Smart Coil 400SMTSTD0410Recalls & reported problems
Penumbra, Inc. · SmtCoil-Std, 4x10 · Class 2
All models of Smart Coil (98)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00814548014609), received between Mar 13, 2017 and Mar 13, 2017 — classified as 2 injuries.
The most frequently reported specific device problem is “Material Protrusion/Extrusion” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Protrusion/Extrusion2
- Unintended Movement2
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMar 13, 2017Material Protrusion/Extrusion; Unintended Movement; Material Protrusion/Extrusion; Unintended Movement
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT A BASILAR TIP ANEURYSM USING PENUMBRA SMART COILS (SMART COIL). DURING THE PROCEDURE, AFTER THE THIRD SMART COIL WAS DEPLOYED AND DETACHED, THE ANGIOGRAPHY SHOWED A PROTRUSION OF A PREVIOUSLY PLACED COIL. THE PHYSICIAN CONTINUED THE PROCEDURE BY PLACING SIX ADDITIONAL SMART COILS. TOWARDS THE END OF THE PROCEDURE, CLOT FORMATION WAS IDENTIFIED ON THE DISPLACED LOOP OF COIL AND AN INTEGRILIN DRIP WAS STARTED. THE PHYSICIAN THEN DECIDED TO PLACE A STENT IN ORDER TO TACK THE PROTRUDING COIL AGAINST THE WALL OF THE VESSEL. THE PATIENT WAS DISCHARGED HOME ON (B)(6) 2017. THE REPORTED CLOT FORMATION WAS ADJUDICATED TO HAVE NO RELATIONSHIP TO THE PENUMBRA DEVICES. HOWEVER, THE REPORTED COIL PROTRUSION WAS POSSIBLY RELATED TO THE ANEURYSM, PROCEDURE AND PENUMBRA SMART COILS. IT IS UNKNOWN WHICH OF THE THREE PREVIOUSLY PLACED SMART COILS PROTRUDED OUT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6401563 ↗InjuryMar 13, 2017Material Protrusion/Extrusion; Unintended Movement; Material Protrusion/Extrusion; Unintended Movement
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT A BASILAR TIP ANEURYSM USING PENUMBRA SMART COILS (SMART COIL). DURING THE PROCEDURE, AFTER THE THIRD SMART COIL WAS DEPLOYED AND DETACHED, THE ANGIOGRAPHY SHOWED A PROTRUSION OF A PREVIOUSLY PLACED COIL. THE PHYSICIAN CONTINUED THE PROCEDURE BY PLACING SIX ADDITIONAL SMART COILS. TOWARDS THE END OF THE PROCEDURE, CLOT FORMATION WAS IDENTIFIED ON THE DISPLACED LOOP OF COIL AND AN INTEGRILIN DRIP WAS STARTED. THE PHYSICIAN THEN DECIDED TO PLACE A STENT IN ORDER TO TACK THE PROTRUDING COIL AGAINST THE WALL OF THE VESSEL. THE PATIENT WAS DISCHARGED HOME ON (B)(6) 2017. THE REPORTED CLOT FORMATION WAS ADJUDICATED TO HAVE NO RELATIONSHIP TO THE PENUMBRA DEVICES. HOWEVER, THE REPORTED COIL PROTRUSION WAS POSSIBLY RELATED TO THE ANEURYSM, PROCEDURE AND PENUMBRA SMART COILS. IT IS UNKNOWN WHICH OF THE THREE PREVIOUSLY PLACED SMART COILS PROTRUDED OUT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6401555 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page