Skip to content

Smart Coil 400SMTHXSFT0308Recalls & reported problems

Penumbra, Inc. · SmtCoil-HelXSft, 3x8 · Class 2 · Implant

All models of Smart Coil (98)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Penumbra Smart Coil is an implant used to block blood flow in brain blood vessels. The 42 reports that name its identifier describe malfunctions with no deaths and no injuries, mostly mentioning coils that would not advance into position or would not detach once in place. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
42
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 42 adverse event reports that list this exact device (DI 00814548015798), received between Mar 3, 2017 and Jul 27, 2023 — classified as 42 malfunctions.

The most frequently reported specific device problem is “Failure to Advance” (25 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Failure to Advance25
  • Physical Resistance/Sticking7
  • Difficult or Delayed Separation6
  • Device Handling Problem2
  • Physical Resistance2
  • Separation Failure2
  • Device Operates Differently Than Expected1
  • Difficult or Delayed Activation1

Patient problems

  • No Consequences Or Impact To Patient25
  • No Clinical Signs, Symptoms or Conditions22
  • No Patient Involvement2

Recent adverse event reports

The 25 most recent of 42 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 27, 2023Difficult or Delayed Separation; Premature Separation; Difficult or Delayed Separation; Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE MIDDLE MENINGEAL ARTERY (MMA) USING PENUMBRA SMART COILS (SMART COIL), SMART COIL DETACHMENT HANDLES (HANDLE), A NON-PENUMBRA MICROCATHETER, AND A GUIDEWIRE. DURING THE PROCEDURE, THE PHYSICIAN ADVANCED A SMART COIL USING A MICROCATHETER INTO THE TARGET VESSEL AND ATTEMPTED TO DETACH THE SMART COIL USING THE HANDLE SEVERAL TIMES; HOWEVER, THE SMART COIL FAILED TO DETACH. THEN THE PHYSICIAN USED A SECOND HANDLE WITHOUT SUCCESS. THEREFORE, THE PHYSICIAN DECIDED TO REMOVE THE SMART COIL. WHILE RETRACTING, THE SMART COIL UNINTENTIONALLY DETACHED WITHIN THE MICROCATHETER. THE PHYSICIAN THEN USED THE DETACHED SMART COIL''S PUSHER WIRE TO ADVANCE THE SMART COIL OUT OF THE MICROCATHETER AND INTO THE TARGET VESSEL. NEXT, THE PHYSICIAN ADVANCED A SECOND SMART COIL INTO THE TARGET VESSEL AND ATTEMPTED TO DETACH THE SMART COIL USING THE SECOND HANDLE; HOWEVER, THE SMART COIL WOULD NOT DETACH. THEREFORE, THE SMART COIL WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NON-PENUMBRA COIL AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17402770 ↗
  • MalfunctionJan 27, 2022Difficult or Delayed Separation; Difficult or Delayed Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE RT CAVERNOUS SINUS USING PENUMBRA SMART COILS (SMART COIL), A PENUMBRA SMART COIL DETACHMENT HANDLE (HANDLE) AND A NON-PENUMBRA MICROCATHETER. IT SHOULD BE NOTED THAT THE PATIENT¿S ANATOMY WAS TORTUOUS. DURING THE PROCEDURE, A SMART COIL COULD NOT BE DETACHED USING THE HANDLE. THEN, THE PHYSICIAN ATTEMPTED TO MANUALLY DETACH THE SMART COIL BY BREAKING THE PUSHER ASSEMBLY; HOWEVER, THE ATTEMPT WAS UNSUCCESSFUL. THEREFORE, THE PHYSICIAN DECIDED TO REMOVE THE SMART COIL. WHILE RETRACTING THE SMART COIL, THE SMART COIL UNINTENTIONALLY DETACHED WITHIN THE MICROCATHETER. SALINE WAS THEN USED TO PUSH THE SMART COIL TO THE TARGET LOCATION. THE SAME ISSUE OCCURRED WITH ADDITIONAL THREE SMART COILS. ALL THREE DETACHED SMART COILS WERE IMPLANTED IN THE CAVERNOUS SINUS BY USING SALINE. THE PROCEDURE WAS COMPLETED USING THIRTY-ONE NON-PENUMBRA COILS AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13367463 ↗
  • MalfunctionNov 24, 2021Difficult or Delayed Separation; Difficult or Delayed Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING PENUMBRA SMART COILS (SMART COILS) AND A PENUMBRA SMART COIL DETACHMENT HANDLE (HANDLE). DURING THE PROCEDURE, AFTER ADVANCING A SMART COIL INTO THE TARGET LOCATION, THE HANDLE WAS USED TO DETACH THE COIL; HOWEVER, THE COIL DID NOT DETACH. IT WAS REPORTED THAT MULTIPLE HANDLES WERE USED AND THE COIL EVENTUALLY DETACHED. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12870472 ↗
  • MalfunctionNov 8, 2021Difficult to Advance; Difficult to Advance

