Smart Coil 400SMTXSFT0104Recalls & reported problems
Penumbra, Inc. · SmtCoil-XSft, 1x4 · Class 2 · Implant
All models of Smart Coil (98)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Penumbra Smart Coil is a small embolization coil used to block blood vessels in the brain during neurovascular procedures. The 29 reports mostly describe the coil failing to advance through its delivery catheter, sometimes with the pusher wire kinking, the coil detaching early inside the catheter, or the coil material unraveling. Reports describe no deaths and no injuries, and most note no effect on the patient. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 29
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 00814548015507), received between Dec 2, 2016 and Nov 22, 2022 — classified as 29 malfunctions.
The most frequently reported specific device problem is “Failure to Advance” (25 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Failure to Advance25
- Physical Resistance5
- Mechanical Jam3
- Physical Resistance/Sticking3
- Device Operates Differently Than Expected2
- Difficult to Advance1
- Difficult to Insert1
- Material Deformation1
Patient problems
- No Consequences Or Impact To Patient24
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 22, 2022Premature Separation; Premature Separation
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE BASILAR TIP USING PENUMBRA SMART COILS (SMART COIL), A PENUMBRA SMART COIL DETACHMENT HANDLE (HANDLE), AND A NON-PENUMBRA MICROCATHETER. IT SHOULD BE NOTED THAT THE PATIENT'S ANATOMY WAS TORTUOUS. DURING THE PROCEDURE, THE PHYSICIAN ADVANCED THE SMART COIL THROUGH THE MICROCATHETER, HOWEVER, THE SMART COIL UNINTENTIONALLY DETACHED WITHIN THE MICROCATHETER. THEREFORE, THE PHYSICIAN USED A SNARE DEVICE TO REMOVE THE DETACHED SMART COIL FROM THE MICROCATHETER. IT WAS REPORTED THAT A SECOND SMART COIL ALSO UNINTENTIONALLY DETACHED WITHIN THE MICROCATHETER AND WAS REMOVED WITH A SNARE DEVICE. THE PROCEDURE WAS COMPLETED WITH ADDITIONAL SMART COILS AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15845188 ↗MalfunctionSep 15, 2021Failure to Advance; Material Deformation; Physical Resistance/Sticking; Failure to Advance; Material Deformation; Physical Resistance/Sticking
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE RIGHT ANTERIOR CHOROIDAL ARTERY USING PENUMBRA SMART COILS (SMART COIL) AND A NON-PENUMBRA MICROCATHETER. DURING THE PROCEDURE, THE LAST SMART COIL WOULD NOT ADVANCE AT ALL WITHIN ITS INTRODUCER. IT WAS REPORTED THAT RESISTANCE WAS ENCOUNTERED AND SUBSEQUENTLY, THE PUSHER ASSEMBLY OF THE SMART COIL KINKED; THEREFORE, THE SMART COIL WAS REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER SMART COIL AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12478576 ↗MalfunctionJun 3, 2021Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE INTERNAL CAROTID ARTERY (ICA) USING PENUMBRA SMART COILS (SMART COIL). DURING THE PROCEDURE, A SMART COIL WOULD NOT ADVANCE AT ALL PAST ITS INITIAL POSITION WITHIN ITS INTRODUCER SHEATH. THEREFORE, THE SMART COIL WAS REMOVED. SUBSEQUENTLY, THE SAME ISSUE OCCURRED WITH FOUR OTHER SMART COILS; THEREFORE, THE FOUR SMART COILS WERE ALSO REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER SMART COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11927512 ↗MalfunctionJun 3, 2021Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE INTERNAL CAROTID ARTERY (ICA) USING PENUMBRA SMART COILS (SMART COIL). DURING THE PROCEDURE, A SMART COIL WOULD NOT ADVANCE AT ALL PAST ITS INITIAL POSITION WITHIN ITS INTRODUCER SHEATH. THEREFORE, THE SMART COIL WAS REMOVED. SUBSEQUENTLY, THE SAME ISSUE OCCURRED WITH FOUR OTHER SMART COILS; THEREFORE, THE FOUR SMART COILS WERE ALSO REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER SMART COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11927499 ↗MalfunctionJun 3, 2021Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE INTERNAL CAROTID ARTERY (ICA) USING PENUMBRA SMART COILS (SMART COIL). DURING THE PROCEDURE, A SMART COIL WOULD NOT ADVANCE AT ALL PAST ITS INITIAL POSITION WITHIN ITS INTRODUCER SHEATH. THEREFORE, THE SMART COIL WAS REMOVED. SUBSEQUENTLY, THE SAME ISSUE OCCURRED WITH FOUR OTHER SMART COILS; THEREFORE, THE FOUR SMART COILS WERE ALSO REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER SMART COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11927495 ↗MalfunctionDec 18, 2020Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE POSTERIOR COMMUNICATION ARTERY (PCOM) USING A PENUMBRA SMART COIL (SMART COIL). DURING THE PROCEDURE, A SMART COIL WOULD NOT ADVANCE AT ALL WITHIN ITS INTRODUCER SHEATH; THEREFORE, THE SMART COIL WAS REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER SMART COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11041089 ↗MalfunctionJul 3, 2020Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING A PENUMBRA SMART COIL (SMART COIL). DURING THE PROCEDURE, THE SMART COIL WOULD NOT ADVANCE WITHIN ITS INTRODUCER SHEATH; THEREFORE, IT WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW SMART COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10228421 ↗MalfunctionApr 3, 2020Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE EXTERNAL CAROTID ARTERY USING PENUMBRA SMART COILS (SMART COILS) AND A NON-PENUMBRA MICROCATHETER. DURING THE PROCEDURE, THE SMART COIL WAS UNABLE TO ADVANCE PAST ITS INITIAL POSITION WITHIN ITS INTRODUCER SHEATH. THEREFORE, THE SMART COIL WAS REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER SMART COIL AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9922424 ↗MalfunctionOct 25, 2019Failure to Advance; Physical Resistance/Sticking; Failure to Advance; Physical Resistance/Sticking
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE ANTERIOR COMMUNICATING ARTERY (ACOM) USING PENUMBRA SMART COILS (SMART COIL). DURING THE PROCEDURE, THE PHYSICIAN EXPERIENCED RESISTANCE AND THE SMART COIL WOULD NOT ADVANCE THROUGH A NON-PENUMBRA MICROCATHETER; THEREFORE, IT WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW SMART COIL AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9239243 ↗MalfunctionSep 5, 2019Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE INTERNAL CAROTID ARTERY (ICA) USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, A SMART COIL WOULD NOT ADVANCE WITHIN ITS INTRODUCER SHEATH; THEREFORE, IT WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW SMART COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8965742 ↗MalfunctionJan 10, 2019Failure to Advance; Physical Resistance/Sticking; Failure to Advance; Physical Resistance/Sticking
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE INTERNAL CAROTID ARTERY (ICA) USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, THE PHYSICIAN SUCCESSFULLY PLACED COILS INTO THE TARGET VESSEL USING A NON-PENUMBRA MICROCATHETER. THE PHYSICIAN THEN ATTEMPTED TO ADVANCE A SMART COIL, HOWEVER, THE SMART COIL COULD NOT ADVANCE FROM ITS STARTING POSITION WITHIN THE INTRODUCER SHEATH. THE SMART COIL WAS THEN REMOVED AND WAS NOT USED IN THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING ADDITIONAL COILS AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8238311 ↗MalfunctionMay 29, 2018Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT A DURAL ARTERIOVENOUS FISTULA USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, THE PHYSICIAN DEPLOYED AND DETACHED 18 SMART COILS IN THE TARGET VESSEL USING A NON-PENUMBRA MICROCATHETER. WHILE ATTEMPTING TO ADVANCE A NEW SMART COIL INTO THE MICROCATHETER, THE SMART COIL WOULD NOT ADVANCE PAST THE FRICTION LOCK; THEREFORE, IT WAS REMOVED. IT WAS REPORTED THAT THE SAME ISSUE OCCURRED WITH ANOTHER SMART COIL. THE