Nevro® LEAD1058-50BRecalls & reported problems
Nevro Corp. · Blue Lead Kit, 50cm with 5mm spacing · Class 3 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Nevro spinal cord stimulator lead has 145 FDA adverse event reports, including 2 deaths and 116 injuries. Many describe infections around the lead site, and some mention spinal fluid leaking or wounds reopening, as well as parts detaching or fracturing. No reports came in the latest 12 months, against 9 in the 12 months before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 145
- Deaths reported
- 2
- Injuries reported
- 116
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 145 adverse event reports that list this exact device (DI 00813426020022), received between Oct 5, 2015 and Oct 20, 2025 — classified as 2 deaths, 116 injuries, 27 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem112
- Detachment of Device or Device Component14
- Fracture7
- Insufficient Device Problem Information5
- Break3
- Material Frayed3
- High impedance2
- Torn Material2
Patient problems
- Unspecified Infection37
- Cerebrospinal Fluid Leakage20
- Pain13
- Wound Dehiscence12
- No Known Impact Or Consequence To Patient11
- No Clinical Signs, Symptoms or Conditions9
- Device Embedded In Tissue or Plaque8
- Skin Erosion6
Recent adverse event reports
The 25 most recent of 145 reports, in the reporter's own words as published by the FDA.
InjuryOct 20, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. A MEDICAL ASSESSMENT FROM THE PHYSICIAN REGARDING THE SPECIFIC NATURE OF THE INFECTION WAS NOT AVAILABLE AT THE TIME OF REPORTING. THE PATIENT WAS HOSPITALIZED AND TREATED WITH INTRAVENOUS (IV) ANTIBIOTICS. IN ADDITION, THEY UNDERWENT DEBRIDEMENT OF AN ABSCESS, AND THE DEVICE WAS REMOVED. NO FURTHER COMPLICATIONS HAVE BEEN REPORTED FOLLOWING THIS EVENT.
Patient problems: Abscess; Insufficient Information; Unspecified Infection
MAUDE report 23341363 ↗InjurySep 9, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT ACQUIRED AN INFECTION AFTER A SPINE SURGERY. ACCORDING TO THE HCP, THE ISSUE WAS NOT RELATED TO THE DEVICE. THE PATIENT WAS TREATED WITH IV ANTIBIOTICS, AND THE DEVICE WAS REMOVED. THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: Insufficient Information; Unspecified Infection
MAUDE report 23001347 ↗InjuryJul 30, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT ACQUIRED AN INFECTION IN HER SPINE. A MEDICAL ASSESSMENT FROM THE PHYSICIAN REGARDING THE NATURE OF THE INFECTION WAS NOT AVAILABLE. THE PATIENT HAD THE DEVICE REMOVED AND THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: Insufficient Information; Unspecified Infection
MAUDE report 22663863 ↗InjuryJun 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED EPIDURAL ABSCESS. AN HCP ASSESSMENT REGARDING THE CAUSE OF THE REPORTED EVENT WAS NOT AVAILABLE. IT WAS NOT MENTIONED THAT THE EVENT IS RELATED TO THE DEVICE. THE PATIENT HAD THE DEVICE EXPLANTED. THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: Abscess
MAUDE report 22215713 ↗InjuryMay 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING PAIN, WHICH WAS ATTRIBUTED TO NERVE ROOT IRRITATION CAUSED BY THE LEAD. THE PATIENT HAD THE LEAD REMOVED AND THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: Pain
MAUDE report 21977259 ↗InjuryJan 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT ACQUIRED AN INFECTION AT THE LEAD INCISION SITE. THE PATIENT RECEIVED ORAL ANTIBIOTICS AND HAD THE DEVICE EXPLANTED. THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 21222993 ↗InjuryDec 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT ACQUIRED AN INFECTION AT THE LEAD INCISION SITE. THE PATIENT WAS HOSPITALIZED AND RECEIVED ORAL ANTIBIOTICS. THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: Bacterial Infection; Unspecified Infection
