICE® Intramedullary Limb Lengthening System P10.7-80D275Recalls & reported problems
Nuvasive Specialized Orthopedics, Inc. · Orthopedic intramedullarylimb lengthening rod, antegrade femur (Trochanteric), 10.7 mm diameter, 275 mm length · Class 2 · Implant
All models of PRECICE Intramedullary Limb Lengthening System (551)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00812258020651), received between Sep 14, 2016 and May 18, 2026 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 29, 2021Class IIOpen, Classified
The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.
Z-0319-2022 · Root cause: Device Design · Recall details · FDA record ↗
- Feb 12, 2021Class IIOpen, Classified
Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.
Z-1349-2021 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Operates Differently Than Expected1
- Failure to Advance1
- Positioning Failure1
Patient problems
- Failure of Implant2
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 18, 2026Positioning Failure
IT WAS REPORTED THAT A PATIENT HAD A LEFT FEMORAL LENGTHENING PROCEDURE WHERE 8 CM LENGTHENING WAS PLANNED VIA NUVASIVE PRECICE MAGNETIC NAIL WHICH WAS PLANNED. AFTER BRIEFLY GAPPING 5 MM, NAIL FAILED TO LENGTHEN AND DETERMINED TO BE DEVICE FAILURE. A REVISION SURGERY PERFORMED WITH NEW NAIL PLACED.
Patient problems: Failure of Implant
MAUDE report 25204460 ↗MalfunctionAug 25, 2025Failure to Advance; Failure to Advance
IT WAS REPORTED THAT A PRECICE NAIL WAS UNABLE TO BE DISTRACTED INTRA-OPERATIVELY, IT WAS REMOVED AND REPLACED.
Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions
MAUDE report 22883201 ↗InjurySep 14, 2016Device Operates Differently Than Expected
A DISTRIBUTOR REPORTED THAT A PATIENT'S PRECICE NAIL WAS REMOVED AFTER APPROXIMATELY ONE (1) MONTH OF IMPLANTATION. THE SURGEON ALLEGED THAT THE NAIL DID NOT APPEAR TO HAVE ACHIEVED EXPECTED TARGET LENGTH.
Patient problems: No Code Available
MAUDE report 5950514 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.