Skip to content

Impella 0550-0008Recalls & reported problems

Abiomed, Inc. · Impella 5.5 with Smart Assist Set, US · Class 3

All models of Impella Heart Pump (34)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Abiomed Impella 5.5 is a heart pump placed through a blood vessel to support the heart. The 1,005 reports often describe false placement alarms and sensor problems, along with bleeding and groin hematomas. Reports include 87 deaths and 410 injuries. 4 recalls are open, covering interactions with other devices, fibers drawn into the pump, and impeller damage causing low flow. Reports fell from 37 in the prior 12 months to 5 in the last 12.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,005
Deaths reported
87
Injuries reported
410
Recalls naming it
4

Summary

The FDA's MAUDE database contains 1,005 adverse event reports that list this exact device (DI 00813502011531), received between Sep 30, 2021 and Mar 11, 2026 — classified as 87 deaths, 410 injuries, 508 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (160 reports).

4 FDA recall records name this device's identifier, 4 still open.

21 other recalls cover devices cleared under the same FDA submission (P140003); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 0 reportsJan 2024: 3 reportsFeb 2024: 2 reportsMar 2024: 57 reportsApr 2024: 19 reportsMay 2024: 14 reportsJun 2024: 4 reportsJul 2024: 3 reportsAug 2024: 26 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 20 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 5 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 5 reportsSep 2025: 4 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202657
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 5, 2024Class IIOpen, Classified

    A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.

    Z-2963-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Dec 27, 2023Class IOpen, Classified

    IFU has been updated to include warnings about the risk of the inlet perforating through the myocardial wall of the left ventricle due to operator handling.

    Z-0980-2024 · Root cause: Labeling Change Control · Recall details · FDA record ↗

  3. Dec 27, 2023Class IIOpen, Classified

    New warnings are being added to the Instructions for Use for all affected products to reduce the possibility of fibers being drawn into the Impella, which may result in low flow of the device.

    Z-1074-2024 · Root cause: Labeling Change Control · Recall details · FDA record ↗

  4. Jun 14, 2023Class IOpen, Classified

    There is a potential risk for unintentional interaction of the Impella motor housing with the distal stent of a transcatheter aortic valve replacement (TAVR) resulting in destruction of the impeller blades. This has resulted in low flow from the damaged Impella system. Systemic embolization of the fractured impeller material is a possibility.

    Z-2114-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem413
  • Fluid/Blood Leak160
  • Activation, Positioning or Separation Problem110
  • Improper Flow or Infusion90
  • Restricted Flow rate71
  • Pumping Stopped47
  • Patient-Device Incompatibility38
  • Increase in Pressure35

Patient problems

  • No Clinical Signs, Symptoms or Conditions578
  • Hemorrhage/Blood Loss/Bleeding157
  • Thrombosis/Thrombus74
  • Hematoma53
  • Cardiovascular Dysfunction/Insufficiency37
  • Tachycardia26
  • Ischemia24
  • Hemolysis23

Recent adverse event reports

The 25 most recent of 1,005 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 11, 2026Adverse Event Without Identified Device or Use Problem

    A 69-YEAR-OLD FEMALE PATIENT WAS ADMITTED WITH ACUTE MYOCARDIAL INFARCTION CARDIOGENIC SHOCK. SHE WAS IMPLANTED WITH AN IMPELLA 5.5 FOR CARDIAC SUPPORT. THE PATIENT DEVELOPED A HEMATOMA AT THE GROIN ACCESS SITE AND HAD TO HAVE IT SURGICALLY EVACUATED. ON (B)(6) 2025, THE IMPELLA WAS REMOVED.

