Skip to content

MotoBAND™ CP DynaForce® DynaBunion Plate 18mm Right Single PackRecalls & reported problems

Crossroads Extremity Systems · The MotoBAND™ CP DynaForce® Implant System is a bone plating system indicated for stabilization and fixation of fresh fractures, revision procedures, joint fusion and reconstruction of small bones of the hand, feet, wrist, ankles, fingers and toes. It consist of plates and screws. Certain plates in the system are compatible with the MotoCLIP/DynaForce Superelastic Fixation System. · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00815432028870), received between Dec 8, 2021 and Sep 29, 2022 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Migration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1
  • Migration1

Patient problems

  • Implant Pain1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2022Migration; Migration

    AT ROUTINE FOLLOW-UP, THE SURGEON IDENTIFIED THE ANTI-DRIFT BOLT WAS THROUGH THE PLATE AT 2 WEEKS POST-OP. AT 6 WEEKS POST-OP, THE SURGEON NOTES THAT THE IM ANGLE IS INCREASED. THE PATIENT IS ASYMPTOMATIC, AND NO REVISION IS PLANNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15501543 ↗
  • MalfunctionDec 8, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    THE SURGEON SUBMITTED A CASE SERIES OF DYNABUNION SYSTEM. THE ANTI-DRIFT BLOT HAS BACKED OUT AS SHOWN IN THE 6-WEEK X-RAY. THE SCREW HEAD PROMINENT ABOVE THE PLATE. THE PATIENT WAS ASYMPTOMATIC AND REVISION SURGERY WAS NOT SCHEDULED AT THE TIME OF THE REPORT. THE SURGEON REPORTED AT A LATER DATE THAT THE SYSTEM WAS REMOVED AFTER A PERIOD OF NON-WEIGHT BEARING. THE PATIENT STATUS AFTER THE REVISION IS UNKNOWN.

    Patient problems: Implant Pain

    MAUDE report 12964836 ↗

Other Plate, Fixation, Bone devices

All Plate, Fixation, Bone devices →

Sources and updates