Penumbra, Inc. · IS ENGINE Canister Assembly · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00814548019703), received between Apr 12, 2019 and Jun 5, 2020 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Suction Problem” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Suction Problem6
  • Pressure Problem3
  • Appropriate Device Problem Term/Code Not Available1
  • Connection Problem1

Patient problems

  • No Consequences Or Impact To Patient11

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 5, 2020Suction Problem; Suction Problem

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE PULMONARY ARTERY USING A PENUMBRA ENGINE CANISTER (CANISTER) AND A PENUMBRA ENGINE (ENGINE). DURING THE PROCEDURE, THE CANISTER WAS PLACED IN THE ENGINE, AND NONE OF THE FOUR INDICATOR LIGHTS ON THE ENGINE ILLUMINATED. THEN, THE PENUMBRA STAFF ATTEMPTED TO TROUBLESHOOT BY RESECURING THE CANISTER AND RESTARTING THE ENGINE. HOWEVER, THE ISSUE WAS NOT RESOLVED; THEREFORE, THE CANISTER WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW CANISTER AND THE SAME ENGINE. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10124536 ↗
  • MalfunctionMar 26, 2020Suction Problem; Suction Problem

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE USING A PENUMBRA ENGINE CANISTER (CANISTER) AND PENUMBRA ENGINE (ENGINE). DURING THE PROCEDURE, THE PHYSICIAN NOTICED THAT THE CANISTER WOULD NOT CREATE VACUUM AFTER BEING SEATED ON THE ENGINE AND NONE OF THE LIGHTS ON THE ENGINE WOULD TURN ON. THEREFORE, THE CANISTER WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW CANISTER AND THE SAME ENGINE. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9885101 ↗
  • MalfunctionMar 25, 2020Connection Problem; Connection Problem

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY (SFA) USING A THE PENUMBRA ENGINE CANISTER (CANISTER) AND A PENUMBRA ENGINE (ENGINE). DURING THE PROCEDURE, THREE INDICATOR LIGHTS ON THE ENGINE WERE NOT ILLUMINATING. THE TECHNICIAN REALIZED THAT THE CANISTER WAS NOT SEALED PROPERLY WITH THE PENUMBRA ENGINE (ENGINE) AND ATTEMPTED TO PUSH THE CANISTER DOWN; HOWEVER, THE ISSUE DID NOT RESOLVE. THEREFORE, THE CANISTER WAS REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER CANISTER AND THE SAME ENGINE. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9881346 ↗
  • MalfunctionDec 20, 2019Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    THE PATIENT WAS UNDERGOING A MEDICAL PROCEDURE USING A PENUMBRA ENGINE CANISTER (CANISTER) AND PENUMBRA ENGINE (ENGINE). DURING THE PROCEDURE, THE HOSPITAL STAFF SEATED THE CANISTER ON THE ENGINE. WHILE THE ENGINE WAS POWERED ON, NOT ALL FOUR OF THE INDICATOR LIGHTS WERE ILLUMINATED; THEREFORE, THE CANISTER WAS REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER CANISTER AND THE SAME ENGINE. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9507230 ↗
  • MalfunctionOct 8, 2019Pressure Problem; Pressure Problem

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE POPLITEAL ARTERY USING A PENUMBRA ENGINE CANISTER (CANISTER). DURING THE PROCEDURE, THE PHYSICIAN CONNECTED THE ASPIRATION TUBING TO THE CATHETER AND ATTEMPTED TO INITIATE THE ASPIRATION; HOWEVER, NO VACUUM WAS PRODUCED IN THE CANISTER. IT WAS ALSO REPORTED THAT THE FOUR INDICATOR LIGHTS ON THE PENUMBRA ENGINE (ENGINE) WERE NOT ILLUMINATING. THEREFORE, THE CANISTER WAS REMOVED AND REPLACED WITH A NEW CANISTER. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE SAME ENGINE AND THE SECOND CANISTER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9169848 ↗
  • MalfunctionJul 26, 2019Pressure Problem; Pressure Problem

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE RIGHT CORONARY ARTERY (RCA) USING A PENUMBRA ENGINE CANISTER (CANISTER). DURING THE PROCEDURE, AFTER MAKING A SUCCESSFUL PASS USING THE INDIGO SYSTEM CATRX ASPIRATION CATHETER (CATRX) WITH THE PENUMBRA ENGINE PUMP (ENGINE), THE PHYSICIAN REMOVED THE CATRX TO TAKE A FEW PICTURES AND FLUSHED IT. HOWEVER, AFTER REATTACHING THE CATRX AND THE ASPIRATION TUBING (TUBING), THE PHYSICIAN NOTICED ONLY TWO LIGHTS WOULD ILLUMINATE ON THE ENGINE. THEREFORE, THE PHYSICIAN VERIFIED TO ENSURE THE CONNECTIONS BETWEEN THE TUBING AND THE CATRX WERE TIGHT; HOWEVER, THERE WAS NO CHANGE IN ASPIRATION POWER. THEREFORE, THE PHYSICIAN DECIDED TO SWITCH OUT FOR ANOTHER CATRX AS PREFERENCE. THE PROCEDURE WAS COMPLETED USING A NEW CATRX WITH THE SAME ENGINE AND CANISTER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8833899 ↗
  • MalfunctionJun 11, 2019Suction Problem; Suction Problem