    Patient problems: Hematoma

    MAUDE report 24569768 ↗
  • MalfunctionDec 9, 2025Device Sensing Problem; Device Sensing Problem

    THE COMPLAINANT REPORTED THAT A PATIENT ON IMPELLA SUPPORT WAS EXPERIENCING PLACEMENT SIGNALING ISSUES DUE TO ALLEGEDLY A DAMAGED SENSOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23758691 ↗
  • MalfunctionDec 8, 2025Device Sensing Problem; Device Sensing Problem

    THE COMPLAINANT REPORTED THAT A PATIENT ON IMPELLA SUPPORT EXPERIENCED SOME INTERMITTENT FALSELY TRIGGERED POSITION IN AORTA ALARM. THE RN REVIEWED THE PATIENT ON BEDSIDE AND FOUND NO IMPELLA ISSUES. NO PATIENT HARM HAS BEEN REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23736880 ↗
  • MalfunctionDec 8, 2025Device Sensing Problem; Device Sensing Problem

    THE COMPLAINANT REPORTED THAT PLACEMENT SIGNAL MONITORING WAS DISABLED DUE TO ONGOING FALSE POSITION ALARMS. SUPPORT WAS MAINTAINED THROUGHOUT THE INCIDENT. NO PATIENT HARM HAS BEEN REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23736709 ↗
  • MalfunctionNov 12, 2025Device Sensing Problem; Device Sensing Problem

    THE COMPLAINANT REPORTED THAT THEIR AUTOMATED IMPELLA CONTROLLER (AIC) FAILED TO DISPLAY A PLACEMENT SIGNAL AND BEGAN TO ALARM WITH A PLACEMENT SIGNAL NOT RELIABLE ALARM. THE AIC WAS SWAPPED, HOWEVER THE PLACEMENT SIGNAL STILL FAILED TO REPOPULATE. PATIENT SUPPORT CONTINUED WITH NO KNOWN PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23532413 ↗
  • MalfunctionOct 21, 2025Device Sensing Problem; Device Sensing Problem

    THE COMPLAINANT REPORTED THAT THE PUMP EXPERIENCED AN OPTICAL SENSOR FAILURE AFTER THE DEVICE WAS REPOSITIONED DURING SUPPORT. BOTH THE PLACEMENT SIGNAL AND LV SIGNAL STOPPED WORKING AS A RESULT OF THE FAILURE. PATIENT SUPPORT CONTINUED WITH THE INSTALLED DEVICE AS THE MOTOR CURRENT, FLOWS, AND PURGE PRESSURE REMAINED WITHIN ACCEPTABLE LIMITS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23348108 ↗
  • DeathSep 12, 2025Device Sensing Problem; Device Sensing Problem

    THE COMPLAINANT REPORTED THAT A PATIENT ON IMPELLA 5.5 SUPPORT ENCOUNTERED A PLACEMENT SIGNALING ALARM. THE ALARM PERSISTED DESPITE CHANGING THE AUTOMATED IMPELLA CONTROLLER. THERE WAS NO HARM TO THE PATIENT AS A RESULT OF THIS INCIDENT, HOWEVER, THE PATIENT EVENTUALLY EXPIRED DURING SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23043188 ↗
  • DeathSep 12, 2025Device Sensing Problem; Device Sensing Problem

    THE COMPLAINANT REPORTED A PLACEMENT SIGNAL UNRELIABLE AND POSITION IN THE AORTA ALARMS OCCURRING ON AND OFF OVER THE WEEKEND FOR A PATIENT ON IMPELLA 5.5 SUPPORT. ECHOCARDIOGRAPHY SHOWED GOOD POSITION, AND STAFF ELECTED TO DISABLE AUDIO. THE DEVICE WAS NOT REPLACED OR REMOVED, AND NO CONSOLE WAS SWITCHED. THE PATIENT EVENTUALLY EXPIRED ON SUPPORT UNRELATED, TO THE ALARMS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23042938 ↗
  • MalfunctionSep 9, 2025Device Sensing Problem; Device Sensing Problem

    THE COMPLAINANT REPORTED THAT AN IMPELLA 5.5 PUMP WAS IMPLANTED FOR CIRCULATORY SUPPORT. DURING SUPPORT, THE PLACEMENT SIGNAL WAS LOST. THE ALARM WAS DISABLED, AND PUMP SUPPORT CONTINUED WITHOUT INTERRUPTION. THERE WAS NO REPORTED HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23006876 ↗
  • MalfunctionSep 9, 2025Device Sensing Problem; Device Sensing Problem

