Cortium Universal OCT Spinal Fixation System UU071-20-4525Recalls & reported problems

Ulrich Medical Usa, Inc. · Thoracic Screw, Ø 4.5mm, length 25mm, high angulation, dual lead · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00814386026819), received between Dec 11, 2024 and Dec 12, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Damaged Thread” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Damaged Thread2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 12, 2024Damaged Thread; Damaged Thread

    THE SURGEON PUT ON A CONNECTOR THAT WAS LATERAL AND SHORT. THE SURGEON THEN NOTICED THE T2 SCREW WAS A LITTLE LATERAL AND NOT ALIGNED PROPERLY. HE PROCEEDED AND PUT THE SET SCREW INTO TO THE CONNECTOR AND TIGHTENED THE FIRST SET SCREW OF THE CONNECTOR TO THE ROD AND HAD NO ISSUES. HE THEN WENT TO TIGHTEN THE SET SCREW TO THE TULIP AND NOTICED IT WAS NOT GOING IN RIGHT. UPON FINAL TIGHTENING HE HEARD A POP, AND THE TULIP SPLAYED AND ONE SIDE THE SET SCREW POPPED OUT OF A THREAD. THEY BACKED THE SCREW OUT AGAIN, THEN REMOVED THE PEDICLE SCREW AND SWITCHED IT OUT WITH A NEW ONE WITHOUT REALIGNING. THE SURGEON LOOSENED THE LATERAL CONNECTOR TO RELIEVE THE FORCES ON THE ONE SIDE OF THE SECOND PEDICLE SCREW WHICH REPLACED THE FIRST SCREW, AND IT SPLAYED AGAIN. THEY THEN REMOVED THE SECOND SCREW THAT HAD ALSO SPLAYED DUE TO THE MISALIGNMENT AND REPLACED IT WITH A THIRD SCREW, BUT THIS TIME ADJUSTED IT TO LINE UP BETTER WITH THE ROD AND CONNECTOR, THEN IT LOCKED DOWN. THE SCREW DID NOT SPLAY THIS TIME AS THE ALIGNMENT WAS CORRECT. THE PATIENT WAS NOT AFFECTED, AND THE CASE WENT FINE OTHERWISE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20912746 ↗
  • InjuryDec 11, 2024Damaged Thread; Damaged Thread

    THE SURGEON PUT ON A CONNECTOR THAT WAS LATERAL AND SHORT. THE SURGEON THEN NOTICED THE T2 SCREW WAS A LITTLE LATERAL AND NOT ALIGNED PROPERLY. HE PROCEEDED AND PUT THE SET SCREW INTO TO THE CONNECTOR AND TIGHTENED THE FIRST SET SCREW OF THE CONNECTOR TO THE ROD AND HAD NO ISSUES. HE THEN WENT TO TIGHTEN THE SET SCREW TO THE TULIP AND NOTICED IT WAS NOT GOING IN RIGHT. UPON FINAL TIGHTENING HE HEARD A POP, AND THE TULIP SPLAYED AND ONE SIDE THE SET SCREW POPPED OUT OF A THREAD. THEY BACKED THE SCREW OUT AGAIN, THEN REMOVED THE PEDICLE SCREW AND SWITCHED IT OUT WITH A NEW ONE WITHOUT REALIGNING. THE SURGEON LOOSENED THE LATERAL CONNECTOR TO RELIEVE THE FORCES ON THE ONE SIDE OF THE SECOND PEDICLE SCREW WHICH REPLACED THE FIRST SCREW, AND IT SPLAYED AGAIN. THEY THEN REMOVED THE SECOND SCREW THAT HAD ALSO SPLAYED DUE TO THE MISALIGNMENT AND REPLACED IT WITH A THIRD SCREW, BUT THIS TIME ADJUSTED IT TO LINE UP BETTER WITH THE ROD AND CONNECTOR, THEN IT LOCKED DOWN. THE SCREW DID NOT SPLAY THIS TIME AS THE ALIGNMENT WAS CORRECT. THE PATIENT WAS NOT AFFECTED, AND THE CASE WENT FINE OTHERWISE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20909008 ↗