Skip to content

Cortium Universal OCT Spinal Fixation System UU071-10-3514Recalls & reported problems

Ulrich Medical Usa, Inc. · Screw, Ø 3.5mm, length 14mm, high angulation · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00814386026352), received between Oct 24, 2025 and Aug 14, 2026 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Detachment of Device or Device Component2
  • Mechanical Problem2

Patient problems

  • Failure of Implant2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 14, 2026Mechanical Problem; Detachment of Device or Device Component

    Tulip disassociated from the polyaxial screw from the polyaxial screw head while the surgeon was implanting it. The screw was removed and replaced without incident.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50255633 ↗
  • InjuryAug 14, 2026Mechanical Problem; Detachment of Device or Device Component

    Tulip disassociated from the polyaxial screw from the polyaxial screw head while the surgeon was implanting it. The screw was removed and replaced without incident.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50255328 ↗
  • MalfunctionJan 8, 2026Break

    DURING SURGERY THE TULIP HEAD DISSOCIATED FROM THE SCREW TAIL. THE SURGEON WAS WORKING AT THE CERVICAL 4 LEVEL WHEN THEY NOTICED THE SCREWDRIVER WAS STICKING SLIGHTLY AND WHEN IT RELEASED THE TULIP HEAD BECAME DISSOCIATED WITH THE SCREWTAIL. THE SCREW WAS REMOVED AND CAME OUT IN TWO PIECES. THE SCREW WAS REPLACED. ALL PARTS WERE RETRIEVED AND THERE WAS NO HARM TO THE PATIENT.

    Patient problems: Failure of Implant

    MAUDE report 24014861 ↗
  • MalfunctionOct 24, 2025Break; Break

    DURING SURGERY THE TULIP HEAD DISSOCIATED FROM THE SCREW TAIL. THE SURGEON DOES NOT TAP THE PREPARATION HOLE PRIOR TO DRIVING THE SCREWS NOR USES THE TISSUE SLEEVE. NOT USING THE TISSUE SLEEVE CAUSED EXCESSIVE PRESSURE TO THE TULIP HEAD WHEN PRESSURE WAS APPLIED TO THE SCREWDRIVER. THIS CAUSED THE TULIP HEAD TO DISSOCIATE FROM THE SCREW TAIL.

    Patient problems: Failure of Implant

    MAUDE report 23382959 ↗

Other Posterior Cervical Screw System devices

All Posterior Cervical Screw System devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page