    THE COMPLAINANT REPORTED THAT AN IMPELLA 5.5 PUMP WAS IMPLANTED FOR CIRCULATORY SUPPORT. DURING SUPPORT, THE PLACEMENT SIGNAL BECAME UNRELIABLE. THE ALARM WAS DISABLED, AND PUMP SUPPORT CONTINUED WITHOUT INTERRUPTION. THERE WAS NO REPORTED HARM OR INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23006787 ↗
  • MalfunctionAug 6, 2025Device Sensing Problem; Device Sensing Problem

    THE COMPLAINANT REPORTED THAT AN IMPELLA 5.5 PUMP WAS IMPLANTED FOR CIRCULATORY SUPPORT. DURING SUPPORT, THE PUMP TRIGGERED PLACEMENT SIGNAL ISSUES. THE PUMP WAS NOT REMOVED. SUPPORT WAS CONTINUED, AND NO HARM OCCURRED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22731865 ↗
  • MalfunctionAug 6, 2025Device Sensing Problem; Device Sensing Problem

    THE COMPLAINANT REPORTED THAT AN IMPELLA 5.5 PUMP WAS IMPLANTED FOR CIRCULATORY SUPPORT. DURING SUPPORT, THE PUMP TRIGGERED MULTIPLE 'IMPELLA IN AORTA' ALARMS DESPITE CONFIRMED PROPER POSITIONING. SUPPORT WAS CONTINUED, AND NO HARM OCCURRED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22731835 ↗
  • MalfunctionAug 6, 2025Device Sensing Problem; Device Sensing Problem

    THE COMPLAINANT REPORTED THAT AN IMPELLA 5.5 PUMP WAS IMPLANTED FOR CIRCULATORY SUPPORT. THE PUMP TRIGGERED A 'PLACEMENT SIGNAL LOST' ALARM AFTER MORE THAN 70 DAYS OF RUNTIME. PUMP POSITION WAS CONFIRMED VIA ECHOCARDIOGRAPHY. SUPPORT WAS CONTINUED, AND NO HARM OCCURRED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22731707 ↗
  • MalfunctionAug 6, 2025Device Sensing Problem; Device Sensing Problem

    THE COMPLAINANT REPORTED THAT AN IMPELLA 5.5 PUMP WAS IMPLANTED FOR CIRCULATORY SUPPORT. THE PUMP TRIGGERED A 'PLACEMENT SIGNAL NOT RELIABLE' ALARM BOTH DURING SETUP AND SUPPORT. SUPPORT WAS CONTINUED, AND NO HARM OCCURRED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22731248 ↗
  • MalfunctionAug 6, 2025Device Sensing Problem; Device Sensing Problem

    THE COMPLAINANT REPORTED THAT AN IMPELLA 5.5 PUMP WAS IMPLANTED FOR CIRCULATORY SUPPORT. DURING USE, THE PUMP TRIGGERED A FALSE 'IMPELLA IN AORTA' ALARM. PROPER PUMP POSITIONING WAS CONFIRMED VIA ECHOCARDIOGRAPHY. NO HARM OCCURRED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22730723 ↗
  • DeathMar 31, 2025Restricted Flow rate; Malposition of Device; Restricted Flow rate; Malposition of Device

