Nuvasive Freedom recallZ-1351-2021 · Nuvasive Specialized Orthopedics Inc · initiated Feb 12, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Nuvasive Specialized Orthopedics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.
FDA root cause category: Device Design
What the company is doing
On 02/11/2021, the firm started emailing an "URGENT FIELD SAFETY NOTICE" Notification to customers informing them that due to not having the full complement of biological assessments, the devices are not indicated for use in individuals under the age of 18 years old. Customers are informed to: -An Recalling Firm representative will be contacting your office or you to help with any questions or concerns. -Review, complete, sign and return the attached Consignee Confirmation Form accompanying this notification in accordance with the directions on the form. -If a patient has been previously implanted with a listed device and was under the age of 18, consultation may be warranted, at the discretion of the provider. -If a patient has been previously implanted with a listed device and is pregnant, becomes pregnant, or intends to become pregnant, consultation may be warranted, at the discretion of the provider. -Forward this notice to anyone in your facility that needs to be informed. -Direct any additional manufacturer inquiries to [email protected] -Report to NSO any adverse effect or product complaints related to the use of these devices to [email protected], whether or not those adverse effects are related to this FSN. As a reminder, the following guidelines according to the Instructions for Use, including, but not limited to: -The IFU should be consulted on an ongoing basis before and throughout patient treatment with Precice products. -These Precice family of devices are contraindicated in patients in which the Precice devices would cross joint spaces or open epiphyseal growth plates. -These Precice family of devices are contraindicated in patients unwilling or incapable of following postoperative care instructions. -Precice System devices should be removed after an implantation period of no more than one year. -Once the physician determines that the Precice/Short/Freedom/Unyte System has achieved its intended use and is no longer requir
Recall details
- Recall number
- Z-1351-2021
- Status
- Open, Classified
- Initiated by the company
- Feb 12, 2021
- Posted by the FDA
- Apr 5, 2021
- Recalling firm
- Nuvasive Specialized Orthopedics Inc
- Quantity
- 59 units
- Product code
- HSB
- FDA submissions
- K141739
- Distribution
- Worldwide distribution, including U.S. Nationwide.
Product description
NUVASIVE FREEDOM: End Cap for PRECICE Intramedullary Limb Lengthening System Implant; Packaged End Cap 8.5/10.7mm, +0mm; Packaged End Cap 8.5/10.7mm, +5mm; Packaged End Cap 8.5/10.7mm, +10mm; Packaged End Cap 8.5/10.7mm, +15mm; Packaged End Cap 8.5/10.7mm, +20mm; Packaged End Cap 12.5mm, +0mm; Packaged End Cap 12.5mm, +5mm; Packaged End Cap 12.5mm, +10mm; Packaged End Cap 12.5mm, +15mm; Packaged End Cap 12.5mm, +20mm; 4.0 mm Locking Screw 20 mm Length; 4.0 mm Locking Screw 25 mm Length; 4.0 mm Locking Screw 30 mm Length; 4.0 mm Locking Screw 35 mm Length; 4.0 mm Locking Screw 40 mm Length; 4.0 mm Locking Screw 45 mm Length; 4.0 mm Locking Screw 50 mm Length; 4.0 mm Locking Screw 55 mm Length; 4.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 20 mm Length; 5.0 mm Locking Screw 25 mm Length; 5.0 mm Locking Screw 30 mm Length; 5.0 mm Locking Screw 35 mm Length; 5.0 mm Locking Screw 40 mm Length; 5.0 mm Locking Screw 45 mm Length; 5.0 mm Locking Screw 50 mm Length; 5.0 mm Locking Screw 55 mm Length; 5.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 65 mm Length; 5.0 mm Locking Screw 70 mm Length; 5.0 mm Locking Screw 75 mm Length; 4.0 mm Locking Screw 20 mm Length; 4.0 mm Locking Screw 25 mm Length; 4.0 mm Locking Screw 30 mm Length; 4.0 mm Locking Screw 35 mm Length; 4.0 mm Locking Screw 40 mm Length; 4.0 mm Locking Screw 45 mm Length; 4.0 mm Locking Screw 50 mm Length; 4.0 mm Locking Screw 55 mm Length; 4.0 mm Locking Screw 60 mm Length; 4.0 mm X 65 mm Locking Screw, Peg; 4.0 mm X 70 mm Locking Screw, Peg; 4.0 mm X 75 mm Locking Screw, Peg; 4.0 mm X 80 mm Locking Screw, Peg; 4.0 mm X 85 mm Locking Screw, Peg; 4.0 mm X 90 mm Locking Screw, Peg; 4.0 mm X 95 mm Locking Screw, Peg; 4.0 mm X 100 mm Locking Screw, Peg; 5.0 mm Locking Screw 20 mm Length; 5.0 mm Locking Screw 25 mm Length; 5.0 mm Locking Screw 30 mm Length; 5.0 mm Locking Screw 35 mm Length; 5.0 mm Locking Screw 40 mm Length; 5.0 mm Locking Screw 45 mm Length; 5.0 mm Locking Screw 50 mm Length; 5.0 mm Locking Screw 55 mm Length; 5.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 65 mm Length; 5.0 mm Locking Screw 70 mm Length; 5.0 mm Locking Screw 75 mm Length; 5.0 mm X 80 mm Locking Screw, Peg; 5.0 mm X 85 mm Locking Screw, Peg; 5.0 mm X 90 mm Locking Screw, Peg; 5.0 mm X 95 mm Locking Screw, Peg; 5.0 mm X 100 mm Locking Screw, Peg; Orthopedic intramedullary limb lengthening rod, residual limb lengthener, 14 mm diameter, 130 mm length; Orthopedic surgical instrument - for use with the intramedullary limb lengthening rod, residual limb lengthener;
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
