Indigo System SEPC4Recalls & reported problems
Penumbra, Inc. · Separator 4 · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00814548017594), received between Jan 11, 2019 and Nov 9, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break2
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 9, 2021Break; Physical Resistance/Sticking; Break; Physical Resistance/Sticking
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE LEFT PERONEAL ARTERY, LEFT ANTERIOR TIBIAL ARTERY AND POSTERIOR TIBIAL ARTERY USING AN INDIGO SYSTEM SEPARATOR 4 (SEP4), AN INDIGO SYSTEM CATRX ASPIRATION CATHETER (CATRX), A NON-PENUMBRA SHEATH AND A GUIDEWIRE. DURING THE PROCEDURE, THE PHYSICIAN SUCCESSFULLY REVASCULARIZED THE LEFT PERONEAL ARTERY BY USING THE CATRX AND THE SEP4. AFTERWARDS, THE PHYSICIAN ADVANCED THE CATRX TO THE LEFT ANTERIOR TIBIAL ARTERY WITH THE HELP OF A GUIDEWIRE AND ADVANCED THE SEP4 THROUGH THE CATRX. THE PHYSICIAN ENCOUNTERED RESISTANCE AS THE SEP4 WAS BEING ADVANCED 1-2 CM PAST THE DISTAL END OF THE CATRX DUE TO SIGNIFICANT CHRONIC THROMBUS IN THE VESSEL. AFTER COMPLETING A FEW PASSES, THE PHYSICIAN NOTICED THAT THE SEP4 BULB WAS NOT MOVING ALONG WITH ITS PUSHER WIRE. HE THEN SUCTIONED THE SEP4 BULB TO THE TIP OF THE CATRX. HE RETRACTED THE CATRX WITH ASPIRATION TURNED ON AND WAS ABLE TO REMOVE THE SEP4 AND THE BULB COMPLETELY FROM THE PATIENT'S VESSEL. THE PROCEDURE WAS COMPLETED USING A NEW SEP4 AND THE SAME CATRX. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12782770 ↗MalfunctionJan 11, 2019Break; Break
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE SUPERIOR MESENTERIC VEIN (SMV) USING AN INDIGO SYSTEM SEPARATOR 4 (SEP4). DURING THE PROCEDURE, THE PHYSICIAN MADE APPROXIMATELY 4 TO 5 PASSES USING THE SEP4 WITHIN AN INDIGO SYSTEM ASPIRATION CATHETER 6 (CAT6). HOWEVER, THE BULB OF THE SEP4 BEGAN TO UNRAVEL AND, THEREFORE, THE SEP4 WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW SEP4 WITH THE SAME CAT6. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8242332 ↗
Other Catheter, Embolectomy devices
- Indigo System CAT6Penumbra, Inc. · CATHETER, EMBOLECTOMY2 deaths408 reports
- Indigo System CATRXKITPenumbra, Inc. · Catheter, Embolectomy1 death339 reports
- Fogarty 120804FPEdwards Lifesciences LLC · Catheter, Embolectomy1 recall1 death200 reports
- LeMaitre Embolectomy Catheter 1651-88Lemaitre Vascular, Inc. · Catheter, Embolectomy1 recall161 reports
- FOGARTY 120404FPEdwards Lifesciences LLC · Catheter, Embolectomy1 recall61 reports
- Indigo System CAT6KIT-APenumbra, Inc. · Catheter, Embolectomy109 reports
- LeMaitre Embolectomy Catheter e1651-88Lemaitre Vascular, Inc. · Catheter, Embolectomy2 recalls4 reports
- Indigo System CAT8XTORQ115Penumbra, Inc. · CATHETER, EMBOLECTOMY2 deaths98 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.