EndoVive Gastrostomy Tube, Right Angle, with ENFit Connector M00510010Recalls & reported problems

Xeridiem Medical Devices, Inc. · EndoVive Gastrostomy Tube, Right Angle, with ENFit Connector, 22FR · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00813939023770), received between Nov 25, 2025 and Nov 25, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 6 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash6

Patient problems

  • Abdominal Pain6
  • Burn(s)6
  • Discomfort6
  • Gastrointestinal Hemorrhage6
  • Inflammation6
  • Swelling/ Edema6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 25, 2025Leak/Splash; Leak/Splash

    ECN EVENT DESCRIPTION: SITUACI6N QUE NOS REPORTAN EN SURA DE LAS SONDAS DE RECAMBIO ENFIT, HAN TENIDO MAS DE 5 PCTES, LAS CUALES HAN PRESENTADO SANGRADO PERIOSTOMAL DESDE EL CAMBIO DE LAS SONDAS, FILTRACIONES, DAI'IO DEL ESTOMA Y PIEL PERILESIONAL, ESTO NO HABIA OCURRIDO ANTES Y SE HA VENIDO EVIDENCIANDO DESDE LAS COLOCACIONES DE SEPTIEMBRE, AGRADEZCO CON LATAM SI PUEDES VALIDAR SI HAN PRESENTADO UNA SITUACI6N SIMILAR. COMPARTO EVIDENCIAS FOTOGRAFICAS DE LAS CASES SITUATION REPORTED TO US AT SURA REGARDING ENFIT REPLACEMENT TUBES: THEY HAVE HAD MORE THAN 5 PATIENTS WHO HAVE PRESENTED PERISTOMAL BLEEDING SINCE THE TUBE CHANGE, LEAKS, DAMAGE TO THE STOMA AND PERILESIONAL SKIN. THIS HAD NOT OCCURRED BEFORE AND HAS BEEN EVIDENT SINCE THE SEPTEMBER PLACEMENTS. I WOULD APPRECIATE IT IF LATAM COULD VALIDATE WHETHER THEY HAVE PRESENTED A SIMILAR SITUATION. I SHARE PHOTOGRAPHIC EVIDENCE OF THE CASES WHAT TROUBLESHOOTING STEPS TOOK PLACE? WHAT TROUBLESHOOTING STEPS, IF ANY, RESOLVED THE ISSUE?: VALORACI6N PAR ESPECIALISTA - SPECIALIST ASSESSMENT WHAT IS THE NEXT COURSE OF ACTION?: VALORACI6N PAR ESPECIALISTA - SPECIALIST ASSESSMENT PATIENT PRESENT AT TIME OF EVENT?: YES PATIENT COMPLICATIONS

    Patient problems: Abdominal Pain; Burn(s); Discomfort; Gastrointestinal Hemorrhage; Inflammation; Swelling/ Edema

    MAUDE report 23652294 ↗
  • MalfunctionNov 25, 2025Leak/Splash; Leak/Splash

    ECN EVENT DESCRIPTION: SITUACIÓN QUE NOS REPORTAN EN SURA DE LAS SONDAS DE RECAMBIO ENFIT, HAN TENIDO MÁS DE 5 PCTES, LOS CUALES HAN PRESENTADO SANGRADO PERIOSTOMAL DESDE EL CAMBIO DE LAS SONDAS, FILTRACIONES, DAÑO DEL ESTOMA Y PIEL PERILESIONAL, ESTO NO HABÍA OCURRIDO ANTES Y SE HA VENIDO EVIDENCIANDO DESDE LAS COLOCACIONES DE SEPTIEMBRE, AGRADEZCO CON LATAM SI PUEDES VALIDAR SI HAN PRESENTADO UNA SITUACIÓN SIMILAR. COMPARTO EVIDENCIAS FOTOGRÁFICAS DE LOS CASOS SITUATION REPORTED TO US AT SURA REGARDING ENFIT REPLACEMENT TUBES: THEY HAVE HAD MORE THAN 5 PATIENTS WHO HAVE PRESENTED PERISTOMAL BLEEDING SINCE THE TUBE CHANGE, LEAKS, DAMAGE TO THE STOMA AND PERILESIONAL SKIN. THIS HAD NOT OCCURRED BEFORE AND HAS BEEN EVIDENT SINCE THE SEPTEMBER PLACEMENTS. I WOULD APPRECIATE IT IF LATAM COULD VALIDATE WHETHER THEY HAVE PRESENTED A SIMILAR SITUATION. I SHARE PHOTOGRAPHIC EVIDENCE OF THE CASES WHAT TROUBLESHOOTING STEPS TOOK PLACE? WHAT TROUBLESHOOTING STEPS, IF ANY, RESOLVED THE ISSUE? VALORACIÓN POR ESPECIALISTA - SPECIALIST ASSESSMENT WHAT IS THE NEXT COURSE OF ACTION? VALORACIÓN POR ESPECIALISTA - SPECIALIST ASSESSMENT PATIENT PRESENT AT TIME OF EVENT? YES, PATIENT COMPLICATIONS: DI