MAUDE report 20921285 ↗MalfunctionNov 20, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
AN MRI TECHNICIAN REPORTED A FAILED IMPEDANCE CHECK, CONFIRMED BY HEARING FOUR SHORT BEEPS AFTER MULTIPLE ATTEMPTS. FURTHER INVESTIGATION REVEALED THE PATIENT HAS A SINGLE PERC LEAD WITH ELECTRODE # 3 OUT OF ACCEPTABLE RANGE. A FRACTURE WAS CONFIRMED ON HIS LEAD. AN HCP ASSESSMENT REGARDING THE CAUSE OF THE REPORTED EVENT WAS NOT AVAILABLE. THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20739489 ↗MalfunctionNov 4, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT WHEN AN HCP DID AN X-RAY IT WAS FOUND BROKEN LEADS. AN HCP ASSESSMENT REGARDING THE CAUSE OF THE REPORTED EVENT WAS NOT AVAILABLE. THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20605682 ↗MalfunctionOct 2, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE EXPLANTING PHYSICIAN THAT TWO CONTACTS WERE MISSING WHEN THE LEADS WERE TAKEN OUT THE PATIENT. AN HCP ASSESSMENT REGARDING THE CAUSE OF THE REPORTED EVENT WAS NOT AVAILABLE. THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20353610 ↗InjuryAug 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT ACQUIRED AN INFECTION AT THE LEAD INCISION SITE. THE PATIENT RECEIVED ORAL ANTIBIOTICS AND THE ELECTRODES WERE REMOVED DURING THE DEVICE SURGERY AFTER IT WAS DISCOVERED THAT THE WOUND HAD BECOME INFECTED. THE ELECTRODES THEMSELVES WERE NOT INFECTED AND WERE ONLY REMOVED FOR SAFETY REASONS. THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 19913402 ↗MalfunctionJul 22, 2024Break; Break
IT WAS REPORTED THAT THE PATIENT'S LEADS ARE BROKEN. THE PATIENT REPORTED HE HAS FALLEN SEVERAL TIMES DUE TO HAVING SO MUCH PAIN IN HIS LOWER BACK AND LEGS THAT HE WAS HAVING BALANCE ISSUES. THE PATIENT MOSTLY USES A WHEEL CHAIR NOW. THE DOCTOR PERFORMED X-RAYS AND REPORTED THE BROKEN LEADS. THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: Balance Problems; Fall; Inadequate Pain Relief; Pain
MAUDE report 19794751 ↗InjuryJun 14, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION INDICATED THAT THE PATIENT HAD AN INFECTION AND THE LEADS WERE STICKING OUT. THE PATIENT HAD THE DEVICE EXPLANTED. THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: Impaired Healing; Skin Erosion; Unspecified Infection
MAUDE report 19536790 ↗MalfunctionFeb 27, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING REVISION SURGERY THE PATIENT'S LEADS WERE FRACTURED. AN HCP ASSESSMENT REGARDING THE CAUSE OF THE REPORTED EVENT WAS NOT AVAILABLE. THE PATIENT'S FULL SYSTEM WAS EXPLANTED AND REPLACED BECAUSE OF THIS ISSUE. THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18791618 ↗MalfunctionFeb 26, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE PATIENT HAD A LEAD BROKEN WHILE REMOVING THE DEVICE DUE TO AN MRI. AN HCP ASSESSMENT REGARDING THE CAUSE OF THE REPORTED EVENT WAS NOT AVAILABLE. THERE ARE NO FRAGMENTS IN THE EPIDURAL BUT THE SOFT TISSUE. THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18784720 ↗MalfunctionDec 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE PATIENT LEAD BROKE INTO 2 DISTINCT PIECES. AN HCP ASSESSMENT REGARDING THE CAUSE OF THE REPORTED EVENT WAS NOT AVAILABLE. THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: Foreign Body In Patient
