Boston Scientific Neuromodulation Corporation · CSK 15cm TC Electrode with 3m Cable · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00813250011548), received between Aug 15, 2017 and May 23, 2019 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Material Separation” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Apr 29, 2026Class IIOpen, Classified

    Reusable electrodes may not meet expected performance levels.

    Z-2402-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Separation3
  • Break2
  • Defective Device1
  • Material Discolored1
  • Material Erosion1

Patient problems

  • No Known Impact Or Consequence To Patient7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 23, 2019Defective Device; Defective Device

    DEVICE ANALYSIS PERFORMED ON THE RETURNED ELECTRODE FOUND THAT THE SHAFT WAS SEPARATED FROM THE HUB. THE DAMAGE WAS LIKELY DUE TO UNINTENDED EXCESSIVE EXTERNAL MECHANICAL FORCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8636428 ↗
  • MalfunctionAug 2, 2018Material Discolored; Material Separation; Material Discolored; Material Separation

    DEVICE ANALYSIS ON THE RETURNED ELECTRODE (L741) REVEALED THAT THE SHAFT OF THE ELECTRODE WAS FOUND TO BE COMPLETELY SEPARATED FROM THE HUB DUE TO EXCESSIVE EXTERNAL MECHANICAL FORCE. THE SHAFT WAS NOT RETURNED FOR ANALYSIS. THE FUNCTIONAL TEST COULD NOT BE PERFORMED DUE TO THE SEPARATED SHAFT. IN ADDITION A SECOND ELECTRODE (21816) WAS ANALYZED AND A VISUAL INSPECTION REVEALED A DISCOLORED EPOXY AND A CHIP OUT. THE COLOR OF NEWLY CURED EPOXY IS AMBER, LIGHT BROWN. IF THE EPOXY IS SUBJECTED TO TOO MANY AUTOCLAVE CYCLES, THE EPOXY BECOMES BRITTLE AND DISCOLORED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7744571 ↗
  • MalfunctionMay 30, 2018Material Separation; Material Separation

    THE DEVICE ANALYSIS ON THE RETURNED ELECTRODE REVEALED SEPARATION OF THE SHAFT FROM THE HUB. THE ROOT CAUSE IS EXCESSIVE EXTERNAL MECHANICAL FORCE PLACED UPON THE SHAFT LED TO THE COMPLETE SEPARATION OF THE SHAFT FROM THE HUB.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7555670 ↗
  • MalfunctionMay 10, 2018Material Separation; Material Separation

    DEVICE EVALUATION PERFORMED ON THE RETURNED ELECTRODE FOUND THAT THE SHAFT WAS COMPLETELY SEPARATED FROM THE HUB DUE TO EXCESSIVE MECHANICAL FORCE. THE SHAFT WAS NOT RETURNED FOR ANALYSIS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7502736 ↗
  • MalfunctionOct 18, 2017Break; Break

    THE RETURNED ELECTRODE WAS ANALYZED AND CONFIRMED THE ELECTRODE SHAFT WAS BROKEN OFF AT THE HUB, DUE TO EXTERNAL MECHANICAL FORCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6960020 ↗
  • MalfunctionOct 18, 2017Break; Break

    A REPORT WAS RECEIVED THAT AN ELECTRODE HAS A BROKEN WIRE FROM THE HUB.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6959987 ↗
  • MalfunctionAug 15, 2017Material Erosion; Material Erosion

    DEVICE ANALYSIS WAS PERFORMED ON THE RETURNED ELECTRODE AND FOUND THAT SOME ADHESIVE HAD MOVED OUT OF THE HUB HOLE AROUND THE ELECTRODE SHAFT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6796057 ↗

FDA clearance and classification

The FDA regulates product code GXD (generator, lesion, radiofrequency) under 21 CFR 882.4400 as a Class II device. 21 CFR 882.4400 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the NA CSK-TC15-3M

Has the NA CSK-TC15-3M been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2402-2026 (Class II), began on Apr 29, 2026: Reusable electrodes may not meet expected performance levels. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the NA CSK-TC15-3M?

The FDA's MAUDE database lists 7 adverse event reports for this device, received between Aug 15, 2017 and May 23, 2019. In those reports, the most common specific device problems are “Material Separation” (3), “Break” (2) and “Defective Device” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA CSK-TC15-3M?

None of the reports are classified as deaths, 0 as injuries and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the NA CSK-TC15-3M, and is it still sold?

The FDA's device database lists Boston Scientific Neuromodulation Corporation as the company responsible for it (brand NA, model CSK-TC15-3M). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other GENERATOR, LESION, RADIOFREQUENCY devices

All GENERATOR, LESION, RADIOFREQUENCY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page