    Patient problems: Abdominal Pain; Burn(s); Discomfort; Gastrointestinal Hemorrhage; Inflammation; Swelling/ Edema

    MAUDE report 23652293 ↗
  • MalfunctionNov 25, 2025Leak/Splash; Leak/Splash

    ECN EVENT DESCRIPTION: SITUACIÓN QUE NOS REPORTAN EN SURA DE LAS SONDAS DE RECAMBIO ENFIT, HAN TENIDO MÁS DE 5 PCTES, LOS CUALES HAN PRESENTADO SANGRADO PERIOSTOMAL DESDE EL CAMBIO DE LAS SONDAS, FILTRACIONES, DAÑO DEL ESTOMA Y PIEL PERILESIONAL, ESTO NO HABÍA OCURRIDO ANTES Y SE HA VENIDO EVIDENCIANDO DESDE LAS COLOCACIONES DE SEPTIEMBRE, AGRADEZCO CON LATAM SI PUEDES VALIDAR SI HAN PRESENTADO UNA SITUACIÓN SIMILAR. COMPARTO EVIDENCIAS FOTOGRÁFICAS DE LOS CASOS SITUATION REPORTED TO US AT SURA REGARDING ENFIT REPLACEMENT TUBES: THEY HAVE HAD MORE THAN 5 PATIENTS WHO HAVE PRESENTED PERISTOMAL BLEEDING SINCE THE TUBE. CHANGE, LEAKS, DAMAGE TO THE STOMA AND PERILESIONAL SKIN. THIS HAD NOT OCCURRED BEFORE AND HAS BEEN EVIDENT SINCE THE (B)(6) PLACEMENTS. I WOULD APPRECIATE IT IF LATAM COULD VALIDATE WHETHER THEY HAVE PRESENTED A SIMILAR SITUATION. I SHARE PHOTOGRAPHIC EVIDENCE OF THE CASES WHAT TROUBLESHOOTING STEPS TOOK PLACE? WHAT TROUBLESHOOTING STEPS, IF ANY, RESOLVED THE ISSUE?: VALORACIÓN POR ESPECIALISTA - SPECIALIST ASSESSMENT WHAT IS THE NEXT COURSE OF ACTION?: VALORACIÓN POR ESPECIALISTA - SPECIALIST ASSESSMENT PATIENT PRESENT AT TIME OF EVENT?: YES PATIENT COMPLICATIONS: DI

    Patient problems: Abdominal Pain; Burn(s); Discomfort; Gastrointestinal Hemorrhage; Inflammation; Swelling/ Edema

