Skip to content

PRECICE® Intramedullary Limb Lengthening System TSA5-065Recalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · 5.0 mm Threaded Locking Screw 65 mm Length · Class 2

All models of PRECICE Intramedullary Limb Lengthening System (551)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00812258027902), received between Mar 30, 2022 and Mar 30, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Oct 29, 2021Class IIOpen, Classified

    The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.

    Z-0319-2022 · Root cause: Device Design · Recall details · FDA record ↗

  2. Feb 12, 2021Class IIOpen, Classified

    Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.

    Z-1349-2021 · Root cause: Device Design · Recall details · FDA record ↗

  3. Feb 12, 2021Class IIOpen, Classified

    Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.

    Z-1350-2021 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fracture1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 30, 2022Fracture; Material Twisted/Bent; Fracture; Material Twisted/Bent

    INFORMATION WAS RECEIVED THAT THE UPPER TITANIUM BONE LOCKING SCREW WAS BENT. THE PATIENT HEARD AN AUDIBLE CRUNCHING NOISE AND DURING FOLLOW UP X-RAYS SHOWED THE SCREW HAD SHEARED OFF COMPLETELY. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13952350 ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page