PROCEDURE WAS COMPLETED USING OTHER COILS AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7550040 ↗MalfunctionMay 29, 2018Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT A DURAL ARTERIOVENOUS FISTULA USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, THE PHYSICIAN DEPLOYED AND DETACHED 18 SMART COILS IN THE TARGET VESSEL USING A NON-PENUMBRA MICROCATHETER. WHILE ATTEMPTING TO ADVANCE A NEW SMART COIL INTO THE MICROCATHETER, THE SMART COIL WOULD NOT ADVANCE PAST THE FRICTION LOCK; THEREFORE, IT WAS REMOVED. IT WAS REPORTED THAT THE SAME ISSUE OCCURRED WITH ANOTHER SMART COIL. THE PROCEDURE WAS COMPLETED USING OTHER COILS AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7550036 ↗MalfunctionJan 10, 2018Failure to Advance; Mechanical Jam; Failure to Advance; Mechanical Jam
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT A RUPTURED ANEURYSM USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, THE PHYSICIAN PLACED SEVERAL COILS INTO THE PATIENT USING A NON-PENUMBRA MICROCATHETER. WHILE ATTEMPTING TO ADVANCE A NEW SMART COIL THROUGH THE MICROCATHETER, THE PHYSICIAN WAS ABLE TO ADVANCE THE COIL INTO THE MICROCATHETER BUT THEN THE COIL STOPPED ADVANCING AFTER SEVERAL CENTIMETERS AND WOULD NOT ADVANCE ANY FURTHER. THEREFORE, THE SMART COIL WAS REMOVED AND THE PROCEDURE WAS COMPLETED USING ADDITIONAL COILS. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7179340 ↗MalfunctionJan 9, 2018Difficult to Advance; Difficult to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, THE PHYSICIAN WAS UNABLE TO ADVANCE THE SMART COIL OUT OF ITS INTRODUCER SHEATH AND INTO THE NON-PENUMBRA MICROCATHETER, AND THEREFORE THE SMART COIL WAS REMOVED. THE PROCEDURE WAS THEN COMPLETED USING ADDITIONAL COILS AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7176580 ↗MalfunctionDec 27, 2017Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE POSTERIOR COMMUNICATING ARTERY (PCOM) USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, THE PHYSICIAN WAS UNABLE TO ADVANCE A SMART COIL OUT OF ITS INTRODUCER SHEATH AND INTO THE HUB OF A NON-PENUMBRA MICROCATHETER; THEREFORE, IT WAS REMOVED. THE PROCEDURE WAS COMPLETED USING ADDITIONAL SMART COILS AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7148056 ↗MalfunctionNov 17, 2017Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT A POSTERIOR COMMUNICATING ARTERY (PCOM) ANEURYSM USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, THE PHYSICIAN SUCCESSFULLY PLACED MULTIPLE SMART COILS USING A NON-PENUMBRA MICROCATHETER. THE PHYSICIAN THEN WAS UNABLE TO ADVANCE A SMART COIL OUT OF ITS INTRODUCER SHEATH AND INTO THE MICROCATHETER, AND THEREFORE THE SMART COIL WAS REMOVED. THE HOSPITAL STAFF ATTEMPTED TO ADVANCE THE SMART COIL OUT OF ITS INTRODUCER SHEATH ON THE BACK TABLE, HOWEVER IT WAS STILL UNABLE TO BE ADVANCED. THE PROCEDURE WAS THEN COMPLETED USING THE SAME MICROCATHETER AND A NEW SMART COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7043180 ↗MalfunctionNov 13, 2017Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT A SUBARACHNOID HEMORRHAGE (SAH) USING PENUMBRA SMART COILS (SMART COILS). AT DIFFERENT TIMES DURING THE PROCEDURE, THE PHYSICIAN WAS UNABLE TO ADVANCE THREE SMART COILS INTO THE HUB OF A NON-PENUMBRA MICROCATHETER. THEREFORE, THE SMART COILS WERE REMOVED AND THE PROCEDURE WAS COMPLETED USING ADDITIONAL SMART COILS AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7025892 ↗MalfunctionSep 28, 2017Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, A SMART COIL WAS UNABLE TO ADVANCE THROUGH THE HUB OF A NON-PENUMBRA MICROCATHETER; THEREFORE, THE SMART COIL WAS REMOVED. THE PROCEDURE WAS COMPLETED USING NEW SMART COILS AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6900747 ↗MalfunctionSep 20, 2017Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT A VEIN OF GALEN MALFORMATION USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, THE SMART COIL WOULD NOT ADVANCE OUT OF ITS INTRODUCER SHEATH AND INTO THE NON-PENUMBRA MICROCATHETER; THEREFORE, THE SMART COIL WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW SMART COIL AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6880977 ↗MalfunctionJul 13, 2017Failure to Advance; Physical Resistance; Failure to Advance; Physical Resistance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, WHILE ATTEMPTING TO ADVANCE FOUR SMART COILS OUT OF THEIR INTRODUCER SHEATHS AND INTO THE NON-PENUMBRA MICROCATHETER, THE PHYSICIAN ENCOUNTERED RESISTANCE AND WAS UNABLE TO ADVANCE THE COILS OUT. THEREFORE, THE INTRODUCER SHEATHS CONTAINING THE COILS WERE REMOVED. IT WAS ALSO REPORTED THAT THE RUBY COILS COULD NOT BE ADVANCED OUT OF THEIR INTRODUCER SHEATHS WHILE ON THE BACK TABLE AND RESISTANCE WAS ENCOUNTERED WITH AT LEAST ONE OF THE COILS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING NEW SMART COILS AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6712821 ↗MalfunctionJul 13, 2017Failure to Advance; Physical Resistance; Failure to Advance; Physical Resistance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, WHILE ATTEMPTING TO ADVANCE FOUR SMART COILS OUT OF THEIR INTRODUCER SHEATHS AND INTO THE NON-PENUMBRA MICROCATHETER, THE PHYSICIAN ENCOUNTERED RESISTANCE AND WAS UNABLE TO ADVANCE THE COILS OUT. THEREFORE, THE INTRODUCER SHEATHS CONTAINING THE COILS WERE REMOVED. IT WAS ALSO REPORTED THAT THE RUBY COILS COULD NOT BE ADVANCED OUT OF THEIR INTRODUCER SHEATHS WHILE ON THE BACK TABLE AND RESISTANCE WAS ENCOUNTERED WITH AT LEAST ONE OF THE COILS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING NEW SMART COILS AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6712786 ↗MalfunctionJul 6, 2017Failure to Advance; Mechanical Jam; Failure to Advance; Mechanical Jam
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE EXTERNAL CAROTID ARTERY USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, THE PHYSICIAN DEPLOYED AND DETACHED THIRTEEN SMART COILS INTO THE TARGET VESSEL USING A NON-PENUMBRA MICROCATHETER. WHILE THE PHYSICIAN WAS ATTEMPTING TO ADVANCE A NEW SMART COIL THROUGH THE MICROCATHETER, THE COIL BECAME STUCK INSIDE THE HUB OF THE MICROCATHETER AND WAS UNABLE TO ADVANCE FURTHER. THE PHYSICIAN REMOVED THE SMART COIL AND OPENED A NEW SMART COIL; HOWEVER, THE SAME ISSUE OCCURRED WITH THIS SMART COIL. THEREFORE, THE SMART COIL WAS REMOVED AND THE PROCEDURE WAS COMPLETED USING TWO NEW SMART COILS AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6690155 ↗MalfunctionJun 6, 2017Failure to Advance; Physical Resistance; Failure to Advance; Physical Resistance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, WHILE ATTEMPTING TO ADVANCE THREE SMART COILS THROUGH A NON-PENUMBRA MICROCATHETER, THE PHYSICIAN EXPERIENCED RESISTANCE AND THE SMART COILS WERE UNABLE TO ADVANCE OUT OF THEIR INTRODUCER SHEATHS AND INTO THE HUB OF THE NON-PENUMBRA MICROCATHETER. THEREFORE, THE SMART COILS WERE REMOVED AND THE PROCEDURE WAS COMPLETED USING THE SAME MICROCATHETER AND NEW LARGER SIZED SMART COILS. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6614827 ↗MalfunctionApr 20, 2017Failure to Advance; Device Operates Differently Than Expected; Failure to Advance; Device Operates Differently Than Expected
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, A SMART COIL WAS RE-SHEATHED AFTER BEING ADVANCED INTO THE MICROCATHETER. LATER IN THE PROCEDURE, WHILE ATTEMPTING TO RE-ADVANCE THE SMART COIL INTO THE MICROCATHETER, THE PHYSICIAN WAS UNABLE TO ADVANCE THE COIL OUT OF ITS INTRODUCER SHEATH. THEREFORE, THE SMART COIL WAS REMOVED AND THE PROCEDURE WAS COMPLETED USING ADDITIONAL SMART COILS. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6507735 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page