    MAUDE report 23652291 ↗
  • MalfunctionNov 25, 2025Leak/Splash; Leak/Splash

    ECN EVENT DESCRIPTION: SITUACI6N QUE NOS REPORTAN EN SURA DE LAS SONDAS DE RECAMBIO ENFIT, HAN TENIDO MAS DE 5 PCTES, LAS CUALES HAN PRESENTADO SANGRADO PERIOSTOMAL DESDE EL CAMBIO DE LAS SONDAS, FILTRACIONES, DAI'IO DEL ESTOMA Y PIEL PERILESIONAL, ESTO NO HABIA OCURRIDO ANTES Y SE HA VENIDO EVIDENCIANDO DESDE LAS COLOCACIONES DE SEPTIEMBRE, AGRADEZCO CON LATAM SI PUEDES VALIDAR SI HAN PRESENTADO UNA SITUACI6N SIMILAR. COMPARTO EVIDENCIAS FOTOGRAFICAS DE LAS CASES SITUATION REPORTED TO US AT SURA REGARDING ENFIT REPLACEMENT TUBES: THEY HAVE HAD MORE THAN 5 PATIENTS WHO HAVE PRESENTED PERISTOMAL BLEEDING SINCE THE TUBE CHANGE, LEAKS, DAMAGE TO THE STOMA AND PERILESIONAL SKIN. THIS HAD NOT OCCURRED BEFORE AND HAS BEEN EVIDENT SINCE THE SEPTEMBER PLACEMENTS. I WOULD APPRECIATE IT IF LATAM COULD VALIDATE WHETHER THEY HAVE PRESENTED A SIMILAR SITUATION. I SHARE PHOTOGRAPHIC EVIDENCE OF THE CASES WHAT TROUBLESHOOTING STEPS TOOK PLACE? WHAT TROUBLESHOOTING STEPS, IF ANY, RESOLVED THE ISSUE? VALORACI6N PAR ESPECIALISTA - SPECIALIST ASSESSMENT WHAT IS THE NEXT COURSE OF ACTION? VALORACI6N PAR ESPECIALISTA - SPECIALIST ASSESSMENT PATIENT PRESENT AT TIME OF EVENT? YES PATIENT COMPLICATIONS: D

    Patient problems: Abdominal Pain; Burn(s); Discomfort; Gastrointestinal Hemorrhage; Inflammation; Swelling/ Edema

    MAUDE report 23649588 ↗
  • MalfunctionNov 25, 2025Leak/Splash; Leak/Splash

    ECN EVENT DESCRIPTION: SITUACIÓN QUE NOS REPORTAN EN SURA DE LAS SONDAS DE RECAMBIO ENFIT, HAN TENIDO MÁS DE 5 PCTES, LOS CUALES HAN PRESENTADO SANGRADO PERIOSTOMAL DESDE EL CAMBIO DE LAS SONDAS, FILTRACIONES, DAÑO DEL ESTOMA Y PIEL PERILESIONAL, ESTO NO HABÍA OCURRIDO ANTES Y SE HA VENIDO EVIDENCIANDO DESDE LAS COLOCACIONES DE SEPTIEMBRE, AGRADEZCO CON LATAM SI PUEDES VALIDAR SI HAN PRESENTADO UNA SITUACIÓN SIMILAR. COMPARTO EVIDENCIAS FOTOGRÁFICAS DE LOS CASOS SITUATION REPORTED TO US AT SURA REGARDING ENFIT REPLACEMENT TUBES: THEY HAVE HAD MORE THAN 5 PATIENTS WHO HAVE PRESENTED PERISTOMAL BLEEDING SINCE THE TUBE CHANGE, LEAKS, DAMAGE TO THE STOMA AND PERILESIONAL SKIN. THIS HAD NOT OCCURRED BEFORE AND HAS BEEN EVIDENT SINCE THE SEPTEMBER PLACEMENTS. I WOULD APPRECIATE IT IF LATAM COULD VALIDATE WHETHER THEY HAVE PRESENTED A SIMILAR SITUATION. I SHARE PHOTOGRAPHIC EVIDENCE OF THE CASES WHAT TROUBLESHOOTING STEPS TOOK PLACE? WHAT TROUBLESHOOTING STEPS, IF ANY, RESOLVED THE ISSUE?: VALORACIÓN POR ESPECIALISTA - SPECIALIST ASSESSMENT WHAT IS THE NEXT COURSE OF ACTION?: VALORACIÓN POR ESPECIALISTA - SPECIALIST ASSESSMENT PATIENT PRESENT AT TIME OF EVENT?: YES PATIENT COMPLICATIONS:

