Nalu Medical, Inc. 71005 (UDI 00812537037189)Recalls & reported problems

Nalu Medical, Inc. · Nalu Neurostimulation Kit (Ported, Dual 8, 40 cm) · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00812537037189), received between Oct 2, 2024 and Jul 27, 2026 — classified as 9 injuries.

The most frequently reported specific device problem is “Intermittent Communication Failure” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Intermittent Communication Failure4
  • Insufficient Device Problem Information2
  • Migration2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Inadequate Pain Relief6
  • Pain2
  • Post Operative Wound Infection1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 27, 2026Insufficient Device Problem Information

    The patient was implanted with a Nalu Spinal Cord Stimulator system on (b)(6) 2023, with the system being replaced in (b)(6) 2025. In (b)(6) 2026, after having used the system for more than a year and a half, the patient reached out to Nalu Product Support to report description of "significant" shocking sensation that was felt only when the Nalu system was active. Per the patient, the sensations began approximately end of (b)(6) 2025, 2 weeks after the system had been replaced. A local Nalu representative met with the patient to assess and reprogram the system during the visit the symptoms reported did not occur, however the patient did note that the symptoms were inconsistent or random. Imaging was completed and found no obvious issues with the implanted components.As no clear cause of the shocking sensation was able to be identified, it was decided that the system would be replaced. On (b)(6) 2026 the IPG was removed and replaced, leaving the original leads in place. The explanted component was not released from the facility for further testing and inspection by Nalu.

    Patient problems: Pain

    MAUDE report 50024263 ↗
  • InjuryMar 30, 2026Intermittent Communication Failure

    THE PATIENT WAS IMPLANTED WITH A NALU PERIPHERAL NERVE STIMULATOR SYSTEM ON (B)(6) 2026 TO TREAT LOWER BACK PAIN. AFTER ACTIVATING THE SYSTEM, THE PATIENT REPORTED SOME CHALLENGES IN MAINTAINING CONNECTION BETWEEN THE IMPLANTABLE PULSE GENERATOR (IPG) AND THE EXTERNAL THERAPY DISCS. A NALU REPRESENTATIVE MET WITH THE PATIENT ON (B)(6) 2026 TO ASSESS THE SYSTEM. IMAGING PERFORMED DURING THE ASSESSMENT CONFIRMED THE IPG HAD MIGRATED WITHIN THE POCKET TO NO LONGER BE SEATED FLAT AND PARALLEL TO THE SKIN SURFACE. ON (B)(6) 2026 A SURGICAL PROCEDURE WAS PERFORMED TO REPOSITION THE IPG INTO A NEW POCKET SO THAT IT WAS PARALLEL TO THE SKIN TO CORRECT CONNECTIVITY ISSUES.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24723461 ↗
  • InjuryDec 13, 2025Migration; Migration

    THE PATIENT WAS IMPLANTED WITH A NALU SPINAL CORD STIMULATOR SYSTEM ON (B)(6) 2025 TO TREAT LOWER BACK PAIN. ON (B)(6) 2025 THE PATIENT WAS DISCOVERED TO HAVE THE IMPLANTABLE PULSE GENERATOR (IPG) MIGRATED FROM THE ORIGINAL IMPLANTED POSITION. A SURGICAL PROCEDURE WAS SCHEDULED TO CORRECT THE POSITION OF THE IPG; HOWEVER, THE PROCEDURE WAS NOT ABLE TO BE COMPLETED. NO FURTHER UPDATES WERE PROVIDED TO THE FIRM REGARDING PLAN OF CARE FOR THIS PATIENT UNTIL A FOLLOW-UP WAS PERFORMED ON (B)(6) 2025 AND THE FIRM BECAME AWARE THAT THE NALU SYSTEM HAD BEEN FULLY EXPLANTED ON AN UNKNOWN DATE AND REPLACED WITH A COMPETITOR SYSTEM.