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE ANTERIOR COMMUNICATING ARTERY (ACOM) USING PENUMBRA SMART COILS (SMART COILS) AND A NON-PENUMBRA MICROCATHETER. DURING THE PROCEDURE, A SMART COIL WAS REPORTED TO BE STIFF AND COULD NOT BE FULLY INSERTED INTO THE ANEURYSM. IT WAS ALSO REPORTED THAT THE SMART COIL KICKED THE MICROCATHETER OUT OF THE TARGET LOCATION. AFTER SEVERAL ATTEMPTS TO IMPLANT THE SAME SMART COIL, THE SAME ISSUE CONTINUED TO OCCUR. THEREFORE, THE SMART COIL WAS REMOVED. NEXT, THE SAME ISSUE OCCURRED WITH THE NEXT TWO SMART COILS. THEREFORE, THESE SMART COILS WERE ALSO REMOVED. THE PROCEDURE WAS COMPLETED USING NON-PENUMBRA COILS AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12771458 ↗
  • MalfunctionOct 12, 2021Failure to Advance; Failure to Advance

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT A CAROTID TERMINUS ANEURYSM USING PENUMBRA SMART COILS (SMART COIL). DURING THE PROCEDURE, A SMART COIL WAS UNABLE TO BE ADVANCED AT ALL PAST ITS INITIAL POSITION WITHIN ITS INTRODUCER SHEATH. THEREFORE, THE SMART COIL WAS REMOVED. SUBSEQUENTLY, THE SAME ISSUE OCCURRED WITH TWO ADDITIONAL SMART COILS. THEREFORE, THE SMART COILS WERE ALSO REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER SMART COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12618303 ↗
  • MalfunctionSep 23, 2021Failure to Advance; Failure to Advance

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE ANTERIOR COMMUNICATING ARTERY USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, TWO SMART COILS WERE UNABLE TO ADVANCE AT ALL FROM THEIR INITIAL POSITIONS WITHIN THEIR RESPECTIVE INTRODUCER SHEATHS. THEREFORE, THE TWO SMART COILS WERE REMOVED. THE PROCEDURE WAS COMPLETED USING NEW SMART COILS. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12523506 ↗
  • MalfunctionSep 20, 2021Material Deformation; Physical Resistance/Sticking; Material Deformation; Physical Resistance/Sticking

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE MIDDLE CEREBRAL ARTERY (MCA) USING PENUMBRA SMART COILS (SMART COIL), A NON-PENUMBRA MICROCATHETER, AND A GUIDEWIRE. DURING THE PROCEDURE, WHILE ADVANCING A SMART COIL THROUGH THE MICROCATHETER, THE PHYSICIAN EXPERIENCED RESISTANCE AND, SUBSEQUENTLY, KINKED THE PUSHER ASSEMBLY OF THE SMART COIL. THEREFORE, THE SMART COIL WAS REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER SMART COIL AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12496836 ↗
  • MalfunctionSep 11, 2021Failure to Advance; Failure to Advance

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE BASILAR ARTERY USING PENUMBRA SMART COILS (SMART COILS) AND A NON-PENUMBRA MICROCATHETER. DURING THE PROCEDURE, THE PHYSICIAN PLACED SOME SMART COILS INTO THE TARGET LOCATION USING THE MICROCATHETER. AFTERWARDS, TWO SMART COILS WOULD NOT ADVANCE AT ALL FROM THEIR INITIAL POSITION WITHIN THEIR RESPECTIVE INTRODUCER SHEATHS. THEREFORE, THE TWO SMART COILS WERE REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW SMART COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12457572 ↗
  • MalfunctionSep 10, 2021Physical Resistance/Sticking; Physical Resistance/Sticking

