Skip to content

Nalu Medical, Inc. 72006-USRecalls & reported problems

Nalu Medical, Inc. · Lead Kit (4 contact, 25 cm, Tined) · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00812537031057), received between Nov 23, 2021 and Jul 23, 2025 — classified as 4 injuries.

The most frequently reported specific device problem is “Device Unsafe to Use in Environment” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Unsafe to Use in Environment1
  • Fracture1
  • Impedance Problem1
  • Migration1

Patient problems

  • Inadequate Pain Relief2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 23, 2025Impedance Problem; Impedance Problem

    PATIENT WAS IMPLANTED WITH A NALU PERIPHERAL NERVE STIMULATOR SYSTEM ON (B)(6) 2021 AND USED THE SYSTEM FOR APPROXIMATELY 4 YEARS. ON 06/21/2025 AN MRI FACILITY CONTACTED NALU PRODUCT SUPPORT TO INQUIRE ABOUT CONDITIONALITY OF THE SYSTEM. DURING A PRE-MRI CHECK THE NALU SYSTEM WAS DISCOVERED TO HAVE IMPEDANCE ISSUES THAT PREVENTED THE PATIENT FROM MOVING FORWARD WITH THE MRI SCAN. THE NALU PRODUCT SUPPORT TEAM WAS NOTIFIED THAT A FULL SYSTEM EXPLANT TOOK PLACE HOWEVER THE EXACT DATE OF THE EXPLANT WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22580013 ↗
  • InjuryApr 27, 2024Fracture; Fracture

    PATIENT WAS IMPLANTED WITH THE NALU PERIPHERAL NERVE STIMULATOR ON (B)(6) 2022 TO TREAT SHOULDER PAIN. IN (B)(6) 2023 THE PATIENT REPORTED A LOSS OF THERAPY. IMAGING CONFIRMED A FRACTURE OF ONE IMPLANTED LEAD. ON (B)(6) 2024 A FULL SYSTEM EXPLANT WAS PERFORMED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 19198983 ↗
  • InjuryJan 10, 2024Device Unsafe to Use in Environment; Device Unsafe to Use in Environment

    PATIENT WAS IMPLANTED WITH THE NALU PERIPHERAL NERVE STIMULATOR SYSTEM ON (B)(6) 2021 AFTER A SUCCESSFUL TRIAL PHASE. AFTER BEING IMPLANTED WITH THE PERMANENT SYSTEM THE PATIENT REPORTED NOT RECEIVING THE SAME PAIN RELIEF AS THE TRIAL AND IS NOT SATISFIED WITH THE RESULTS OF REPROGRAMMING THE SYSTEM. PATIENT REFUSED FURTHER PROGRAMMING AND REQUESTED THE SYSTEM BE REMOVED. A FULL SYSTEM EXPLANT WAS PERFORMED ON (B)(6) 2023.

    Patient problems: Inadequate Pain Relief

    MAUDE report 18490066 ↗
  • InjuryNov 23, 2021Migration; Migration

    PATIENT WAS IMPLANTED ON (B)(6) 2021. PATIENT PERFORMED EXERCISE ACTIVITY IMMEDIATELY AFTER IMPLANT PROCEDURE. FLUOROSCOPY IMAGING SHOWED THE LEAD MIGRATED FORWARD, FURTHER INTO THE POPLITEAL FOSSA. ON (B)(6) 2021 THE PATIENT HAD A REVISION SURGERY TO REMOVE THE ORIGINAL TINED LEAD AND ANOTHER LEAD WITH THE SAME SPECIFICATIONS WAS PLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12862850 ↗

Other Stimulator, Peripheral Nerve, Implanted (Pain Relief) devices

All Stimulator, Peripheral Nerve, Implanted (Pain Relief) devices →

Sources and updates