Skip to content

Nalu Medical, Inc. 71005-US (UDI 00812537033518)Recalls & reported problems

Nalu Medical, Inc. · Nalu Neurostimulation Kit (Ported, Dual 8, 40 cm) · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00812537033518), received between Mar 19, 2022 and Mar 19, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jun 3, 2024Class IIOpen, Classified

    Neurostimulation Systems have a manufacturing non-conformance that may reduce the efficiency of the radiofrequency signals between the Implantable Pulse Generator (IPG) and the Therapy Disc, causing the IPG to be non-functional, which causes the device to be non-functional so it will no longer inhibit the transmission of pain signals to the brain.

    Z-2257-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Insufficient Device Problem Information1

Patient problems

  • Post Operative Wound Infection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 19, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    PATIENT PRESENTED WITH INFECTION AT LEAD INCISION SITE 3 WEEKS POST OP. PATIENT WAS EVACUATING FROM HURRICANE IDA AND DIDN'T KEEP INCISION OR DRESSING CLEAN BECAUSE OF NO POWER OR WATER DURING HURRICANE.

    Patient problems: Post Operative Wound Infection

    MAUDE report 13826900 ↗

Other Stimulator, Peripheral Nerve, Implanted (Pain Relief) devices

All Stimulator, Peripheral Nerve, Implanted (Pain Relief) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page