    Patient problems: Inadequate Pain Relief

    MAUDE report 23797649 ↗
  • InjuryDec 8, 2025Intermittent Communication Failure; Intermittent Communication Failure

    THE PATIENT WAS IMPLANTED WITH A NALU PERIPHERAL NERVE STIMULATOR SYSTEM ON (B)(6) 2025 TO TREAT LOWER BACK PAIN. APPROXIMATELY 2-3 MONTHS AFTER THE IMPLANT THE PATIENT REPORTED ONSET OF CONNECTIVITY ISSUES BETWEEN THE IMPLANTABLE PULSE GENERATOR (IPG) AND THE EXTERNAL THERAPY DISCS. IT WAS DETERMINED THAT THE IPG HAD MIGRATED WITHIN THE POCKET TO NO LONGER BE SEATED FLAT AND PARALLEL TO THE SKIN SURFACE, CAUSING THE CONNECTIVITY CHALLENGES. ON (B)(6) 2025, THE IPG WAS REMOVED FROM THE EXISTING POCKET ON THE RIGHT FLANK AREA AND A NEW IPG WAS IMPLANTED IN THE LEFT FLANK AREA.

    Patient problems: Inadequate Pain Relief

    MAUDE report 23739139 ↗
  • InjuryOct 23, 2025Migration; Migration

    THE PATIENT WAS IMPLANTED WITH A NALU SPINAL CORD STIMULATOR SYSTEM ON (B)(6) 2025 AFTER PARTICIPATING IN A SUCCESSFUL TRIAL PHASE WITH NALU. THE PERMANENT SYSTEM WAS IMPLANTED WITH THE LEADS IN APPROXIMATELY THE SAME LOCATION AS THE TRIAL LEADS. ALTHOUGH THE PATIENT HAD REPORTED APPROXIMATELY 90% PAIN REDUCTION DURING THE TRIAL PHASE, THE PERMANENT SYSTEM WAS REPORTED TO HAVE PROVIDED NO PAIN RELIEF AT ALL. TROUBLESHOOTING IN THE FIELD INCLUDED 5 DIFFERENT SESSIONS OF REPROGRAMMING THE SYSTEM TO OPTIMIZE THERAPY, WITH ONGOING REPORTS OF NO PAIN RELIEF. XRAY IMAGING WAS PERFORMED AND CONFIRMED THAT THE LEADS HAD MIGRATED DOWN A VERTEBRAL LEVEL, THUS WERE NO LONGER STIMULATING THE AREA OF THE PATIENT'S PAIN SYMPTOMS. ON (B)(6) 2025 A SURGICAL PROCEDURE WAS PERFORMED TO REPOSITION THE IMPLANTED LEADS AND PLACE AN ANCHOR TO HOLD THE LEADS IN PLACE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 23375769 ↗
  • InjuryOct 12, 2025Intermittent Communication Failure; Intermittent Communication Failure

    THE PATIENT WAS IMPLANTED WITH A NALU SPINAL CORD STIMULATOR SYSTEM ON (B)(6) 2024 TARGETING THE THORACIC NERVE. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PLACED IN THE PATIENT'S LOWER BACK AREA. DURING THE HEALING PROCESS, PRIOR TO THE IMPLANTED COMPONENTS DEVELOPING SCAR TISSUE, THE PATIENT NOTED INCONSISTENT CONNECTIVITY BETWEEN THE IPG AND THE EXTERNAL THERAPY DISCS AND IT WAS DETERMINED THAT THE IPG HAD ROTATED WITHIN THE POCKET SO THAT IT WAS NO LONGER FLAT AND PARALLEL TO THE SKIN SURFACE. TROUBLESHOOTING WAS PERFORMED IN THE FIELD TO ENCOURAGE THE IPG TO REMAIN FLAT, HOWEVER WAS NOT SUCCESSFUL. BOTH LEADS LATER RETURNED IMPEDANCE ERRORS IN ADDITION TO THE CONNECTIVITY ISSUES FROM THE IPG MIGRATION. ON (B)(6) 2025 A SURGICAL REVISION WAS PERFORMED TO PLACE A NEW IPG IN A NEW POCKET LOCATION AND BOTH MALFUNCTIONING LEADS WERE REPLACED WITH NEW LEADS.

