PRECICE® Intramedullary Limb Lengthening System P8.5-50M165Recalls & reported problems
Nuvasive Specialized Orthopedics, Inc. · Orthopedic Intramedullary Limb Lengthening Rod, Retrograde Femur Trochanteric, 10° Bend, 8.5 mm Diameter, 165 mm · Class 2 · Implant
All models of PRECICE Intramedullary Limb Lengthening System (551)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00812258028237), received between Mar 10, 2021 and Apr 9, 2024 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 29, 2021Class IIOpen, Classified
The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.
Z-0319-2022 · Root cause: Device Design · Recall details · FDA record ↗
- Feb 12, 2021Class IIOpen, Classified
Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.
Z-1349-2021 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Break1
- Use of Device Problem1
Patient problems
- Failure of Implant3
- Bone Fracture(s)1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryApr 9, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
NO ADDITIONAL INFORMATION AVAILABLE. COMPLAINT 2 OF 3.
Patient problems: Bone Fracture(s)
MAUDE report 19075183 ↗InjuryAug 24, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
INFORMATION WAS RECEIVED THAT THE NAIL DISASSEMBLED DURING EXPLANTATION AND SUBSEQUENTLY THE FEMORAL BONE NEEDED TO BE FENESTRATED DUE TO BONE DAMAGE INCURRED DURING THE REMOVAL PROCESS. THE DISTAL PORTION OF THE DISASSEMBLED NAIL WAS UNABLE TO BE REMOVED FROM THE INTRAMEDULLARY CANAL AS IT WAS REPORTEDLY TOO TIGHT IN THE BONE. DUE TO THE BONE FENESTRATION, THE PATIENT'S HOSPITAL STAY WAS EXTENDED SIX WEEKS AND REQUIRED THE PATIENT TO WALK WITH FOREARM SUPPORT. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 17614745 ↗MalfunctionAug 9, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 17492669 ↗MalfunctionMar 10, 2021Use of Device Problem; Use of Device Problem
INFORMATION WAS RECEIVED THAT AN ANTEGRADE NAIL WAS PUT IN RETROGRADE AND IT WAS NOTICED DURING A FOLLOW UP X-RAY THAT THE NAIL WAS NOT WORKING PROPERLY. AS A RESULT A REVISION PROCEDURE WAS PERFORMED AND A NEW NAIL WAS IMPLANTED. IT WAS NOTED THE NEW NAIL IS WORKING AS INTENDED.
Patient problems: Failure of Implant
MAUDE report 11457763 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page