PRECICE® Intramedullary Limb Lengthening System P12.5-80E245Recalls & reported problems
Nuvasive Specialized Orthopedics, Inc. · Orthopedic intramedullary limb lengthening rod; retrograde femur straight, 12.5 mm diameter, 245 mm length · Class 2 · Implant
All models of PRECICE Intramedullary Limb Lengthening System (551)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00812258020897), received between Nov 24, 2020 and Dec 7, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (2 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Oct 29, 2021Class IIOpen, Classified
The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.
Z-0319-2022 · Root cause: Device Design · Recall details · FDA record ↗
- Feb 12, 2021Class IIOpen, Classified
Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.
Z-1349-2021 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Mechanical Problem2
Patient problems
- Failure of Implant2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 7, 2020Mechanical Problem; Mechanical Problem
INFORMATION WAS RECEIVED THAT THE NAIL WAS DISTRACTED 80MM PRIOR TO BEING IMPLANTED AND AFTER BEING DISTRACTED THE NAIL WOULD NOT RETRACT. REPORTEDLY, THE NAIL MADE A "CLICK" NOISE AND THE ONLY WAY THE NAIL WOULD WOULD RETRACT WAS TO PUT EXTRA FORCE BY HAND IN ADDITION TO UTILIZING THE FAST DISTRACTION. ONCE THIS OCCURRED, THE NAIL MADE A "CLICK" NOISE AGAIN AND THE NAIL WAS FUNCTIONAL. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: Failure of Implant
MAUDE report 10956395 ↗MalfunctionNov 24, 2020Mechanical Problem; Mechanical Problem
REPORT HAS BEEN UPDATED FOR CORRECTION OF REPORTED EXPLANT DATE. INFORMATION WAS RECEIVED THAT THE NAIL WAS NOT EXPLANTED DURING THE REVISION PROCEDURE. THE PATIENT IS STILL TRANSPORTING AND IS IN GOOD CONDITION.
Patient problems: Failure of Implant
MAUDE report 10893011 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page