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT A BASILAR TIP ANEURYSM USING PENUMBRA SMART COILS (SMART COIL) AND A NON-PENUMBRA MICROCATHETER. DURING THE PROCEDURE, THE SMART COIL WOULD NOT ADVANCE PAST THE FRICTION LOCK WITHIN ITS INTRODUCER SHEATH. THEREFORE, THE SMART COIL WAS REMOVED. NEXT, WHILE ATTEMPTING TO PLACE ANOTHER SMART COIL INTO THE ANEURYSM, THE PHYSICIAN EXPERIENCED RESISTANCE AND THE MICROCATHETER AND THE SMART COIL KICKED BACK. IT WAS REPORTED THAT THE SMART COIL WAS ABLE TO FORM IN THE ANEURYSM BUT THE PHYSICIAN DID NOT FEEL SAFE TO ADVANCE THE SMART COIL FURTHER. THEREFORE, THE SMART COIL WAS REMOVED INTACT. THE PROCEDURE WAS COMPLETED USING ADDITIONAL SMART COILS AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12456347 ↗
  • MalfunctionJun 15, 2021Failure to Advance; Failure to Advance

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE CAVERNOUS SINUS USING PENUMBRA SMART COILS (SMART COILS) AND A NON-PENUMBRA MICROCATHETER. DURING THE PROCEDURE, THE SMART COIL WOULD NOT ADVANCE PAST ITS INITIAL POSITION WITHIN ITS INTRODUCER SHEATH. THEREFORE, THE SMART COIL WAS REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER SMART COIL AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12004347 ↗
  • MalfunctionJun 3, 2021Failure to Advance; Failure to Advance

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE INTERNAL CAROTID ARTERY (ICA) USING PENUMBRA SMART COILS (SMART COIL). DURING THE PROCEDURE, A SMART COIL WOULD NOT ADVANCE AT ALL PAST ITS INITIAL POSITION WITHIN ITS INTRODUCER SHEATH. THEREFORE, THE SMART COIL WAS REMOVED. SUBSEQUENTLY, THE SAME ISSUE OCCURRED WITH FOUR OTHER SMART COILS; THEREFORE, THE FOUR SMART COILS WERE ALSO REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER SMART COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11927501 ↗
  • MalfunctionMay 10, 2021Failure to Advance; Failure to Advance

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE ANTERIOR COMMUNICATING ARTERY USING PENUMBRA SMART COILS (SMART COILS), A PENUMBRA SMART COIL DETACHMENT HANDLE (HANDLE), AND A NON-PENUMBRA MICROCATHETER. DURING THE PROCEDURE, THE PHYSICIAN SUCCESSFULLY DETACHED A SMART COIL IN THE TARGET VESSEL USING THE HANDLE. HOWEVER, AFTER THE DETACHMENT, THE PUSHER ASSEMBLY OF THE SMART COIL BECAME STUCK IN THE HANDLE. SUBSEQUENTLY, THE PUSHER ASSEMBLY WAS REMOVED FROM THE HANDLE. THE PHYSICIAN THEN ATTEMPTED TO ADVANCE ANOTHER SMART COIL INTO THE MICROCATHETER; HOWEVER, THE SMART COIL WOULD NOT ADVANCE OUT OF ITS INTRODUCER SHEATH. THEREFORE, THE SMART COIL WAS REMOVED. THE PROCEDURE WAS COMPLETED USING ADDITIONAL SMART COILS, THE SAME HANDLE, AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11802891 ↗
  • MalfunctionApr 23, 2021Separation Failure; Physical Resistance/Sticking; Separation Failure; Physical Resistance/Sticking

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE INTERNAL CAROTID ARTERY (ICA) USING PENUMBRA SMART COILS (SMART COILS), PENUMBRA SMART COIL DETACHMENT HANDLES (HANDLES), AND TWO NON-PENUMBRA MICROCATHETERS. DURING THE PROCEDURE, THE FIRST SMART COIL WOULD NOT DETACH AFTER MULTIPLE ATTEMPTS WERE MADE WITH THREE HANDLES. WHILE RETRACTING THE SMART COIL, THE PHYSICIAN EXPERIENCED RESISTANCE, AND THE SMART COIL WAS STUCK IN THE FIRST MICROCATHETER. THE SMART COIL WAS THEN REMOVED. SUBSEQUENTLY, THE PHYSICIAN USED A SECOND SMART COIL, THE THIRD HANDLE, AND THE SAME MICROCATHETER. HOWEVER, THE PHYSICIAN WAS ALSO UNABLE TO DETACH THE SMART COIL WITH THE HANDLE AFTER MULTIPLE ATTEMPTS WERE MADE. WHILE RETRACTING THE SMART COIL, THE PHYSICIAN EXPERIENCED RESISTANCE, AND THE SMART COIL WAS STUCK IN THE MICROCATHETER. THE SMART COIL AND MICROCATHETER WERE THEN REMOVED. THE PHYSICIAN RESUMED THE PROCEDURE AND SUCCESSFULLY PLACED THE THIRD SMART COIL WITH THE HANDLE AND SECOND MICROCATHETER. NEXT, THE PHYSICIAN USED A FOURTH SMART COIL, THE SAME HANDLE, AND THE SAME SECOND MICROCATHETER. HOWEVER, THE PHYSICIAN WAS ALSO UNABLE TO DETACH THE SMART COIL WITH THE HANDLE AFTER MULTIP

