Momentum UU052-30-0200Recalls & reported problems
Ulrich Medical Usa, Inc. · Screwdriver, SR20, 1/4"-square · Class 2
All models of Momentum (19)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00814386025041), received between Jun 2, 2026 and Jun 2, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 2, 2026Break
DURING A CASE, THE SURGEON BROKE 5 DRIVERS IN THE ATTEMPT TO INSTALL AND THEN REMOVE A SCREW. THE SURGEON DID NOT TAP LINE TO LINE PRIOR TO IMPLANTING THE SCREW. AFTER BREAKING ALL DRIVERS, THE SURGEON USED ANOTHER MANUFACTURER'S INSTRUMENT TO REALIGN THE TRAJECTORY OF A SCREW AND BEGAN HAMMERING ON THE INSTRUMENT WHICH CAUSED THE SCREW TO BREAK. ALL PIECES OF THE SCREW WERE RETRIEVED BUT WERE NOT RETURNED TO UUSA. THE PATIENT WAS NOT AFFECTED. NO FURTHER INFORMATION WAS AVAILABLE FOR THE MANUFACTURER OF THE SCREW USED. THE AVAILABLE INFORMATION DID NOT ALLOW US TO DETERMINE WHETHER A SERIOUS INJURY COULD RESULT FROM THE DRIVER BREAKAGES, A MEDICAL DEVICE REPORT WAS SUBMITTED FOR EACH SCREWDRIVER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25362693 ↗
Other Thoracolumbosacral Pedicle Screw System devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page