Smart Coil 400SMTSTD0715Recalls & reported problems
Penumbra, Inc. · SmtCoil-Std, 7x15 · Class 2 · Implant
All models of Smart Coil (98)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00814548014487), received between Nov 21, 2016 and Feb 25, 2022 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Difficult or Delayed Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Difficult or Delayed Separation2
- Premature Separation2
- Device Handling Problem1
- Failure to Advance1
- Mechanical Jam1
- Physical Resistance1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 25, 2022Difficult or Delayed Separation; Premature Separation; Difficult or Delayed Separation; Premature Separation
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT A BASILAR ANEURYSM USING PENUMBRA SMART COILS (SMART COILS), PENUMBRA SMART COIL HANDLES (HANDLES), AND A NON-PENUMBRA MICROCATHETER. DURING THE PROCEDURE, THE PHYSICIAN ADVANCED A SMART COIL THROUGH THE MICROCATHETER TO THE TARGET VESSEL AND ATTEMPTED TO DETACH IT USING THE HANDLE; HOWEVER, THE SMART COIL FAILED TO DETACH. THEREFORE, THE PHYSICIAN ATTEMPTED TO DETACH THE SMART COIL WITH A NEW SECOND HANDLE; HOWEVER, THE SMART COIL FAILED TO DETACH. THE PHYSICIAN ALSO ATTEMPTED TO MANUALLY DETACH THE SMART COIL BUT WAS UNSUCCESSFUL. THEREFORE, THE SMART COIL WAS REMOVED. NEXT, THE PHYSICIAN SUCCESSFULLY IMPLANTED TWO NEW SMART COILS USING THE SECOND HANDLE AND THE MICROCATHETER. SUBSEQUENTLY, THE PHYSICIAN ADVANCED A NEW SMART COIL THROUGH THE MICROCATHETER TO THE TARGET VESSEL AND ATTEMPTED TO DETACH IT USING THE SECOND HANDLE; HOWEVER, THE SMART COIL FAILED TO DETACH. THEREFORE, THE PHYSICIAN ATTEMPTED TO DETACH THE SMART COIL WITH A NEW THIRD HANDLE; HOWEVER, THE SMART COIL FAILED TO DETACH. THE PHYSICIAN ALSO ATTEMPTED TO MANUALLY DETACH THE SMART COIL BUT WAS UNSUCCESSFUL. THE PHYSICIAN THEN REMOVED THE PUSHER A
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13604355 ↗MalfunctionFeb 25, 2022Difficult or Delayed Separation; Premature Separation; Difficult or Delayed Separation; Premature Separation
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT A BASILAR ANEURYSM USING PENUMBRA SMART COILS (SMART COILS), PENUMBRA SMART COIL HANDLES (HANDLES), AND A NON-PENUMBRA MICROCATHETER. DURING THE PROCEDURE, THE PHYSICIAN ADVANCED A SMART COIL THROUGH THE MICROCATHETER TO THE TARGET VESSEL AND ATTEMPTED TO DETACH IT USING THE HANDLE; HOWEVER, THE SMART COIL FAILED TO DETACH. THEREFORE, THE PHYSICIAN ATTEMPTED TO DETACH THE SMART COIL WITH A NEW SECOND HANDLE; HOWEVER, THE SMART COIL FAILED TO DETACH. THE PHYSICIAN ALSO ATTEMPTED TO MANUALLY DETACH THE SMART COIL BUT WAS UNSUCCESSFUL. THEREFORE, THE SMART COIL WAS REMOVED. NEXT, THE PHYSICIAN SUCCESSFULLY IMPLANTED TWO NEW SMART COILS USING THE SECOND HANDLE AND THE MICROCATHETER. SUBSEQUENTLY, THE PHYSICIAN ADVANCED A NEW SMART COIL THROUGH THE MICROCATHETER TO THE TARGET VESSEL AND ATTEMPTED TO DETACH IT USING THE SECOND HANDLE; HOWEVER, THE SMART COIL FAILED TO DETACH. THEREFORE, THE PHYSICIAN ATTEMPTED TO DETACH THE SMART COIL WITH A NEW THIRD HANDLE; HOWEVER, THE SMART COIL FAILED TO DETACH. THE PHYSICIAN ALSO ATTEMPTED TO MANUALLY DETACH THE SMART COIL BUT WAS UNSUCCESSFUL. THE PHYSICIAN THEN REMOVED THE PUSHER A
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13604344 ↗MalfunctionJun 20, 2017Device Handling Problem; Device Handling Problem
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION TO TREAT A SUBARACHNOID HEMORRHAGE USING A PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, THE SCRUB TECHNOLOGIST INADVERTENTLY DROPPED THE END OF THE COIL PUSHER ASSEMBLY ONTO THE FLOOR AND THE COIL BECAME UNSTERILE. THEREFORE, THE SMART COIL WAS REMOVED AND THE PROCEDURE WAS COMPLETED USING A TOTAL OF (B)(4) SMART COILS. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6654956 ↗MalfunctionNov 21, 2016Failure to Advance; Physical Resistance; Mechanical Jam; Failure to Advance; Physical Resistance; Mechanical Jam
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION TO TREAT A CEREBRAL ANEURYSM USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, THE PHYSICIAN ADVANCED A SMART COIL IN THE TARGET VESSEL USING ANOTHER MANUFACTURER¿S MICROCATHETER AND REALIZED THAT THE SIZE OF THE COIL DID NOT MATCH THE SIZE OF THE ANEURYSM. THEREFORE, THE SMART COIL WAS RETRACTED AND RE-SHEATHED. THE PHYSICIAN THEN ATTEMPTED TO PLACE ANOTHER SMART COIL INTO THE TARGET VESSEL; HOWEVER, THE PHYSICIAN REALIZED THAT THE COIL SIZE DID NOT MATCH THE SIZE OF THE ANEURYSM. THEREFORE, THE SMART COIL WAS RETRACTED AND RE-SHEATHED. IT WAS REPORTED THAT THE PHYSICIAN SOAKED THE SMART COILS IN HEPARINIZED SALINE AFTER THEY WERE RETRACTED. THEREAFTER, THE PHYSICIAN DEPLOYED AND DETACHED A NEW SMART COIL AND ANOTHER MANUFACTURER¿S COIL IN THE ANEURYSM USING THE SAME MICROCATHETER. WHILE ATTEMPTING TO USE ONE OF THE PREVIOUS SMART COILS THAT WERE REMOVED FROM THE PATIENT, THE PHYSICIAN EXPERIENCED RESISTANCE AND THE SMART COIL WAS UNABLE TO ADVANCE OUT OF ITS INTRODUCER SHEATH. BOTH SMART COILS WERE THEN DISCARDED BEFORE BEING RETRIEVED AGAIN. HOWEVER, AFTER RETRIEVAL FROM THE GARBAGE, ONE OF THE SMART COILS WAS DETACHED
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6117041 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page