All Lots manufactured between April 2015 to February 2021. NUVASIVE Freedom Product Codes: End Caps: CPA1-000; CPA2-000; CPA2-005; CPA2-010; CPA2-015; CPA2-020; CPA3-000; CPA3-005; CPA3-010; CPA3-015; CPA3-020; Locking screws: LSB4-020; LSB4-025; LSB4-030; LSB4-035; LSB4-040; LSB4-045; LSB4-050; LSB4-055; LSB4-060; LSB5-020; LSB5-025; LSB5-030; LSB5-035; LSB5-040; LSB5-045; LSB5-050; LSB5-055; LSB5-060; LSB5-065; LSB5-070; LSB5-075; LSC4-020; LSC4-025; LSC4-030; LSC4-035; LSC4-040; LSC4-045; LSC4-050; LSC4-055; LSC4-060; LSC4-065; LSC4-070; LSC4-075; LSC4-080; LSC4-085; LSC4-090; LSC4-095; LSC4-100; LSC5-020; LSC5-025; LSC5-030; LSC5-035; LSC5-040; LSC5-045; LSC5-050; LSC5-055; LSC5-060; LSC5-065; LSC5-070; LSC5-075; LSC5-080; LSC5-085; LSC5-090; LSC5-095; LSC5-100; Limb lengthening intramedullary rod, residual limb lengthener: R14-100A130; Orthopedic surgical instrument - for use with the intramedullary limb lengthening rod, residual limb lengthener: RLL1-000; GTIN Numbers: 00812258020002; 00887517001139; 00887517001146; 00887517001153; 00887517001160; 00887517001177; 00887517001184; 00887517001191; 00887517001207; 00887517001214; 00887517001221; 00856719002763; 00856719002770; 00856719002787; 00856719002794; 00856719002800; 00856719002817; 00856719002824; 00856719002831; 00856719002848; 00856719002886; 00856719002893; 00856719002909; 00856719002916; 00856719002923; 00856719002930; 00856719002947; 00856719002954; 00856719002961; 00856719002978; 00856719002985; 00856719002992; 00812258021429; 00812258021436; 00812258021443; 00812258021450; 00812258021467; 00812258021474; 00812258021481; 00812258021498; 00812258021504; 00887517000958; 00887517000965; 00887517000972; 00887517000989; 00887517000996; 00887517001009; 00887517001016; 00887517001023; 00812258021511; 00812258021528; 00812258021535; 00812258021542; 00812258021559; 00812258021566; 00812258021573; 00812258021580; 00812258021597; 00812258021603; 00812258021610; 00812258021627; 00887517001030; 00887517001047; 00887517001054; 00887517001061; 00887517001078; 00812258024239; 00812258026479;
Devices named in this recall
68 devices in the FDA device registry whose identifier appears in the recall's code information.
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-020Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-025Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-030Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-035Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-055Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-060Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-065Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-070Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-020Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-025Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-030Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-035Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-055Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-060Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-065Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-070Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-080Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-085Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-090Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-095Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-100Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System CPA2-005Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System CPA1-000Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-020Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-025Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-030Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-035Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-055Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-060Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB4-020Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB4-025Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB4-030Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB4-035Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB4-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB4-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB4-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB4-055Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE Intramedullary Limb Lengthening System LSB4-060Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-065Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-070Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-080Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-085Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-090Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-095Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System LSC4-100Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System CPA2-000Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System CPA2-010Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System CPA2-015Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System CPA2-020Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System CPA3-000Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System CPA3-005Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System CPA3-010Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System CPA3-015Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- PRECICE® Intramedullary Limb Lengthening System CPA3-020Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories3 recalls0 reports
- FREEDOM Residual Limb Lengthening System R14-100A130Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories2 recalls1 report
- FREEDOM Residual Limb Lengthening System RLL1-000Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 87353).
- Z-1349-2021 · Open, ClassifiedNUVASIVE PRECICE Intramedullary Limb Lengthening (IMLL) and (Short) System End Cap for PRECICE Intramedullary Limb Lengthening System Implant; Packaged End Cap 8.5/10.7mm, +0mm…
- Z-1350-2021 · Open, ClassifiedNUVASIVE PRECICE UNYTE System (Humerus and Tibia): PRECICE Trauma Nail System/PRECICE Unyte-Humerus Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 165 mm lengt…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page