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11716626 ↗
  • MalfunctionApr 22, 2021Failure to Advance; Physical Resistance/Sticking; Failure to Advance; Physical Resistance/Sticking

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE ANTERIOR COMMUNICATING ARTERY (ACOM) USING A SMART COIL. DURING THE PROCEDURE, WHILE ADVANCING A SMART COIL WITHIN ITS INTRODUCER SHEATH, THE PHYSICIAN EXPERIENCED RESISTANCE. SUBSEQUENTLY, THE SMART COIL BECAME STUCK AND WOULD NOT ADVANCE FURTHER WITHIN ITS INTRODUCER SHEATH. THEREFORE, THE SMART COIL WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW SMART COIL AND A NON-PENUMBRA COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11712235 ↗
  • MalfunctionMar 25, 2021Difficult or Delayed Separation; Difficult or Delayed Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE INTERNAL CAROTID ARTERY (ICA) USING A PENUMBRA SMART COIL (SMART COIL). DURING THE PROCEDURE, THE PHYSICIAN ADVANCED A SMART COIL INTO THE TARGET VESSEL USING A MICROCATHETER AND ATTEMPTED TO DETACH IT USING A PENUMBRA SMART COIL DETACHMENT HANDLE (HANDLE). HOWEVER, THE SMART COIL FAILED TO DETACH. THEREFORE, THE PHYSICIAN MANUALLY DETACHED THE SMART COIL. THE PROCEDURE ENDED AT THIS POINT. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11570712 ↗
  • MalfunctionJan 5, 2021Failure to Advance; Failure to Advance

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING A PENUMBRA SMART COIL (SMART COIL). DURING THE PROCEDURE, A SMART COIL WOULD NOT ADVANCE AT ALL WITHIN ITS INTRODUCER SHEATH; THEREFORE, IT WAS REMOVED. THE PROCEDURE WAS COMPLETED USING ADDITIONAL SMART COILS. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11120457 ↗
  • MalfunctionNov 20, 2020Failure to Advance; Failure to Advance

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING IN THE RIGHT MIDDLE CEREBRAL ARTERY (MCA) PENUMBRA SMART COILS (SMART COILS) AND A NON-PENUMBRA MICROCATHETER. DURING THE PROCEDURE, A SMART COIL WOULD NOT ADVANCE FROM THE STARTING POSITION WITHIN THE INTRODUCER SHEATH. THEREFORE, THE SMART COIL WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW SMART COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10878764 ↗
  • MalfunctionNov 6, 2020Physical Resistance/Sticking; Physical Resistance/Sticking

    THE PATIENT WAS A COIL EMBOLIZATION IN THE POSTERIOR COMMUNICATING ARTERY (PCOM) USING PENUMBRA SMART COILS (SMART COILS) AND A NON-PENUMBRA MICROCATHETER. DURING THE PROCEDURE, A SMART COIL WAS STUCK AT THE STARTING POSITION OF THE INTRODUCER SHEATH. THEREFORE, THE SMART COIL WAS REMOVED. IT WAS REPORTED THAT THE FRICTION LOCK OF THE SMART COIL APPEARED TO BE TIGHT. THE PROCEDURE WAS COMPLETED USING A NEW SMART COIL AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10797628 ↗
  • MalfunctionOct 21, 2020Failure to Advance; Failure to Advance

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, THE PHYSICIAN REPORTED TWO SMART COILS WOULD NOT DEPLOY. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10712722 ↗
  • MalfunctionOct 16, 2020Failure to Advance; Failure to Advance