MAUDE report 18414970 ↗MalfunctionDec 6, 2023Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT THE HCP PERFORMED A LEAD REVISION TO CORRECT THE ORIGINAL IMPLANTED LEAD THAT WAS LEFT OUTSIDE OF EPIDURAL SPACE. DURING THE CASE, THE HCP STATED HE FELT HE COULDN'T FULLY REMOVE THE OLD LEAD BECAUSE HE FELT HE WOULD¿VE DAMAGED THE GOOD LEAD BY PULLING IT OUT. THE HCP CUT THE ELECTRODES OFF OF BOTH SIDES OF THE LEAD AND THE LEAD WAS LEFT IN THE PATIENT. THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18279782 ↗MalfunctionOct 26, 2023Material Frayed; Insufficient Device Problem Information; Material Frayed; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT HAD A FRAYED LEAD INSULATION NEAR THE IPG CONNECTION END OF THE LEAD. AN HCP ASSESSMENT REGARDING THE CAUSE OF THE REPORTED EVENT WAS NOT AVAILABLE. THE DEVICE WAS REPLACED AND THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18014580 ↗MalfunctionOct 26, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT HAD A FRAYED LEAD INSULATION NEAR THE IPG CONNECTION END OF THE LEAD. AN HCP ASSESSMENT REGARDING THE CAUSE OF THE REPORTED EVENT WAS NOT AVAILABLE. THE DEVICE WAS REPLACED AND THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18014336 ↗InjurySep 4, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH VOLUNTARY MEDWATCH MW5119976 THAT THE PATIENT HAS EXPERIENCED INCREASING THORACIC PAIN FOLLOWING THE IMPLANT PROCEDURE. THE PATIENT HAS BEEN TREATED WITH IV MORPHINE, NERVE BLOCKS, EPIDURAL INJECTIONS AND A LEAD WAS REMOVED TO ATTEMPT TO RESOLVE THE PAIN. THE PATIENT SINCE CONTINUES TO USE THE DEVICE AND THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: Chest Pain
MAUDE report 17676544 ↗InjuryJun 30, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED WOUND HEALING COMPLICATIONS AT THE LEAD INCISION SITE. A MEDICAL ASSESSMENT FROM THE PHYSICIAN REGARDING THE NATURE OF THE REPORTED SYMPTOMS WAS NOT AVAILABLE. THE INCISION SITE WAS DRAINED, AND THE PATIENT WILL BE SCHEDULED FOR A DEVICE REMOVAL. THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: Impaired Healing
MAUDE report 17241108 ↗InjuryMay 4, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT'S SYSTEM WAS REMOVED DUE TO WOUND DEHISCENSE AT THE LEAD INCISION, AND INFECTION AT THE POCKET AND LEAD INCISION SITES. A MEDICAL ASSESSMENT FROM THE PHYSICIAN REGARDING THE NATURE OF THE INFECTIONS AND WOUND HEALING ISSUES WAS NOT AVAILABLE. THE DEVICES WERE REMOVED AND THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: Unspecified Infection; Wound Dehiscence
MAUDE report 16867935 ↗InjuryApr 24, 2023Migration; Migration
IT WAS REPORTED THAT THE PATIENT EXPERIENCED SHOCKING SENSATIONS AND LEAD MIGRATION. THE PHYSICIAN BELIEVED THE SHOCKING SENSATIONS WERE DEVICE RELATED. THE PATIENT WAS HOSPITALIZED. THE PERCUTANEOUS LEADS WERE REPLACED WITH A PADDLE LEAD AND THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 16802428 ↗InjuryMar 23, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
ADDITIONAL INFORMATION INDICATED THAT : "THE LEAD WAS EXPLANTED ON (B)(6) 2023 AND IS NOT AVAILABLE FOR RETURN."
Patient problems: Hematoma
MAUDE report 16600466 ↗MalfunctionFeb 10, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT'S TOP LEAD CONTACT IS APPROXIMATELY AT A 90 DEGREE ANGLE AND IS POSSIBLY FRACTURED. THE PHYSICIAN BELIEVES THIS WAS LIKELY CAUSED BY A FALL INVOLVING IMPACT TO THE NECK. THE PHYSICIAN DOES NOT BELIEVE THIS IS CAUSING THE PATIENT NECK PAIN AND IS NOT PLANNING ANY SURGICAL INTERVENTION. THERE HAVE BEEN NO REPORTS OF FURTHER COMPLICATIONS REGARDING THIS EVENT.
Patient problems: Fall
MAUDE report 16358631 ↗
Other Stimulator, Spinal-Cord, Totally Implanted For Pain Relief devices
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page