Skip to content

PRECICE® Intramedullary Limb Lengthening System P8.5-50M165Recalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · Orthopedic Intramedullary Limb Lengthening Rod, Retrograde Femur Trochanteric, 10° Bend, 8.5 mm Diameter, 165 mm · Class 2 · Implant

All models of PRECICE Intramedullary Limb Lengthening System (551)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
2

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00812258028237), received between Mar 10, 2021 and Apr 9, 2024 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 29, 2021Class IIOpen, Classified

    The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.

    Z-0319-2022 · Root cause: Device Design · Recall details · FDA record ↗

  2. Feb 12, 2021Class IIOpen, Classified

    Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.

    Z-1349-2021 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Break1
  • Use of Device Problem1

Patient problems

  • Failure of Implant3
  • Bone Fracture(s)1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 9, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION AVAILABLE. COMPLAINT 2 OF 3.

    Patient problems: Bone Fracture(s)

    MAUDE report 19075183 ↗
  • InjuryAug 24, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    INFORMATION WAS RECEIVED THAT THE NAIL DISASSEMBLED DURING EXPLANTATION AND SUBSEQUENTLY THE FEMORAL BONE NEEDED TO BE FENESTRATED DUE TO BONE DAMAGE INCURRED DURING THE REMOVAL PROCESS. THE DISTAL PORTION OF THE DISASSEMBLED NAIL WAS UNABLE TO BE REMOVED FROM THE INTRAMEDULLARY CANAL AS IT WAS REPORTEDLY TOO TIGHT IN THE BONE. DUE TO THE BONE FENESTRATION, THE PATIENT'S HOSPITAL STAY WAS EXTENDED SIX WEEKS AND REQUIRED THE PATIENT TO WALK WITH FOREARM SUPPORT. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 17614745 ↗
  • MalfunctionAug 9, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 17492669 ↗
  • MalfunctionMar 10, 2021Use of Device Problem; Use of Device Problem

    INFORMATION WAS RECEIVED THAT AN ANTEGRADE NAIL WAS PUT IN RETROGRADE AND IT WAS NOTICED DURING A FOLLOW UP X-RAY THAT THE NAIL WAS NOT WORKING PROPERLY. AS A RESULT A REVISION PROCEDURE WAS PERFORMED AND A NEW NAIL WAS IMPLANTED. IT WAS NOTED THE NEW NAIL IS WORKING AS INTENDED.

    Patient problems: Failure of Implant

    MAUDE report 11457763 ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →

Sources and updates