    THE COMPLAINANT REPORTED A PATIENT IN NON-ST ELEVATION MYOCARDIAL INFARCTION WITH HYPOTENSION, ACUTE SEVERE MITRAL REGURGITATION, AND LOW EJECTION FRACTION WAS IMPLANTED WITH AN IMPELLA 5.5 DEVICE FOR MECHANICAL CIRCULATORY SUPPORT TO STABILIZE AND UNLOAD THE PATIENT'S LEFT HEART. DURING DELIVERY OF THE IMPELLA INTO THE LEFT VENTRICLE, THE IMPELLA 5.5 WAS PLACED DIRECTLY INFO THE PAPILLARY WALL AND WAS UNABLE TO BE MOVED. THE MEDICAL TEAM WAS ADVISED TO PULL THE GUIDEWIRE, AND THEN TURN THE IMPELLA AS THERE WAS SUSPICION THE GUIDEWIRE WAS IN THE CHORDAE. THE PATIENT THEN WENT INTO BRADYCARDIA AND LOST BLOOD PRESSURES. THE IMPELLA 5.5 WAS SUCCESSFULLY TURNED TOWARDS THE APEX AFTER REMOVAL OF THE GUIDEWIRE AND SUPPORT WAS THEN INITIATED. IT WAS NOTED BY THE PHYSICIAN THEY BELIEVE THAT THE PATIENT'S PRESSURES DROPPED DUE TO THE RIGHT CORONARY ARTERY BEING COMPLETELY BLOCKED AND HAVING NO PERFUSION. DURING SUPPORT, THERE WERE SUCTION ALARMS. THE POWER LEVEL OF THE IMPELLA 5.5 WAS DROPPED FROM P9 TO P8 AND THE PATIENT WAS GIVEN FLUID VOLUME TO RESOLVE THE ALARMS. THE PATIENT REMAINED ON SUPPORT FOR 5 DAYS, AND IT WAS DETERMINED THE PATIENT WAS UNLIKELY TO SURVIVE ANY FURTHER REVASCULARI

    Patient problems: Bradycardia

    MAUDE report 21732384 ↗
  • DeathMar 31, 2025Increase in Pressure; Increase in Pressure

    THE COMPLAINANT REPORTED A PATIENT IN ACUTE MYOCARDIAL INFARCTION/CARDIOGENIC SHOCK WAS IMPLANTED WITH AN IMPELLA CP DEVICE FOR MECHANICAL CIRCULATORY SUPPORT. THE PATIENT WAS THEN ESCALATED FROM THE IMPELLA CP TO AN IMPELLA 5.5 FOR INCREASED CIRCULATORY SUPPORT AFTER 4 DAYS OF SUPPORT ON THE IMPELLA CP. WHEN THE PATIENT WAS BROUGHT TO THE OPERATING ROOM (OR) TO REPLACE THE IMPELLA CP WITH IMPELLA 5.5 THE DECISION WAS MADE FOR THE PATIENT TO HAVE CORONARY ARTERY BYPASS (CABG). THE PATIENT WAS PLACED ON BYPASS, AND THE IMPELLA WAS PLACE DIRECTLY THROUGH THE AORTA AFTER AN AORTIC VALVE REPLACEMENT WAS PERFORMED. CABG WAS PERFORMED SUCCESSFULLY AND WITHOUT COMPLICATIONS. THE IMPELLA 5.5 PLACEMENT WAS CONFIRMED WITH TRANSESOPHAGEAL ECHOCARDIOGRAPHY (TEE), SUPPORT WAS INITIATED AND THE PATIENT WAS WEANED OFF BYPASS. TEE OF THE RIGHT VENTRICLE REVEALED WORSENING RIGHT HEART FUNCTION. THE PATIENT WAS GIVEN AN INCREASED INOTROPE SUPPORT AND TRANSPORTED BACK TO THE ICU. THE FOLLOWING DAY, ON (B)(6) 2022, THE PATIENT WAS TAKEN BACK TO THE OR FOR VENOARTERIAL EXTRACORPOREAL MEMBRANE OXYGENATION CANNULATION (VA ECMO) DUE TO THE FAILING RIGHT VENTRICLE AND ASSOCIATED DECOMPENSATION. DURING THIS

    Patient problems: Insufficient Information

    MAUDE report 21732344 ↗
  • MalfunctionMar 17, 2025Failure to Advance; Patient-Device Incompatibility; Failure to Advance; Patient-Device Incompatibility