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE RIGHT VERTEBRAL ARTERY USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, A SMART COIL WOULD NOT ADVANCE WITHIN ITS INTRODUCER SHEATH; THEREFORE, IT WAS REMOVED. WHILE ATTEMPTING TO ADVANCE ANOTHER SMART COIL, THE SAME ISSUE OCCURRED. THEREFORE, THE SMART COIL WAS ALSO REMOVED. THE PROCEDURE WAS COMPLETED USING ADDITIONAL SMART COILS. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THIS PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10693544 ↗
  • MalfunctionOct 16, 2020Failure to Advance; Failure to Advance

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE VERTEBRAL ARTERY USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, THE SMART COIL WOULD NOT ADVANCE PAST ITS INITIAL POSITION WITHIN ITS INTRODUCER SHEATH AFTER SEVERAL ATTEMPTS. THEREFORE, THE SMART COIL WAS REMOVED. THE PROCEDURE WAS COMPLETED USING ADDITIONAL SMART COILS AND PAC400 COILS. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10691411 ↗
  • MalfunctionOct 2, 2020Difficult or Delayed Separation; Difficult or Delayed Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING A PENUMBRA SMART COIL (SMART COIL) AND A PENUMBRA SMART COIL DETACHMENT HANDLE (HANDLE). DURING THE PROCEDURE, THE PHYSICIAN DETACHED A SMART COIL INTO THE TARGET LOCATION USING THE HANDLE AFTER CLICKING A FEW TIMES. AFTERWARDS, THE PHYSICIAN ATTEMPTED TO REMOVE THE PUSHER ASSEMBLY OF THE SMART COIL FROM THE HANDLE; HOWEVER, THE PUSHER ASSEMBLY WAS STUCK. SUBSEQUENTLY, THE PHYSICIAN WAS UNABLE TO REMOVE THE PUSHER ASSEMBLY OF THE SMART COIL FROM THE HANDLE. THE PROCEDURE ENDED AT THIS POINT. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10624025 ↗
  • MalfunctionAug 11, 2020Failure to Advance; Failure to Advance

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE BASILAR ARTERY USING PENUMBRA SMART COILS (SMART COILS). DURING THE PROCEDURE, TWO SMART COILS WOULD NOT ADVANCE WITHIN THE INTRODUCER SHEATHS AND WERE STUCK IN THEIR INITIAL POSITION. THEREFORE, THE PHYSICIAN PULLED THE SMART COILS OUT FROM THE BACKEND OF THE INTRODUCER SHEATHS AND BACKLOADED THE SMART COILS INTO THE INTRODUCER SHEATHS. THE PROCEDURE WAS COMPLETED USING THE SAME TWO SMART COILS. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10394485 ↗
  • MalfunctionJul 9, 2020Failure to Advance; Failure to Advance

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE LEFT INTERNAL CAROTID ARTERY (ICA) USING A PENUMBRA SMART COIL (SMART COIL). DURING THE PROCEDURE, THE SMART COIL WOULD NOT ADVANCE WITHIN ITS INTRODUCER SHEATH; THEREFORE, IT WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW SMART COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10255461 ↗
  • MalfunctionMar 13, 2020Failure to Advance; Failure to Advance

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE FOR A VESSEL SACRIFICE IN THE LEFT VERTEBRAL ARTERY USING PENUMBRA SMART COILS (SMART COILS) AND NON-PENUMBRA MICROCATHETER. DURING THE PROCEDURE, THE PHYSICIAN PLACED MULTIPLE SMART COILS IN THE TARGET VESSEL USING THE MICROCATHETER. THE PHYSICIAN THEN ATTEMPTED TO ADVANCE ANOTHER SMART COIL; HOWEVER, THE COIL WOULD NOT ADVANCE AT ALL WITHIN ITS INTRODUCER SHEATH. THEREFORE, THE SMART COIL WAS REMOVED. THE PHYSICIAN ATTEMPTED TO ADVANCE A NEW SMART COIL, HOWEVER, THE SAME ISSUE OCCURRED, AND THE COIL WOULD NOT ADVANCE WITHIN ITS INTRODUCER SHEATH; THEREFORE, THE SMART COIL WAS REMOVED. THE PROCEDURE WAS COMPLETED USING NEW SMART COILS. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9831178 ↗

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices

All DEVICE, NEUROVASCULAR EMBOLIZATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page