Nalu Medical, Inc. 71007Recalls & reported problems

Nalu Medical, Inc. · Nalu Neurostimulation Kit (Ported, Single 4) · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00812537037356), received between Jan 14, 2025 and Feb 19, 2026 — classified as 3 injuries.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Detachment of Device or Device Component1

Patient problems

  • Post Operative Wound Infection2
  • Discomfort1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 19, 2026Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS IMPLANTED WITH A NALU PERIPHERAL NERVE STIMULATOR SYSTEM ON (B)(6) 2025 TO TREAT FOOT PAIN. AT THE TIME OF THE INITIAL SYSTEM ACTIVATION THE PATIENT REPORTED GOOD HEALING AND RECEIVING 100% PAIN RELIEF FROM THE SYSTEM. ON (B)(6) 2025 THE PATIENT NOTICED SOME REDNESS AND DISCOMFORT AT THE INCISION SITE FOR THE IMPLANTABLE PULSE GENERATOR (IPG). PATIENT WAS EVALUATED BY THE PRIMARY DOCTOR (NOT THE IMPLANTING DOCTOR) AND INFORMED AN INFECTION MAY BE DEVELOPING. THE PRIMARY DOCTOR CLEANED THE SITE AND PRESCRIBED ORAL ANTIBIOTICS. AFTER BEING SEEN BY THE PRIMARY DOCTOR, THE PATIENT THEN NOTIFIED THE IMPLANTING DOCTOR OF THE POSSIBLE INFECTION. THE IMPLANTING DOCTOR ADVISED TO REMOVE THE SYSTEM AND CULTURE TO DETERMINE INFECTION TYPE. ON (B)(6) 2025 THE IPG WAS REMOVED, HOWEVER THE DOCTOR WAS UNCOMFORTABLE REMOVING THE LEADS AS THEY WERE SCARRED IN CLOSE PROXIMITY TO THE TIBIAL NERVE.

    Patient problems: Post Operative Wound Infection

    MAUDE report 24393123 ↗
  • InjuryJul 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS IMPLANTED WITH TWO NALU PNS SYSTEMS ON (B)(6) 2025 TO TREAT BILATERAL LOWER EXTREMITY PAIN. EACH SYSTEM WAS IMPLANTED SUCH THAN AN IMPLANTABLE PULSE GENERATOR (IPG) WAS IN THE LATERAL THIGH AREA AND IMPLANTED LEADS TARGETED THE SCIATIC NERVE. ON (B)(6) 2025 A NALU REPRESENTATIVE WAS NOTIFIED THAT THE PATIENT HAD DEVELOPED AN INFECTION IN THE RIGHT LOWER EXTREMITY AT THE SURGICAL INCISION SITE. A FULL SYSTEM EXPLANT TOOK PLACE ON (B)(6) 2025.

    Patient problems: Post Operative Wound Infection

    MAUDE report 22683273 ↗
  • InjuryJan 14, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PATIENT WAS IMPLANTED WITH A NALU PERIPHERAL NERVE STIMULATOR SYSTEM ON (B)(6) 2024 TO TREAT RIGHT LOWER EXTREMITY PAIN. WHEN A NALU REPRESENTATIVE ATTEMPTED TO ACTIVATE THE SYSTEM DURING A POST-OP VISIT, THE PATIENT REPORTED NO SENSATION FROM THE DEVICE AND TESTING RETURNED LOW IMPEDANCE READINGS. DURING THE ACTIVATION, THE PHYSICIAN NOTED SOME PAIN AND SLIGHT REDNESS AT THE INCISION SITE AND SUSPECTED INFECTION. PHYSICIAN RECOMMENDED TO EXPLANT THE SYSTEM DUE TO THE SUSPECTED INFECTION. ON (B)(6) 2024 THE SYSTEM WAS EXPLANTED AND THE PHYSICIAN REPORTED NO SIGNS OF INFECTION UPON OPENING THE SITE. DURING THE EXPLANT PROCEDURE IT WAS DISCOVERED THAT THE IMPLANTED LEAD HAD BECOME DISCONNECTED FROM THE IMPLANTABLE PULSE GENERATOR (IPG) AND THAT HAD BEEN THE CAUSE OF THE PATIENT'S DISCOMFORT AT THE INCISION SITE.

    Patient problems: Discomfort

    MAUDE report 21147626 ↗

Other Stimulator, Peripheral Nerve, Implanted (Pain Relief) devices

All Stimulator, Peripheral Nerve, Implanted (Pain Relief) devices →