Bio-Med Devices, Inc. CROSSVENT (REF 2200IC)Recalls & reported problems

Bio-Med Devices, Inc. · Ventilator, Transport · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00813841020805), received between Jun 6, 2016 and May 27, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1
  • Device Displays Incorrect Message1
  • Device Inoperable1
  • Device Operates Differently Than Expected1
  • Loss of Power1
  • Premature Discharge of Battery1
  • Pressure Problem1

Patient problems

  • Low Oxygen Saturation1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1
  • Respiratory Insufficiency1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 27, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    USER NARRATIVE: "THIS VENT IS USED ON OUR TRANSPORT STRETCHER. A THERAPIST WENT TO PICK UP AN INTUBATED BABY. WHILE HE WAS IN THE REFERRING HOSPITAL, HE PLACED THE BABY ON THE CROSSVENT WITH THE AIR AND OXYGEN HOOKED TO THE WALL. BABY WAS FINE. HE SWAPPED GASES TO THE TANKS AND TOOK THE BABY TO THE AMBULANCE. ONCE HE GOT ON THE AMBULANCE THE BABY BEGAN TO DESAT, CHEST MOVEMENT DECREASED, AND POOR BREATH SOUNDS NOTED. HE HAD TO GASES HOOKED TO THE WALL OUTLET OF THE AMBULANCE. BOTH CYLINDERS ON THE AMBULANCE WERE FULL. ET TUBE WAS CONFIRMED IN PLACE. THEY TOOK THE BABY OFF THE VENT AND BAGGED. THE SATS CAME UP AND THE BREATH SOUNDS WERE GOOD. HE PUT THE BABY BACK ON THE VENT AND WITHIN A COUPLE OF MINS THE SAME SCENARIO HAPPENED, DESAT, POOR BREATH SOUNDS AND DECREASED CHEST MOVEMENT. HE ENDED UP BAGGING THE BABY FOR THE REMAINDER OF THE TRANSPORT. HE SAID THE VENT WAS SHOWING THE SET PIP AND PEEP. NOTHING LOOKED OFF ON THE VENT SCREEN. ONCE HE GOT BACK, WE LEFT THE CIRCUIT ON AND TRIED TO TROUBLESHOOT. WE PLACED A TEST LUNG AND STARTED THE VENT. NOTHING LOOKED OUT OF PLACE. THE TEST LUNG HELD THE PEEP AND IT WAS NOTICEABLE WHEN A BREATH WAS DELIVERED. YOU COULD TAKE THE TEST LUNG O

    Patient problems: Respiratory Insufficiency

    MAUDE report 22088473 ↗
  • MalfunctionAug 3, 2023Premature Discharge of Battery; Premature Discharge of Battery

    UNIT STOPPED VENTILATING, WITH ALARMS, WHILE ON PATIENT. NO PATIENT INJURY. ASSUMEDLY, BACKUP VENTILATION WAS USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17452209 ↗
  • MalfunctionApr 7, 2018Device Inoperable

    TRANSPORT VENTILATOR "IS SAID TO HAVE STOPPED WORKING DURING PATIENT TRANSPORT." STILL TRYING TO OBTAIN INCIDENT DETAILS. NOT CLEAR YET IF THE VENTILATOR WAS ACTUALLY APPLIED TO A PATIENT AT THE TIME.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7405605 ↗
  • MalfunctionJun 6, 2016Pressure Problem

    NARRATION FROM USER OF TRANSPORT VENTILATOR: AFTER 20 MINUTES ON A TRANSPORT VENTILATOR, 'PATIENT O2 SATS DROPPED. COULD NOT DIAL IN A PIP OR PEEP PRESSURE. COULD NOT HEAR THE VENT CYCLING.' PATIENT WAS THEN HAND-VENTILATED INSTEAD, AND THERE WAS NO INJURY TO THE PATIENT. VENTILATOR WAS REMOVED FROM SERVICE, AND HOSPITAL TECHNICIANS COULD FIND NOTHING WRONG. TECHNICIANS DID NOTICE THAT ONE OF THE SUPPLY OXYGEN TANKS TAKEN ON THE TRANSPORT WAS EMPTY. IT WAS SUSPECTED THAT RUNNING OUT OF PRESSURIZED SUPPLY GAS MAY HAVE CAUSED THIS INCIDENT.

    Patient problems: Low Oxygen Saturation

    MAUDE report 5701969 ↗
  • Jun 6, 2016Loss of Power; Device Displays Incorrect Message; Device Operates Differently Than Expected

    TRANSPORT VENTILATOR "TURNS OFF WHILE ON PATIENTS" WITH ALARMS SOUNDING. NO DETAILS ON ANY SPECIFIC INCIDENTS PROVIDED BY USER, AND THERE WERE NO PATIENT INJURIES. DEVICE WAS RUN ON BATTERY POWER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5701319 ↗

Other Ventilator, Continuous, Facility Use devices

All Ventilator, Continuous, Facility Use devices →