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE USING A PENUMBRA SYSTEM ENGINE CANISTER (CANISTER). DURING THE PROCEDURE, IT WAS REPORTED THAT ONLY TWO LIGHTS WERE ILLUMINATING ON THE PENUMBRA SYSTEM ENGINE (ENGINE). THE TECHNICIAN THEN APPLIED PRESSURE TO THE CANISTER AND ALL FOUR LIGHTS ILLUMINATED. THE PROCEDURE WAS THEN CONTINUED IN THIS MANNER. WHEN THE PHYSICIAN WAS MAKING THE THIRD PASS, IT WAS REPORTED THAT ONLY THREE LIGHTS WERE ILLUMINATING ON THE ENGINE; HOWEVER, THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8689139 ↗
  • MalfunctionMay 30, 2019Suction Problem; Suction Problem

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE RIGHT ILIAC VEIN USING A PENUMBRA ENGINE CANISTER (CANISTER). DURING THE PROCEDURE, THE CANISTER WOULDN'T HOLD PRESSURE. IT WAS REPORTED THAT ONLY TWO OF THE FOUR INDICATOR LIGHTS WOULD ILLUMINATE. THEREFORE, THE CANISTER WAS REPLACED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8656575 ↗
  • MalfunctionMay 16, 2019Suction Problem; Suction Problem

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE USING A PENUMBRA ENGINE CANISTER (CANISTER) AND AN INDIGO SYSTEM ASPIRATION CATHETER 8 (CAT8). DURING THE PROCEDURE, THE PENUMBRA ENGINE (ENGINE) WAS UNABLE TO BUILD VACUUM WITH THE CANISTER AND, THEREFORE, THE PHYSICIAN SWITCHED TO A NEW CANISTER. THE PROCEDURE WAS THEN CONTINUED, AND THE PHYSICIAN USED A CAT8 TO ASPIRATE WITHIN THE PATIENT. AFTER MAKING A PASS, THE CAT8 WAS REMOVED AND FLUSHED ON THE BACK TABLE. HOWEVER, THE PROXIMAL END OF THE CAT8 WAS INADVERTENTLY KINKED WHILE BEING FLUSHED. THE PROCEDURE WAS THEN COMPLETED USING A NEW CAT8. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8615414 ↗
  • MalfunctionApr 24, 2019Pressure Problem; Pressure Problem

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE USING A PENUMBRA ENGINE CANISTER (CANISTER). DURING THE PROCEDURE, THE HOSPITAL TECHNOLOGIST PLACED THE CANISTER ON THE PENUMBRA ENGINE (ENGINE), CONNECTED THE ASPIRATION TUBING AND THEN ATTEMPTED TO BUILD NEGATIVE PRESSURE. HOWEVER, THE ENGINE DID NOT BUILD PRESSURE. IT WAS REPORTED THAT ONLY TWO OF THE FOUR LIGHT BARS LIT UP, DESPITE ATTEMPTING TWICE. THEREFORE, THE CANISTER WAS REPLACED AND THE PROCEDURE WAS COMPLETED USING THE SAME ENGINE AND AN INDIGO SYSTEM CAT RX ASPIRATION CATHETER (CATRX). THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8544902 ↗
  • MalfunctionApr 12, 2019Suction Problem; Suction Problem

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE RIGHT CORONARY ARTERY (RCA) USING AN INDIGO SYSTEM CATRX ASPIRATION CATHETER (CATRX) AND INDIGO SYSTEM ASPIRATION TUBING (TUBING) FROM AN INDIGO SYSTEM CATRX KIT, AND A PENUMBRA ENGINE CANISTER (CANISTER). DURING THE PROCEDURE, THE PHYSICIAN MADE ONE PASS USING THE CATRX AND REPORTED THAT THE ASPIRATION HAD STOPPED WORKING. HE DECIDED TO REMOVE THE CATRX AND TUBING FROM USE IN THE PATIENT AND CONFIRMED THAT THERE WAS NO ASPIRATION USING THE SYSTEM. SUBSEQUENTLY, THE PHYSICIAN REMOVED THE CATRX, TUBING, AND CANISTER, AND COMPLETED THE PROCEDURE USING A NEW CATRX, NEW TUBING, AND A NEW CANISTER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8511254 ↗

FDA clearance and classification

Cleared through Special 510(k) K180105 on Mar 8, 2018 (decision: Substantially Equivalent). Submitted by Penumbra, Inc. as “Indigo Aspiration System - Penumbra Engine Pump, Indigo Aspiration System - Penumbra Engine Pump Canister”. The FDA review took 51 days. FDA record ↗

The FDA regulates product code DXE (catheter, embolectomy) under 21 CFR 870.5150 as a Class II device. 21 CFR 870.5150 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Indigo Aspiration System IAPS3-A

Has the Indigo Aspiration System IAPS3-A been recalled?

No FDA recall record in our data names this exact device identifier (DI 00814548019703), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Indigo Aspiration System IAPS3-A?

The FDA's MAUDE database lists 11 adverse event reports for this device, received between Apr 12, 2019 and Jun 5, 2020. In those reports, the most common specific device problems are “Suction Problem” (6), “Pressure Problem” (3) and “Connection Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Indigo Aspiration System IAPS3-A?

None of the reports are classified as deaths, 0 as injuries and 11 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Indigo Aspiration System IAPS3-A FDA approved?

It was cleared through Special 510(k) K180105 on Mar 8, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Indigo Aspiration System IAPS3-A, and is it still sold?

The FDA's device database lists Penumbra, Inc. as the company responsible for it (brand Indigo Aspiration System, model IAPS3-A). Its record shows it as in commercial distribution.