    Patient problems: Abdominal Pain; Burn(s); Discomfort; Gastrointestinal Hemorrhage; Inflammation; Swelling/ Edema

    MAUDE report 23649211 ↗
  • MalfunctionNov 25, 2025Leak/Splash; Leak/Splash

    ECN EVENT DESCRIPTION: SITUACIÓN QUE NOS REPORTAN EN SURA DE LAS SONDAS DE RECAMBIO ENFIT, HAN TENIDO MÁS DE 5 PCTES, LOS CUALES HAN PRESENTADO SANGRADO PERIOSTOMAL DESDE EL CAMBIO DE LAS SONDAS, FILTRACIONES, DAÑO DEL ESTOMA Y PIEL PERILESIONAL, ESTO NO HABÍA OCURRIDO ANTES Y SE HA VENIDO EVIDENCIANDO DESDE LAS COLOCACIONES DE SEPTIEMBRE, AGRADEZCO CON LATAM SI PUEDES VALIDAR SI HAN PRESENTADO UNA SITUACIÓN SIMILAR. COMPARTO EVIDENCIAS FOTOGRÁFICAS DE LOS CASOS SITUATION REPORTED TO US AT SURA REGARDING ENFIT REPLACEMENT TUBES: THEY HAVE HAD MORE THAN 5 PATIENTS WHO HAVE PRESENTED PERISTOMAL BLEEDING SINCE THE TUBE CHANGE, LEAKS, DAMAGE TO THE STOMA AND PERILESIONAL SKIN. THIS HAD NOT OCCURRED BEFORE AND HAS BEEN EVIDENT SINCE THE SEPTEMBER PLACEMENTS. I WOULD APPRECIATE IT IF LATAM COULD VALIDATE WHETHER THEY HAVE PRESENTED A SIMILAR SITUATION. I SHARE PHOTOGRAPHIC EVIDENCE OF THE CASES WHAT TROUBLESHOOTING STEPS TOOK PLACE? WHAT TROUBLESHOOTING STEPS, IF ANY, RESOLVED THE ISSUE?: VALORACIÓN POR ESPECIALISTA - SPECIALIST ASSESSMENT WHAT IS THE NEXT COURSE OF ACTION?: VALORACIÓN POR ESPECIALISTA - SPECIALIST ASSESSMENT PATIENT PRESENT AT TIME OF EVENT?: YES PATIENT COMPLICATIONS:

    Patient problems: Abdominal Pain; Burn(s); Discomfort; Gastrointestinal Hemorrhage; Inflammation; Swelling/ Edema

    MAUDE report 23647770 ↗

FDA clearance and classification

The FDA regulates product code PIF (gastrointestinal tubes with enteral specific connectors) under 21 CFR 876.5980 as a Class II device. 21 CFR 876.5980 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the EndoVive Gastrostomy Tube, Right Angle, with ENFit Connector M00510010

Has the EndoVive Gastrostomy Tube, Right Angle, with ENFit Connector M00510010 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00813939023770), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the EndoVive Gastrostomy Tube, Right Angle, with ENFit Connector M00510010?

The FDA's MAUDE database lists 6 adverse event reports for this device, received on Nov 25, 2025. In those reports, the most common specific device problem is “Leak/Splash” (6); the most common patient problems are “Abdominal Pain” (6), “Burn(s)” (6) and “Discomfort” (6). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the EndoVive Gastrostomy Tube, Right Angle, with ENFit Connector M00510010?

None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the EndoVive Gastrostomy Tube, Right Angle, with ENFit Connector M00510010 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the EndoVive Gastrostomy Tube, Right Angle, with ENFit Connector M00510010, and is it still sold?

The FDA's device database lists Xeridiem Medical Devices, Inc. as the company responsible for it (brand EndoVive Gastrostomy Tube, Right Angle, with ENFit Connector, model M00510010). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Gastrointestinal Tubes With Enteral Specific Connectors devices

All Gastrointestinal Tubes With Enteral Specific Connectors devices →