    THE COMPLAINANT REPORTED THAT AN IMPELLA 5.5 PUMP WAS UNABLE TO PASS THROUGH THE PATIENT'S VASCULATURE DURING ATTEMPTED DELIVERY. AS A RESULT OF THE RESISTANCE CAUSED BY THE PATIENT'S ANATOMY, THE DECISION WAS MADE TO ABORT THE IMPLANT. AN INTRA-AORTIC BALLOON PUMP WAS SUBSEQUENTLY IMPLANTED FOR PATIENT SUPPORT. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21623635 ↗
  • MalfunctionMar 17, 2025Failure to Advance; Patient-Device Incompatibility; Failure to Advance; Patient-Device Incompatibility

    THE COMPLAINANT REPORTED THAT AN IMPELLA 5.5 PUMP WAS UNABLE TO PASS THROUGH THE PATIENT'S TORTUOUS VASCULATURE DURING ATTEMPTED DELIVERY. AS A RESULT OF THE NOTED RESISTANCE CAUSED BY THE PATIENT'S ANATOMY, THE DECISION WAS MADE TO ABORT THE IMPLANT. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21620452 ↗
  • MalfunctionMar 17, 2025Failure to Advance; Patient-Device Incompatibility; Failure to Advance; Patient-Device Incompatibility

    THE COMPLAINANT REPORTED THAT THE MEDICAL TEAM WAS HAVING DELIVERY ISSUES WITH THE IMPELLA 5.5 PUMP BECAUSE THE AXILLARY ARTERY WAS TOO SMALL SO THE PUMP WAS ABORTED. THERE WAS NO HARM TO THE PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21619988 ↗
  • InjuryFeb 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED VIA THE IMPACT EU STUDY A PATIENT EXPERIENCED A MEDIASTINAL HEMORRHAGE WITH A DEFINITE RELATIONSHIP TO IMPELLA DEVICE. THE ADVERSE EVENT NARRATIVE EXPLAINED THE PATIENT WAS GOING TO BE DISCHARGED THE DAY OF THE EVENT (WHICH WAS FOUR DAYS AFTER EXPLANT OF THE IMPELLA 5.5 DEVICE; THE PATIENT RECOVERED TO NATIVE HEART FUNCTION). THE PATIENT WAS UNRESPONSIVE AFTER THE SECOND WALK TO THE ROOM ON THE UNIT. THE PATIENT'S PULSE DROPPED, AND CODE BLUE WAS ACTIVATED. CARDIOPULMONARY WAS ADMINISTERED AND RETURN OF SPONTANEOUS CIRCULATION ACHIEVED. THERE WAS A SIGNIFICANT GROWING LEFT PLEURAL EFFUSION. LEFT CHEST TUBE WAS PLACED WITH RETURN OF 1500ML OF BLOOD. THE PATIENT RECEIVED 4 UNITS OF PACKED RED BLOOD CELLS, 1 FRESH FROZEN PLASMA AND 1 PLATELET. ARTERIAL BLOOD GAS REVEALED ACIDOSIS, 4 AMPS. BICARBONATE WAS GIVEN. THE PATIENT WAS TAKEN TO THE OPERATING ROOM FOR MEDIASTINAL WASHOUT. SMALL BLEEDING FROM THE SIDE OF THE RIGHT CORONARY BUTTON WAS FOUND AND STITCH WAS PLACED. THE PATIENT WAS NOTED TO HAVE SURVIVED THE EVENT.

    Patient problems: Cardiac Arrest; Cardiac Tamponade; Hematoma; Hemorrhage/Blood Loss/Bleeding; Hemothorax; Intraoperative Cardiac Injury; Pleural Effusion; Thrombosis/Thrombus; Ventricular Fibrillation

