PRECICE® Intramedullary Limb Lengthening System P12.5-80D305Recalls & reported problems
Nuvasive Specialized Orthopedics, Inc. · Orthopedic intramedullary limb lengthening rod, antegrade femur (Trochanteric), 12.5 mm diameter, 305 mm length · Class 2 · Implant
All models of PRECICE Intramedullary Limb Lengthening System (551)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00812258020736), received between Jul 30, 2020 and Jul 8, 2024 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Failure to Advance” (4 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 29, 2021Class IIOpen, Classified
The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.
Z-0319-2022 · Root cause: Device Design · Recall details · FDA record ↗
- Feb 12, 2021Class IIOpen, Classified
Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.
Z-1349-2021 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Advance4
- Defective Device2
- Collapse1
- Crack1
- Fracture1
Patient problems
- Failure of Implant5
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 8, 2024Failure to Advance; Failure to Advance
NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19695309 ↗MalfunctionJul 8, 2024Failure to Advance; Failure to Advance
NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 19693490 ↗MalfunctionJun 20, 2023Failure to Advance; Failure to Advance
INFORMATION WAS RECEIVED THAT AT THE FIRST FOLLOW UP APPOINTMENT IT WAS NOTED VIA X-RAYS THAT THE NAIL DID NOT CONTINUE TO LENGTHEN AT THE APPROPRIATE RATE AND DISTANCE. CHANGES SEEN ON THE X-RAY WERE INDICATIVE THAT THE NAIL WAS NOT FUNCTIONING PROPERLY; RATHER THAN MOVING DISTALLY, ONE OF THE INTERNAL SCREWS WAS MOVING PROXIMALLY INSIDE THE ROD. THE PATIENT UNDERWENT A REVISION TO REPLACE THE MALFUNCTIONING NAIL AND THE CORRECTIVE PROCEDURE WAS PERFORMED 10 DAYS LATER WITHOUT COMPLICATIONS.
Patient problems: Failure of Implant
MAUDE report 17164591 ↗InjuryApr 24, 2023Failure to Advance; Defective Device; Failure to Advance; Defective Device
INFORMATION WAS RECEIVED THAT THERE WAS AN UNSPECIFIED IMPLANT FAILURE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 16798709 ↗InjuryAug 17, 2022Fracture; Fracture
INFORMATION WAS RECEIVED THAT THE NAIL FRACTURED IN-SITU. THE PATIENT WAS REPORTEDLY ENGAGING IN FULL WEIGHT-BEARING ON THE NAIL WITHOUT BONE CONSOLIDATION STRONG ENOUGH TO SUPPORT NORMAL GAIT MOVEMENT. THE PATIENT CLAIMED THEY FELT A POP DURING MOVEMENT AND THE NAIL WAS SUBSEQUENTLY FOUND TO HAVE A CLEAN BREAK THROUGH THE CENTER OF THE NAIL AT THE SITE OF ORIGINAL OSTEOTOMY. THE PATIENT UNDERWENT A REVISION SURGERY; HOWEVER, NO FURTHER ADVERSE IMPACT TO THE PATIENT WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 15246143 ↗MalfunctionFeb 16, 2022Collapse; Crack; Collapse; Crack
INFORMATION WAS RECEIVED THAT THE PATIENT HAD FINISHED LENGTHENING AND WAS IN THE CONSOLIDATION PHASE WHEN THE CROWN CRACKED AND THE NAIL COLLAPSED. NO PATIENT ADVERSE EVENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13545962 ↗MalfunctionJul 30, 2020Defective Device; Defective Device
INFORMATION WAS RECEIVED THAT A PROCEDURE WAS PERFORMED POST SUCCESSFUL LENGTHENING. AS PER THE REPORTER, AFTER REMOVAL IT WAS OBSERVED THAT THE NAIL HAD CRACKED/FRACTURED NEAR THE MOST DISTAL PART OF THE FEMALE PORTION OF THE NAIL. THE NAIL WILL NOT BE RETURNING FOR EVALUATION.
Patient problems: Failure of Implant
MAUDE report 10345215 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page