Indigo System CATRXKIT160Recalls & reported problems
Penumbra, Inc. · 044 Coronary Aspiration Catheter 160 cm Kit · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00814548017587), received between Aug 3, 2017 and Sep 21, 2017 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Aspiration Issue” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Aspiration Issue2
- Bent1
- Kinked1
- Material Deformation1
- Physical Resistance1
- Sticking1
Patient problems
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 21, 2017Kinked; Aspiration Issue; Kinked; Aspiration Issue
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE ANTERIOR TIBIAL ARTERY USING AN INDIGO SYSTEM CAT RX KIT (KIT). DURING THE PROCEDURE, WHILE ADVANCING DOWN THE WIRE JUST PROXIMAL TO WHERE THE RAPID EXCHANGE PART BEGINS, THE PHYSICIAN DID NOT MENTION RESISTANCE; HOWEVER, THE INDIGO SYSTEM CAT RX ASPIRATION CATHETER (CATRX) BECAME KINKED AND CONSEQUENTLY, THE CATRX WOULD NOT ASPIRATE. THEREFORE, THE CATRX WAS REMOVED AND THE PROCEDURE WAS COMPLETED USING MANUAL ASPIRATION WITH ANOTHER CATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6884515 ↗MalfunctionAug 10, 2017Bent; Aspiration Issue; Bent; Aspiration Issue
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE ULNAR ARTERY USING AN INDIGO SYSTEM CAT RX KIT (KIT). DURING THE PROCEDURE, THE PHYSICIAN ADVANCED THE INDIGO SYSTEM CAT RX ASPIRATION CATHETER (CATRX) INTO THE TARGET VESSEL THROUGH A NON-PENUMBRA SHEATH AND OVER A NON-PENUMBRA WIRE. AFTER ASPIRATION WAS PERFORMED FOR A TOTAL OF THREE MINUTES IN THE ULNAR ARTERY, THE CATRX WAS REMOVED. THE PHYSICIAN DID NOT MENTION RESISTANCE DURING THE PROCEDURE; HOWEVER, ASPIRATION STOPPED. WHILE REMOVING THE CATRX FROM THE SHEATH, THE PHYSICIAN NOTICED THAT IT WAS BENT. THEREFORE, THE CATRX WAS SET ASIDE AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING AN INDIGO SYSTEM CAT3 ASPIRATION CATHETER (CAT3) THROUGH THE SAME SHEATH AND OVER THE SAME WIRE. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6785429 ↗MalfunctionAug 3, 2017Sticking; Physical Resistance; Material Deformation; Sticking; Physical Resistance; Material Deformation
THE PATIENT WAS UNDERGOING THROMBECTOMY PROCEDURE USING AN INDIGO SYSTEM CAT RX KIT (KIT). DURING THE PROCEDURE, THE PHYSICIAN COMPLETED TWO PASSES IN THE PERONEAL ARTERY USING AN INDIGO SYSTEM CAT RX ASPIRATION CATHETER (CAT RX) WITH A NON-PENUMBRA DESTINATION SHEATH AND A NON-PENUMBRA WIRE. IT WAS REPORTED THAT THE CAT RX WAS FLUSHED AFTER EACH PASS. WHILE ADVANCING THE CAT RX TO THE TARGET VESSEL FOR THE THIRD PASS UNDER ASPIRATION, THE PHYSICIAN EXPERIENCED RESISTANCE. IT WAS ALSO REPORTED THAT THERE WAS CONSTANT BLOOD FLOW BUT THE CLOT WAS NEVER ENGAGED; THEREFORE, THE PHYSICIAN DECIDED TO REMOVE THE CAT RX. HOWEVER, WHILE REMOVING THE CAT RX, THE CAT RX BEGAN STICKING ON THE WIRE AND THE PHYSICIAN EXPERIENCED RESISTANCE; THEREFORE, THE PHYSICIAN PULLED THE CAT RX WITH THE WIRE ALL TOGETHER. IT WAS ALSO REPORTED THAT THE CAT RX PORT WAS BENDING LIKE AN ACCORDION WHEN REMOVING IT OVER THE WIRE OUTSIDE THE PATIENT¿S BODY. THE PROCEDURE WAS COMPLETED USING PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) ONCE. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6764522 ↗
Other Catheter, Embolectomy devices
- Indigo System CAT6Penumbra, Inc. · CATHETER, EMBOLECTOMY2 deaths408 reports
- Indigo System CATRXKITPenumbra, Inc. · Catheter, Embolectomy1 death339 reports
- Fogarty 120804FPEdwards Lifesciences LLC · Catheter, Embolectomy1 recall1 death200 reports
- LeMaitre Embolectomy Catheter 1651-88Lemaitre Vascular, Inc. · Catheter, Embolectomy1 recall161 reports
- FOGARTY 120404FPEdwards Lifesciences LLC · Catheter, Embolectomy1 recall61 reports
- Indigo System CAT6KIT-APenumbra, Inc. · Catheter, Embolectomy109 reports
- LeMaitre Embolectomy Catheter e1651-88Lemaitre Vascular, Inc. · Catheter, Embolectomy2 recalls4 reports
- Indigo System CAT8XTORQ115Penumbra, Inc. · CATHETER, EMBOLECTOMY2 deaths98 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.