    MAUDE report 21420301 ↗
  • InjuryDec 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ABIOMED JAPAN FORWARDED 7 PUBLICATIONS. ONE FROM TEAM IN CT/(B)(6) HOSPITAL WHICH WAS ENTITLED: "ONE PUMP FOR TWO HEARTS: USING AN IMPELLA 5.5 MICRO-AXIAL PUMP IN PERIPARTUM CARDIOGENIC SHOCK" WHICH DETAILS A CP SUPPORT IN AN EXPECTANT MOTHER WHO HAD AN ACCESS SITE OOZE. UNTIL THE DAY BEFORE DELIVERY, SHE WAS ANTICOAGULATED WITH HEPARIN, ALONG WITH A HEPARIN-BASED PURGE FLUID. A DAY BEFORE DELIVERY, SHE DEVELOPED EPISTAXIS AND OOZING FROM THE IMPELLA PLACEMENT SITE. ANTICOAGULATION WAS TEMPORARILY STOPPED, AND THE HEPARIN-BASED PURGE FLUID WAS SUBSTITUTED WITH 5% DEXTROSE BUT OOZING CONTINUED. ULTIMATELY A DECISION TO PROCEED WITH DELIVERY WAS MADE DUE TO PERSISTENT OOZING DESPITE HOLDING HEPARIN AND SEVERE FETAL GROWTH RESTRICTION. THE PROCEDURE WAS SUCCESSFUL WITH ANOTHER DEVICE AND THE PROCEDURAL OUTCOME WAS THE PATIENT SURVIVED SURGERY.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 21021688 ↗
  • DeathNov 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE COMPLAINANT REPORTED A PATIENT WITH CARDIOMYOPATHY WAS IMPLANTED WITH AN IMPELLA 5.5 FOR MECHANICAL CIRCULATORY SUPPORT. THE PATIENT HAD BEEN ESCALATED FROM AN IMPELLA CP TO THE IMPELLA 5.5 AS HEMODYNAMIC SUPPORT NEED HAD INCREASED. AFTER BEING IMPLANTED WITH THE IMPELLA 5.5 IT WAS REPORTED THAT THE PATIENT WAS BLEEDING FROM MULTIPLE SITES AND WAS TRANSFUSED WITH BLOOD PRODUCTS AS A RESULT OF THE BLEEDING. ON (B)(6) 2022, THE PATIENT WAS DIAGNOSED WITH HEPARIN INDUCED THROMBOCYTOPENIA. IT WAS REPORTED THE PATIENT'S PLATELET COUNT WAS 19,000 CELLS PER MICROLITER. THE PATENT WAS TRANSFUSED WITH PLATELETS AND BLOOD PRODUCTS. ADDITIONALLY, THE PATIENT HAD THREE EPISODES OF VENTRICULAR FIBRILLATION AND REQUIRED DEFIBRILLATION TO RESOLVE EACH EPISODE. A FAMILY MEETING WAS CALLED TO DISCUSS WITHDRAWING CARE. IT IS UNKNOWN IF CARE WAS WITHDRAWN, HOWEVER IT IS NOTED THAT THE PATIENT EXPIRED WHILE ON SUPPORT. THERE IS NOT ENOUGH EVIDENCE TO EXCLUDE IMPELLA AS AN ASSOCIATED FACTOR IN THE PATIENT'S DEMISE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Thrombocytopenia; Ventricular Fibrillation

    MAUDE report 20814125 ↗
  • DeathNov 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE COMPLAINANT REPORTED A PATIENT IN ACUTE MYOCARDIAL INFARCTION/CARDIOGENIC SHOCK WAS IMPLANTED WITH AN IMPELLA 5.5 DEVICE FOR MECHANICAL CIRCULATORY SUPPORT. WHILE ON SUPPORT IT WAS NOTED THAT THERE WAS BLEEDING AT THE ACCESS SITE THAT WAS BEING MANAGED WITH WOUND VAC AND TRANSFUSIONS. ADDITIONAL INFORMATION WAS PROVIDED THAT NOTED THE PATIENT EXPIRED. THERE IS NOT ENOUGH INFORMATION TO EXCLUDE THE IMPELLA DEVICE AS AN ASSOCIATED FACTOR IN THE PATIENT'S DEMISE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 20769001 ↗
  • DeathNov 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED A PATIENT WITH CARDIOGENIC SHOCK FROM AN ACUTE MYOCARDIAL INFARCTION WAS IMPLANTED WITH AN IMPELLA 5.5 DEVICE FOR MECHANICAL CIRCULATORY SUPPORT AND EXPERIENCED A HEMORRHAGIC STROKE WITH A DEMISED OUTCOME. THE PATIENT WAS ESCALATED FROM A IMPELLA CP TO THE IMPELLA 5.5 FOR HEMODYNAMIC SUPPORT. SIX DAYS INTO SUPPORT, THE PATIENT REMAINED INTUBATED AND SEDATED BUT FOLLOWED COMMANDS, AND THE PATIENT WAS NOTED TO HAVE EXPERIENCED A SUBARACHNOID BLEED. HEPARIN WAS PUT BACK INTO PURGE AFTER REPEAT HEAD COMPUTED TOMOGRAPHY SCAN. OF NOTE, UNRELATED, IT WAS DOCUMENTED THAT THE PATIENT WAS BEING TREATED FOR A FUNGAL INFECTION IN LUNGS. CARE WAS EVENTUALLY WITHDRAWN FOUR DAYS LATER, AND THE PATIENT EXPIRED. NO FURTHER INFORMATION WAS NOTED.

