Impella catheters - Intravascular micro axial blood pumps… recallZ-0980-2024 · Abiomed, Inc. · initiated Dec 27, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abiomed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

IFU has been updated to include warnings about the risk of the inlet perforating through the myocardial wall of the left ventricle due to operator handling.

FDA root cause category: Labeling Change Control

What the company is doing

Firm notified affected consignees of the updated warnings in the IFU beginning on December 27, 2023. Customers were provided with details of the IFU modifications. Customers should notify everyone at their facility who needs to be informed of this correction. If products have been forwarded to another facility, please notify that facility of the updated IFUs. Post a copy of the recall notice in a visible area for awareness. If you have questions or concerns regarding this notice, please contact [email protected] and/or your local clinical field staff.

Recall details

Recall number
Z-0980-2024
Status
Open, Classified
Initiated by the company
Dec 27, 2023
Posted by the FDA
Feb 9, 2024
Recalling firm
Abiomed, Inc.
Quantity
91,914 (65,857 US; 26,075 OUS)
Product code
OZD
FDA submissions
P140003, P170011
Distribution
Worldwide distribution - US Nationwide and the countries of AU, CA, DE, FR, IN, MX, TW, and VI.

Product description

Impella catheters - Intravascular micro axial blood pumps that support a patient's circulatory system. (1) Product Code 005042 - Impella 2.5 (2) Product Code 005062 - Impella 5.0 (3) Product Code 005082- Impella LD (7) Product Codes 0550-0008 and 1000100 - Impella 5.5 with SmartAssist (8) Product Code 0048-0032 - Impella CP (9) Product Codes 1000080 and 0048-0045 - Impella CP with SmartAssist ***Updated February 2024*** (10) 004413 - Impella 2.5 Set (12) 0046-0026 - Impella 5.0 Pump Set ROW (13) 0046-0037 - Impella 5.0 Pump Set APAC (15) 004680-AU - 5.0 Pump Set AU (16) 0048-0002 - Impella CP Pump Set, EU (17) 0048-0002-BR - Impella CP Pump Set BR (18) 0048-0004 - Impella CP Pump Set, Canada (19) 0048-0014 - Impella CP Smart Assist Set, EU (20) 0048-0024-JP - Impella CP Smart Assist Set, JP (21) 0048-0044 - Impella CP Smart Assist Set, Canada (22) 0048-0047 - Impella CP Smart Assist Set APAC (23) 005040 - Impella 2.5 IMC Pump Set EU (24) 005048-JP - Impella 2.5 Pump Set, Japan (25) 005060 - Impella 5.0 IMC Pump Set EU (26) 005064 - Impella 5.0 IMC Pump Set Canada (27) 005066-JP - Impella 5.0 Pump Set, Japan (28) 0550-0002 - Impella 5.5 with SmartAssist Set, EU (29) 0550-0004 - Impella 5.5 with SmartAssist Set, CA (30) 1000115 - Impella CP Pump set, APAC (31) 1000211 - Impella 5.5 SmartAssist Set, JP (32) 1000302 - Impella CP with SmartAssist APAC (33) 1000361 - Impella 5.5 Set AU (34) 1000402 - Impella CP Smart Assist Set ***Updated March 15, 2024*** The following products were removed from the list of affected products because the Right-side devices are not in the scope of the warning to avoid left ventricle perforation: (4) Product Code 004334 - Impella RP (5) Product Code 0046-0035 - Impella RP with SmartAssist (6) Product Code 1000323 - RP Flex with SmartAssist (11) 0046-0011 - Impella RP Pump Set, EU (14) 0046-0039 - Impella RP Set APAC

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

UDI-DI (1) 00813502011081 (2) 00813502011180 (3) 00813502011227 (7) 00813502011531 and 00813502012828 (8) 00813502011388 (9) 00813502012279 and 00813502011876 All product IFUs include the Update ***Added February 2024*** (10) 813502010947 (12) 813502011821 (13) 813502011937 (15) N/A (16) 4260113630242 (17) 813502011777 (18) 813502011265 (19) 4260113630280 (20) 813502011609 (21) 813502012200 (22) 813502011944 (23) 4260113630136 (24) 813502010046 (25) 4260113630174 (26) 813502011197 (27) 813502010053 (28) 813502011630 (29) 813502010466 (30) 813502010473 (31) 813502012453 (32) 813502012767 (33) 813502012873 (34) 813502012873 ***Removed from scope March 15, 2024*** (4) 00813502011029 (5) 00813502011869 (6) 00813502012811 (11) 4260113630273 (14) 813502011951

Devices named in this recall

12 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

7 devices cleared under P140003, P170011 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.