Smart Coil 400SMTHXSFT1H01Recalls & reported problems
Penumbra, Inc. · SmtCoil-HelXSft, 1.5x1 · Class 2 · Implant
All models of Smart Coil (98)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00814548015637), received between Nov 25, 2019 and Mar 11, 2021 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Failure to Advance” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Failure to Advance6
Patient problems
- No Consequences Or Impact To Patient5
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 11, 2021Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE LEFT VERTEBRAL ARTERY USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, A SMART COIL WOULD NOT ADVANCE PAST ITS INITIAL POSITION WITHIN ITS INTRODUCER SHEATH. THEREFORE, THE SMART COIL WAS REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER SMART COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11464389 ↗MalfunctionAug 1, 2020Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE MIDDLE CEREBRAL ARTERY. DURING THE PROCEDURE, THE PATIENT¿S VESSEL WAS PERFORATED USING A NON-PENUMBRA GUIDEWIRE. THE PHYSICIAN ATTEMPTED TO USE PENUMBRA SMART COILS (SMART COILS) TO STOP THE BLEEDING; HOWEVER, SIX SMART COILS WOULD NOT ADVANCE PAST THEIR INITIAL POSITIONS WITHIN THE INTRODUCER SHEATHS. THEREFORE, THE SMART COILS WERE NOT USED IN THE PROCEDURE. THE NEXT THREE SMART COILS WERE IMPLANTED IN THE TARGET LOCATION USING THE PENUMBRA SMART COIL DETACHMENT HANDLE (HANDLE). WHILE ATTEMPTING TO IMPLANT THE NEXT SMART COIL USING THE HANDLE, IT WOULD NOT DETACH. THE PHYSICIAN ATTEMPTED TO MANUALLY DETACH THE SMART COIL WITHOUT SUCCESS; THEREFORE, THIS SMART COIL WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED USING FIVE SMART COILS AND THE SAME HANDLE. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10354066 ↗MalfunctionAug 1, 2020Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE MIDDLE CEREBRAL ARTERY. DURING THE PROCEDURE, THE PATIENT¿S VESSEL WAS PERFORATED USING A NON-PENUMBRA GUIDEWIRE. THE PHYSICIAN ATTEMPTED TO USE PENUMBRA SMART COILS (SMART COILS) TO STOP THE BLEEDING; HOWEVER, SIX SMART COILS WOULD NOT ADVANCE PAST THEIR INITIAL POSITIONS WITHIN THE INTRODUCER SHEATHS. THEREFORE, THE SMART COILS WERE NOT USED IN THE PROCEDURE. THE NEXT THREE SMART COILS WERE IMPLANTED IN THE TARGET LOCATION USING THE PENUMBRA SMART COIL DETACHMENT HANDLE (HANDLE). WHILE ATTEMPTING TO IMPLANT THE NEXT SMART COIL USING THE HANDLE, IT WOULD NOT DETACH. THE PHYSICIAN ATTEMPTED TO MANUALLY DETACH THE SMART COIL WITHOUT SUCCESS; THEREFORE, THIS SMART COIL WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED USING FIVE SMART COILS AND THE SAME HANDLE. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10354061 ↗MalfunctionMay 19, 2020Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE MIDDLE CEREBRAL ARTERY (MCA) USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, A SMART COIL WOULD NOT ADVANCE AT ALL WITHIN ITS INTRODUCER SHEATH; THEREFORE, IT WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW SMART COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10072208 ↗MalfunctionDec 26, 2019Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING PENUMBRA SMART COILS. DURING THE PROCEDURE, A SMART COIL WOULD NOT ADVANCE AT ALL WITHIN ITS INTRODUCER SHEATH; THEREFORE, IT WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW SMART COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9522032 ↗MalfunctionNov 25, 2019Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE VERTEBRAL ARTERY USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, A SMART COIL WOULD NOT ADVANCE WITHIN ITS INTRODUCER SHEATH; THEREFORE, IT WAS REMOVED. THE PHYSICIAN OPENED ANOTHER SMART COIL, AND IT WOULD NOT ADVANCE WITHIN ITS INTRODUCER SHEATH; THEREFORE, IT WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW SMART COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9373278 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page