    Patient problems: Hemorrhagic Stroke

    MAUDE report 20768908 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P140003. One submission often covers many models, so check the product and lot details in each record.

  • Z-3148-2026 · Aug 12, 2026 · Open, Classified
    Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella
  • Z-3150-2026 · Aug 12, 2026 · Open, Classified
    Impella Optical Controller, Automated Impella Controller, Impella Connect Package with the below product descriptions and corresponding Product Codes. 1. Optical, AIC, Impella Conn
  • Z-3149-2026 · Aug 12, 2026 · Open, Classified
    Impella Optical Controller with the below product descriptions and corresponding Product Codes. 1. Impella Optical Controller, Packaged, AU; Product Code: 0042-0010-AU. 2. Impella
  • Z-2457-2026 · May 18, 2026 · Open, Classified
    Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.
  • Z-2360-2026 · May 14, 2026 · Open, Classified
    Automated Impella Controller (AIC), for use with left heart support blood pump. Labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, AU; P
  • Z-2211-2026 · Apr 20, 2026 · Open, Classified
    Automated Impella Controller (AIC); Product Code: 0042-0000-US;
  • Z-1589-2026 · Feb 18, 2026 · Open, Classified
    Purge Cassette, sold within Impella Pump Sets and Individually Packaged. Labeled and distributed in the below configurations and corresponding product codes. 1. Purge Cassette, 5
  • Z-1591-2026 · Feb 16, 2026 · Open, Classified
    Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Contr
  • Z-1030-2026 · Dec 4, 2025 · Open, Classified
    Impella 5.5 with SmartAssist S2, Impella 5.5 with SmartAssist S2 Set, AU.
  • Z-0069-2026 · Oct 1, 2025 · Open, Classified
    Automated Impella Controller (AIC), used for left heart support blood pump, labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, AU; Product
  • Z-0136-2026 · Sep 16, 2025 · Open, Classified
    Automated Impella Controller (AIC) labeled as the following with corresponding Product Codes: 1. AIC w/Impella Connect for ECP, Product Code: 1000432. 2. Dbl optical, AIC Impella
  • Z-2164-2025 · Jun 23, 2025 · Open, Classified
    Automated Impella Controller (AIC), used in left heart support blood pump, including the following: 1. Impella Controller, Packaged, AU; Product Code: 0042-0000-AU. 2. Impella
  • Z-2319-2024 · May 31, 2024 · Open, Classified
    Impella CP with SmartAssist, Product No.: 1000080 (REF 0048-0003). left heart support blood pump
  • Z-0636-2024 · Dec 6, 2023 · Open, Classified
    Impella Connect, software portal for online device management that allows clinicians and Impella support staff to remotely view Impella data.
  • Z-0441-2024 · Oct 10, 2023 · Terminated
    Automated Impella Controller (AIC), product number 0042-040-US with software versions V8.4 and V8.4.1.

…and 6 more.

Other Temporary Non-Roller Type Left Heart Support Blood Pump devices

All Temporary Non-Roller Type Left Heart Support Blood Pump devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page