Nalu Medical, Inc. 72012-USRecalls & reported problems

Nalu Medical, Inc. · Peripheral Nerve Stimulation Lead Kit (8 contact, 60 cm) · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00812537035642), received between Feb 23, 2023 and Dec 26, 2024 — classified as 5 injuries.

The most frequently reported specific device problem is “Migration” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration4
  • Adverse Event Without Identified Device or Use Problem1
  • Off-Label Use1

Patient problems

  • Inadequate Pain Relief5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 26, 2024Migration; Migration

    PATIENT WAS IMPLANTED WITH A NALU PERIPHERAL NERVE STIMULATOR SYSTEM ON (B)(6)2024 TO TREAT HEAD PAIN. IN (B)(6) 2024 THE PATIENT REPORTED CHANGES IN THE EFFECTIVENESS OF THERAPY. XRAY IMAGING SUGGESTED THAT BOTH OF THE IMPLANTED LEADS HAD MIGRATED AWAY FROM THE INITIAL PLACEMENT. ON (B)(6)2024 A SURGICAL REVISION WAS PERFORMED IN WHICH IT WAS CONFIRMED THAT ONLY ONE OF THE LEADS HAD MIGRATED AND REQUIRED REPLACEMENT.

    Patient problems: Inadequate Pain Relief

    MAUDE report 21024100 ↗
  • InjuryApr 4, 2024Migration; Migration

    ON (B)(6) 2023 THE PATIENT WAS IMPLANTED WITH A NALU PERIPHERAL NERVE STIMULATOR SYSTEM TARGETING THE OCCIPITAL NERVE TO TREAT HEAD PAIN AND HEADACHES. PATIENT HAD PREVIOUSLY COMPLETED A TRIAL PHASE WITH THE NALU SYSTEM AND HAD REPORTED PAIN WAS REDUCED FROM 7/10 TO 1/10 DURING THE TRIAL. AFTER COMPLETING THE PERMANENT SYSTEM IMPLANT AND INITIAL ACTIVATION OF THE SYSTEM, THE PATIENT REPORTED INADQUATE PAIN RELIEF AND DISCOMFORT WITH THE PLACEMENT OF THE IMPLANTED LEADS. XRAY IMAGING FOUND ONE OF THE IMPLANTED LEADS HAD MIGRATED. PATIENT WAS OFFERED REPROGRAMMING TO ATTEMPT TO ATTAIN ADEQUATE PAIN RELIEF WITH THE LEAD AS-IS AND PATIENT REFUSED ALL ATTEMPTS AT TROUBLESHOOTING. A FULL SYSTEM EXPLANT WAS PERFORMED ON (B)(6) 2024 PER THE PATIENT REQUEST.

    Patient problems: Inadequate Pain Relief

    MAUDE report 19047825 ↗
  • InjuryAug 2, 2023Migration; Migration

    PATIENT WAS IMPLANTED WITH THE NALU PERIPHERAL NERVE STIMULATOR SYSTEM ON (B)(6) 2021. AFTER IMPLANT THE PATIENT EXPERIENCED LOSS OF THERAPY WHICH WAS DETERMINED TO BE DUE TO MIGRATION OF THE IMPLANTED LEADS. FIELD OBSERVATIONS CONCLUDED THE ANCHORING TECHNIQUE CONTRIBUTED TO LEAD MIGRATION AND SUBSEQUENT LOSS OF PAIN RELIEF THERAPY. SURGICAL REVISION WAS PERFORMED ON (B)(6) 2022 TO REPOSITION THE MIGRATED LEADS. DURING THE PROCEDURE THE IMPLANTABLE PULSE GENERATOR (IPG) AND IMPLANTED LEADS WERE DAMAGED AND REQUIRED REPLACEMENT.

    Patient problems: Inadequate Pain Relief

    MAUDE report 17443743 ↗
  • InjuryMar 17, 2023Migration; Migration

    PATIENT WAS IMPLANTED WITH THE NALU PERIPHERAL NERVE STIMULATOR ON (B)(6) 2022. PATIENT SUBSEQUENTLY REPORTED LOSS OF THERAPY WHICH WAS DETERMINED TO BE DUE TO AXIAL LEAD MIGRATION. SURGICAL REVISION WAS PERFORMED ON (B)(6) 2023 TO MOVE THE MIGRATED LEAD TO A MORE OPTIMAL LOCATION. NO COMPONENTS WERE REMOVED AND NO NEW COMPONENTS WERE IMPLANTED. PATIENT REPORTS RECEIVING STIMULATION TO THE DESIRED LOCATION AFTER THE PROCEDURE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 16565205 ↗
  • InjuryFeb 23, 2023Off-Label Use; Adverse Event Without Identified Device or Use Problem; Off-Label Use; Adverse Event Without Identified Device or Use Problem

    ON 24JAN2023 SURGICAL REVISION WAS PERFORMED TO INTRODUCE A SECONDARY LEAD IN ORDER TO PROVIDE ADDITIONAL PAIN RELIEF COVERAGE. NO REPORTS OF DEVICE MALFUNCTION.

    Patient problems: Inadequate Pain Relief

    MAUDE report 16433606 ↗