    Patient problems: Inadequate Pain Relief

    MAUDE report 23275742 ↗
  • InjuryJul 21, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    AFTER PARTICIPATING IN A SUCCESSFUL TRIAL PHASE WITH NALU, THE PATIENT WAS IMPLANTED WITH A NALU SPINAL CORD STIMULATION SYSTEM ON (B)(6) 2025, TARGETING THE LUMBAR NERVE TO TREAT LOWER BACK AND LOWER EXTREMITY PAIN THE SYSTEM WAS ACTIVATED ON (B)(6) 2025 ON (B)(6) 2025 A NALU REPRESENTATIVE WAS CONTACTED BY THE PATIENT'S DAUGHTER WHO STATED THAT THE PATIENT WAS COMPLAINING OF GROIN PAIN THAT HAD BEGUN AFTER THE NALU SYSTEM WAS IMPLANTED. THE PATIENT'S DAUGHTER REQUESTED RECORDS OF THE IMPLANT PROCEDURE AS THE FAMILY WAS OPTING TO SEEK A SECOND OPINION REGARDING THE NEW PAIN SYMPTOMS. ON (B)(6) 2025 THE PATIENT'S DAUGHTER REPORTED TO A NALU REPRESENTATIVE THAT THE SYSTEM HAD BEEN EXPLANTED THAT DAY DUE TO PAIN AT THE SITE OF THE IMPLANTABLE PULSE GENERATOR (IPG).

    Patient problems: Pain

    MAUDE report 22552665 ↗
  • InjuryMay 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS IMPLANTED WITH A PERIPHERAL NERVE STIMULATOR SYSTEM ON (B)(6) 2025 TO TREAT KNEE PAIN. DURING A POST-OPERATIVE FOLLOW-UP VISIT AT THE MEDICAL CLINIC, THE PATIENT WAS NOTED TO HAVE VISIBLE SIGNS OF INFECTION WHICH DEVELOPED ALMOST IMMEDIATELY AFTER IMPLANT. ON (B)(6) 2025 A FULL SYSTEM EXPLANT WAS PERFORMED DUE TO THE SUSPECTED INFECTION.

    Patient problems: Post Operative Wound Infection

    MAUDE report 21982143 ↗
  • InjuryOct 2, 2024Intermittent Communication Failure; Intermittent Communication Failure

    PATIENT WAS IMPLANTED WITH A NALU SPINAL CORD STIMULATOR SYSTEM ON (B)(6) 2024 TO TREAT LOWER BACK PAIN. IMMEDIATELY UPON ACTIVATING THE SYSTEM IT WAS NOTED THAT THE COMMUNICATION BETWEEN THE IMPLANTABLE PULSE GENERATOR (IPG) AND EXTERNAL THERAPY DISCS WAS NOT IDEAL AND WOULD CAUSE INTERMITTENT DISRUPTIONS TO THERAPY FOR THE PATIENT. THE COMMUNICATION ISSUES FAILED TO IMPROVE AFTER ADDITIONAL HEALING TIME AND XRAY IMAGING CONFIRMED THE IPG HAD MIGRATED BEYOND THE RECOMMENDED DEPTH AND HAD ALSO ROTATED TO NO LONGER BE PARALLEL TO THE SKIN SURFACE. ON (B)(6) 2024 A SURGICAL PROCEDURE WAS PERFORMED TO REPOSITION THE EXISTING EQUIPMENT TO BRING THE IPG TO THE RECOMMENDED DEPTH AND POSITION FOR OPTIMAL COMMUNICATION. THE SYSTEM WAS SUCCESSFULLY REACTIVATED AND PATIENT HAS REPORTED GETTING GOOD PAIN RELIEF FROM THE SYSTEM.

    Patient problems: Inadequate Pain Relief

    MAUDE report 20356236 ↗

Other Stimulator, Peripheral Nerve, Implanted (Pain Relief) devices

All Stimulator, Peripheral Nerve, Implanted (Pain Relief) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page