Tiger 200-30-024Recalls & reported problems
Trilliant Surgical, LLC · 3.0mm x 24mm Tiger Cannulated Screw · Class 2 · Implant
All models of Tiger (20)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00812926020990), received between Mar 3, 2021 and Mar 3, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Patient Device Interaction Problem1
Patient problems
- Pain1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMar 3, 2021Patient Device Interaction Problem; Patient Device Interaction Problem
ON (B)(6) 2021, A TRILLIANT SURGICAL INDEPENDENT SALES REPRESENTATIVE SENT IN A FIELD INCIDENT REPORT (FIR) WITH A 200-30-024 (3.0MM X 24MM TIGER CANNULATED SCREW) TO TRILLIANT SURGICAL CORPORATE. A TRILLIANT SURGICAL SALES SUPPORT SPECIALIST CALLED THE SALES REPRESENTATIVE TO OBTAIN FURTHER DETAILS FOR THE PART RETURN AND FIR FORM. THE SALES REPRESENTATIVE STATED THAT ON (B)(6) 2021, DOCTOR 1 HAD SCHEDULED A REMOVAL AT FACILITY 1, ON A (B)(6)-YEAR-OLD FEMALE PATIENT (DOB AND PATIENT WEIGHT REMAINS UNKNOWN AFTER REVIEW OF CASE WRITE UP) WHO WAS EXPERIENCING LOCALIZED PAIN / IRRITATION ON HER RIGHT FOOT. IT IS UNKNOWN WHEN THE PAIN INITIALLY BEGAN WITH THE PATIENT, HOWEVER, DOCTOR 1 SCHEDULED TO REMOVE THE PREVIOUS HARDWARE ORIGINALLY IMPLANTED. THE ORIGINAL IMPLANTATION OCCURRED ON (B)(6) 2020 AND THE SALES REPRESENTATIVE RECALLED THAT DOCTOR 1 PERFORMED A CHEILECTOMY / 1ST MPJ ARTHROPLASTY WITH A 101-00-002 (3S HEMI IMPLANT, MEDIUM, NON-STERILE) WHILE ALSO INSERTING A 200-30-024 (3.0MM X 24MM TIGER CANNULATED SCREW) TO FIXATE THE PATIENT'S BUNION. THE SURGERY WAS REPORTED TO BE SUCCESSFUL. AT TIME OF REMOVAL, DOCTOR 1 NOTICED THE 200-30-024 WAS INTERFERING WITH THE PATIENT'S SESAM
Patient problems: Pain
MAUDE report 11406300 ↗
Other Screw, Fixation, Bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-200Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page